DuraGen™ Dural Graft Matrix is indicated as a dura substitute for the repair of dura mater.
Device Story
DuraGen™ Dural Graft Matrix is an absorbable, porous, Type I collagen matrix derived from bovine achilles tendon. Designed for surgical repair of dura mater; device is soft, pliable, and nonfriable. Implanted by surgeons during neurosurgical procedures to serve as a scaffold for tissue repair; undergoes gradual biodegradation. Manufacturing process includes sodium hydroxide treatment for viral inactivation and BSE risk mitigation. Supplied sterile for single use. Clinical benefit includes effective dural closure without graft encapsulation, neomembrane formation, or foreign body reactions.
Clinical Evidence
Clinical evaluation involved over 1,000 patients. Biopsies from 100 patients underwent macroscopic and histological examination. Results showed no incidences of graft encapsulation, neomembrane formation, delayed hemorrhage, or foreign body reactions. Device demonstrated safe and effective performance as a dura substitute.
Technological Characteristics
Absorbable Type I collagen matrix derived from bovine achilles tendon. Porous, soft, pliable, nonfriable form factor. Sterilized for single use. Manufacturing utilizes sodium hydroxide treatment for viral inactivation. Complies with ISO 9001/EN 46001 standards. Non-pyrogenic.
Indications for Use
Indicated for the repair of dura mater in patients undergoing neurosurgical procedures requiring dural substitution.
Regulatory Classification
Identification
A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).
{0}------------------------------------------------
K.982180
JUL - 6 1999
Integra LifeSciences Corporation 510(k) K982180 DuraGen™ Dural Graft Matrix
## 510(k) PREMARKET NOTIFICATION SUMMARY OF SAFETY AND EFFECTIVENESS In Accordance with the Safe Medical Device Act of 1990
## DuraGen™ Dural Graft Matrix
- Submitted by: 1.
Integra LifeSciences Corporation 105 Morgan Lane Plainsboro, NJ 08536 USA 609-275-0500 Telephone: 609-799-3297 Fax:
### Contact Person:
Judith E. O'Grady, RN, MSN, RAC Sr. Vice President, Regulatory Affairs (609) 936-2311
- Date Prepared: June 1, 1999 2.
### Name of the Medical Device: 3.
DuraGen™ Dural Graft Matrix Proprietary Name: Classification Name: Dura Substitute
### Device Classification: 4.
- Class II Regulation Number 21 CFR 882.5910
### ડ. Indications for Use
- DuraGen™ Dural Graft Matrix is indicated as a dura substitute for the repair ◆ of dura mater.
#### Statement of Substantial Equivalence Q-
DuraGen™ Dural Graft Matrix is substantially equivalent in function and intended use to the following devices:
- I yoplant Dura Substitute, (Bovine Pericardium), 510(k) K970851 તે. Manufactured for AESCULAP, Inc.
- Dura-Guard® Dural Repair Patch, (Bovine Pericardium), 510(k) K973706 b. Manufactured by BioVascular, Inc.
- Preclude® Dura Substitute, (Expanded polytetrafluroethylene), C. 510(k) K953967. Manufactured by W.L. Gore & Associates
{1}------------------------------------------------
:
Integra LifeSciences Corporation 510(k) K982180 DuraGen™ Dural Graft Matrix
- TutoPlast® Processed Dura Mater, (Human Dura Mater), 510(k) K910555 d. Manufactured by Biodynamics International, Inc.
- Lyodura®, (Human Dura Mater) e. Manufactured for AESCULAP, Inc.
### 7. Device Description
DuraGen™ Dural Graft Matrix is an absorbable implant for repair of dura mater. DuraGen [M is an easy to handle, soft, white, pliable, nonfriable, porous collagen matrix. DuraGen™ is supplied sterile, nonpyrogenic, for single use in double peel packages in a variety of sizes.
DuraGen™ is manufactured from bovine achilles tendon, which is classified by European Standards as Class IV material, no detectable infectivity for Bovine Spongiform Encephalopathy (BSE). The bovine tendon is known to be one of the purest sources of Type I Collagen that is commercially available. The collagen used to manufacture DuraGen™ is currently used in the manufacture of an artificial skin, absorbable hemostatic sponges, and absorbable wound dressings. The manufacturing process for DuraGen134 meets the United States Food and Drug Administration's Guidance document and European Standards for animal tissue sourcing, handling and inactivation of BSE. This process involves a treatment with sodium hydroxide that is a recognized method of inactivation of Spongiform Encephalopathy (SE). The bovine tendon is only obtained from facilities in the United States of America that have been inspected by the United States Department of Agriculture (USDA). Bovine Spongiform Encephalopathy (BSE) is not known to exist in the United States of America, therefore, the United States of America is considered by the European Union as a low risk country for bovine sourced materials.
A viral inactivation study for the DuraGen™ manufacturing process was conducted by an independent certified laboratory. In this study, the sodium hydroxide treatment was evaluated for its ability to inactivate the following viral strains: I luman Immunodeficiency Virus Type I (HIV), Bovine Viral Diarrhea (BVD), Infectious Bovine Rhinotracheitis (IBR), Parainfluenza Virus Type 3 (P13). Vesicular Stomatitis (VSV). For these viruses, the sodium hydroxide treatment reduced the viral titer to non-detectable levels.
### 8. Performance Data
DuraGen™ Dural Graft Matrix has been evaluated by a number of tests for safety, biocompatibility, toxicity, pyrogenicity, sterility and preclinical and clinical evaluations. Summaries and reports of all data are contained in the Premarket Notification submission.
{2}------------------------------------------------
Integra LifeSciences Corporation 510(k) K982180 DuraGen™ Dural Graft Matrix
# Biocompatibility
The biocompatibility studies have demonstrated DuraGen™ Dural Graft Matrix to be: noncytotoxic, nonpyrogenic, nonmutagenic, nonsensitizing and be absorbed with no cellular or allergic responses. The following studies were conducted:
- Cytotoxicity Irritation/Intracutaneous Reactivity Pyrogenicity Genotoxicity Sensitization Study Chronic Toxicity Study Implantation Studies Intramuscular Toxicity Study Subcutaneous Implantation Study Immunological/Sensitivity Studies Mcchanical Testing
## Clinical Studies
The collagen sponge was evaluated in a long-term clinical study to examine its suitability as a dural graft. Clinical evaluations involving more than 1,000 patients demonstrated that DuraGen™ Dural Graft Matrix performed safely and effectively as a dura substitute. During the course of the study, biopsies of the collagen implant were available from 100 patients for macroscopic and histological examination. There were no incidences of graft encapsulation, neomembrane formation, delayed hemorrhage, or foreign body reactions.
### Summary/Comparison of Technological Characteristics 9.
## Technological Characteristics
DuraGen™ and its predicates have similar technological characteristics.
The technological differences between DuraGen™ and the predicate devices do not raise new types of safety or effectiveness issues. The technological differences have been assessed by valid scientific and clinical methods. Physical, laboratory, animal and human clinical test methods have been performed and have proven the ability of DuraGen™ to function as a dura substitute.
Table 1 describes the properties of DuraGen™ Dural Graft Matrix compared to properties of the predicate devices.
### 10. Manufacturing
DuraGen™ Dural Graft Matrix is manufactured by Integra LifeSciences Corporation at its facility in Plainsboro, NJ in compliance with FDA Quality System Regulations. The facility is registered as a United States Food and Drug Administration Medical Device manufacturing facility. The Integra LifeSciences facility is also ISO 9001/EN 46001/Medical Device Directive certified.
{3}------------------------------------------------
Integra LifeSciences Corporation 510(k) K982180 DuraGen™ Dural Graft Matrix
> The collagen used to manufacture DuraGen™ is used in other implantable medical devices manufactured by Integra LifeSciences such as hemostatic sponges, implantable wound dressings, and artificial skin.
### Quality Assurance 11.
DuraGen™ is fully tested according to an established Quality Control procedure and is only released if the product conforms to specifications.
The test data from extensive in vivo preclinical testing and extensive human clinical studies presented in this submission establish that DuraGen™ Dural Graft Matrix is equivalent to predicate devices in biocompatibility and safety and effectiveness as a dura substitute.
{4}------------------------------------------------
| Feature | DuraGen™<br>Dural Graft Matrix | Lyoplant Dura<br>Substitute¹ | Lyodura®¹ | Dura-Guard®<br>Dural Repair<br>Patch¹ | Preclude Dura<br>Substitute¹ | Tutoplast®<br>Dura<br>Substitute¹ |
|------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|------------------------------------------------------------------------|
| Manufacturer/<br>Distributor | Integra LifeSciences<br>Corporation | B. Braun /Aesculap | B. Braun/Aesculap | Bio-Vascular, Inc. | W.L. Gore &<br>Associates, Inc. | Biodynamics,<br>Inc. |
| Indications for<br>Use | Dura Substitute<br>Used as a dura<br>substitute for the repair<br>of dura mater. | Dura Substitute<br>Used as a dura mater<br>substitute in<br>neurological<br>procedures for soft<br>tissue reconstruction<br>of damaged,<br>impaired, or missing<br>tissue. | Dura Substitute<br>Indicated for the<br>replacement and<br>enlargement of<br>connective-tissue<br>structures. | Dura Substitute<br>Used as a dural<br>substitute for the<br>closure of dura<br>mater during<br>neurosurgery. | Dura Substitute<br>Used as a dura<br>substitute and for<br>spinal membrane | Dura Substitute<br>Used as a<br>substitute for<br>human dura<br>mater. |
| Materials | Bovine Type I Collagen | Bovine pericardium | Human Dura Mater | Bovine<br>pericardium | Expanded<br>polytetrafluoro-<br>ethylene | Human Dura<br>Mater |
| Absorption<br>Time | Gradual biodegradation | Gradual<br>biodegradation | Information not<br>available | Gradual<br>biodegradation | Non-resorbable | Gradual<br>biodegradation |
| Non-Pyrogenic | YES | YES | YES | Not claimed | YES | YES |
# TABLE 1: Substantial Equivalence Comparison Chart
'Based on information from company brochure
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of a human figure with outstretched arms, resembling an eagle or bird-like shape.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 6 1999
Ms. Judith E. O'Grady, RN, MSN Senior Vice President Regulatory Affairs, Quality Assurance and Clinical Affairs Integra Life Sciences Corporation 105 Morgan Lane · Plainsboro, New Jersey 08536
Re: K982180
> Trade Name: DuraGen™ Dural Graft Matrix Regulatory Class: II Product Code: GXQ Dated: June 1, 1999 Received: June 3, 1999
Dear Ms. O'Grady:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
## Page 2 - Ms. Judith E. O'Grady, RN, MSN
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
.
:
Page 1 of 1
510(k) Number: K982180 Device Name: DuraGen™ Dural Graft Matrix
# Indications for Use:
- DuraGen™ Dural Graft Matrix is indicated as a dura substitute for the . repair of dura mater.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|-------------------------------------------------------------------|
| Prescription Use | <div style="display:inline-block; vertical-align:middle;">X</div> |
| (Per 21 CFR 801.109 | |
| Optional Format 1-2-96) | |
| | Or |
| Over-the-Counter Use | ________________ |
| (Division Sign-Off) | |
| Division of General Restorative Devices | |
| 510(k) Number | K982180 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.