K982508 · Boston Scientific Corp · FGE · Oct 16, 1998 · Gastroenterology, Urology
Device Facts
Record ID
K982508
Device Name
FLEXIMA
Applicant
Boston Scientific Corp
Product Code
FGE · Gastroenterology, Urology
Decision Date
Oct 16, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Flexima™ Nasobiliary Catheter is indicated for prolonged access to the biliary ductal system via the nose, including: - Decompression of an obstructed bile duct in acute suppurative cholangitis. - Temporary or short-term decompression of the common bile duct, similar to that which follows unsuccessful stone extraction after endoscopic sphincterotomy. - Prevention of stone impaction after endoscopic sphincterotomy. - Infusion of contrast medium for repeat cholangiography. - Instillation of various therapeutic solutions, including antibiotics (for acute bacterial cholangitis) or saline (to flush sludge or small stones after endoscopic sphincterotomy). - Preoperative biliary decompression to decrease jaundice (thereby decreasing perioperative complications) in patients undergoing elective biliary tract surgery. - Temporary biliary decompression in patients who are septic or who have severe coagulopathy. Once infection has been controlled or coagulopathy corrected, sphincterotomy can be performed, and a large stent can be inserted for long-term therapy. - Access for intraluminal irradiation therapy, using iridium. - Aspiration of bile for chemical and bacteriologic studies (e.g., to identify causative agents in bacterial cholangitis or to determine the lithogenicity of bile in patients with cholestasis). - Facilitating the healing process in traumatic or surgical biliary fistulae.
Device Story
Flexima™ Nasobiliary Catheter provides prolonged access to biliary ductal system via nasal route. Used for biliary decompression, therapeutic solution instillation (antibiotics/saline), contrast infusion, bile aspiration, and intraluminal irradiation. Operated by clinicians during endoscopic procedures. Device facilitates drainage and diagnostic sampling; aids in managing obstructive cholangitis, post-sphincterotomy stone prevention, and preoperative jaundice reduction. Benefits include improved patient outcomes through controlled biliary access and reduced perioperative complications.
Clinical Evidence
Bench testing only. Biocompatibility assessment performed on patient- and fluid-contact materials with satisfactory results. No clinical data presented.
Technological Characteristics
Nasobiliary catheter; patient-contact materials subjected to biocompatibility testing. Packaging consists of Tyvek® and extrusion-coated polyethylene on polyester. Sterilized via ethylene oxide gas per AAMI protocol. Periodic pyrogenicity testing for bacterial endotoxin levels.
Indications for Use
Indicated for patients requiring prolonged biliary ductal access via the nose for decompression, therapeutic infusion, diagnostic aspiration, or fistula management. Contraindicated in patients with coagulopathy or other contraindications to esophagogastroduodenoscopy.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Microvasive® Biliary Stent System
Medi-Tech® Flexima™ Biliary Drainage Catheter
Wilson-Cook® Nasal Biliary Drainage Set
Submission Summary (Full Text)
{0}------------------------------------------------
K982508
Page 1 of 2
SECTION 9 510(K) SUMMARY
# FOI RELEASABLE
Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Boston Scientific Corporation is required to submit with this Premarket Notification "...adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Boston Scientific Corporation chooses to submit a summary of information respecting safety and effectiveness.
- > Common/Usual Names: Nasobiliary Catheter; Nasobiliary Tube
- > TRADE/PROPRIETARY NAME: Flexima™
| | > CLASSIFICATION NAME & | | |
|--|-------------------------|--|--|
|--|-------------------------|--|--|
| DEVICE CLASSIFICATION: | Class II | | |
|------------------------|----------------------------------------------------------------------------------------------------------------|--------|-------------|
| | Name | Number | 21 CFR Ref. |
| | Biliary Catheter and Accessories | 78 FGE | 876.5010 |
| ➤ DEVICE PANEL/BRANCH: | Gastroenterology-Urology (GU)<br>Gastro-Renal (GRDB) | | |
| ➤ OWNER/OPERATOR: | Boston Scientific Corporation<br>One Boston Scientific Place<br>Natick, MA 01760<br>Owner/Operator No. 9912058 | | |
| ➤ CONTACT PERSON: | Daniel J. Dillon, Regulatory Affairs Project Manager | | |
#### INDICATIONS FOR USE
The Flexima™ Nasobiliary Catheter is indicated for prolonged access to the biliary ductal system via the nose, including:
- > Decompression of an obstructed bile duct in acute suppurative cholangitis.
- > Temporary or short-term decompression of the common bile duct, similar to that which follows unsuccessful stone extraction after endoscopic sphincterotomy.
- > Prevention of stone impaction after endoscopic sphincterotomy.
- > Infusion of contrast medium for repeat cholangiography.
- > Instillation of various therapeutic solutions, including antibiotics (for acute bacterial cholangitis) or saline (to flush sludge or small stones after endoscopic sphincterotomy).
- > Preoperative biliary decompression to decrease jaundice (thereby decreasing perioperative complications) in patients undergoing elective biliary tract surgery.
{1}------------------------------------------------
K982508
Page 2 of 2
- > Temporary biliary decompression in patients who are septic or who have severe coagulopathy. Once infection has been controlled or coagulopathy corrected, sphincterotomy can be performed, and a large stent can be inserted for long-term therapy.
- > Access for intraluminal irradiation therapy, using iridium.
- > Aspiration of bile for chemical and bacteriologic studies (e.g., to identify causative agents in bacterial cholangitis or to determine the lithogenicity of bile in patients with cholestasis).
- > Facilitating the healing process in traumatic or surgical biliary fistulae.
### Contraindications
Patients with coagulopathy or any other contraindication to esophagogastroduodenoscopy are contraindicated for the Flexima™ Nasobiliary Catheter.
### POTENTIAL COMPLICATIONS
Possible complications include, but may not be limited to: dislodgment of the catheter, nose and throat irritation, and excess bile loss due to external drainage.
### SUBSTANTIAL EQUIVALENCE
Boston Scientific Corporation believes that the Flexima™ Nasobiliary Catheter is substantially equivalent to the currently-marketed Microvasive® Biliary Stent System, Medi-Tech® Flexima™ Biliary Drainage Catheter and the Wilson-Cook® Nasal Biliary Drainage Set.
# Performance Characteristics
Laboratory testing regarding characteristics was performed on Flexima™ Nasobiliary Catheter to verify its safety and performance. A biocompatibility assessment was performed on the patient- and fluid-contact materials of the Flexima™ Nasobiliary Catheter with satisfactory results.
# Packaging, Sterilization, And Pyrogenicity
The Flexima™ Nasobiliary Catheter will be packaged in Tyvek® and extrusion coated-polyethylene on polyester. The Flexima™ Nasobiliary Catheter will be sterilized using ethylene oxide gas using the AAMI protocol for ethylene oxide sterilization. Pyrogenicity testing will be performed on a periodic basis to monitor bacterial endotoxin levels.
# CONCLUSION
Boston Scientific Corporation believes that Flexima™ Nasobiliary Catheter is substantially equivalent to the currently-marketed Microvasive® Biliary Stent System, Medi-Tech® Flexima™ Biliary Drainage Catheter and Wilson-Cook® Nasal Biliary Drainage Set. Boston Scientific Corporation has presented laboratory testing and biocompatibility information. The information presented provides assurance that the Flexima™ Nasobiliary Catheter will meet the minimum requirements that are considered acceptable for its intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human profile. The symbol is composed of three curved lines that suggest movement or flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 16 1998
Mr. Daniel J. Dillon Regulatory Affairs Project Manager Boston Scientific Corporation One Boston Scientific Place Natick, MA 01760-1537
Re: K982508 Flexima™ Nasobiliary Catheter Regulatory Class: II Product Code: 78 FGE 21 CFR 876.5010 Dated: September 3, 1998 Received: September 4, 1998
Dear Mr. Dillon:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2 - Mr. Daniel J. Dillon
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Vivid G. de Yaram
Lillian Yin. Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number: To Be Determined
Device Name:
Flexima™ Nasobiliary Catheter
Indication for Use:
The Flexima™ Nasobiliary Catheter is indicated for prolonged access to the biliary ductal system via the nose, including:
- > Decompression of an obstructed bile duct in acute suppurative cholangitis.
- > Temporary or short-term decompression of the common bile duct, similar to that which follows unsuccessful stone extraction after endoscopic sphincterotomy.
- > Prevention of stone impaction after endoscopic sphincterotomy.
- > Infusion of contrast medium for repeat cholangiography.
- > Instillation of various therapeutic solutions, including antibiotics (for acute bacterial cholangitis) or saline (to flush sludge or small stones after endoscopic sphincterotomy).
- > Preoperative biliary decompression to decrease jaundice (thereby decreasing perioperative complications) in patients undergoing elective biliary tract surgery.
- > Temporary biliary decompression in patients who are septic or who have severe coagulopathy. Once infection has been controlled or coagulopathy corrected, sphincterotomy can be performed, and a large stent can be inserted for long-term therapy.
- > Access for intraluminal irradiation therapy, using iridium.
- > Aspiration of bile for chemical and bacteriologic studies (e.g., to identify causative agents in bacterial cholangitis or to determine the lithogenicity of bile in patients with cholestasis).
- > Facilitating the healing process in traumatic or surgical biliary fistulae.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Stypnon
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K982508
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.