K982144 · Mega Motion, Inc. · INI · Apr 14, 1999 · Physical Medicine
Device Facts
Record ID
K982144
Device Name
MEGA 4
Applicant
Mega Motion, Inc.
Product Code
INI · Physical Medicine
Decision Date
Apr 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
The Mega 4 scooter is intended to help the user achieve a more independent lifestyle by increasing their personal mobility. The Mega 4 scooter is intended for use by people who due to medical reasons, accident, or injury do not have adequate mobility to lead the active lifestyle they want. It is also intended for any person who would simply rather ride a scooter than walk.
Device Story
Four-wheel power scooter designed for outdoor use on uneven terrain; provides personal mobility for individuals with limited mobility. Input: user operation via throttle lever and speed potentiometer. Operation: sealed transaxle with pneumatic tires; dynamic regenerative braking system; electronic parking brake engages 0.5 seconds after throttle release. Output: mechanical propulsion; maximum forward speed 5.0 mph; reverse speed 3.0 mph. Features: adjustable tiller, 360-degree swivel seat, anti-tip wheels, and high-visibility reflectors. Operated by the user; intended for personal use. Benefits: increased independence and mobility. Safety features include EMI immunity (20 V/m) and water-resistant electronic casing.
Clinical Evidence
No clinical data submitted. Bench testing performed in accordance with ANSI/RESNA Standard, Wheelchair - Testing of Power and Control Systems for Electric Wheelchair (WC/14, Dec 1991), including parts 00, 01, 02, 03, 05, and 10. EMI testing passed at 20 V/m.
Technological Characteristics
Four-wheel power scooter; sealed transaxle; pneumatic tires (9"x4" front, 10"x4" rear); two 12-Volt 31 Amp sealed lead acid batteries; on-board charger. Speed regulated by throttle lever and speed potentiometer. EMI immunity 20 V/m. ANSI/RESNA WC/14 compliant.
Indications for Use
Indicated for individuals with limited mobility due to medical conditions, accidents, or injuries, or those preferring a scooter for personal mobility assistance.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
Predicate Devices
Pace Saver Eclipse Premier
Submission Summary (Full Text)
{0}------------------------------------------------
APR .1 4 1999
K982144
P.O. Box 3744 Scranton, PA 18505-0744 Phone: 1-888-415-1200 Fax: 1-888-415-1210
# 510(k) Summary
a)
Contact person: (1) Submitter's name & address: Mega Motion, Inc. Paul Gronka P.O. Box 3744 Official Correspondent Scranton, PA 18505-3744 1-888-415-1200 June 15, 1998 Date of preparation of this summary: Mega 4 scopter (2) Device trade or proprietary name 4-wheel power scooter Device common or usual name or classification name: 89INI Classification Number:
#### (3) Substantial Equivalence:
The Mega 4 power scooter is substantially equivalent in safety, efficacy, technology, and intended use to the Pace Saver Eclipse Premier, currently marketed by Leisure Lift.
(4) Description of the new device:
The Mega 4 is a four-wheel power scooter designed to help people with poor mobility lead a more active lifestyle. The fact that this is a 4-wheel scooter with a two wheel front end provides for a very stable ride. The Mega 4 scooter is an ideal outdoor scooter capable of traversing bumpy or uneven terrain.
The Mega 4 is a four-wheel scooter with exceptional handling and is quiet due to the use of pneumatic tires and a sealed transaxle as opposed to a separate motor, differential, and chain application. It is equipped with two kinds of brakes: dynamic regenerative and a parking brake. The electronic parking brake is normally on and is released when the driver moves the throttle lever. When the driver releases the throttle lever, the regenerative braking will slow the scooter to a stop and the parking brake will be activated 1/2 second thereafter, holding the scooter, preventing any movement. The electronic controller of the Mega 4 has a water-resistant casing with no exposed wires. This feature protects the electronics from environmental damage and corrosion.
The Mega 4 Scooter's maximum forward speed is 5.0mph. The maximum speed in reverse is 60% of the maximum forward speed or 3.0mph. The speed in reverse is reduced for safety purposes as it is more difficult to negotiate a scooter in reverse motion than in forward motion. The speeds are regulated by two devices: a throttle lever and a speed potentiometer.
The Mega 4 has several user-friendly features. It comes equipped with an external freewheeling release lever, external battery charger jack, and an external circuit breaker reset button all for easy accessibility. It has a special light-weight-seat (21 lbs.) for easy
tions, Inc. "Products for Better Livi
{1}------------------------------------------------
June 15, 1998 Page 2
> disassembly and transportation. The seat is adjustable up and down, swivels 360°, and has flip up arms. The Mega 4 is equipped with two lateral and two rear high visibility reflectors to increase the scooter's visibility. This feature, as well as the anti-tip wheels located 1 ¼" from the ground on the rear of the scooter, provide added safety for the passenger. The Mega 4 has an adjustable tiller for user comfort and easy transportation.
Electromagnetic interference (EMI) is the energy signals given off by any transmitting device such as CB radios and Cellular phones. These signals may cause erratic behavior in other electronic equipment operating nearby. This includes scooters. There is a caution sticker on the tiller to remind the user to use caution when riding the Mega 4 scooter and not to use communication equipment while the scooter is turned on. The Mega 4 has been EMI tested and passed at an immunity level of 20 V/m. While there is no way to tell if the scooter is totally safe, an immunity level of 20V/m (May '94) is generally achievable and useful. The results of the EMI testing performed on the Mega 4 scooter are included in section 6 of the 510(k)-application package.
The Mega 4 scooter has an on board charger as standard equipment. The charger has passed the minimum voltage ground leakage as required by the FDA/U.L. Certification from the charger supplier is included in section 4 of the 510(k)-application package.
The Mega 4 uses two 12-Volt 31 Amp sealed lead acid batteries. Certifications from the battery supplier that the batteries used on the Mega 3 meet the requirements of 49CFR173.159(d) and Special Provision A67 of the IATA Dangerous Goods Regulations for non-spillable batteries and are therefore, unrestricted for transportation by any means are enclosed in section 4 of this 510(k)-application package.
### (5) Intended use of the device:
The Mega 4 scooter is intended to help the user achieve a more independent lifestyle by increasing their personal mobility. The Mega 4 scooter is intended for use by people who due to medical reasons, accident, or injury do not have adequate mobility to lead the active lifestyle they want. It is also intended for any person who would simply rather ride a scooter than walk.
## (6) Mega 4 Specifications
| Length: | 47" |
|---------------------|-----------------------------|
| Width: | 24" |
| Turning radius: | 42" |
| Maximum Speed: | 5.0 mph |
| Weight capacity: | 250 lbs. |
| Maximum range: | 20-25 miles |
| Tire type (front): | 9" x 4" pneumatic |
| Tire type (rear): | 10" x 4" pneumatic |
| Rear weight: | 45 lbs. |
| Seat weight: | 21 lbs. |
| Base/Tiller weight: | 39 lbs. |
| Battery weight: | 24 lbs. (each) |
| Battery type: | Two 12-Volt 31 amp (sealed) |
| Battery charger: | On-board |
{2}------------------------------------------------
June 15, 1998 Page 3
- (1) Summary of nonclinical tests submitted with the premarket notification for the device. b) The Mega 3 power scooter was tested in accordance with the ANSI/RESNA Standard, Wheelchair -Testing of Power and Control Systems for Electric Wheelchair. WC/14, December 1991, parts 00, 01, 02, 03, 05, and 10 and passed. The results of these tests are included in section 6 of this 510(k) application package.
- (2) Summary of clinical tests submitted with the premarket notificationfor the device. b) No clinical tests were submitted with the premarket notification for the Mega 4 power scooter.
#### (3) Conclusions drawn from the clinical and nonclinical trials. b)
- Analysis of the comparison of design, function, and features of the Mega 4 power scooter to other devices currently legally marketed for the same intended use, together with the results of testing conducted to access the Mega 4 power scooter's compliance with existing ANSI/RESNA standards for powered wheelchairs demonstrates the Mega 4 to be substantially equivalent to these predicate devices in terms of safety, efficacy, intended use and technology.
{3}------------------------------------------------
APR 1 4 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Paul Gronka Official Correspondent Mega Motion, Inc. P.O. Box 3744 Scranton, Pennsylvania 18505-0744
K982144 Re : Mega 4 Trade Name: K982145 Mega 3 Trade Name: Regulatory Class: II Product Code: INI January 19 and March 19, 1999 Dated: January 21 and March 23, 1999 Received:
Dear Mr. Gronka:
We have reviewed your Section 510(k) notifications of intent to market the devices referenced above and we have determined these devices are substantially equivalent (for the indications for use stated in the enclosures) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major requlations affecting your devices can be found in the Code of a Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. · Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your devices in the Federal Register. Please note: this response to your premarket notification submission does
{4}------------------------------------------------
Page 2 - Mr. Paul Gronka
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Whitton, Ph.D., M.D.
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name:
Indications For Use:
To Whom It May Concern:
Indications for Use
hank vou
The Mega 4 scooter is intended to help the user achieve a more independent lifestyle by increasing their personal mobility. The Mega 4 is intended for use by people who due to medical reasons, accident, or injury do not have adequate mobility to lead the active lifestyle they want. It is also intended for any person who would simply rather ride a scooter than walk.
James D. Hill
Assistant to the Official Correspondent
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off
(Division Sign-Off) Division of General Restorativ 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use ✓
:
(Optional Format 1-2-96)
2007 HADD
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.