← Product Code [BYE](/productcode/BYE) · K982283

# WHISPERFLOW OXYGEN FLOW GENERATOR - ADJUSTABLE MODEL WF 8500 WHISPERFLOW OXYGEN FLOW GENERATOR - FIXED MODEL WF 8530 WH (K982283)

_Caradyne, Ltd. · BYE · Mar 9, 1999 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K982283

## Device Facts

- **Applicant:** Caradyne, Ltd.
- **Product Code:** [BYE](/productcode/BYE.md)
- **Decision Date:** Mar 9, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5965
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

To provide CPAP to spontaneously breathing patients in the hospital and pre-hospital (EMS) environment.

## Device Story

Whisperflow Oxygen Flow Generator is a venturi-based device delivering oxygen/air mixtures to spontaneously breathing patients. It connects to a wall oxygen source (60 psi) to generate high flows (up to 140 Lpm) and CPAP pressure. The device utilizes standard in-line PEEP valves to establish circuit pressure and interfaces with the patient via a face mask or ET tube. It is operated by clinicians in hospital or EMS settings. The device allows adjustment of oxygen flow and FiO2 (28-100%). By providing high-flow CPAP, it supports respiratory function and may assist in weaning patients from mechanical ventilators. The system includes accessories such as CPAP masks, tubing, filters, and connectors.

## Clinical Evidence

Bench testing only. Performance specifications compared to predicates include oxygen input pressure (60 psi), output flow range (up to 140 Lpm), and FiO2 ranges.

## Technological Characteristics

Venturi-based oxygen/air mixture delivery. Materials: Stainless steel, PVC, silicone. Dimensions: 22 mm standard fittings. Energy source: Pneumatic (wall oxygen source). Sterilization: EO gas. Connectivity: None (mechanical/pneumatic).

## Regulatory Identification

A positive end expiratory pressure (PEEP) breathing attachment is a device attached to a ventilator that is used to elevate pressure in a patient's lungs above atmospheric pressure at the end of exhalation.

## Predicate Devices

- Vital Signs - Downs Flow Generators ([K811393](/device/K811393.md), K83150)
- Vital Signs CPAP mask ([K801883](/device/K801883.md))
- EMS Mask ([K871851](/device/K871851.md))
- Vital Signs Model 9250 ([K821503](/device/K821503.md))
- Vital Signs Model 9200 ([K811393](/device/K811393.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

3/9/99

Image /page/0/Picture/1 description: The image shows the number "K 982283" on the left side. To the right of the number is a logo of three interlocking circles. Next to the logo is the word "caradyne" in bold, black letters. 

Caradyne Limited Parkmore Business Centre Parkmore West Galway, Ireland.

## Non-Confidential Summary of Safety and Effectiveness

page 1 of 5 June 25, 1998 Tel: +353 (0)91 709010 Fax: +353 (0)91 758929 Email: info@caradyne.com Internet: www.caradyne.com

| Caradyne, Ltd.             |                                                        |
|----------------------------|--------------------------------------------------------|
| Parkmore Business Park     |                                                        |
| Parkmore West              |                                                        |
| Galway, Ireland            |                                                        |
|                            | Tel - 011-353-91-709010                                |
|                            | Fax - 011-353-91-758929                                |
| Official Contact:          | John O'Dea, Ph.D., General Manager                     |
| Proprietary or Trade Name: | Whisperflow Oxygen Flow Generators and accessories     |
| Common/Usual Name:         | CPAP generator                                         |
| Classification Name:       | Breathing Attachment, Positive End Expiratory Pressure |
| Device:                    | Whisperflow Oxygen Flow Generators and accessories     |
| Predicate Devices:         | Vital Signs - Downs Flow Generators - K811393, K83150  |
|                            | Vital Signs CPAP mask - K801883                        |
|                            | EMS Mask - K871851                                     |

Device Description: 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -

The Whisperflow Oxygen Flow Generator is a venturi type oxygen / air mixture delivery device which provides CPAP pressure at high flows to a spontaneously breathing patient. It can deliver up to 140 Lpm flow at a FiO2 between 28- 100%. It utilizes standard in-line PEEP valves to set the prescribed pressure and interfaces with the patient via a face mask or ET tube.

| Indicated Use --      | To provide CPAP to a spontaneously breathing adult patient. |
|-----------------------|-------------------------------------------------------------|
| Environment of Use -- | Hospital and pre-hospital.                                  |
| Patient population -- | Spontaneously breathing adult patients.                     |

Comparison to Predicate Devices:

| Attribute                                                                                        | Variable / Adjustable<br>Vital Signs |                                        |                              |
|--------------------------------------------------------------------------------------------------|--------------------------------------|----------------------------------------|------------------------------|
|                                                                                                  | Whisperflow<br>WF 8500               | Model 9250<br>K821503<br>Vital Flow100 |                              |
| Use                                                                                              |                                      |                                        |                              |
| Intended for delivering of CPAP                                                                  | Yes                                  | Yes                                    |                              |
| Attribute                                                                                        | Variable / Adjustable                |                                        |                              |
|                                                                                                  | Whisperflow<br>WF 8500               | Model 9250<br>K821503                  | Vital Signs<br>Vital Flow100 |
| Used with PEEP valves, CPAP mask<br>and circuit tubing                                           | Yes                                  | Yes                                    | Yes                          |
| Environment Hospital /<br>pre-hospital (EMS)                                                     | Yes                                  | Yes                                    | Yes                          |
| <b>Design</b>                                                                                    |                                      |                                        |                              |
| Works by a venturi method to create a<br>vacuum to provide high flows                            | Yes                                  | Yes                                    | Yes                          |
| Has oxygen inlet fitting which attached<br>to wall oxygen source standard CGA or<br>DISS fitting | Yes                                  | Yes                                    | Yes                          |
| Has an On / Off valve                                                                            | Yes                                  | Yes                                    | Yes                          |
| Can adjust oxygen flow                                                                           | Yes                                  | Yes                                    | Yes                          |
| Has an air entrainment port with 22 mm ID inlet                                                  | Yes                                  | Yes                                    | Yes                          |
| Air entrainment port can have a standard<br>particulate / bacterial filter attached              | Yes                                  | Yes                                    | Yes                          |
| Has a valve which adjusts the flow to<br>the venturi nozzle                                      | Yes                                  | No                                     | Yes                          |
| Outlet port (22 mm)                                                                              | Yes                                  | Yes                                    | Yes                          |
| Recommend an in-line oxygen analyzer                                                             | Yes                                  | Yes                                    | Yes                          |
| Recommend an in-line safety valve PEEP valve                                                     | Yes                                  | Yes                                    | Yes                          |
| Circuit is standard 22 mm tubing                                                                 | Yes                                  | Yes                                    | Yes                          |
| Can have a humidifier placed in-line                                                             | Yes                                  | Yes                                    | Yes                          |
| Connects to patient interface - mask or ET tube                                                  | Yes                                  | Yes                                    | Yes                          |
| Utilizes a standard PEEP valve to<br>establish the circuit pressure                              | Yes                                  | Yes                                    | Yes                          |
| Patient can entrain room air should<br>oxygen flow fail                                          | Yes                                  | Yes                                    | Yes                          |
| Attribute                                                                                        | Variable / Adjustable                | Vital Signs                            |                              |
|                                                                                                  | Whisperflow<br>WF 8500               | Model 9250<br>K821503                  | Vital Flow100                |
| Mask has a one-way valve to prevent<br>rebreathing, if no oxygen flow                            | Yes                                  | Yes                                    | Yes                          |
| Flow generator can be EO gas sterilized                                                          | Yes                                  | Yes                                    | Yes                          |
| Accessories required -                                                                           |                                      |                                        |                              |
| CPAP mask                                                                                        | Yes                                  | Yes                                    | Yes                          |
| Particulate filter at air entrainment port                                                       | Yes                                  | Yes                                    | Yes                          |
| 22 mm tubing                                                                                     | Yes                                  | Yes                                    | Yes                          |
| Head strap for mask                                                                              | Yes                                  | Yes                                    | Yes                          |
| Various connectors                                                                               | Yes                                  | Yes                                    | Yes                          |
| PEEP valves                                                                                      | Yes                                  | Yes                                    | Yes                          |
| Performance Standards / Specifications                                                           |                                      |                                        |                              |
| Oxygen input pressure 60 psi                                                                     | Yes                                  | Yes                                    | Yes                          |
| Output flow range up to 140 Lpm                                                                  | Yes                                  | Yes                                    | Yes                          |
| Fittings 22 mm                                                                                   | Yes                                  | Yes                                    | Yes                          |
| Range of Oxygen (FiO2)                                                                           | 28-100%                              | 33-100%                                | 33-100%                      |

Registered in Ireland: 270891 Registered Office 2 Harbourmaster Place Custom House Dock Dublin 1

{1}------------------------------------------------

## (continued)

page 2 of 5

June 25, 1998

{2}------------------------------------------------

page 3 of 5

## June 25, 1998

| Materials and the comments of the comments of the comments of |     |     |     | and the contraction of the comments of the country of the count |
|---------------------------------------------------------------|-----|-----|-----|-----------------------------------------------------------------|
| Whisperflow generator - Stainless steel, PVC                  | Yes | Yes | Yes |                                                                 |
| CPAP mask - PVC and silicone                                  | Yes | Yes | Yes |                                                                 |

{3}------------------------------------------------

page 4 of 5

## June 25, 1998

| Attribute                                                                                        | Whisperflow<br>WF 8530               | Vital Signs<br>Model 9200<br>K811393 |
|--------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------|
| Use                                                                                              |                                      |                                      |
| Intended for delivering of CPAP                                                                  | Yes                                  | Yes                                  |
| Used with PEEP valves, CPAP mask<br>and circuit tubing                                           | Yes                                  | Yes                                  |
| Environment Hospital /<br>pre-hospital (EMS)                                                     | Yes                                  | Yes                                  |
| Design                                                                                           |                                      |                                      |
| Works by a venturi method to create a<br>vacuum to provide high flows                            | Yes                                  | Yes                                  |
| Has oxygen inlet fitting which attached<br>to wall oxygen source standard CGA or<br>DISS fitting | Yes                                  | Yes                                  |
| Has an air entrainment port with 22 mm ID inlet                                                  | Yes                                  | Yes                                  |
| Air entrainment port can have a standard<br>particulate / bacterial filter attached              | Yes                                  | Yes                                  |
| Outlet port which is 22 mm                                                                       | Yes                                  | Yes                                  |
| Recommend an in-line oxygen analyzer                                                             | Yes                                  | Yes                                  |
| Recommend an in-line safety valve PEEP valve                                                     | Yes                                  | Yes                                  |
| Circuit is standard 22 mm tubing                                                                 | Yes                                  | Yes                                  |
| Can have a humidifier placed in-line                                                             | Yes                                  | Yes                                  |
| Connects to patient interface - mask or ET tube                                                  | Yes                                  | Yes                                  |
| Utilizes a standard PEEP valve to<br>establish the circuit pressure                              | Yes                                  | Yes                                  |
| Patient can entrain room air should<br>oxygen flow fail                                          | Yes                                  | Yes                                  |
| Mask has a one-way valve to prevent<br>rebreathing if no oxygen flow                             | Yes                                  | Yes                                  |
| Attribute                                                                                        | Fixed Flow<br>Whisperflow<br>WF 8530 | Vital Signs<br>Model 9200<br>K811393 |
| Flow generator can be EO gas sterilized                                                          | Yes                                  | Yes                                  |
| Accessories required -                                                                           |                                      |                                      |
| CPAP mask                                                                                        | Yes                                  | Yes                                  |
| Particulate filter for air entrainment port                                                      | Yes                                  | Yes                                  |
| 22 mm tubing                                                                                     | Yes                                  | Yes                                  |
| Head strap for mask                                                                              | Yes                                  | Yes                                  |
| Various connectors                                                                               | Yes                                  | Yes                                  |
| PEEP valves                                                                                      | Yes                                  | Yes                                  |
| Performance Standards / Specifications                                                           |                                      |                                      |
| Oxygen input pressure 60 psi                                                                     | Yes                                  | Yes                                  |
| Output flow range up to 140 Lpm dependent                                                        |                                      |                                      |
| of PEEP pressure                                                                                 | Yes                                  | Yes                                  |
| Fittings 22 mm                                                                                   | Yes                                  | Yes                                  |
| Oxygen (FiO2) %                                                                                  | 28%                                  | 33%                                  |
| Materials                                                                                        |                                      |                                      |
| Whisperflow generator - Stainless steel, PVC                                                     | Yes                                  | Yes                                  |
| CPAP mask - PVC and silicone                                                                     | Yes                                  | Yes                                  |
|                                                                                                  |                                      |                                      |

{4}------------------------------------------------

#### page 5 of 5

#### June 25, 1998

# Differences between Other Legally Marketed Predicate Devices =================================================================================================================

The differences between the Whisperflow generators and CPAP mask and the predicates are:

- 1. CPAP mask is a ribbed cushion design versus an air cushion design.
- 2. The Whisperflow generators have a wider range of FiO2 which may help in weaning patients from the ventilator.

There are no other significant differences between the intended device and the predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure with three faces in profile, connected by flowing lines. The figure is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a sans-serif font.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 9 1999

Mr. Paul E. Dryden Caradyne, Ltd. c/o Promedic, Inc. 6329 W. Waterview Court McCordsville, IN 46055

K982283 Re: Whisperflow Oxygen Flow Generator - Adjustable Model WF 85 Regulatory Class: II (two) Product Code: 73 BYE January 26, 1999 Dated: January 27, 1999 Received:

Dear Mr. Dryden:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP requlation may result in assumptions. In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Paqe 2 - Mr. Paul E. Dryden

This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

#### TONSFORTIST

Page 1 of 1

Pursuant to the Notice of February 6, 1996 regarding listing of Indications for Use on a separate sheet, the following is per that request.

K982283_______________________________________________________________________________________________________________________________________________________________________ (To be assigned) 510(k) Number:

Device Name:

Whisperflow Oxygen Flow Generator system

Intended Use :

To provide CPAP to spontaneously breathing patients in the hospital and pre-hospital (EMS) environment.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Lurk Modoo

Lath Modon 9-2

9-21- 78

**Prescription Use** 
(Per CFR 801.109) 
✓

or

Over-the-counter use

---

**Source:** [https://fda-staging.innolitics.com/device/K982283](https://fda-staging.innolitics.com/device/K982283)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).