NAFCILLIN ANTIMICROBIAL SUSCEPTIBILITY TEST DISC

K982528 · Oxoid , Ltd. · JTN · Sep 9, 1998 · Microbiology

Device Facts

Record IDK982528
Device NameNAFCILLIN ANTIMICROBIAL SUSCEPTIBILITY TEST DISC
ApplicantOxoid , Ltd.
Product CodeJTN · Microbiology
Decision DateSep 9, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1620
Device ClassClass 2

Indications for Use

Antimicrobial Susceptibility Test Discs are for the semi-quantitative susceptibility testing by agar diffusion test procedure of rapidly growing micro-organisms. Nafcillin Susceptibility Test Discos are only for use with Staphylococcus spp.

Device Story

Device consists of paper discs impregnated with Nafcillin; used in clinical microbiology laboratories by trained personnel. Discs are placed on agar plates inoculated with Staphylococcus spp. cultures. As antibiotic diffuses into agar, it inhibits bacterial growth, creating zone of inhibition. Healthcare providers measure zone diameter to determine susceptibility of organism to Nafcillin. Results assist clinicians in selecting appropriate antibiotic therapy for patients with staphylococcal infections.

Clinical Evidence

No clinical data provided; device relies on established agar diffusion susceptibility testing principles.

Technological Characteristics

Paper discs impregnated with Nafcillin; intended for agar diffusion susceptibility testing. Class II device, product code JTN.

Indications for Use

Indicated for semi-quantitative susceptibility testing of rapidly growing microorganisms, specifically Staphylococcus spp., using agar diffusion test procedures.

Regulatory Classification

Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure with three faces in profile, one behind the other. The figure is positioned to the right of a circular seal that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. ## SEb 0 1008 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Andy Hollingsworth Quality Systems Manager Oxoid. Ltd. Wade Road Basingstoke Hants RG24 8PW England Re: K982528 > Trade Name: Nafcillin Antimicrobial Susceptibility Test Disc Regulatory Class: II Product Code: JTN Dated: July 13, 1998 Received: July 20, 1998 Dear Mr. Hollingsworth: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1}------------------------------------------------ Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours. Steven Autman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## OXOID LTD 510(K) Number : Unknown Oxoid Nafcillin Susceptibility Test Disc Device Name : Indication for Use : Antimicrobial Susceptibility Test Discs are for the semi-quantitative susceptibility testing by agar diffusion test procedure of rapidly growing micro-organisms. Nafcillin Susceptibility Test Discos are only for use with Staphylococcus spp. Woody Dubois 510(k) Number PRESCRIPTION USE X
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...