K982352 · Columbia Bioscience, Inc. · LSE · Nov 25, 1998 · Microbiology
Device Facts
Record ID
K982352
Device Name
VCA IGM ELISA TEST SYSTEM
Applicant
Columbia Bioscience, Inc.
Product Code
LSE · Microbiology
Decision Date
Nov 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3235
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K982352 · Nov 25, 1998
VCA IGM ELISA TEST SYSTEM
Columbia Bioscience, Inc.
Retrospective patient serum samples
Retrospective clinical sera were used to evaluate the relative sensitivity and specificity of the EBV-VCA IgM ELISA Kit compared to existing commercial serological assays.
176 patients (102 convalescent, 32 seronegative, 42 current infection); Sample Size: 176; Number of Sites: 1
Commercially available kits for VCA IgM, VCA IgG, EBNA IgG and heterophile antibodies
Relative sensitivity and specificity
Indications for Use
For the qualitative determination of IgM antibodies in human serum to Epstein Barr (recombinant) Viral Capsid antigen (EBV-VCA) antigen. The EBV-VCA IgM assay should be used in conjunction with other Epstein-Barr serologies (EBV-VCA IgG, EBNA-1 IgG, EA-D IgG, EA-D IgM, EBNA-1 IgM and heterophile) as an aid in the diagnosis of infectious monomucleosis. The test can be performed either manually or in conjunction with the MAGO PLUS ™ Automated EIA Processor.
Device Story
EBV-VCA IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for detecting IgM antibodies to Epstein-Barr Viral Capsid antigen in human serum. Recombinant EBV-VCA antigen is immobilized on microtiter wells; patient serum is added; if specific IgM antibodies are present, they bind to the antigen. After washing, enzyme-labeled anti-human immunoglobulin conjugate is added, followed by a substrate solution. Color intensity is measured photometrically, providing an indirect detection of specific antibodies. The test is performed manually or using the MAGO PLUS Automated EIA Processor in a clinical laboratory setting. Results assist clinicians in diagnosing infectious mononucleosis when used alongside other EBV serological markers. The device provides qualitative results to aid clinical decision-making.
Clinical Evidence
Clinical evaluation used 176 retrospective patient sera characterized by commercial kits. Results: Relative sensitivity for current infection was 95.0% (38/40; 95% CI 83.1-99.4); relative specificity for convalescent sera was 97.0% (98/101; 95% CI 91.6-99.4); relative specificity for seronegative sera was 100% (32/32; 95% CI 89.1-100). Overall agreement was 97.1%. Precision studies (intra-assay, inter-assay, inter-site) were conducted across three sites. Cross-reactivity testing showed some reactivity with VZV, CMV, and HSV IgM positive sera.
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) using recombinant EBV-VCA antigen immobilized on microtiter wells. Detection via enzyme-labeled anti-human immunoglobulin conjugate and photometric measurement of substrate color intensity. Compatible with manual or automated (MAGO PLUS) processing. No specific material standards or software algorithm class provided.
Indications for Use
Indicated for qualitative detection of IgM antibodies to Epstein-Barr Viral Capsid antigen in human serum as an aid in diagnosing infectious mononucleosis. Used in conjunction with other EBV serologies.
Regulatory Classification
Identification
Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).
Reference Devices
Is-EBV-VCA IgM Test Kit
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 2 5 1998
16982352
## 510k Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
#### Applicant Information:
| Date Prepared: | June 11, 1998 |
|----------------|------------------------------------------------------|
| Name: | Columbia Bioscience, Inc. |
| Address: | 8775 M Centre Park Drive, #559<br>Columbia, MD 21045 |
| Contact Person: | Norman Jenkins |
|-----------------|----------------|
| PhoneNumber: | 410-995-0450 |
| Fax Number: | 410-995-0448 |
#### Device Information:
| Trade Name: | EBV-VCA IgM ELISA Kit |
|----------------------|-----------------------|
| Common Name. | EBV-VCA IgM EIA Test |
| Classification Name; | Epstein-Barr Virus |
Equivalent Device: EBV Serology
Device Description: The & EBV-VCA IgM ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgM antibodies to Epstein-Barr Viral Capsid antigen in human serum.
Intended Use: For the qualitative determination of IgM antibodies in human serum to Epstein Barr (recombinant) Viral Capsid antigen (EBV-VCA) antigen. The EBV-VCA IgM assay should be used in conjunction with other Epstein-Barr serologies (EBV-VCA IgG, EBNA-1 IgG, EA-D IgG, EA-D IgM, EBNA-1 IgM and heterophile) as an aid in the diagnosis of infectious monomucleosis. The test can be performed either manually or in conjunction with the MAGO PLUS ™ Automated EIA Processor.
Principle of Procedure:
The & EBV-VCA IgM ELISA Kit is an enzyme-linked immunosorbent assay to detect IgM to EBV-VCA in human serum. Recombinant EBV-VCA antigen is attached to a solid phase (microtiter well). Diluted test sera are added to each well. If antibodies which recognize the EBV-VCA antigen are present in the patient sample they will bind to the antigen in the well. After incubation, the wells are washed to remove unbound antibody. An enzyme labeled anti-human immunoglobulin (conjugate) is added to each test well. If antibody is present the enzyme-linked antibody will bind to it. After incubation, the wells are washed to remove unbound conjugate. A substrate solution is then added to each well. If enzyme is present from prior step, the reaction is stopped and the color intensity is measured photometrically producing an indirect detection of the specific antibody present in the patient sample.
{1}------------------------------------------------
# Performance Characteristics
## A. Clinical Sensitivity and Specificity Using Characterized Sera
Frozen retrospective sera from one hundred and seventy-six patients were characterized using commercially available kits for VCA IgM, VCA IgG, EBNA IgG and heterophile antibodies. Based on the results of this testing, the patient sera were characterized as follows:
- * 102 sera were characterized as convalescent (past infection). These were positive for VCA IgG and/or EBNA IgG antibodies and negative for VCA IgM and heterophile antibody.
- * 32 scra were characterized as scronegative. These were negative for VCA IgG, VCA IgM, EBNA IgG and beterophile antibody.
- * 42 sera were characterized as having a current (recent) infection. These were positive for VCA IgM and/or heterophile antibody and were negative for EBNA IgG.
All 176 sera were then tested by an independent clinical commercial laboratory using the Is-EBV-VCA IgM Test Kit. The results obtained are shown in Table 2:
| EBV Serological Status | | | | |
|--------------------------------------------------------|--------------|-------------------|---------------------|----|
| | Convalescent | Current Infection | Seronegative | |
| POSITIVE | 3 | 38 | 0 | |
| Is-EBV-VCA IgM | NEGATIVE | 98 | 2 | 32 |
| *EQUIVOCAL | 1 | 2 | 0 | |
| Relative Specificity (Convalescent) 98/101 = 97.0% | | | 95% CI<br>91.6-99.4 | |
| Relative Sensitivity (Current Infection) 38/40 = 95.0% | | | 83.1-99.4 | |
| Relative Specificity (Seronegative) 32/32 = 100% | | | 89.1-100 | |
| Overall Agreement 168/173 = 97.1% | | | 93.4-99.1 | |
* Equivocal results were excluded from calculations
NOTE : Please be advised that 'relative' refers to the comparison of the assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison's accuracy to predict disease. Since the above studies were performed on a pre-selected, retrospective, population, no calculations for the assay's positive predictive value may be done or inforred.
{2}------------------------------------------------
### B. Precision
To determine the precision of the Is-EBV-VCA IgM Test Kit, four positive and two negative sera were assayed ten times each in three different runs at three different sites. The 3 sites include: the manufacturer, a research and develooment laboratory, and a clinical commercial laboratory. The intra- and interassay precision obtained at each site is shown in Tables 3, 4 and 5. The Inter-Site precision is shown in Table 6.
TABLE 3 : Site #1 - Intra-Assay and Interassay Precision
| SERUM | MEAN<br>INDEX | INTRA-ASSAY RUN 1<br>CV% | MEAN<br>INDEX | INTRA-ASSAY RUN 2<br>CV% | MEAN<br>INDEX | INTRA-ASSAY RUN 3<br>CV% | MEAN<br>INDEX | INTERASSAY<br>CV% | |
|---------|---------------|--------------------------|---------------|--------------------------|---------------|--------------------------|---------------|-------------------|-------|
| A (POS) | 1.41 | 8.08 | 1.77 | 6.07 | 1.67 | 8.45 | 1.62 | 12.05 | |
| B (POS) | 2.17 | 7.97 | 2.78 | 7.01 | 2.35 | 6.26 | 2.44 | 12.67 | |
| C (POS) | 4.01 | 9.92 | 5.05 | 7.15 | 4.22 | 6.78 | 4.43 | 12.86 | |
| D (POS) | 2.70 | 8.91 | 3.32 | 13.74 | 2.81 | 4.35 | 2.94 | 13.69 | |
| E (NEG) | 0.35 | 17.21 | 0.54 | 8.51 | 0.45 | 21.03 | 0.44 | 22.75 | |
| F (NEG) | 0.09 | 49.37 | 0.17 | 26.31 | 0.18 | 14.93 | 0.14 | 39.31 | |
| | | | | | | | CAL | 1.03 | 10.81 |
| | | | | | | | PC | 1.59 | 25.45 |
| | | | | | | | NC | 0.49 | 18.70 |
TABLE 4 : Site #2- Intra-Assay and Interassay Precision
| SERUM | INTRA-ASSAY RUN 1 | | INTRA-ASSAY RUN 2 | | INTRA-ASSAY RUN 3 | | INTERASSAY | |
|---------|-------------------|-------|-------------------|-------|-------------------|-------|---------------|-------|
| | MEAN<br>INDEX | CV% | MEAN<br>INDEX | CV% | MEAN<br>INDEX | CV% | MEAN<br>INDEX | CV% |
| A (POS) | 1.47 | 3.96 | 1.34 | 4.14 | 1.31 | 4.13 | 1.37 | 6.53 |
| B (POS) | 2.25 | 6.35 | 1.99 | 4.31 | 2.06 | 4.90 | 2.10 | 7.34 |
| C (POS) | 3.92 | 3.95 | 3.45 | 2.43 | 3.50 | 4.35 | 3.62 | 6.96 |
| D (POS) | 2.57 | 6.26 | 2.26 | 5.36 | 2.41 | 3.06 | 2.41 | 7.31 |
| E (NEG) | 0.35 | 7.41 | 0.32 | 8.33 | 0.33 | 11.03 | 0.33 | 9.38 |
| F (NEG) | 0.14 | 17.42 | 0.12 | 10.23 | 0.12 | 9.31 | 0.12 | 14.80 |
| | | | | | | CAL | 1.00 | 3.79 |
| | | | | | | PC | 1.43 | 3.91 |
| | | | | | | NC | 0.19 | 22.90 |
| | | | TABLE 5 : Site #3 - Intra-assay and Interassay Precision |
|--|--|--|----------------------------------------------------------|
|--|--|--|----------------------------------------------------------|
| SERUM | INTRA-ASSAY RUN 1 | | INTRA-ASSAY RUN 2 | | INTRA-ASSAY RUN 3 | | INTERASSAY | | |
|---------|-------------------|-------|-------------------|------|-------------------|-------|------------|-------|-------|
| | MEAN INDEX | CV% | MEAN INDEX | CV% | MEAN INDEX | CV% | MEAN INDEX | CV% | |
| A (POS) | 1.48 | 4.94 | 1.43 | 4.97 | 1.48 | 5.15 | 1.46 | 5.08 | |
| B (POS) | 2.19 | 6.22 | 2.20 | 6.44 | 2.13 | 4.55 | 2.17 | 5.80 | |
| C (POS) | 3.78 | 5.99 | 3.58 | 4.80 | 3.70 | 5.25 | 3.68 | 5.69 | |
| D (POS) | 2.64 | 5.67 | 2.52 | 5.75 | 2.56 | 4.15 | 2.57 | 5.40 | |
| E (NEG) | 0.45 | 12.82 | 0.38 | 9.77 | 0.40 | 10.27 | 0.41 | 13.26 | |
| F (NEG) | 0.20 | 9.43 | 0.16 | 7.16 | 0.18 | 11.82 | 0.18 | 12.70 | |
| | | | | | | CAL | 1.00 | 3.22 | n=9 |
| | | | | | | PC | 1.21 | 3.74 | n=3 |
| | | | | | | NC | 0.62 | 1.85 | n = 3 |
| | | | | | | TABLE 6 : Inter-Site Precision |
|--|--|--|--|--|--|--------------------------------|
|--|--|--|--|--|--|--------------------------------|
| SERUM<br>(n=90) | INTER-SITE | |
|-----------------|---------------|-------|
| | MEAN<br>INDEX | CV% |
| A (POS) | 1.48 | 11.13 |
| B (POS) | 2.24 | 11.41 |
| C (POS) | 3.91 | 13.43 |
| D (POS) | 2.64 | 13.09 |
| E (NEG) | 0.40 | 20.69 |
| F (NEG) | 0.15 | 28.77 |
| CAL (n=36) | 1.01 | 6.22 |
| LPC (n=18) | 1.42 | 13.16 |
| NC (n=18) | 0.31 | 60.11 |
{3}------------------------------------------------
### C. Specificity with Potentially Cross-Reactive Sera
Thirteen scra, reactive for IgM antibodies to varicella zoster, cylomegalovirus and herpes simplex virus by EIA were cested in the Is-EBV-VCA IgM Test Kit. 3/4 anti-VZV IgM positive sera were non-reactive for anti-VCA IgM; 3/5 anti-CMV IgM positive scra were non-reactive for anti-VCA IgM and 4/4 anti-HSV positive scra were non-reactive for anti-VCA IgM. This suggests that some specific cross-reactivity should be expected with the Is-VCA IgM Test Kit from these analytes.
#### D. Correlation of Manual and MAGO Plus Results
The Is-EBV-VCA IgM Test Kit has been developed for automated as well as manual use. To demonstrate the equivalence of the manual and MAGO Plus procedures, the results of 120 serum samples tested by both methods were plotted. A scattergram and regression line of the results obtained with 95% confidence intervals is shown in Figure 3. The data indicate good correlation with a Pearson Correlation Coefficient of 0.990.
Image /page/3/Figure/4 description: The image is a scatter plot comparing MAGO PLUS INDEX VALUES to MANUAL INDEX VALUES. The x-axis represents MANUAL INDEX VALUES, ranging from 0 to 7, while the y-axis represents MAGO PLUS INDEX VALUES, ranging from -0.5 to 5.5. The plot includes a regression line with the equation Y = 0.0045 + 0.7074X, and the correlation coefficient r is 0.990, indicating a strong positive correlation between the two variables.
FIGURE 3 : Manual and MAGO Plus Result Correlation
#### D. MAGO Plus Precision
The precision of the assay when performed on the MAGO Plus Automated EIA Processor was determined by assaying six sera ten times each in three different runs. Table 7 shows the intra-and interassay precision obtained using the MAGO Plus.
| SERUM | INTRA-ASSAY RUN 1 | | INTRA-ASSAY RUN 2 | | INTRA-ASSAY RUN 3 | | INTERASSAY | | |
|---------|-------------------|-------|-------------------|-------|-------------------|-------|------------|-------|-------|
| | MEAN | CV% | MEAN | CV% | MEAN | CV% | MEAN | CV% | |
| | INDEX | | INDEX | | INDEX | | INDEX | | |
| A (POS) | 1.15 | 4.58 | 1.28 | 7.18 | 1.16 | 9.27 | 1.20 | 8.64 | |
| B (POS) | 1.74 | 5.55 | 1.80 | 6.42 | 1.69 | 7.08 | 1.74 | 6.68 | |
| C (POS) | 3.19 | 4.30 | 3.34 | 4.72 | 3.22 | 3.53 | 3.25 | 4.55 | |
| D (POS) | 1.96 | 4.93 | 2.07 | 6.85 | 2.06 | 5.22 | 2.03 | 6.09 | |
| E (NEG) | 0.28 | 22.59 | 0.31 | 10.20 | 0.29 | 19.57 | 0.29 | 17.76 | |
| F (NEG) | 0.10 | 0.00 | 0.10 | 0.00 | 0.10 | 0.00 | 0.10 | 0.00 | |
| | | | | | | CAL | 1.00 | 4.54 | n=S |
| | | | | | | PC | 1.26 | 3.20 | n = 3 |
| | | | | | | NC | 0.39 | 9.08 | n = 3 |
TABLE 7 : Site #2- Intra-Assay and Interassay Precision - MAGO Plus
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 2 5 1998
Norman Jenkins President Columbia Bioscience, Inc. 8775 M Centre Park Drive, #559 Columbia, Maryland 21045
Re: K982352 Device: VCA IgM ELISA Test System Dated: September 22, 1998 Received: September 28, 1998
Dear Mr. Jenkins:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number: Not Known
Device Name: 7 VCA IgM ELISA
Indications For Use: For the qualitative determination of IgM antibodies in human serum to Epstein Barr (recombinant) Viral Capsid antigen (VCA) antigen. The 78 VCA IgM assay should be used in conjunction with other Epstein-Barr serologies (VCA IgG, EBNA-1 IgG, EA-D IgM, EA-D IgG, EBNA-1 IgM and heterophile) as an aid in the diagnosis of infectious mononucleosis. The test can be performed either manually or in conjunction with the MAGO PLUS TM Automated EIA Processor
#### PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) 】 【 【 】 】
___________________________________________________________________________________________________________________________________________________________________
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109) OR
Over-The Counter Use (Optional Format 1-2-96)
Woody Dubois
Division of Clinical Laboratory Devices 510(k) Number_1 982352
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.