← Product Code [DXC](/productcode/DXC) · K982365

# ATRAUMATIC VASCULAR CLAMPS, MODEL #'S 22-XXXX, 23-XXXX, 03-60XX, 25-XXXX MICRO VASCULAR CLAMPS MODEL # 75-0XXX, ATRAUMAT (K982365)

_Geister Medizin Technik GmbH · DXC · Apr 8, 1999 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K982365

## Device Facts

- **Applicant:** Geister Medizin Technik GmbH
- **Product Code:** [DXC](/productcode/DXC.md)
- **Decision Date:** Apr 8, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

To temporarily occlude an artery or vein with correct tension to produce minimal trauma to vessels. The different sizes and models are designed according to the type and size of the blood vessels, anatomical sites, and surgical techniques. Bulldog clamps are self-closing and are used for small vessels. The closing pressure of the GLOVER clamps can be adjusted using a nut and spring. Applying forceps are used to apply bulldog clamps in deep surgical wounds. To secure vascular occlusion of small vessels during surgery and to secure vascular occlusion during pediatric surgery.

## Device Story

Manual surgical instruments for vascular occlusion; includes atraumatic clamps, bulldog clamps, micro vascular clamps, and applying forceps. Bulldog clamps are self-closing; GLOVER clamps feature adjustable closing pressure via nut and spring. Micro vascular clamps available as single/double, with/without frames or approximators. Applying forceps facilitate placement of bulldog clamps in deep surgical wounds. Used by surgeons in clinical/OR settings to temporarily occlude blood vessels during procedures. Benefit: enables controlled vascular occlusion with minimal trauma to vessel walls.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Manual surgical instruments; stainless steel construction; various jaw configurations (straight, curved, angled); lengths 10-335 mm; jaw widths 0.8-10 mm; jaw lengths 5-200 mm. Bulldog clamps are self-closing; GLOVER clamps utilize mechanical nut/spring adjustment. Micro vascular clamps accommodate vessel sizes 0.3-5.0 mm.

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three horizontal lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 8 1999

Ms. Daqmar S. Maeser C/O Geister Medizintechnik GMBH Business Support International Amstel 320-I 1017 AP Amsterdam The Netherlands

Re : K982365

Trade Name: Atraumatic Clamps, Bulldog Clamps, Micro Vascular Clamps and Applying Forceps Requlatory Class: II Product Code: DXC January 27, 1999 Dated: Received: January 29, 1999

Dear Ms. Maeser:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to

{1}------------------------------------------------

Page 2 - Ms. Dagmar S. Maeser

your premarket notification submission does not affect any obliqation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callehan

Thomas J. Callaha Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

4/8/99

K982365

# Geister Medizintechnik GmbH

510(k) Number:

Device Name:

Vascular Clamp

Classification Name:

Vascular Clamp

Product Code:

scular Clamp

21 CFR 870.4450 74 DXC Class II

# INDICATIONS FOR USE:

## Atraumatic Clamps and Bulldog Clamps

### Intended Use:

To temporarily occlude an artery or vein with correct tension to produce minimal trauma to vessels. The different sizes and models are designed according to the type and size of the blood vessels, anatomical sites, and surgical techniques. Bulldog clamps are self-closing and are used for small vessels. The closing pressure of the GLOVER clamps can be adjusted using a nut and spring.

- Various jaw types (straight, curved, or angled) Varieties:

#### Dimensional Ranges:

| Lengths:     | 10.0 - 335 mm |
|--------------|---------------|
| Jaw Widths:  | 0.8 - 10 mm   |
| Jaw Lengths: | 5.0 - 200 mm  |

#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

Qitor M. H. for TTC

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 982365 510/k) Numb

1/3

{3}------------------------------------------------

# Geister Medizintechnik GmbH

510(k) Number:

Device Name:

Classification Name:

Vascular Clamp Vascular Clamp

Product Code:

21 CFR 870.4450 74 DXC Class II

# INDICATIONS FOR USE (continued):

Applying Forceps

## Intended Use:

Applying forceps are used to apply bulldog clamps in deep surgical wounds.

- Varieties: Various mouth pieces
- Length: 190 – 240 mm

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

II

Prescription Use _ (Per 21 CFR 801.109) OR V

Over-The-Counter Use

(Optional Format 1-2-96)

Chitoph. M. th for TJC
(Division Sign-Off)

Division of Cardiovascular, Respiratory, and Neurological Devices ﺳﻠ

{4}------------------------------------------------

# Geister Medizintechnik GmbH

510(k) Number:

| Device Name:         | Vascular Clamp                  |
|----------------------|---------------------------------|
| Classification Name: | Vascular Clamp                  |
| Product Code:        | 74 DXC Class II 21 CFR 870.4450 |

#### INDICATIONS FOR USE: (continued):

## Micro Vascular Clamps

#### Intended Use

To secure vascular occlusion of small vessels during surgery and to secure vascular occlusion during pediatric surgery.

Varieties: Single clamps or double clamps

- With or without frame
- With or without approximator (movable)
- Different types for veins or arteries
- Two models (Acland and Acland mod.)
- For various vessel sizes
- Plain or black (Acland mod.)

#### Dimensional Ranges:

For Vessel Sizes: 0.3 - 5.0 mm Widths (without frames): 1.0 - 3.0 mm 7.0 - 36.0 mm Lengths:

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

\checkmark

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

Chitylm Al for TJC

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices

982365

---

**Source:** [https://fda-staging.innolitics.com/device/K982365](https://fda-staging.innolitics.com/device/K982365)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
