BROWSER (EVENT SOFTWARE) MODEL # 4500E

K982550 · Respironics, Inc. · BZQ · Oct 5, 1998 · Anesthesiology

Device Facts

Record IDK982550
Device NameBROWSER (EVENT SOFTWARE) MODEL # 4500E
ApplicantRespironics, Inc.
Product CodeBZQ · Anesthesiology
Decision DateOct 5, 1998
DecisionST
Submission TypeTraditional
Regulation21 CFR 868.2375
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

Browser is a software program that is used by a HCP (Healthcare Professional) to view, analyze, and edit patient data gathered with a SmartMonitor® Infant Apnea Monitor. Information that is collected includes patient information, equipment parameters, compliance logs, patient events, equipment events, and waveforms. The data is downloaded to a PC, either by direct connection to the monitor or by modem. The data can also be retrieved from a computer file or memory disk such as a floppy or card. The software allows the user to review and manipulate event information, produce reports, save changes, and forward data via e-mail.

Device Story

Browser is a Windows 95-based software application for reviewing physiological data recorded by the SmartMonitor Infant Apnea Monitor. Data inputs include patient information, equipment parameters, compliance logs, event logs (apnea, bradycardia, tachycardia, oxygen desaturation), and waveforms, transferred via direct cable, modem, or storage media (floppy/card). The software enables healthcare professionals to visualize, analyze, and edit these events and waveforms, generate reports, and transmit data via email. It serves as a clinical tool for reviewing recorded patient data to assist in monitoring infant respiratory status. The device does not perform real-time monitoring but processes stored data for clinical review.

Clinical Evidence

Clinical comparison performed by comparing reports generated by the predicate 9500E software and the Browser software using actual patient data. No further clinical trials reported.

Technological Characteristics

Software application running on Windows 95 platform. Interfaces with SmartMonitor via direct cable, modem, or storage media. Provides data visualization, report generation, and data editing capabilities.

Indications for Use

Indicated for use by healthcare professionals to view, analyze, and edit patient data (including apnea, bradycardia, tachycardia, and oxygen desaturation events and associated waveforms) collected by the SmartMonitor Infant Apnea Monitor.

Regulatory Classification

Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 5 1898 OCT 1 Image /page/0/Picture/1 description: The image shows the logo for Respironics. The logo consists of a stylized human figure inside a triangle, with the word "RESPIRONICS" written below it in a bold, sans-serif font. The human figure appears to be in motion, and the triangle is a simple, geometric shape. The logo is black and white. 1255 Kennestone Circle Marietta, Georgia 30066-6029 USA ার প্রায় ## 510(k) SUMMARY K982550 Establishment Registration Number: 1040777 Submitter: Betsy Cortelloni Respironics, Inc. 1255 Kennestone Circle Marietta, GA 30066 Phone: 770-429-2894 Fax: 770-499-1139 | Date Prepared: | July 20, 1998 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name of Contact: | Betsy Cortelloni, Regulatory Affairs Associate | | Device Name: | Browser | | Device Model Number: | 4500E | | Classification Name: | Breathing Frequency Monitor, 21 CFR 868.2375 | | Device Classification: | Class II | | Predicate Devices: | Healthdyne Technologies, Inc. 9500E software.<br>Originally cleared under K892006 (5/23/89), and a<br>subsequent modification K903287 (8/17/90). | | Device Description: | Browser is a Windows 95 software program used for<br>reviewing events recorded by a Healthdyne<br>Technologies SmartMonitor. Browser allows the<br>operator to observe the time and duration of<br>physiological events such as apnea, bradycardia,<br>tachycardia and oxygen desaturation. The operator<br>can also observe the waveforms associated with<br>these events. These events and waveforms are<br>downloaded from the monitor using a directly<br>connected cable or via a modem. | {1}------------------------------------------------ 510(k) Summary. continued: Browser is a software program which is used by Intended Use: Healthcare Professionals to view, analyze and edit patient data gathered with a SmartMonitor® Infant Apnea Monitor. Comparison of Technological Characteristics: The predecessor to Browser is the Healthdyne Technologies 9500SE software. Since both software packages interface to the SmartMonitor, they both have the same essential user features. Both allow the user to observe physiological events and equipment events and both allow the user to view waveforms associated with events. Browser runs on the Windows 95 platform whereas the 9500E ran on the DOS platform. However, Browser is not a port of the 9500E software to Windows. Browser is a completely new program that shares no source code with the 9500E. The Browser Software Test Plan consisted of system level and module/integration level testing to verify all the defined software requirements. The tests utilized actual patient data as well as simulated patient data. A clinical comparison of actual patient data was performed by comparing reports generated by the predicate 9500E software and the Browser software. The cumulative test results demonstrated the functionality and safety and effectiveness of the Browser software, as well as its substantial equivalence to the predicate device. Non-Clinical Testing: Conclusion: {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three stripes representing the feathers. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 5 1898 Ms. Betsy Cortelloni Respironics, Inc. 1255 Kennestone Circle Marietta, GA 30066 Re: K982550 Browser (Event Software) Model 4500E Regulatory Class: II (two) Product Code: 73 BZQ Dated: July 21, 1998 Received: July 22, 1998 Dear Ms. Cortelloni: : 11 0 We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {3}------------------------------------------------ ## Page 2 - Ms. Betsy Cortelloni This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ enter for Devices and Radiological Health Page Lof | 510(k) Number (if known): | K982550 | |---------------------------|---------| | Device Name: | BROWSER | Indications for Use: Browser is a software program that is used by a HCP (Healthcare Professional) to view, analyze, and edit patient data gathered with a SmartMonitor® Infant Apnea Monitor. Information that is collected includes patient information, equipment parameters, compliance logs, patient events, equipment events, and waveforms. The data is downloaded to a PC, either by direct connection to the monitor or by modem. The data can also be retrieved from a computer file or memory disk such as a floppy or card. The software allows the user to review and manipulate event information, produce reports, save changes, and forward data via e-mail. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark Kramer (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number _ Prescription Use (Per 21 CFR 801.109) OR II II Over-The-Counter Use (Optional Formal 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...