QUICK SCREEN BARBITURATE SCREENING TEST MODELS 9020 AND 9021, QUICK SCREEN PRO MULTI DRUG SCREENING TEST MODL 9152

K982152 · Phamatech · DIS · Sep 21, 1998 · Clinical Toxicology

Device Facts

Record IDK982152
Device NameQUICK SCREEN BARBITURATE SCREENING TEST MODELS 9020 AND 9021, QUICK SCREEN PRO MULTI DRUG SCREENING TEST MODL 9152
ApplicantPhamatech
Product CodeDIS · Clinical Toxicology
Decision DateSep 21, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3150
Device ClassClass 2

Indications for Use

The QuickScreen™ Barbiturates Screening Test is a rapid, qualitative immunoassay for the detection of barbiturates compounds in urine at a cutoff concentration of 200 ng/mL of Secobarbital. This assay is intended for professional use. This test provides only a preliminary test result. A more specific atternate test method must be used in order to obtain a confirmed analytical result. Gas chromatography / mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are observed.

Device Story

QuickScreen™ Barbiturates Screening Test is a rapid, qualitative, one-step immunoassay for urine drug screening. Device utilizes visual, color sandwich immunochemical technology to detect barbiturate compounds. Input is a urine sample; output is a visual color change indicating presence or absence of barbiturates at a 200 ng/mL cutoff. Intended for professional use in clinical settings. Results are preliminary and require confirmation via GC/MS or other chemical methods. Clinical judgment is required for interpretation. Device aids healthcare providers in identifying potential drug abuse, facilitating subsequent diagnostic or clinical management decisions.

Clinical Evidence

Performance evaluated via clinical sample correlation study and blind-labeled spiked sample study. Correlation study against Behring EMIT II showed 97% agreement. No further statistical metrics provided.

Technological Characteristics

Visual, color sandwich, one-step immunoassay. Principle: immunochemical sandwich assay (antibody/barbiturate/antibody complex formation). Form factor: rapid test kit. No electronic components, software, or external energy source required.

Indications for Use

Indicated for professional use as an in vitro qualitative screening test for barbiturates in human urine at a 200 ng/mL cutoff concentration of Secobarbital to identify potential drug abuse.

Regulatory Classification

Identification

A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 2 1 1998 K982152 ## 510 (k) SUMMARY AS REQUIRED BY SECTION 807,92(C) Identification: QuickScreen™ Barbiturates Screening Test (9020/9021) Description: Immunoassay for the qualitative detection of barbiturates in urine Phamatech Name of Manufacturer: 9265 Activity Road #112 San Diego, California 92126, USA Intended Use; The QuickScreen™ Barbiturates Screening Test is a rapid, qualitative immunoassay for the detection of barbiturates compounds in urine at a cutoff concentration of 200 ng/mL of Secobarbital. This assay is intended for professional use. This test provides only a preliminary test result. A more specific atternate test method must be used in order to obtain a confirmed analytical result. Gas chromatography / mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are observed. Barbiturates form a large class of abused pharmaceuticals. These products are anxiolytics, sedatives/hypnotics, anticonvulsants and anesthetic drugs. As CNS depressants, the barbiturates exert effects on excitatory and inhibitory synaptic neurotransmission. The ultra short acting barbiturates used for anesthesia, such as Thiopental, depress excitatory neuronal transmission to a greater extent than the anticonvulsant barbiturates such as Pentobarbital. Barbiturates are rapidly and completely absorbed with nearly 100% bioavailability. Short-acting barbiturates are generally excreted in unne as metabolites, while long-acting barbiturates are orimarily excreted unchanged. Ratios of drugs to metabolites vary, dependent upon duration of action. Technology: The QuickScreen"" Test, like many commercially available drug screening test kits, qualitatively measures the presence of barbiturates by visual, color sandwich, one-step immunoassay technology. Examples of such predicate devices include the ABMC Rapid Drug Screen (Ancramdale, NY) and the Applied Biotech SureStep Test (San Diego, CA 92121). All of the above devices rely on the basic immunochemical sandwich assay principle of recognition and formation of specific antibody / barbiturates / antibody complexes. Performance: The product performance characteristics of the QuickScreen™ Barbiturates Screening Test were evaluated in a clinical sample correlation study and a blind-labeled spiked sample study. The results of these studies demonstrate the Phamatech QuickScreen™ Test to be substantially equivalent to the reported performance characteristics of other commercially available tests for the qualitative detection of barbiturates in unne. Correlation studies, using clinical specimens, produced a 97% agreement when compared to the Behring EMIT II (Cupertino, CA 95014). Conclusion: For the reasons mentioned above, it may be concluded that the Phamatech QuickScreen™ Barbiturates Screening Test is substantially equivalent to a variety of barbiturate detection tests currently in commercial distribution. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SEP 2 1 1998 Carl Mongiovi Director of Operations Phamatech 9265 Activity Road #112-113 San Diego, California 92126 Re : K982152 QuickScreen Barbiturates Screening Test Requlatory Class: II Product Code: DIS Dated: August 14, 1998 Received: Auqust 18, 1998 Dear Mr. Mongiovi: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਖੇ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {2}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Litman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## INDICATIONS FOR USE Applicant: Phamatech 510(k) Number (if known): K982152 Device Name: QuickScreen™ Barbiturates Screening Test Indications for Use: An in vitro test for the qualitative identification of barbiturates in urine at a cutoff concentration of 200 ng/mL of Secobarbital. Measurements that are obtained by this device are used in the screening for drug abuse. PLEASE DO NOT WRITE BELOW THIS LINE Concurrence of CDRH Office of Device Evaluation (ODE) Ci --- (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number . Division Sign-Off Division of Clinical Laboratory Devices 510(k) Number:_K9 5 215 2 Prescription Use: / OR Over the Counter: _ Per 21 CFR 801.109
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...