# BARD PROFORMA AND APOLLO PRE-CURVED PAPILLOTOMES (K982557)

_Bard Interventional Products Div. · KNS · Sep 10, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K982557

## Device Facts

- **Applicant:** Bard Interventional Products Div.
- **Product Code:** [KNS](/productcode/KNS.md)
- **Decision Date:** Sep 10, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Bard papillotomes are designed and recommended for transendoscopic cannulation and sphincterotomy of the Papilla of Vater and/or Sphincter of Oddi.

## Device Story

The Bard ProForma and Apollo3 Papillotomes are endoscopic electrosurgical accessories used for cannulation and sphincterotomy of the Papilla of Vater or Sphincter of Oddi. The device is inserted through the biopsy channel of a side-viewing duodenoscope, with or without fluoroscopic guidance. It features a handle, double or triple lumen shafts, and a precurved tip with a cutting wire. A diathermic current is applied through the wire to perform the incision. The device is operated by a physician (biliary endoscopist) in a clinical setting. The output is a controlled incision of the sphincter, facilitating therapeutic biliary or pancreatic procedures. The ergonomic handle is constructed from ABS plastic, and the device is compatible with 0.035-inch guidewires.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including bow orientation, joint integrity, radiopacity, electrical safety, and biocompatibility testing.

## Technological Characteristics

Device consists of a handle, double or triple lumen shafts, and a cutting wire. Handle material is ABS. Tips are precurved, beveled, or tapered (8mm to 20mm lengths). Compatible with 0.035-inch guidewires. Operates via diathermic current. No software or electronic processing components.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Wiltek Papillotome ([K894861](/device/K894861.md))
- Accuratome3 ([K894861](/device/K894861.md))
- Accuratome ([K894861](/device/K894861.md))
- Double Lumen Papillotome ([K894861](/device/K894861.md))

## Reference Devices

- Wilson Cook papillotome

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K92551

Bard Interventional Products Division C.R. Bard, Inc. 129 Concord Road P.O. Box 7031 Billerica, MA 01821-7031 978-663-8989

SEP 1 0 1998

Image /page/0/Picture/2 description: The image shows the word "BARID" in a bold, sans-serif font. The letters are all capitalized and have a slightly geometric design. The image is in black and white, with the letters appearing in black against a white background.

# VI 510(k) SUMMARY FOR THE BARD® ProForma™ and Apollo3™ PAPILLOTOMES

As required by the Safe Medical Devices Act of 1990, codified under Section 513, Part (I)(3)(A) of the Food Drug and Cosmetic Act, a summary of the safety and effectiveness information upon which substantial equivalence determination is based follows.

## A. General Information

-

- Name and address of submitter: -Bard Interventional Products Division, C.R. Bard. Inc. 129 Concord Road, Building 3 Billerica, MA 01821
Contact: Marion Gordon R.A.C. Senior Regulatory Affairs Coordinator Telephone (978) 262-4867 Fax (978) 262-4878

- Date of Summary: -July 21, 1998

| Name of Device:      |                                                 |
|----------------------|-------------------------------------------------|
| Trade Name:          | Bard® ProForma™ and Apollo3™ Papillotomes       |
| Common/Usual Name:   | Papillotome                                     |
| Classification Name: | Endoscopic electrosurgical unit and accessories |

{1}------------------------------------------------

- Predicate Device: Wiltek Papillotome Wiltek Medical, Inc.
- .. Description and Intended Use of Device:

The Bard® ProForma™ and Apollo3™ Papillotomes are designed and recommended for transendoscopic cannulation and sphincterotomy of the Papilla of Vater and/or Sphincter of Oddi. A papillotome is placed under direct vision of a side-viewing duodenoscope, with or without fluoroscopic aid. Both the double lumen and triple lumen designs will have a variety of precurved tip configurations and cutting wire lengths. The Bard papillotomes are compatible with a 0.035 inch O.D. guidewire.

## B. Summary of Similarities and Differences

The Bard® ProForma™ and Apollo3™ Papillotomes are substantially equivalent to the following legally marketed papillotomes manufactured by Wiltek Medical, Inc. (K894861) and commercialized under the trade names of:

- Accuratome3 *, Accuratome" , and Double Lumen * Papillotome .
#### Similarities

The Bard" ProForma™ and Apollo3™ Papillotomes have equivalent intended uses as the Wiltek papillotomes. The general design characteristics and functionality are also similar. They all have a one (1) piece system construction which includes a handle, either double or triple lumen shafts, various cutting wire lengths and construction with precurved tips in multiple lengths.

Both the Wiltek and Bard devices may be advanced with or without using an .035 inch guidewire through the biopsy channel of a duodenoscope with optional use of fluoroscopy. A diathermic current is applied through the cutting wire to incise the Papilla of Vater or Sphincter of Oddi. A Touhy Borst may be attached to the guidewire lumen of the double lumen device to assist in placement and minimize

{2}------------------------------------------------

contrast leakage.

## Differences

The major design difference between the Bard and Wiltek papillotomes is the ProForma™ and Apollo³™ will have an ergonomic handle made from ABS instead of the polycarbonate handle used in the Wiltek design. The precurved tips on the Bard papillotomes will be either beveled or tapered, where the Wiltek tips are only beveled. A full range of tip lengths, from 8mm to 20mm, are provided by both manufacturers.

Additionally, Wiltek does not include the term 'cannulation' in their intended use. However, this is a common practice among billary endoscopists and a term found in the labeling accompanying other competitive papillotomes, such as the Wilson Cook papillotome supplied in Appendix C.

Although there may be minor differences in the manufacture and packaging of the Bard papillotomes, they are considered substantially equivalent to the Wiltek papillotomes. Any difference between the products raises no issue of safety or effectiveness based upon bow orientation, joint integrity, radiopacity, electrical and biocompatibility testing.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## SEP 1 0 1998

Mr. Marion Gordon Regulatory Affairs Coordinator C. R. Bard, Inc. 129 Concord Road, Bldg. 3 P.O. Box 7031 Billerica, MA 01821-7031

Re: K982557

Bard® ProForma™ and Apollo³™ Pre-Curved Papillotomes Dated: July 21, 1998 Received: July 22, 1998 Regulatory Class: II 21 CFR 876.4300/Procode: 78 KNS

Dear Mr. Gordon:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (2) CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmadsmam.html".

Sincerely yours

Lillian Yin, Ph.D.

Director, Division of Reproducti Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

. 510(k) Number (if known): _________________________________________________________________________________________________________________________________________________ Bard® ProForma™ and Apollo3™ Papillotomes_____________________________________________________________________________________________________________________________________ Device Name: Indications For Use:

Bard papillotomes are designed and recommended for transendoscopic cannulation and sphincterotomy of the Papilla of Vater and/or Sphincter of Oddi.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Vand C. Szyrm

---

Division Sign Off

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological De 510(k) Number

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K982557](https://fda-staging.innolitics.com/device/K982557)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).