K982010 · Philips Medical Systems (Cleveland), Inc. · JAK · Nov 19, 1998 · Radiology
Device Facts
Record ID
K982010
Device Name
FACTS
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
JAK · Radiology
Decision Date
Nov 19, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The FACTS is intended to produce routine cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles of the entire human body. In addition, this device's fluoroscopic mode is intended to be used in CT interventional procedures and allows some minimally invasive surgical procedures to be implemented with the CT system. Procedures possible with this equipment include tumor biopsies, abscess drainages, bone intervention, visceral, neck trauma evaluation and catheter placement for organ assessment.
Device Story
FACTS (Fluoroscopic Assisted Computed Tomography System) produces cross-sectional images of human body via computer reconstruction of x-ray transmission data; utilizes solid state x-ray imager (flat panel) to convert x-radiation into visible images. System includes signal analysis/display equipment, patient/equipment supports, and accessories. Operates in standard CT mode and fluoroscopic mode for interventional procedures. Used by clinicians in clinical settings to guide minimally invasive procedures like biopsies, drainages, and catheter placements. Output allows visualization of anatomical structures to assist in clinical decision-making and procedural guidance.
Clinical Evidence
Bench testing only. The device was designed and manufactured in accordance with FDA 21 CFR 820, 1020.30-1020.33, UL 187, and IEC 601-1 standards to ensure safety and effectiveness.
Technological Characteristics
Computed tomography x-ray system with solid state x-ray imager (flat panel). Components include signal analysis/display equipment and patient supports. Designed to conform to IEC 601-1, UL 187, and FDA 21 CFR 1020.30-1020.33. Year 2000 ready.
Indications for Use
Indicated for patients requiring cross-sectional imaging of the entire body and those undergoing CT-guided interventional procedures, including tumor biopsies, abscess drainages, bone intervention, visceral/neck trauma evaluation, and catheter placement.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
NOV 1 9 1998
ﺤﺼ
### Picker 510(k) Notice
## FACTS
#### SUBSTANTIAL EQUIVALENCE SUMMARY
The following information is being submitted in accordance with 21 CFR 807.92(a) and in the order specified in that section.
(1) The submitter's name, address, telephone number, a contact person, and the date the summary was prepared;
| Submitter: | Picker International, Inc.<br>595 miner Road<br>Highland Heights, Ohio 44143<br>(440)473-3000 |
|------------|--------------------------------------------------------------------------------------------------------------------|
| Contact: | Robert L. Turocy<br>Picker International, Inc.<br>595 Miner Road<br>Highland Heights, Ohio 44143<br>(440) 473-3528 |
Date of Summary: June 5, 1998
(2) The name of the device, including the trade name or proprietary name if applicable, the common or usual name, and the classification name, if known;
| Device Name<br>(Proprietary Name): | FACTS |
|------------------------------------|--------------------------------------------------------------------------|
| Classification Name: | Computed Tomography X-Ray System<br>and Solid State X-Ray Imaging Device |
| Common Name: | Computed Tomography X-Ray System<br>and Solid State X-Ray Imaging Device |
The FDA has classified the FACTS as a combination of a CT System as Class II in 21 CFR 892.1750 (ProCode 90JAK) plus the FACTS Monoblock Diagnostic Source Assembly as Class II in 21 CFR 892.1610 (Product Code 90 IZX/90 ITY/90IZO) plus a SSXI Class II in 892.1660 (Product Code 90MQB) based on the recommendations of the Radiology Devices Panel.
{1}------------------------------------------------
# Continued FACTS SUBSTANTIAL EQUIVALENCE SUMMARY
(3) An identification of the legally marketed device to which the submitter claims equivalence. A legally marketed device to which a new device may be compared for a determination regarding substantial equivalence is a device that was legally marketed prior to May 28, 1976, or a device which has been reclassified from Class III to Class II or 1 the predicate) or a device which has been found to be substantially equivalent through the 510(k) Premarket Notification process.
Picker claims substantially equivalence of the FACTS to the combination of legally marketed devices with a SSXI. The intended use of the FACTS is the same as legally marketed device PQ2000+ CT System granted marketing permission in K955268 combined with a C-Arm including the DynaRad Diagnostic Source Assembly cleared in K981267 and a SSXI subject of this 510(k) manufactured by Varian for Picker that is equivalent to the image intensifier of the Orbitor cleared in K800639. No new questions of safety or effectiveness are raised with the combination and introduction of this device.
(4) A description of the device that is the subject of the Premarket Notification submission, such as might be found in the labeling or promotional material for the device, including an explanation of how the device functions, the scientific concepts that form the basis for the device, and the significant physical and performance characteristics of the device such as device design, material used, and physical properties:
The FACTS X-Ray System, is a device intended to produce cross-sectional images of the human body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. The solid state x-ray imager is a device intended to be used to visualize anatomical structures by using a flat panel to convert x-radiation into a visible image. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. For additional information about the FACTS, please refer to the preliminary Promotional Literature, Product Specification in Attachment "B" and the preliminary Operator's Manual in Attachment "C."
Picker FACTS is designed and manufactured in accordance with FDA 21CFR 820, 1020.30 through 1020.33, and voluntary standards for safety and effectiveness (UL 187) all of which mandate that components are tested to minimize hazards (electrical, mechanical, and radiation). In addition, the system is designed to conform to IEC 601-1. The subject device has the same technological characteristics as a legally marketed predicate devices. Specifically, the features, specifications, materials, and mode of operation are equivalent. The FACTS is year 2000 ready,
{2}------------------------------------------------
## Continued FACTS SUBSTANTIAL EQUIVALENCE SUMMARY
8
(5) A statement of the intended use of the device that is subject of the Premarket Notification submission, including a general description of the diseases or conditions that the device will diagnose, treat, prevent, cure, or will mitigate, including a description, where appropriate, of the patient population for which the device is intended. If the indication statements are different from those of the legally marketed device identified in paragraph (a)(3) of this section, the 510(k) summary shall contain an explanation as to why the differences are not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, and why the differences do not affect the safety and effectiveness of the device when used as labeled; and
The FACTS is intended to produce routine cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles of the entire human body. In addition, this device's fluoroscopic mode is intended to be used in CT interventional procedures and allows some Reastwoon and in a rail in procedures to be implemented with the CT system. Procedures possible with this equipment include tumor biopsies, abscess drainages, bone intervention, visceral, neck trauma evaluation and catheter placement for organ assessment.
(6) If the device has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device identified in paragraph (a)(3) of this section, a summary of the technological characteristics of the new device in comparison to those of the predicate device. If the device has different technological characteristics from the predicate device, a summary of how the technological characteristics of the device compare to a legally marketed device identified in paragraph (a)(3) of this section.
Claims: The FACTS is a comparable type and substantially equivalent to legally marketed devices. The intended use of the The FACTS X-Ray System, is a device intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. The solid state x-ray imager is a device intended to be used to visualize anatomical structures by using a flat panel to convert x-radiation into a visible image. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. Specifically, the features, specifications, materials, mode of operation are equivalent.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo includes the department's emblem on the left, which features a stylized human figure. To the right of the emblem, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is displayed in a clear, sans-serif font.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Robert L. Turocy Regulatory Affairs & Compliance Manager Picker International. Inc. World Headquarters 595 Miner Road CLEVELAND OH 44143
AUG 2 3 2013
Re: K982010
Trade/Device Name: FACTS (fluoroscopic assisted computed tomography system) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK, OWB, and JAA Dated: September 18, 1998 Received: September 21. 1998
Dear Mr. Turocy:
This letter corrects our substantially equivalent letter of November 19. 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Janine M. Morri Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
page 1 of 1
510(K) Number (if known): K98 20/0
Device Name: FACTS
Indications for Use:
The FACTS is intended to produce routine cross-sectional images of the body by The FACTS is Thended to prouded found the same axial plane taxen computer reconstraction of x ray transmoody. In addition, this device's fluoroscopic mode is intended to be used in CT interventional procedures and allows some minimally invasive surgical procedures to be implemented with the CT system. millimally invasive "Surgion" provinment include tumor biopsies, abscess drainage's, Frocedures possible with this equipment eveluation and catheter placement for organ assessment.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
David A. Seyssm
(Division Sign-Off)
Division of Reproductive, Abdominal, El and Radiological Dev 510(k) Number
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-The Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.