DPX-RX BONE DENSITOMETER

K982267 · Lunar Corp. · KGI · Aug 18, 1998 · Radiology

Device Facts

Record IDK982267
Device NameDPX-RX BONE DENSITOMETER
ApplicantLunar Corp.
Product CodeKGI · Radiology
Decision DateAug 18, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1170
Device ClassClass 2

Indications for Use

The DPX-RX Bone Densitometer provides an estimate of BMD at the lumbar Spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician. The use of the DPX-RX Bone Densitometer is restricted to prescription use only. The operator's manual for the DPX-RX system contains the following statement: "United States Federal Law restricts this device to the sale, distribution, and use by or on the order of a physician."

Device Story

DPX-RX Bone Densitometer estimates bone mineral density (BMD) in g/cm2 at lumbar spine and proximal femur. Device performs 30-second scan; utilizes ionizing radiation (2.5 mrem dose). Operated by trained personnel in clinical settings. Output consists of BMD values; physician compares these values against reference populations to aid clinical decision-making regarding bone health. Benefits include non-invasive assessment of bone density.

Clinical Evidence

Bench testing and in vivo clinical comparison. In vivo study of 50 subjects showed BMD estimations correlated with r>0.98 to the DPX-L predicate. Short-term in vivo precision (%CV) was approximately 1.0% for lumbar spine and proximal femur. In vitro precision (%CV) was <0.5%. Radiation exposure is 2.5 mrem per measurement.

Technological Characteristics

Bone densitometer utilizing ionizing radiation for BMD estimation. 30-second scan time. Radiation dose: 2.5 mrem per measurement. Form factor: clinical imaging system. Connectivity/software details not specified.

Indications for Use

Indicated for the estimation of bone mineral density (BMD) at the lumbar spine and proximal femur regions in patients requiring clinical assessment of bone density. Restricted to prescription use by or on the order of a physician.

Regulatory Classification

Identification

A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/3 description: The image contains a handwritten string of alphanumeric characters. The string appears to be "K982267". The characters are written in a cursive style, with varying stroke thicknesses. The image is in black and white. 313 W. BELTLINE HIGHWAY MADISON, WI 53713 AUG 25 1998 (608) 274-2663 ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS 10.0 This summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92(c). | Contact Person: | Kenneth D. Buroker<br>LUNAR Corporation<br>313 West Beltline Highway<br>Madison, WI 53713 | |----------------------|------------------------------------------------------------------------------------------------| | Phone: | (608) 288-6460 | | Fax: | (608) 274-0853 | | Date: | June 18, 1998 | | Device/Trade Name: | DPX-RX Bone Densitometer | | Common Name: | Bone Densitometer | | Classification Name: | Bone Densitometer<br>21CFR 892.1170 | | Predicate Device: | LUNAR DPX-Alpha and DPX-L<br>510(k) Accession Number K904980<br>LUNAR EXPERT<br>510(k) K945526 | #### 10.1 DESCRIPTION OF THE DEVICE: The DPX-RX Bone Densitometer provides an estimation of Bone Mineral Density (BMD in g/cm2) of the lumbar spine and the proximal femur. #### SUMMARY OF TECHNICAL CHARACTERISTICS 10.2 The DPX-RX Bone Densitometer performs a 30-second scan, with a total skin exposure dose of 2.5 mrem per measurement. The radiation exposure of 2.5 mrem is comparable to the predicate DPX-L, and is ten times lower than predicate fanbeam densitometers such as the EXPERT. The BMD estimations in vivo provided by the DPX-RX correlate r>0.98 with the DPX-L. The average BMD values obtained in 50 subjects in vivo were very similar with DPX-L and DPX-RX. The average short-term precision (%CV) in vitro was <0.5%. The short-term %CV in vivo was approximately 1.0% for lumbar spine and proximal femur BMD. These values are comparable to those shown on currently marketed devices. {1}------------------------------------------------ ## CONCLUSION 10.3 The DPX-RX Bone densitometer is substantially equivalent to currently marketed devices. No new safety and offectiveness questions are raised with the DPX-RX Bone Densitometer. ``` Signed Kenneth D. Buroker Name Director, Regulatory Affairs Title Summary of Safety, Effectiveness Page 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 25 1998 Kenneth D. Buroker Lunar Corporation 313 West Beltline Highway Madison, WI 53713 Re: K982267 Trade Name: DPX-RX Bone Densitometer Regulatory Class: II Product Code: 90 KGI Dated: June 26, 1998 Received: June 29, 1998 Dear Mr. Buroker: This letter corrects our substantially equivalent letter of August 18, 1998, regarding the name of your device. We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General (QS) regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product -Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 - Kenneth Buroker If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address, "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, David A. Segerson Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### 3.0 INDICATION FOR USE FORM 501(k) Number (if known) . DPX-RX Bone Densitometer Device name: . Indications For Use: . The DPX-RX Bone Densitometer provides an estimate of BMD at the lumbar Spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician. The use of the DPX-RX Bone Densitometer is restricted to prescription use only. The operator's manual for the DPX-RX system contains the following statement: > "United States Federal Law restricts this device to the sale, distribution, and use by or on the order of a physician." # PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED. Concurrence of CDRH, Office of Device Evaluation (ODE) David la heppner (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K982267 **Prescription Use** (Per 21 CRF 801.109) ್ಕೆ. OR Over-the-Counter Use (Optional Format 1-2-96) Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...