CANOGEN PORTABLE OXYGEN GENERATOR MODEL 615

K982243 · Canogen International , Ltd. · CAW · Nov 19, 1998 · Anesthesiology

Device Facts

Record IDK982243
Device NameCANOGEN PORTABLE OXYGEN GENERATOR MODEL 615
ApplicantCanogen International , Ltd.
Product CodeCAW · Anesthesiology
Decision DateNov 19, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5440
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Canogen Portable Oxygen Generator, Model 615 is intended to produce oxygen for emergency use.

Device Story

Canogen Portable Oxygen Generator, Model 615 is a portable device for emergency oxygen production. It utilizes a chemical reaction within a reusable reaction vessel to generate oxygen. The system includes a chemical cartridge, reaction vessel, plastic tubing, and a gas mask. Operation involves inserting the chemical cartridge into the vessel and adding tap water to initiate the reaction. The device produces humidified oxygen at >99% purity at an average flow rate of 6.5 liters per minute for at least 15 minutes. The generated gas is delivered to the patient via tubing and a face mask. It is intended for emergency use, providing a portable source of oxygen when traditional supplies are unavailable.

Clinical Evidence

Bench testing only. Performance was monitored using flow gauges, temperature sensors, and data loggers. Results demonstrated the device produces >99% pure humidified oxygen at an average flow rate of 6.5 L/min for a minimum of 15 minutes.

Technological Characteristics

Chemical oxygen generator; reaction vessel with replaceable chemical cartridges; produces humidified oxygen via chemical reaction; output flow 6.5 L/min; purity >99%; duration 15 minutes; manual operation; no electronic components or software.

Indications for Use

Indicated for emergency oxygen production for patients requiring supplemental oxygen.

Regulatory Classification

Identification

A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K982143 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA - I. General Information | A. | Submitted By: | Canogen International, Ltd.<br>P.O. Box 766<br>Syosset, New York 11791<br>Tel: (516) 228-6600<br>Fax: (516) 228-6664 | |----|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | | Contact Person: | Steven R. Broder at address above | | B. | Device Trade Name:<br>Common Name:<br>Classification Name: | Canogen Portable Oxygen Generator, Model 615<br>Portable Oxygen Generator<br>Portable Oxygen Generator | | C. | Predicate Device: | Jet Research Center, Model 415 Oxygen Canister<br>and Dispenser Model 3445-c, 510(k) #K780020 | #### D. Device Description: The Canogen Portable Oxygen Generator, Model 615 provides an average of 6.5 liters a minute for a minimum of 15 minutes, through a chemical reaction. The Canogen Portable Oxygen Generator, Model 615 consists of three parts. The first part is a reaction vessel in which the chemical reaction takes place. The reaction vessel may be reused. The second part is the cartridge containing the chemicals required to generate the oxygen produced by this device. The third part of the generator consists of the plastic tubing and gas mask which are attached to the outlet of the generator. - E. Indications for Use: ! The Canogen Portable Oxygen Generator. Model 615 is intended to produce oxygen for emergency use. - F. Technological Comparison: The Canogen Portable Oxygen Generator, Model 615 and the Model 415 Oxygen Canister are designed to generate oxygen. Both devices are provided in {1}------------------------------------------------ kit form with the necessary containers. chemicals and accessories to provide oxygen for emergency use (with the exception of tap water for the Portable Oxygen Generator). As with the Jet Research Model 415, the chemicals used in the Canogen Portable Oxygen Generator are packaged in cartridges which are inserted into the reaction vessel to contain the reaction and port the gas to the face mask tubing. Page 2 of 2 #### II. Testing Bench test runs were conducted while using flow gauges, temperature sensors, data logger and pressure gauges to monitor performance characteristics during each run. The cartridge was inserted into the reaction vessel. The gas generated by the reaction was ported through plastic tubing to the flow sensor and manual flow gauges. For each test run the data acquisition system was used to log the flow rate, and temperatures. The Canogen Portable Oxygen Generator. Model 615 was determined to produce humidified oxygen at a purity of greater than 99% for a minimum of fifteen (15) minutes at an average flow rate of six and a half (6.5) liters per minute. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the United States. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ### NOV 1 9 1998 Ms. Melissa Mahall Canogen International Ltd. c/o Bio-Reg Associates, Inc. 14900 Sweitzer Lane, Suite 200 Laurel, MD 20707 Re: K982243 Canogen Portable Oxygen Generator, Model 615 Requlatory Class: II (two) Product Code: 73 CAW Dated: October 7, 1998 Received: October 13, 1998 Dear Ms. Mahall: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set {3}------------------------------------------------ Page 2 -- Ms. Melissa Mahall forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Cella Thomas J / Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # INDICATIONS FOR USE STATEMENT 510(k) Number (if known): Canogen Portable Oxygen Generator, Model 615 Device Name: Canogen International Ltd. Sponsor Name: ### Indications for Use The Canogen Portable Oxygen Generator, Model 615 is intended to produce oxygen for emergency use. Do Not Write Below This Line - Continue on Another Page if Needed Concurrence of CDRH, Office of Device Evaluation (ODE) Mark kramer (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number ា Prescription Use Over-The-Counter Use ਸ {
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...