AMIRA MEDICAL ATLAST BLOOD GLUCOSE MONITORING SYSTEM
K982076 · Amira Medical, Inc. · CGA · Dec 10, 1998 · Clinical Chemistry
Device Facts
Record ID
K982076
Device Name
AMIRA MEDICAL ATLAST BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Amira Medical, Inc.
Product Code
CGA · Clinical Chemistry
Decision Date
Dec 10, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ATLAST Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in whole blood. It is intended for use by people with diabetes mellitus in the home as an aid to monitor the effectiveness of diabetes control. It is not intended for use in the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates. The ATLAST Blood Glucose Monitoring System is specifically indicated for the quantitative measurement of glucose in whole blood samples obtained from the finger, forearm, upper arm and the thigh.
Device Story
System measures capillary whole blood glucose; integrates reflectance photometer and lancing device. User inserts sterile lancet into sampler head; device pierces skin; user performs sliding motion to draw blood. Test strip inserted into meter; capillary action draws blood into strip; glucose oxidase/peroxidase reaction produces color change proportional to glucose concentration. Meter measures color intensity via reflectance photometry; provides digital readout in 10-30 seconds. Used by patients with diabetes at home; aids in monitoring diabetes control. Output informs patient of glucose levels for self-management.
Clinical Evidence
Multi-center, subject-controlled clinical study evaluated accuracy in lay users with diabetes and trained technicians. Compared ATLAST system to predicate device and laboratory reference method. Results analyzed via Clarke Error Grid; demonstrated clinically acceptable accuracy and correlation with laboratory reference. Performance equivalent between lay users and trained technicians.
Technological Characteristics
Reflectance photometer; glucose oxidase/peroxidase reagent chemistry. Integrated lancing device. Dimensions/form factor: handheld. Connectivity: none (standalone). Sterilization: sterile lancets provided. Software: embedded firmware for signal processing and display.
Indications for Use
Indicated for quantitative blood glucose measurement in patients with diabetes mellitus for home monitoring of glycemic control. Not for diagnosis, screening, or use on neonates. Samples may be obtained from finger, forearm, upper arm, or thigh.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
LifeScan One Touch Basic Blood Glucose Monitoring System (K922888)
Boehringer Mannheim Accu-Chek Easy Blood Glucose Monitoring System (K923048)
Submission Summary (Full Text)
{0}------------------------------------------------
PAGE 04
DEC 1 0 1998
L982076
# Amira Medical ATLAST Blood Glucose Monitoring System 510(k) Premarket Notification
# 510(k) SUMMARY
# Submitter's Name, Address, Telephone Number, and Contact Person
Amira Medical 4742 Scotts Valley Road Scotts Valley, CA 95065 (408) 408-439-0700 X305 Phone: (408) 439-0907 Facsimile:
Nina Peled, Ph.D. Contact Person: Vice President, Scientific Affairs
## Date Prepared: June 10, 1998
## Name of Device and Name/Address of Sponsor
Trade name: ATLAST Blood Glucose Monitoring System
Amira Medical 4742 Scotts Valley Road Scotts Valley, CA 95065
#### Classification Names:
Colorimeter, photometer, spectrophotometer for clinical use (21 C.F.R. § 862.2300); Glucose oxidase, glucose test system (21 C.F.R. § 862.1345) Single analyte control (21 C.F.R. § 862.1660) Blood lancet (21 C.F.R. § 878.4800)
## Predicate Devices
- 1. LifeScan One Touch Basic Blood Glucose Monitoring System (K922888)
- 2. Boehringer Mannheim Accu-Chek Easy Blood Glucose Monitoring System (K923048)
Amira Medical Confidential
Revised: 11/24/98
{1}------------------------------------------------
#### Intended Use/Indications
The ATLAST Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in whole blood. It is intended for use by people with diabetes mellitus in the home as an aid to monitor the effectiveness of diabetes control. It is not intended for use in the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates. The ATLAST Blood Glucose Monitoring System is specifically indicated for the quantitative measurement of glucose in whole blood samples obtained from the finger, forearm, upper arm and the thigh.
#### Device Description
The ATLAST Blood Glucose Monitoring System measures the amount of glucose in capillary whole blood. When whole blood is applied to the ATLAST Test Strip, reagents on the Test Strip react with the blood and a color is formed, the intensity of which is measured by the ATLAST System. The ATLAST Blood Glucose Monitoring System consists of the following components: (1) a reflectance photometer which incorporates a lancing device ("ATLAST System"), (2) glucose reagent test strips ("ATLAST Test Strips") and calibration chip, and (3) sterile lancets ("ATLAST Lancets").
#### Principle of Operation
The user obtains a blood sample using the ATLAST System. A sterile lancet is inserted into the lancing component of the ATLAST System by inserting a sterile lancet into the sampler head and re-attaching it to the ATLAST System. The unit is cocked by pulling the cocking knob until it latches, then the user releases the knob. With the sampler head positioned firmly on the forearm, the user fully depresses the body of the unit which releases the lancet to pierce the skin. Immediately following the lancing action, the lancet is automatically retracted. A sliding motion is required to obtain an adequate blood sample. While maintaining the lancet and sampler head in position on the sample site, the body of the unit is depressed and released several times to draw blood from the incision point. After obtaining a sample, the user is ready to touch the test strip to the blood sample to begin the testing process.
The user inserts the test strip into the test strip holder of the ATLAST System and turns the ATLAST System on. The test strip is then brought to the lancing site and the user touches the capillary tip of the ATLAST Test Strip to the blood sample. Blood is drawn into the test strip by capillary action. The meter "beeps" when an adequate blood sample has been received and the testing sequence begins.
{2}------------------------------------------------
The glucose in the blood sample reacts with glucose oxidase to yield gluconic acid. This reaction is catalyzed by the peroxidase present in the test strip. The hydrogen peroxide formed then oxidizes the dye couple in the test strip. This reaction, which is catalyzed by the enzyme peroxidase, oxidizes the dye couple to form a dye that intensifies in color with increasing glucose concentrations. The color change on the reagent test strip pad is proportional to the concentration of the glucose present in the whole blood sample.
The user obtains blood glucose results in approximately 10-30 seconds, depending on glucose levels. The ATLAST System will "beep" when the test is complete and the blood glucose results will appear on the meter display.
### Data Demonstrating Substantial Equivalence
Performance testing on the ATLAST Blood Glucose Monitoring System demonstrated that the System meets or exceeds the performance requirements for the intended clinical use of the device. Laboratory/Bench Testing was conducted in accordance with FDA's draft guidance, Review Criteria for Assessment of Portable Blood Glucose Monitoring In-Vitro Diagnostic Devices Using Glucose Oxidase Methodology. The results demonstrated that the ATLAST Blood Glucose Monitoring System satisfies all of its performance specifications, which are designed to ensure that the System is safe and effective for its intended use.
A multi-center, subject controlled clinical study was conducted to evaluate the accuracy of the ATLAST Blood Glucose Monitoring System when used by lay users with diabetes mellitus and by experienced technicians trained in blood glucose testing techniques. In addition, the ATLAST Blood Glucose Monitoring System was compared to a predicate blood glucose monitoring system. Blood glucose results obtained with the ATLAST Blood Glucose Monitoring System and predicate device were compared to results obtained using a laboratory reference method for measuring blood glucose levels.
The clinical data demonstrate that the performance of the ATLAST Blood Glucose Monitoring System correlates well with the laboratory blood glucose reference test method. When the blood glucose test results were analyzed by the Error Grid Analysis of Clarke et al., the System provided results within the range of clinically acceptable accuracy. The data also demonstrate that the ATLAST Blood Glucose Monitoring System's performance is substantially equivalent to that of a predicate device. The clinical data demonstrate that the ATLAST Blood Glucose Monitoring System performs equivalently in the hands of the diabetic lay user and when used by a trained technician.
{3}------------------------------------------------
# Conclusion
.
Laboratory/Bench Testing and Clinical Correlation Studies demonstrate that the Labordvery/Bener February and Glucose Monitoring System is safe and effective for its intended use.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three heads, representing the department's commitment to health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
# DEC 10 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Nina Peled, Ph.D. Vice President, Scientific Affairs Amira Medical 4742 Scotts Valley Road Scotts Valley, California 95065
Re : K982076 Trade Name: ATLAST Blood Glucose Monitoring System Requlatory Class: II Product Code: CGA Dated: June 10, 1998 Received: June 12, 1998
Dear Dr. Peled:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K982076 510(k) Number (if known):
Device Name: Amira Medical ATLAST Blood Glucose Monitoring System
Indications For Use:
The Amira Medical ATLAST Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in whole blood. It is intended for use by people with diabetes mellitus in the home as an aid to monitor the effectiveness of diabetes control. It is not intended for use in the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates (newborn children).
The ATLAST Blood Glucose Monitoring System is specifically indicated for the quantitative measurement of glucose in whole blood samples obtained from the finger, forearm, upper arm and the thigh.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NERDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off)
Division of Clinical Laboratory Devices 12982076 510(k) Number.
Prescription Use
OR (Per 21 CFR 801.109)
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.