K982472 · Genicon, LC · GEA · Feb 4, 1999 · General, Plastic Surgery
Device Facts
Record ID
K982472
Device Name
GENICON TROCAR
Applicant
Genicon, LC
Product Code
GEA · General, Plastic Surgery
Decision Date
Feb 4, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4800
Device Class
Class 1
Indications for Use
The intended use of this device is for laparoscopic surgery by general, thoracic, gynecologic and urological surgeons.
Device Story
Trocar/cannula system for laparoscopic surgery; used by general, thoracic, gynecologic, and urological surgeons to establish port of entry for endoscopic instruments. Device features modified pyramidal or conical tip for tissue incision; enlarged handle for grip; clear cannula casing for visualization; reverse trapezoidal fascia threads for anchoring; extended CO2 port; double-wall valve lock; tapered membrane valve assembly to reduce friction and improve seal. Operates as manual surgical instrument; provides access for instrumentation; facilitates specimen retrieval via valve pull-tab. Benefits include easier insertion, improved instrument manipulation, and enhanced valve performance compared to predicate devices.
Clinical Evidence
Bench testing only. Engineering performance testing conducted in a simulated human abdominal cavity compared the device to Apple and Core trocar valve assemblies. Metrics included pressure testing, ease of use, and insertion performance.
Technological Characteristics
Manual surgical instrument. Materials include stainless steel (tip) and silicone (valve membrane). Features: modified pyramidal/conical tip, reverse trapezoidal fascia threads, double-wall valve lock, tapered membrane valve, clear cannula casing. No energy source or software.
Indications for Use
Indicated for use in gynecologic, general, thoracic, and urology endoscopic procedures to establish a port of entry for instrumentation. Available in 5, 7/8, 10, and 12mm diameters with pyramidal or conical tips.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
Predicate Devices
Apple Medical Hunt/Reich Trocar
Core Dynamics Entree II
Submission Summary (Full Text)
{0}------------------------------------------------
K982472
## Section II. 510(k) Summary
GeniCon L.C. Contact: Gary Haberland 573 Waterscape Wav Orlando, FL 32828 Phone: (407) 273-7619 Fax: (407) 306-9356
Date Prepared: October 20, 1998
Trade Name: GeniCon Trocar/Cannula System Common Name: Trocar Classification Name: Trocar (per CFR 870.1390)
The intended use of this device is for laparoscopic surgery by general, thoracic, gynecologic and urological surgeons.
The legally marketed device to which we are claiming previous acceptance of predicate device is the Apple Medical Hunt/Reich Trocar and the Core Dynamics Entree II both in sizes 5, 7/8, 10 and 12mm systems.
Engineering testing has demonstrated that the valve assembly performs better than that of the Apple and Core trocar valve assembly. This testing incorporated a simulated human abdominal cavity and the method of insertion of the device and use with various common laparoscopic instruments. In all areas of performance, this device performed equal to if not superior in pressure testing, ease of use and insertion.
This device incorporates several design features which facilitate the surgeons use both during insertion and while instrumentation is being inserted and manipulated.
Trocar Tip:
- A modified pyramidal tip to provide easier insertion for the surgeon by enabling the tip to . minimally incise the tissue while providing a lower profile transition from the stainless tip to the cannula.
Trocar Handle:
- Enlarged to provide a more positive fit in the surgeon's palm during insertion, and provides a . groove to provide a more positive fit for the thumb during removal or re-insertion.
## Cannula:
- Clear casing for visualization of instrument and tissue passage. .
- Reverse trapezoidal fascia threads, they provide an easier insertion while a positive anchoring . during instrument use.
- . Extended CO2 port with male luer cap provides a "finger grip" during device insertion while the male luer cap is easier to remove and replace than the Apple silicone cap.
- . Double wall valve lock, assists in maintaining the valve position and to alleviate the premature release sometimes experienced by the Apple product.
Valve:
- . A tapered membrane assembly provides a more positive seal around the instrument, while the silicon membrane was designed to reduce friction during instrument use.
- . A pull-tab was added to facilitate removal of the valve during specimen retrieval.
Image /page/0/Picture/22 description: The image shows the logo for Genicon. The logo is in black and white and features the company name in a stylized font. Above the "o" in Genicon are two surgical instruments. Below the company name is the tagline "New Ideas for Modern Health Care".
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and features the department's name around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that suggest movement and progress.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 1999 EEB
Mr. Gary Haberland Product Manager Genicon 573 Waterscape Way Orlando, Florida 32828
Re: K982472
> Trade Name: Genicon Trocar/Cannula System Regulatory Class: II Product Code: GEA Dated: January 19, 1999 Received: January 21, 1999
Dear Mr. Haberland:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 -- Mr. Gary Haberland
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely vours.
Styph Rhodes
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Section I. Indications for Use
510 (k) Number: K982472
Device Name: Trocar
Indications for Use:
The GeniCon trocar is available in 5, 7/8, 10 and 12mm diameter with either a pyramidal or conical tip. This trocar has application in gynecologic, general, thoracic and urology endoscopic procedures to establish a port of entry for instrumentation.
## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Prescription Use_ X OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Styt Durbe
(Division Sign-Off) (Division of General Restorative Devices K9P2472 510(k) Number -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.