← Product Code [DQO](/productcode/DQO) · K982299

# SPYGLASS 5 FR ANGIOGRAPHIC CATHETER (K982299)

_Daig Corp. · DQO · Oct 19, 1998 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K982299

## Device Facts

- **Applicant:** Daig Corp.
- **Product Code:** [DQO](/productcode/DQO.md)
- **Decision Date:** Oct 19, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1200
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The 5Fr Spyglass™ Angiographic Catheter is designed for delivery of radiopaque contrast media during angiographic procedures.

## Device Story

The Spyglass™ 5Fr Angiographic Catheter is a radiopaque, polymer-based catheter reinforced with braided stainless steel. It is designed for use by clinicians during angiographic procedures to deliver radiopaque contrast media into the vascular system. The device features a soft, atraumatic tip to minimize vessel damage and preformed tip shapes to facilitate access to specific anatomical sites. Its construction allows for a small diameter while maintaining structural strength and high flow rate capabilities. The device is used in clinical settings by physicians to assist in diagnostic imaging, enabling the visualization of blood vessels through the administration of contrast agents.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Radiopaque polymer catheter reinforced with braided stainless steel; soft atraumatic tip; preformed tip shapes; 5Fr diameter; intended for contrast media delivery; sterilization method not specified.

## Regulatory Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

## Predicate Devices

- Spyglass™ Angiographic Catheter ([K944284](/device/K944284.md))
- Spyglass™ Angiographic Catheter ([K962805](/device/K962805.md))
- Spyglass™ Angiographic Catheter ([K965249](/device/K965249.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 1 9 1998

510(k) Summary _______________________________________________________________________________________________________________________________________________________________

B.

## 510(k) Summary (As required by 21 CFR 807.92)

#### A. Submitter Information

| Submitter's Name     | Daig Corporation, a St. Jude Medical Company                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address              | 14901 DeVeau Place<br>Minnetonka, Minnesota 55345-2126 U.S.A.                                                                                                                                                                                                                                                                                                                                                               |
| Telephone Number     | (612) 933-4700                                                                                                                                                                                                                                                                                                                                                                                                              |
| Contact Person       | Michael G. Schultz                                                                                                                                                                                                                                                                                                                                                                                                          |
| Submission Prepared  | June 30, 1998                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Information   |                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Common or Usual Name | SpyglassTM 5Fr Angiographic Catheter                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Name  | Angiographic Catheter                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate Device     | SpyglassTM Angiographic Catheter Family<br>(K944284, K962805, and K965249)                                                                                                                                                                                                                                                                                                                                                  |
| Device Description   | Daig SpyglassTM angiographic catheters are radiopaque<br>polymer reinforced with braided stainless steel. This<br>construction yields a small diameter catheter, which retains<br>strength and high flow rate capabilities. A soft atraumatic<br>tip provides good memory and protects against vessel<br>damage. Preformed tip shapes matching or complementing<br>specific anatomy allow access into a variety of vessels. |
| Intended Use         | The Daig SpyglassTM angiographic catheters are designed<br>for delivery of radiopaque contrast media during<br>angiographic procedures.                                                                                                                                                                                                                                                                                     |

#### C. Comparison of Required Technological Characteristics

All technological characteristics of the Spyglass ™ 5Fr Angiographic Catheter are identical to the predicate device (K962805) including product design, packaging, sterilization, and labeling.

{1}------------------------------------------------

510(k) Summary________________________________________________________________________________________________________________________________________________________________

### Support of the Substantial Equivalence D.

Daig Corporation considers the Spyglass™ 5Fr Angiographic Catheter to be substantially equivalent to the following predicate devices: Spyglass™ Angiographic Catheter (formerly called Control™ Angiographic Catheter) received marketing clearance on March 3, 1995 (K944284) and Spyglass™ Angiographic Catheter received marketing clearance on October 28, 1996 (K962805).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 19 1998

Mr. Michael G. Schultz Quality Systems Manager DAIG Corporation 14901 Deveau Pl. Minnetonka, MN 5345-2126

Re: K982299 5FR Spyglass™ Angiographic Catheter Trade Name: Regulatory Class: II Product Code: DOO Dated: September 25, 1998 Received: September 28, 1998

Dear Mr. Schultz:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act

{3}------------------------------------------------

Page 2 - Mr. Michael G. Schultz

for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

|  |  |  | more was<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| 510(k) Number (if known): | K98 2299 |
|---------------------------|----------|
|---------------------------|----------|

5Fr Spyglass™ Angiographic Catheter __________________________________________________________________________________________________________________________________________ Device Name:

Indications for Use:

. •

The 5Fr Spyglass™ Angiographic Catheter is designed for delivery of radiopaque contrast media during angiographic procedures.

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

![Signature](signature.png)
(Division Sign-Off)

Division of Cardiovascular, Respiratory,

and Neurological Devices

| 510(k) Number | K982299 |
|---------------|---------|
|---------------|---------|

| Prescription Use<br>(Per 21 CFR 801.109) | <div> <img alt="Check Mark" src="check_mark.png"/> </div> | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|-----------------------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|-----------------------------------------------------------|----|--------------------------------------------------|

105

---

**Source:** [https://fda-staging.innolitics.com/device/K982299](https://fda-staging.innolitics.com/device/K982299)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com