The CARESIDE™ Total Cholesterol cartridge is intended for in vitro diagnostic use in conjunction with the Exigent Diagnostics CARESIDE™ Analyzer to quantitatively measure total cholesterol concentration in anti-coagulated whole blood, plasma or serum.
Device Story
CARESIDE™ Total Cholesterol is a single-use disposable cartridge used with the CARESIDE™ Analyzer. It accepts anti-coagulated whole blood, plasma, or serum. The analyzer heats the cartridge to 37°C and spins it to separate blood cells from plasma/serum. Plasma is metered and dispensed onto a multi-layer reagent film. Lipoprotein lipase and surfactant dissociate cholesterol from protein; cholesterol esterase hydrolyzes esters; cholesterol oxidase produces hydrogen peroxide; peroxidase reacts with a diarylimidazole leuco dye to form a blue chromogen. A photodiode measures reflectance at 505 nm. The analyzer uses lot-specific standard curves to calculate total cholesterol concentration. Used in professional laboratory settings to assist clinicians in diagnosing and monitoring lipid metabolism disorders and cardiovascular risk. Benefits include direct whole blood testing without manual separation and rapid results.
Clinical Evidence
Method comparison study against the predicate device yielded a correlation of r = 0.94 (y = 1.04x + 23 mg/dL). Precision testing showed a total CV of 5.2% at 203 mg/dL. Accuracy demonstrated a mean recovery of 106%. Interference testing showed no significant impact from ascorbic acid (1 mg/dL), bilirubin (10 mg/dL), hemoglobin (250 mg/dL), protein (5-9 g/dL), or triglyceride (600 mg/dL).
Technological Characteristics
Dry film reagent cartridge containing lipoprotein lipase, cholesterol esterase, cholesterol oxidase, peroxidase, and diarylimidazole leuco dye. Sensing via reflectance photometry at 505 nm. Operates at 37°C. Features on-board barcode calibration. Connectivity via analyzer-integrated system. Quantitative measurement of total cholesterol in whole blood, plasma, or serum.
Indications for Use
Indicated for in vitro diagnostic use to measure total cholesterol from anti-coagulated whole blood, plasma, or serum to aid in the diagnosis and treatment of patients with disorders of lipid and lipoprotein metabolism.
Regulatory Classification
Identification
A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
Predicate Devices
Johnson and Johnson's Vitros CHOL Slides for Vitros DT 60 (K912844/A)
Submission Summary (Full Text)
{0}------------------------------------------------
# 111 2 / 1998
Exigent Diagnostics, Inc. Page 10
May 29, 1998
K982/118
# 510(K) SUMMARY: CARESIDE™ TOTAL IV. CHOLESTEROL SAFETY AND EFFECTIVENESS
### Applicant Information I.
- Applicant Name A.
- Applicant/Manufacturer Address B.
- C. Telephone Number
- D. Contact Person
- E. FAX Number
- F. e-Mail Address
- Date 510(k) Summary prepared G.
#### II. Device Information
- A. Device Name (Trade)
- Device Name (Classification) B.
- C. Device Classification
### D. Special controls and performance standards
#### III. Substantial Equivalence Claim
- A. General equivalency claim
The ability to monitor analyte-specific biochemical reactions in dry film and other formats is widely recognized and has gained widespread acceptance for use in chemistry assays.
Total cholesterol in vitro diagnostic products, in both dry film and other formats, are already on the U.S. market, including total cholesterol products which utilize enzymatic (cholesterol oxidase) generation of hydrogen peroxide from free and liberated cholesterol which reacts with chromogens in a peroxidase catalyzed reaction to form a dye.
- B. Specific equivalency claim
This CARESIDE™ Total Cholesterol test is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of total cholesterol on the Vitros DT 60 II.
| Name of Predicate Device: | Johnson and Johnson's (formerly Eastman Kodak,<br>Inc.) Vitros CHOL Slides for Johnson and<br>Johnson's Vitros DT 60 (formerly Eastman Kodak's<br>DT 60 II). |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device 510K number: | K912844/A |
| Product Code: | 75CHH |
Exigent Diagnostics, Inc. 6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Kenneth B. Asarch, Pharm.D., Ph.D. 310-338-6789 asarchk@worldnet.att.net May 29, 1998
## CARESIDE™ Total Cholesterol
Total Cholesterol test system Clinical chemistry panel Total cholesterol test system Regulation Number: 21 CFR 862.1175 Regulatory Class I Classification Number: 75CHH None applicable
{1}------------------------------------------------
#### IV. Device Description
CARESIDE™ Total Cholesterol cartridges are used with the Exigent Diagnostics CARESIDE™ Analyzer to measure total cholesterol concentration in anti-coagulated whole blood, plasma or serum specimens. The CARESIDE™ Total Cholesterol cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma or serum to a dry film to initiate the measurement of total cholesterol concentration. The film cartridge (patent pending) contains all reagents necessary to measure total concentration of cholesterol.
#### A. Explanation of Device Function
Each Exigent Diagnostics CARESIDE™ Total Cholesterol cartridge consists of a total cholesterol-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the anti-coagulated whole blood, serum, or plasma specimen into the cartridge sample deposition well, closes the lid and inserts the cartridge into the Exigent Diagnostics CARESIDE™ Analyzer.
Once loaded, the CARESIDE™ Analyzer scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers. As the cartridge continues to spin, the blood cells are separated from the plasma/serum and the cells accumulate in the separation well. Approximately ten microliters of plasma (or serum, as applicable) remain in the metering passage. Any excess sample flows into an overflow well.
The plasma (or serum, as applicable) is automatically dispensed onto the multi-layer reagent film. The spreading layer distributes the sample evenly on the film, and the sample moves through a reflection layer into the detection layer. Cholesterol is dissociated from protein by the action of lipoprotein lipase (LPL) and a surfactant. Cholesterol esterase (CHE) hydrolyzes the cholesterol esters (CHOL-ester) into free cholesterol. Endogenous cholesterol and the liberated cholesterol react with oxygen and water to produce hydrogen peroxide in a cholesterol oxidase (CO) catalyzed reaction. Peroxidase (POD) then catalyzes the reaction of the hydrogen peroxide with a diarylimidazole leuco dye to produce a blue chromogen. The color intensity of the resulting blue dye, as measured by the amount of reflected light at 505 nanometers, directly relates to the total cholesterol concentration of the specimen.
Test Reaction Sequence: Total Cholesterol ------------------------------------------------------------------------------------------------------------------------------------------------------------CHE -> CHOL (free) CHOL-Ester -CHOL + O2 + H2O2 + Cholesten-3one POD Fe .... > Blue dye Leuco dye + H2O2 -
As the cartridges spin, a photodiode measures reflectance of light emitted by a wavelength-specific light emitting diode (LED) at a fixed time. The analyzer uses the reflectance measurements and the lot-specific standard curve to calculate total cholesterol concentration.
{2}------------------------------------------------
Exigent Diagnostics, Inc. Page 12
#### B. Test Summary
Total cholesterol in blood comprises all of the cholesterol found in various lipoproteins. Cholesterol is the major component of the low density lipoprotein (LDL) fraction, and a minor component of the very low density lipoprotein (VLDL) and high density lipoprotein (HDL) fractions. Cholesterol measurements are used in classifying patients according to risk of coronary heart disease, in the diagnosis and treatment of various primary or secondary hyperlipidemias, and to monitor changes in cholesterol levels resulting from treatment. Elevated LDL-cholesterol has consistently been associated with incidence of atherosclerosis. There is also a strong correlation between considerably elevated cholesterol levels and an increased tendency for atherosclerosis. Conversely, HDL-cholesterol concentration and cardiovascular disease risk are inversely related. Measurement of total and HDL-cholesterol in serum is useful in evaluating cardiovascular disease risk.
#### V. Intended Use
- A. Intended Use
The CARESIDE™ Total Cholesterol cartridge is intended for in vitro diagnostic use in conjunction with the Exigent Diagnostics CARESIDE™ Analyzer to quantitatively measure total cholesterol concentration in anti-coagulated whole blood, plasma or serum.
- B. Indications for Use
This product is indicated for use in the diagnosis and treatment of patients with disorders of lipid and lipoprotein metabolism.
{3}------------------------------------------------
i
### VI. Technological Characteristics
- Similarities A.
| | CARESIDE™ Total<br>Cholesterol | Vitros CHOL DT Slides |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Primarily to aid in the diagnosis<br>and treatment of patients with<br>disorders of lipid and<br>lipoprotein metabolism. | Used in classifying patients<br>according to risk of coronary<br>heart disease, in the diagnosis<br>and treatment of various primary<br>or secondary hyperlipidemias,<br>and to monitor changes in<br>cholesterol and levels resulting<br>from treatment. |
| Indications | For <i>in vitro</i> diagnostic use.<br>For professional laboratory use. | For <i>in vitro</i> diagnostic use |
| Measurement | Quantitative | Same |
| Method Principle | Dry film based already on the<br>U.S. market, including total<br>cholesterol products which<br>utilize enzymatic (cholesterol<br>oxidase) generation - of<br>hydrogen peroxide from free<br>and liberated cholesterol<br>which reacts with chromogens<br>in a peroxidase catalyzed<br>reaction to form a dye. | Same |
| Specimen dilution | Not required | Same |
| Materials | Lipoprotein lipase, surfactant,<br>cholesterol esterase,<br>cholesterol oxidase,<br>peroxidase, and<br>diarylimidazole leuco dye | Surfactant, cholesterol<br>esterase, cholesterol oxidase,<br>peroxidase, and leuco dye |
| Detector | Reflectance (505 nm) | Reflectance (555 nm) |
| Test time | Approx. 4 minute warm-up (on-<br>board) plus 6 minute test time. | 15 minutes slide warm-up (off-<br>line) plus 5 minutes test time. |
| Reference Method | Abell Kendall | Abell Kendall |
| Sample Type | Serum, plasma, whole blood<br>[whole blood applied sample,<br>plasma test sample] | serum, plasma |
| Specimen volume | 10 µl test volume<br>(85 ± 15 µl applied volume) | 10 µl |
| Calibration | Calibration information bar-<br>coded on each cartridge.<br>Calibration information may<br>change with each lot. | Run Vitros DT II calibrators<br>whenever a new slide lot is<br>used or when necessary. |
| Quality Control | 2 levels | Same |
| Reporting Units | mg/dL or mmol/L | Same |
| Reaction Temp. | 37 °C | Same |
| | CARESIDE™ Total<br>Cholesterol | Vitros CHOL DT Slides |
| Direct blood<br>specimen | Yes, whole blood | No, requires separation of<br>whole blood prior to sample<br>application |
| Reportable range | 50 to 450 mg/dL | 50 to 325 mg/dL |
| Accurate<br>pipetting | Not required | Required |
| Reagent pre-<br>warming | Not required | Required |
{4}------------------------------------------------
### B. Differences
C. Comparative Performance Characteristics
| | CARESIDE™ Total<br>Cholesterol | Vitros CHOL DT Slides |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Detection limit | 50 mg/dL | 50 mg/dL |
| Reportable range | 50 to 450 mg/dL | 50 to 325 mg/dL |
| Accuracy | Mean recovery 106% | Not provided |
| Precision | Total CV, 203 mg/dL, 5.2% | Total CV, 202mg/dL, 3.4% |
| Method comparison | CARESIDE™ = 1.04 (Vitros CHOL DT) + 23 mg/dL, r = 0.94 | |
| Linearity | Linearity by mixing and by<br>dilution yielded slope and<br>correlation coefficient within<br>acceptable limits. | Not provided |
| Interference | No significant interference observed at tested concentration of interferent:<br><br>Ascorbic acid 1 mg/dL<br>Bilirubin 10 mg/dL<br>Hemoglobin 250 mg/dL<br>Protein 5 to 9 g/dL<br>Triglyceride 600 mg/dL | Very high protein > 10 mg/dL |
D. Conclusion
> The nonclinical and clinical data provided demonstrate that the CARESIDE™ Total Cholesterol product is as safe, effective, and performs as well as or better than the legally marketed predicate device
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around an emblem. The emblem is a stylized representation of a human figure embracing a bird, symbolizing care and protection.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUL 24 1998
Kenneth B. Asarch, Ph.D. . VP Quality Systems and Regulatory Affairs Exigent Diagnostics Inc. 6100 Bristol Parkway Culver City, California 90230
K982118 Re: CARESIDE™ Total Cholesterol Regulatory Class: I Product Code: CHH Dated: May 29, 1998 June 3, 1998 Received:
Dear Dr. Asarch:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ದ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## INDICATIONS FOR USE VI.
510(k) Number:
Device Name:
CARESIDE™ Total Cholesterol
Indications for use:
For in vitro diagnostic use with Exigent Diagnostics' CARESIDE™ Analyzer to measure total cholesterol from anti-coagulated whole blood, plasma or serum specimens to aid in the diagnosis and treatment of patients with disorders of lipid and lipoprotein metabolism
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number k982118
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.