K982064 · Invacare Corp. · ITI · Oct 5, 1998 · Physical Medicine
Device Facts
Record ID
K982064
Device Name
ACTION STORM RANGER II, ACTION STORM POWER 9000
Applicant
Invacare Corp.
Product Code
ITI · Physical Medicine
Decision Date
Oct 5, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Invacare Action Storm and Ranger II Front Wheel Drive Power Wheelchairs is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair.
Device Story
Battery-powered, motor-driven, folding front-wheel-drive power wheelchairs; provide mobility to users limited to a sitting position. Input: user operation via joystick-controlled motor controller. Operation: DC batteries power motor/gearbox drive mechanism; adjustable motor/gearbox position compensates for user-dependent weight/balance. Output: physical mobility. Used in various environments by patients capable of operating the device. Healthcare providers use the device to restore mobility; patient benefits from increased independence. Construction: welded steel frame; U240 Nylon or reinforced vinyl upholstery.
Clinical Evidence
Bench testing only. Devices tested for electromagnetic compatibility in accordance with ISO EMC Draft Standard 7176-14. All units met required performance criteria.
Technological Characteristics
Battery-powered (two 12 VDC Size 22NF batteries); front-wheel-drive; welded steel construction; U240 Nylon or reinforced vinyl upholstery (meeting California 117 and Boston BFD-1 fire retardancy specs); joystick-operated motor controller (Invacare Model MCC-MKIV RII).
Indications for Use
Indicated for persons limited to a seated position who have the capability of operating a powered wheelchair.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
K982064
文 1 P Sac Fifth
#### 510(k) SUMMARY
Invacare Corporation's
Models Action Storm Ranger II and Power 9000 Front Wheel Drive Power Wheelchairs
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared.
Invacare Corporation One Invacare Way PO Box 4028 Elyria, Ohio 44036 Phone: (440) 329-6000 Facsimile: (440) 365-4558
Contact Person:
. ............................................................................................................................................................................
Edward A. Kroll Director, TQM and Regulatory Affairs
Date Prepared:
June 10, 1998
## Name of Device and Name/Address of Sponsor
Models Action Storm Ranger II and Action Storm Power 9000 Power Wheelchairs
Invacare Corporation One Invacare Way Elyria, Ohio 44036-2028 Phone: (440) 329-6000 Facsimile: (440) 365-4558
## Common or Usual Name
Power Wheelchair
### Classification Name
Wheelchair, Powered
{1}------------------------------------------------
## Predicate Devices
The Models Ranger II and Power 9000 Front Wheel Drive power wheelchairs are substantially equivalent to Permobil's Chairman (K960951, April 4, 1997) Front Wheel Drive Power Wheelchair and Invacare Power 9000 (K900565, February 2, 1990) power wheelchairs. The wheelchairs are controlled using the Invacare Model MCC-MKIV RII Motor Controller and Joystick. This controller was granted marketing clearance by FDA on June 2, 1994, under 510(k) Accession Number K940972.
#### Intended Use
The intended use of the Invacare Action Storm and Ranger II Front Wheel Drive Power Wheelchairs is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair.
## Technological Characteristics and Substantial Equivalence
### A. Device Description
The Invacare Models Ranger II and Power 9000 front wheel drive power wheelchairs are battery powered, motor driven devices with the intended function and use of providing mobility to those persons limited to a sitting position that have the capability of operating a power wheelchair. They are non rigid, or "folding" type, front wheel drive power wheelchairs.
The wheelchairs are controlled using the Invacare Model MCC-MKIV RII Motor Controller and Joystick. They are powered by two Size 22NF 12 VDC batteries. Access to the batteries is gained from underneath the chairs will travel approximately 17 to 22 miles between charges, depending on use.
The wheelchairs consist of two basic sub-sections. These are the base section of the wheelchair, and the seating section of the wheelchair. Both are of welded steel construction. The base section includes the base frame, front drive wheels and axles, rear pivoting casters, motor/gearbox drive mechanism and batteries. The motor/gearbox location is adjustable, relative to the upper seating frame. This adjustment is used to compensate for weight and balance shift, which are both user dependent.
The seating, or upper section of the wheelchair, includes the seating upholstery, front foot rests, side arm rests, and joystick operating control. Seating sizes range from 14" wide to 20" wide, depending on user needs. The wheelchair motor controller is mounted to the side arm and under the backrest of the seat. The upholstery material is fabricated from either U240 Nylon, or reinforced vinyl, depending upon the user's preference. Both of these materials meet California 117 and Boston Fire Department BFD-1 specifications for fire retardancy.
{2}------------------------------------------------
#### B. Substantial Equivalence
Products which are substantially equivalent to these devices are Permobil's Chairman Front Wheel Drive Power Wheelchair (K960951, April 4, 1997) and Invacare Corporations' Power 9000 (K900565, February 2, 1990) power wheelchairs.
Each of these products are battery powered, motor driven powered wheelchairs with the same intended function and use which is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair. Similarities include large wheels with attached motor/gearbox drive mechanisms, smaller pivoting casters for turning, and joystick operated motor controllers to engage system motion and steer the wheelchair. They are all · constructed from the same basic materials, have the same basic operational principles and all use DC batteries as their source of power.
#### PERFORMANCE DATA
As required by FDA's July 26, 1995, draft publication entitled "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three- Wheeled Vehicles", the Invacare Action Storm and Ranger II Front Wheel Drive Power Wheelchairs were tested in accordance with ISO EMC Draft Standard 7176-14 (Titled "Draft ISO EMC Group Proposal: Electromagnetic Compatibility Addition" And Dated April 3, 1995) for powered wheelchairs and motorized scooters. In all instances, the Invacare Action Storm and Ranger II Front Wheel Drive Power Wheelchairs met the required performance criteria and functioned as intended.
15"
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The caduceus is black, and the text is also black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
,一个
OCT 2
Mr. Edward A. Kroll Director, TQM and Regulatory Affairs Invacare Corporation One Invacare Way P.O. Box 4028 Elyria, Ohio 44036
Re: K982064 Action Storm Ranger II, Action Storm Power 9000 Requlatory Class: II ITI Product Code: Dated: August 25, 1998 Received: September 11, 1998
Dear Mr. Kroll:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2 - Mr. Edward A. Kroll
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Whitten, Ph.D., M
Celia M. Witten, Ph.D., M.D. Director
Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 510(k) Number (if known): TBD
Device Name: Invacare Models Ranger II and Power 9000 Front Wheel Drive Power Wheelchair
## Indications For Use:
To provide mobility to persons limited to a seated position.
的意思想,
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEI
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
**Over-The-Counter Use**\$\triangle\$
(Optional Format I
(Division Sia) Division of General Restorative Dev 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.