K982444 · Biosound Esaote, Inc. · IYO · Oct 6, 1998 · Radiology
Device Facts
Record ID
K982444
Device Name
7250 ULTRASOUND IMAGING SYSTEM
Applicant
Biosound Esaote, Inc.
Product Code
IYO · Radiology
Decision Date
Oct 6, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ESAOTE's Mod. 7250 is a compact ultrasound system used to perform non-invasive diagnostic general ultrasound studies.
Device Story
Mod. 7250 is a compact ultrasound system for non-invasive diagnostic imaging; supports B-Mode, M-Mode, Doppler, Color Flow Mapping, and Amplitude Doppler. System drives annular, phased, convex, and linear array probes. Features volumetric scan converter for simultaneous control of two independent scan planes (Bi-Scan); enables volumetric datasets for omni-directional 2D imaging and volume computation via Simoson rule. Operated by clinicians in clinical settings; provides on-screen data entry and annotations. Includes 3.5" floppy drive for software/upgrades, optional optical disk, and LAN port for data storage. Output displayed on LCD or 15" monitor; supports recording via VCR or printer. Facilitates rapid examination and efficient workflow through mode selection keys and pop-up menus. Benefits include non-invasive diagnostic capability and volumetric analysis of irregular structures without geometric assumptions.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by compliance with EN60601-1 electrical safety standards and Track 3 acoustic output display requirements.
Technological Characteristics
Compact ultrasound system; supports annular, phased, convex, and linear array probes. Imaging frequencies: 2.5-10 MHz; Doppler frequencies: 2.0-6.6 MHz. Features volumetric scan converter for Bi-Scan probes. Connectivity: LAN port, 3.5" floppy drive, optional optical disk. Display: SVGA/RGB. Software-based calculations and measurements. Sterilization: Not specified (probes are patient-contacting).
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, abdominal, pediatric, small organ (thyroid, breast, testicles), neonatal cephalic, adult cephalic, cardiac, transesophageal, transrectal, transvaginal, and peripheral vascular applications.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K982444
892.1550
892.1560
#### OCT 6 1998
Safety and Effectiveness Summary 7250 Ultrasound Imaging System Biosound Esaote
# Safety and Effectiveness Summary
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR 807.92(a).
807.92(a)(1)
# Submitter Information
Colleen Hittle, Official Correspondent 8000 Castleway Drive Indianapolis, IN 46250 (317) 849-1916 Phone: (317) 577-9070 Facsimile:
Colleen Hittle Contact Person:
Date: July 2, 1998
807.92(a)(2)
7250 Ultrasound Imaging System Trade Name: Ultrasound Imaging System Common Name: Classification Name(s): Ultrasonic pulsed doppler imaging system Ultrasonic pulsed echo imaging system
> 90IYN 90IYO
7050 (AU3)
Classification Number:
807.92(a)(3)
Predicate Device(s)
Esaote
K944287
Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table.
{1}------------------------------------------------
Safety and Effectiveness Summary 7250 Ultrasound Imaging System Biosound Esaote
807.92(a)(4)
# Device Description
ESAOTE's Mod. 7250 is a compact ultrasound system used to perform non-invasive diagnostic general ultrasound studies. Its primary modes of operation are the following: B-Mode, M-Mode, Doppler, Color Flow Mapping and Amplitude Doppler, The MOD.7250 can be equipped with an LCD Color Display (Portable Configuration) or with a 15" Color Monitor and a cart (Mainframe Configuration). The full alphanumeric keyboard allows complete on-screen data entry of patient information and on-screen annotations.
The MOD. 7250 can drive annular array, phased, convex and linear array probes. In addition. Mod.7250 is equipped with a volumetric scan converter and the ability to control simultaneously two independent scan planes. This technological characteristic allows to manage Bi-Scan Probes, to obtain volumetric datasets; these datasets can then be used to display "omni-directional" 2D images (anyplane sectioning) as well as multiple parallel equidistant sections (paraplane sectioning). This Bi-Scan approach is ideal for volume computations through the Simoson rule: specifically, it can be used to measure volumes of irregular structures since it does not reguire a geometrical assumption.
The MOD.7250 is designed for ease of use. The user interface allows the operator to perform an examination quickly and efficiently. Clearly labeled mode selection keys are easily accessed, and the system's "pop-up" menus allow the operator to change parameters with ease. The user may also access special function menus and perform calculations with a minimal number of key strokes. The MOD.7250 offers a vast selection of calculations and measurements which can be performed quickly and easily. The MOD.7250 is equipped with a 3.5" floppy disk drive to simplify software modifications and provides fast, cost effective system upgrades. This drive (or an optional Optical Disk Drive) can also be used for image storage. Moreover, this unit can store data directly to a Personal Computer via a LAN (Network) port. In addition, MOD.7250 can be equipped with recording devices, including a S-VHS video recorder and a black-and-white or color page printer, which are controlled through
#### 807.92(a)(5)
the keyboard.
## Intended Use(s)
ESAOTE's Mod. 7250 is a compact ultrasound system used to perform non-invasive diagnostic general ultrasound studies.
{2}------------------------------------------------
K982444 P3 of 3
Safety and Effectiveness Summary 7250 Ultrasound Imaging System Biosound Esaote
| | Comparison Chart for Substantial Equivalence | |
|---------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|
| | 7250 | 7050 |
| Electrical Safety | EN60601-1 | EN60601-1 |
| Ultrasound Safety | Track 3 (Acoustic Output Display) | Track 3 (Acoustic Output Display) |
| Intended Use | | |
| • Cardiac (Transthoracic) | YES | YES |
| • Cardiac (Transesophageal) | YES (Multiplane) | YES (Monoplane) |
| • Vascular | YES | YES |
| • Abdominal | YES | YES |
| • Fetal | YES | YES |
| • Adult Cephalic | YES | YES |
| • Neonatal Cephalic/Small organ | YES | YES |
| • Endovaginal | YES (Sagittal & Transverse Planes) | YES (Sagittal & Transverse Planes) |
| • Endorectal | YES (Sagittal & Transverse Planes | YES (Sagittal & Transverse Planes |
| Probe Technology | | |
| • Annual Array | YES | YES |
| • Electronical Array | YES | YES |
| • Bi-Scan Probes | YES | NO |
| • Doppler Probes | YES | YES |
| Modes of operation | 2D, M-Mode, PW, CW, CFM, Amplitude Doppler | 2D, M-Mode, PW, CW, CFM, Amplitude Doppler |
| Imaging Frequencies | 2.5, 3.5, 5.0, 7.5, 10 MHz | 2.5, 3.5, 5.0, 7.5, 10 MHz |
| CFM/Doppler Frequencies | 2.0, 2.5, 3.3, 5.0, 6.6 MHz | 2.0, 2.5, 3.3, 5.0, 6.6 MHz |
| Biopsy Guidance | YES | YES |
| • ABS11 & ABS13 use | General Purpose | General Purpose |
| • ABS12 use | Transrectal/transvaginal | Transrectal/transvaginal |
| • Biopsy Line Depth marker | 1 cm | 1 cm |
| • Needle guide angle | ABS11: 25-45°<br>ABS12: fixed<br>ABS13: 30-50° | ABS11: 25-45°<br>ABS12: fixed<br>ABS13: 30-50° |
| Display Type | SVGA | RGB |
| Digital Archival Capabilities | YES | YES |
| VCR/Page Printer | YES | YES |
| M&A Capabilities | Cardiac, Vascular, Fetal and general purpose measurements | Cardiac, Vascular, Fetal and general purpose measurements |
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three wing-like shapes. The bird is positioned in the center of a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
#### OCT ୧ 1098
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Colleen Hittle Official Correspondent Biosound Esaote 8000 Castleway Drive Indianapolis, IN 46250
Re: K982444 7250 Ultrasound Imaging System Regulatory Class: II Product Code: 90 IYO/21 CFR 892.1560 Dated: July 2, 1998 Received: July 14, 1998
Dear Ms. Hittle:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Mod. 7250, as described in your premarket notification:
Transducer Model Number
PA020 CA11 7510 AA-C LA13 BLA023 5 MHz TEE TRT12
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2 - Colleen Hittle
Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 Therefore, your product labeling must be consistent with FDA's CFR, Part 812. position on this use.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your The FDA finding of substantial equivalence of your ... . premarket notification. device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 ... CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97) . . Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
{5}------------------------------------------------
Page 3 - Colleen Hittle
If you have any questions regarding the content of this letter, please contact Paul Gammell, Ph.D. at (301) 594-1212.
Sincerely yours,
David A. Szymon
for Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{6}------------------------------------------------
#### Mod.7250
# Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | | |
|-----------------------------|---|-----------------------------------------------------------------------------------------------------------------------|---|-----|-----|-------------------------|----------------------------|------------------------------|-----------------------|-----------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler<br>CFM | Amplitude<br>Doppler<br>PD | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | | See<br>comments | | |
| Abdominal | | N | N | N | | N | N | | See<br>comments | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | | See<br>comments | | |
| Small Organ (specify) [1] | | N | N | N | N | N | N | | See<br>comments | | |
| Neonatal Cephalic | | N | N | N | N | N | N | | See<br>comments | | |
| Adult Cephalic | | N | N | N | N | N | N | | See<br>comments | | |
| Cardiac | | N | N | N | N | N | N | | See<br>comments | | |
| Transesophageal | | N | N | N | N | N | N | | See<br>comments | | |
| Transrectal | | N | N | N | | N | N | | See<br>comments | | |
| Transvaqinal | | N | N | N | | N | N | | See<br>comments | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | N | N | N | | See<br>comments | N[2] | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | |
| Conventional | | | | | | | | | | | |
| Musculoskeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
| | | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:<br>Mode of Operation | | | | | | | | | |
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler<br>CFM | Amplitude<br>Doppler<br>PD | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | N | N | N | N | N | N | | See<br>comments | | |
| Cardiac | | N | N | N | N | N | | | See<br>comments | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | |
| Conventional | | | | | | | | | | | |
| Musculoskeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E= added under Appendix E
Additional Comments: [1] Small organs include Thyroid, Breast and Testicles. [2] This unit manages Bi-scan
probes for volumetric acquisitions from which any 2D tomographic image can be displayed.
Applicable combined modes: B+M+PW+CW+CFM+PD
Daniel A. Segarra
(Division Sign-Off)
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF sion of Reproductive, Abdominal, ENT, concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
510(k) Number
11
{7}------------------------------------------------
#### Transducer: PA020
### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body or follows:
N=new indication; P=previously cleared by FDA; E= added under Appendix E
Additional Comments: / Applicable combined modes: B+M+PW+CW+CFM+PD
(PLEASE DO NOT WRITE BELOW THIS JINE. CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K982444
{8}------------------------------------------------
K982444
#### Transducer: CA11
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler<br>CFM | Amplitude<br>Doppler<br>PD | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) |
|-----------------------------|---|---|---|-----|-----|-------------------------|----------------------------|------------------------------|-----------------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | P | P | P | P | | P | P | See<br>comments | | |
| Abdominal | P | P | P | P | | P | P | See<br>comments | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculoskeletal Superficial | | | | | | | | | | |
| Other (specify)<br>: | | | | | | | | | | |
### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N=new indication; P=previously cleared by FDA; E= added under Appendix E
Additional Comments: This transducer has been previously cleared by FDA with the AU3 unit (K944287) for uses indicated as "P". Applicable combined modes: B+M+PW+CFM+PD
(PLEASE DO NOT WRITE BELOW THIS LINE, CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Sijim
---
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT. and Radiological Devices
510(k) Number K482444
{9}------------------------------------------------
# K982494
# Diagnostic Ultrasound Indications for Use Form
### Transducer: 7510 AA-C
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler<br>CFM | Amplitude<br>Doppler<br>PD | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) |
|-----------------------------|---|---|---|-----|-----|-------------------------|----------------------------|------------------------------|-----------------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | N | N | N | N | N | N | | | See<br>comments | |
| Small Organ (specify) [1] | P | P | P | P | P | P | | | See<br>comments | |
| Neonatal Cephalic | P | P | P | P | P | P | | | See<br>comments | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | N | N | N | N | N | N | | | See<br>comments | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | | | See<br>comments | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculoskeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E= added under Appendix E
Additional Comments: [1] Small organs include Thyroid, Breast and Testicles. This transducer has been previously cleared by FDA with the AU3 unit (K944287). Applicable combined modes: B+M+PW+CFM
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Office of Device Evaluation (ODE
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K982444.
Warr
{10}------------------------------------------------
### Transducer: LA13
| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | |
|-----------------------------|--------------------------------------------------------------------------------------------------|---|---|-----|-----|-------------------------|----------------------------|------------------------------|-----------------------|-----------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler<br>CFM | Amplitude<br>Doppler<br>PD | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) [1] | N | N | N | N | | N | N | | See<br>comments | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | P | | See<br>comments | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | |
| Conventional | | | | | | | | | | | |
| Musculoskeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as for
N=new indication; P=previously cleared by FDA; E= added under Appendix E
Additional Comments: [1] Small organs include Thyroid, Breast and Testicles. This transducer has been previously
cleared by FDA with the AU3 unit (K944287). Applicable combined modes: B+M+PW+CFM+PD
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Lopson
(Division Sign-Off) (Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
{11}------------------------------------------------
#### Transducer: BLA023
# Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | | |
|-----------------------------|---|-------------------|---|-----|-----|-------------------------|----------------------------|------------------------------|-----------------------|-----------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler<br>CFM | Amplitude<br>Doppler<br>PD | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | | See<br>comments | N [1] | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | |
| Conventional | | | | | | | | | | | |
| Musculoskeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E= added under Appendix E
Additional Comments: [1]: This probe can be used to acquire volumetric datasets from which any 2D tomographic
plane can be then visualized. Applicable combined modes: B+M+PW+CFM+PD
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Office of Device Evaluation (ODE)
David C. Severson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
{12}------------------------------------------------
#### Transducer: 5 MHz TEE
| Clinical Application | | Mode of Operation | | | | | | | | | |
|-----------------------------|---|-------------------|---|-----|-----|-------------------------|----------------------------|------------------------------|-----------------------|-----------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler<br>CFM | Amplitude<br>Doppler<br>PD | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | N | N | N | N | N | | | See<br>comments | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | |
| Conventional | | | | | | | | | | | |
| Musculoskeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
# : 1 ต
N=new indication; P=previously cleared by FDA; E= added under Appendix E
Additional Comments: / Applicable combined modes: B+M+PW+CW+CFM
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Legerson
(Division Sign-Off) (Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number
{13}------------------------------------------------
#### Transducer: TRT12
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler<br>CFM | Amplitude<br>Doppler<br>PD | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) |
|-----------------------------|---|---|---|-----|-----|-------------------------|----------------------------|------------------------------|-----------------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) [1] | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | P | | P | P | | See<br>comments | |
| Transvaginal | | P | P | P | | P | P | | See<br>comments | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculoskeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E= added under Appendix E
Additional Comments: This transducer has been previously cleared by FDA with the AU3 unit (K953716).
Applicable combined modes: B+M+PW+CFM+PD
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Leggett
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K982444
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.