CYTO-STAT TRICHROME CD45-FITC/CD56-RD1/CD3-PC5 MONOCLONAL ANTIBODY REAGENT WITH CYTO-STAT TRICHROME CD45-FITC/MSIGG1-RD1
Applicant
Coulter Corp.
Product Code
GKZ · Hematology
Decision Date
Nov 5, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
CYTO-STAT® triCHROME™ CD45-FITC/CD56-RD1/CD3-PC5 Monoclonal Antibody Reagent is a three-color fluorescent reagent comprised of three murine monoclonal antibody is labeled with a different color fluorochrome. The reagent allows simultaneous identification and enumeration of total CD3+CD56+ lymphocytes in whole blood by flow cytometry. An isotypic control, CYTO-STAT® triCHROME™ CD45-FITC/MsIgG1-PC5, is used to monitor nonspecific staining. CYTO-STAT® triCHROME™ CD45-FITCMsIgG1-RD1/MsIgG1-PC5 Isotypic Control is a three-color fluoresent reagent comprised of three murine monoclonal antibody is labeled with a different color fluorochrome. This product is intended for use as a quality control reagent to monitor the levels of nonspecific antibody binding in cell surface staining procedures which use CYTO-STAT® triCHROME™ Monoclonal Antibody Reagents comprised of CD45-FITC and two MsIgG1 subclass monoclonal antibodies conjugated to RD1 and PC5. CD3+ lymphocyte percentages and absolute counts may be used as aids to evaluate immune competency underlying known or unknown disease states and to monitor lymphocvte levels following organ transplantation. To illustrate, identification of abnormal levels of CD3+ lymphocytes may aid in the diagnosis and/or progrosis of unidentified disease conditions in patients with low white blood cell counts. Measurement of CD3+ lymphocytes, in conjunction with CD4+ (induce) and CD8+ (suppressor/cytoxic) T lymphocytes and corresponding T478 ratios, may aid in the diagnois and/or prognosis of immunodeficiency disease such as infection with human immunodeficiency virus (HIV), the etiologic agent of acquired immunodeficiency syndrome (AIDS). Altered percentages of CD3+ lymphocytes recorded following organ (for example, kidney, heart, liver, lung) transplantation suggests T lymphocyte quantitation may be useful as an aid in monitoring these cellular populations. NK (Natural Killer) lymphocyte populations have been functionally defined as a lymphocyte population capable of mediating non-MHC restricted cytoxicity against targets such as certain tumor and virus-infected cells. As part of a Three Color Lymphocyte Immunophenotyping Panel which includes the B lymphocyte reagent, CYTO-STAT® triCHROME™ CD45-FITC/CD19-RD1/CD3-PC5, CYTO-STAT® triCHROME™ CD45-FITC/CD56-RD1/CD3-PC5 provides the ability to comprehensively identify and enumerate an individual's major lymphocyte subsets: T, B and NK. The reagent also functions as a quality control check for a specimen in terms of total lymphocyte percentage and CD3+ lymphocyte measurements across the panel.
Device Story
Reagent kit uses three murine monoclonal antibodies (CD45, CD56, CD3) conjugated to fluorochromes (FITC, RD1, PC5) to stain whole blood samples. Operates via flow cytometry; identifies and enumerates lymphocyte subsets (T and NK cells) by gating on CD45+ cells. Used in clinical laboratories by trained technicians. Output consists of percentage and absolute counts of lymphocyte populations. Healthcare providers use these metrics to assess immune status, monitor post-transplant patients, and aid in diagnosing immunodeficiency conditions. Includes an isotypic control reagent to monitor nonspecific antibody binding.
Clinical Evidence
Bench testing only. Accuracy evaluated using normal and abnormal whole blood specimens (n=diverse population, ages 19-84) compared against the predicate (CD3/CD56). Results demonstrated comparable identification and enumeration of CD3+ and CD3-/CD56+ lymphocytes. Linearity confirmed via serial dilution of concentrated cells. Precision (within-run and interlaboratory) demonstrated via replicate measurements with acceptable mean, SD, and CV values.
Indicated for the identification and enumeration of total CD3+ and CD3-/CD56+ lymphocyte populations in whole blood specimens via flow cytometry. Used to evaluate immune competency in patients with known or unknown disease states, monitor lymphocyte levels post-organ transplantation, and aid in the diagnosis/prognosis of immunodeficiency diseases (e.g., HIV/AIDS).
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
CYTO-STAT® CD3-FITC/CD56-RD1 Monoclonal Antibody Reagent with CYTO-STAT®/COULTER CLONE® MsIgG1-RD1/MsIgG1-FITC Isotypic Control (K982172)
{0}------------------------------------------------
NOV COULTER CORPORATION P.O. BOX 169015 Date: November 4, 1998 Miami Florida 33116-9015 USA Customer Service: (800) 526-769 Summary of Safety and Effectiveness Information For 510(k) Premarket Notification Title: Product Information: (800) 526-6932 (800) 327-6531 (305) 357-653 www.coulicr.com Product: CYTO-STAT® triCHROME™ CD45-FITC/CD56-RD1/CD3-PC5 Monoclonal Antibody Reagent with CYTO-STAT® triCHROME™ CD45-FITC/MsIgG1-RDJ/MsIgG1-PC5 Isotypic Control Coulter Corporation Company; 10.17 11 11 11 11 11 11 11 11 11800 SW 147 Avenue Miami, FL 33196-2500 ... ... . . Contact: Dr. Marion S. Gaide (M/C: 31-B06) Senior Regulatory Affairs Specialist 1. . . . . ( ' ' . . . Sun 10 11 11 Corporate Regulatory Affairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Yollars for note to be and 12. I mades! (, addrivers , trans , Telephone: (305) 380-2594 Mary College Comments of : . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Lymphocyte Immunophenotyping Monoclonal Antibody Reagent with Isotypic Control Kray at to marker Classified on the 2014 Product Classification: Product Code: GKZ; C.F.R. Section: 864.5220; Classification Hematology and Pathology Devices; Device Class: II Panel: いた。 Produkt K
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 Come of Moscore Matther CYTO STATE Conceller CYTO-STAT® triCHROME™ CD45-FITC/CD56-RD1/CD3-PC5 is a three-color fluorescont reagent · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · time in thorochrome. The reagent allows simultancous identification and enumeration of total CD3+ and CD3-. ... .. .. .. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . No do C V a la triCHROMETM CD45-FITC/MsIgG1-PCS, is used to monitor nonspecific staining. range of the count of the triCHROME CD45-FITCMsIgGT+RD1/MsgG1-PCS is a threesed in filloresear color in we lighted of three inutine monoclonal antibodies. Each antibody is labeled with a different store of the color fluorochrome. This product is intended for use as a quality control reagent to monitor the levels of Course In nonspecific antibody binding in cell surface staining procedures which use CYTO-STAT® a monoloriCHROME™ Monoclonal Antibody Reagents comprised of CD45-FITC and two monoclonal Colletter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Caracta, Vanchestia Substantial Equivalence: 510(k) Premarket Notification: K982172 CYTO-STAT® CD3-FITC/CD56-RD1 Monoclonal Antibody Reagent with CYTO-STAT®/COULTER CLONE® MsIgG1-RD1/MsIgG1-FITC Isotypic Control Product Comparison: The CD45/CD56/CD3 and CD3/CD56 systems are essentially identical with respect to features and principles of operation. Each liquid reagent allows simultaneous identification and enumeration of more than one lymphocyte population (CD3+ and CD3-/CD56+) in a single specimen using a single reagent. Bach system also requires an isotypic control to monitor nonspecific binding. The difference between the systems is that CD45/CD36/CD3 contains CD45 to identify a lymphocyte gate for making CD3+ and CD3-/CD56+ measurcments. CD3/CD56 rcquires a scparate reagent, CYTO-STAT®/COULTER CLONE® Mo2-RD1/KCS6 (T-200)-FITC, for this purpose. MAb Conjugation: CD45: FITC
(Fluorescein Isothiocyanate). CD56: RD1 (Phycoerythrin). CD3: PCS (Phycoerythrin-Cy5),
{1}------------------------------------------------
- Product testing to assess the performance of CDAS/CD3 is described below. Studies were Product Testing: designed in line with instructions for use in the product package insert and performance specifications, Specimens were assayed with CD3/CD56 for comparison purposes. The results of product testing demonstrated that CD45/CD56/CD3 meets all performance specifications and provides mature T (CD3+) and NK (CD3-/CD56+) lymphocyte values comparable to those of CD3/CD56.
- l . Accuracy:
Normal and abnormal whole blood specimens were collected from geographically diverse populations of males and females unselected as to race and ranging in age from 19 to 84 years. Specimens were divided, processed as lysed preparations and assayed in parallel with CD45/CD56/CD3 and CD3/CD56. CD3+ and CD3-/CD56+ percentages expressed in terms of the total lymphocyte count and absolute counts (cells/uL) were determined with COULTER® EPICS® XL-MCL ™ flow cytomerers gated on lymphocytes. White blood cell counts and 5-part differentials were obtained for all specimens.
Results analyzed in terms of minimums, maximums, means ± 1 SD, confidence intervals, regression analyses and analyses of variance demonstrated that CD45/CD3 and CD3/CD56 identify and enumerate essentially identical numbers of the targeted lymphocytes in whole blood specimens.
- 2. Linearity:
Three replicate measurements were made on a concentrated COULTER™ CYTO-TROL Cclls sample serially diluted to achieve a range of CD3+ and CD3-/CD56+ lymphocyte concentrations, Samples were assayed with CD45/CD56/CD3 and analyzed on a COULTER® EPICS® XL-MCL™ flow cytometer gated on lymphocytes. Values were expressed in terms of absolute count (cclls/uL).
Results analyzed in terms of regression and correlation analyses for recovered vs. expected absolute counts demonstrated Linearity of the assay.
- Precision: Within Run (Intralaboratory): 3.
Ten replicate measurements were made for each of three levels of CD3+ and CD3-/CD56+ lymphocyte concentrations on the same day using a COULTER® EPICS® XL-MCL™ flow cytometer gated on lymphocytes. Levels were obtained by selective screening of normal whole blood specimens and assayed with CD45/CD56/CD3. Values were expressed in terms of percentage of the total lymphocyte count.
Results analyzed in terms of mean ± 1 SD and CV demonstrated Within Day (Intralaboratory) Precision of the assay.
- 4. Precision (Interlaboratory):
Ton replicate measurements on were made on the same day using different laboratories and COULTER® EPICS® XL-MCL™ flow cytomcters. All measurements were made on a single normal whole blood specimen divided and assayed with CD45/CD56/CD3, Values were expressed in terms of percentage of the total lymphocyte count.
Results analyzed in terms of mean ± 1 SD and CV demonstrated Interlaboratory Precision of the assay.
Marion S. Gaido, Ph.D.
Senior Regulatory Affairs Specialist Corporate Regulatory Affairs
November 4, 1998
Date
{2}------------------------------------------------
Received: October 1, 1998
Image /page/2/Picture/2 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes extending from its back, symbolizing the department's commitment to health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Marion S. Gaide, Ph.D. Senior Regulatory Affairs Specialist Coulter Corporation P.O. Box 169015 NOV રે 1998 Miami, Florida 33116-9015 Re : к982166/S1 Trade Name: CYTO-STAT® triCHROME™ CD45-FITC/CD56-RD1/CD3-PC5 Monoclonal Antibody Reagent with CYTO-STAT® triCHROME™ CD45-FITC/MsIqG1-RD1/MsIgG1-PC5 Isotypic Control Regulatory Class: II Product Code: GKZ Dated: September 29, 1998
Dear Dr. Gaide:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, throuqh periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
## Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
## INDICATIONS FOR USE
510(k) Number (if known): Not Yet Assigned
## CYTO-STAT® triCHROME™ CD45-FITC/CD56-RD1/CD3-PC5 Monoclonal Antibody Reagent Device Name: with CYTO-STAT® triCHROME™ CD45-FITC/MsIgG1-RD1/MsIgG1-PC5 Isotypic Control
## Indications For Use:
CYTO-STAT® triCHROME™ CD45-FITC/CD56-RD1/CD3-PC5 Monoclonal Antibody Reagent is a three-color fluorescent reagent comprised of three murine monoclonal antibody is labeled with a different color fluorochrome. The reagent allows simultaneous identification and enumeration of total CD3+CD56+ lymphocytes in whole blood by flow cytometry. An isotypic control, CYTO-STAT® triCHROME™ CD45-FITC/MsIgG1-PC5, is used to monitor nonspecific staining.
CYTO-STAT® triCHROME™ CD45-FITCMsIgG1-RD1/MsIgG1-PC5 Isotypic Control is a three-color fluoresent reagent comprised of three murine monoclonal antibody is labeled with a different color fluorochrome. This product is intended for use as a quality control reagent to monitor the levels of nonspecific antibody binding in cell surface staining procedures which use CYTO-STAT® triCHROME™ Monoclonal Antibody Reagents comprised of CD45-FITC and two MsIgG1 subclass monoclonal antibodies conjugated to RD1 and PC5.
CD3+ lymphocyte percentages and absolute counts may be used as aids to evaluate immune competency underlying known or unknown disease states and to monitor lymphocvte levels following organ transplantation.
To illustrate, identification of abnormal levels of CD3+ lymphocytes may aid in the diagnosis and/or progrosis of unidentified disease conditions in patients with low white blood cell counts. Measurement of CD3+ lymphocytes, in conjunction with CD4+ (induce) and CD8+ (suppressor/cytoxic) T lymphocytes and corresponding T478 ratios, may aid in the diagnois and/or prognosis of immunodeficiency disease such as infection with human immunodeficiency virus (HIV), the etiologic agent of acquired immunodeficiency syndrome (AIDS). Altered percentages of CD3+ lymphocytes recorded following organ (for example, kidney, heart, liver, lung) transplantation suggests T lymphocyte quantitation may be useful as an aid in monitoring these cellular populations.
NK (Natural Killer) lymphocyte populations have been functionally defined as a lymphocyte population capable of mediating non-MHC restricted cytoxicity against targets such as certain tumor and virus-infected cells.
As part of a Three Color Lymphocyte Immunophenotyping Panel which includes the B lymphocyte reagent, CYTO-STAT® triCHROME™ CD45-FITC/CD19-RD1/CD3-PC5, CYTO-STAT® triCHROME™ CD45-FITC/CD56-RD1/CD3-PC5 provides the ability to comprehensively identify and enumerate an individual's major lymphocyte subsets: T, B and NK. The reagent also functions as a quality control check for a specimen in terms of total lymphocyte percentage and CD3+ lymphocyte measurements across the panel.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div>✓</div> | Over-The-Counter Use | |
|----------------------|--------------|----------------------|--|
| (Per 21 CFR 801.109) | | | |
(Division Sign-Off)Division Sian-Om Division of Clinical Laboratory Devices 510/k) Nimber
510i4us2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.