K981959 · Philips Medical Systems (Cleveland), Inc. · MOS · Jul 31, 1998 · Radiology
Device Facts
Record ID
K981959
Device Name
OUTLOOK LARGE PHASED ARRAY NECK COIL
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
MOS · Radiology
Decision Date
Jul 31, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Picker Outlook system is intended for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and 3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis. The Large Phased Array Neck Coil is indicated for use in providing coverage of the neck, from the cerebellum to the upper thoracic spine region.
Device Story
Large Phased Array Neck Coil; receive-only MRI accessory for Picker 0.23T Outlook system. Captures hydrogen proton signals from neck anatomy (cerebellum to upper thoracic spine). Two-channel phased array design; optimizes cervical spine imaging. Operated by trained clinicians in clinical MRI settings. Device does not transmit RF power; decoupling circuitry isolates elements during transmission. Output consists of raw signal data processed by the host MRI system into diagnostic images. Benefits include improved soft tissue visualization for physician diagnosis. Safety features include non-conductive enclosure and automatic tuning.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing and comparison of technical specifications, materials, and operational parameters with predicate devices.
Technological Characteristics
Two-channel receive-only phased array coil; optimized for 0.23T field strength. Enclosure material: Royalite R59 ABS/PVC alloy. Features switching diode decoupling and automatic patient tuning. Receive-only operation; no RF transmission. Compatible with Picker 0.23T Outlook system.
Indications for Use
Indicated for imaging the neck region, specifically from the cerebellum to the upper thoracic spine, in patients requiring MRI diagnostics.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
JUL 3 / 1998
K981959
# SUMMARY OF SAFETY AND EFFECTIVENESS (As required by 21 CFR 807.92)
#### General Information 1.
| Classification: | Class II<br>Magnetic Resonance Imaging (MRI) Accessory |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Magnetic Resonance Imaging (MRI) Coil |
| Proprietary Name: | Outlook Large Phased Array Neck Coil |
| Establishment Registration: | Picker International, Inc.<br>World Headquarters<br>595 Miner Road<br>Highland Heights, Ohio 44143<br>FDA Owner Number: #1580240<br>FDA Registration Number: #1525965 |
| Performance Standards. | Not Applicable |
#### Intended Uses 2.
The Picker Outlook system is intended for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and 3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The Large Phased Array Neck Coil is indicated for use in providing coverage of the neck, from the cerebellum to the upper thoracic spine region.
| Anatomic Regions: | Neck |
|-------------------|----------|
| Nuclei Excited: | Hydrogen |
#### 3. Device Description
The Large Phased Array Neck Coil is a receive-only coil for the 0.23T Outlook system, and is optimized to image the cervical spine. The sensitive region of the coil covers the base of the cerebellum to the third thoracic vertebra for greater than 98% of patients.
| PICKER INTERNATIONAL, INC. | (LNECK) | 5/22/98 | I - |
|----------------------------|---------|---------|-----|
|----------------------------|---------|---------|-----|
{1}------------------------------------------------
#### Safety and Effectiveness 4.
୍କ ଦି
The Picker Large Phased Array Neck Coil is substantially equivalent in safety and effectiveness to the USA Instruments (USAI) Quadrature Volume Neck Coil and the Picker Phased Array Neck Coil. The following chart has been compiled to demonstrate the substantial equivalence of these devices.
| Parameter | Large Phased Array<br>Neck Coil | Predicate Device<br>Picker - Outlook Neck Coil (K964678)<br>USAI - Profile 7000 Neck Coil (K964531) |
|--------------------------------|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Compatibility | Picker 0.23T Outlook system | Hitachi 1.5T MRH-1500 system<br>(See K964531) |
| Enclosure Material | Same. | Royalite R59 ABS/PVC alloy<br>(See K964531) |
| Coil Design | Two-channel receive-only<br>phased array coil. | Two-channel receive-only quadrature coil<br>(See K964531). |
| Decoupling | Same. | Switching Diode decoupling.<br>(See K964531) |
| Formation of Resonant<br>Loops | Same. | Length of cable and stiffness does not<br>permit looping (See K964531 or K964678) |
| Potential for RF burns | Same. | Limited by the following:<br>1) Does not transmit RF power<br>2) Coil elements and circuitry encased in<br>non-conductive material.<br>3) Decoupling isolates the coil elements<br>from transmitted RF.<br>(See K964531 or K964678) |
| Radio Frequency<br>Absorption | Same. | Coil is receive-only and does not transmit<br>RF power. Power deposition during<br>imaging is limited by the system SAR<br>algorithm. (See K964531 or K964678) |
| Tuning Scheme | Same. | Automatically tuned for each patient.<br>(See K964678) |
(LNECK)
5/22/98
I - 2
{2}------------------------------------------------
| Parameter | Large Phased Array<br>Neck Coil | Predicate Device<br>Picker - Outlook Neck Coil (K964678)<br>USAI - Profile 7000 Neck Coil (K964531) |
|---------------------|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Same. | The Phased Array Neck Coil is indicated for<br>use in providing coverage of the neck, from<br>the cerebellum to the upper thoracic spine<br>region.<br><br>Anatomic Regions: Neck<br>Nuclei Excited: Hydrogen<br>(See K964678) |
| Intended Use | Same. | The Picker Outlook system is intended for<br>use as a NMR device that produces images<br>that: (1) correspond to the distribution of<br>protons exhibiting NMR, (2) depend upon<br>the NMR parameters (proton density, flow<br>velocity, spin-lattice relaxation time (T1),<br>and spin-spin relaxation time (T2)) and 3)<br>display the soft tissue structure of the head<br>and whole body. When interpreted by a<br>trained physician, these images yield<br>information that can be useful in the<br>determination of a diagnosis.<br>(See K964678) |
PICKER INTERNATIONAL, INC.
ﻤﺴﺘﺸﻔﻪ
(LNECK)
:
11:40 PM IST
5/22/98
I - 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three curved lines.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 31 1000
Elaine K. Keeler, Ph.D. Manager, Clinical Science Picker International, Inc. 595 Miner Road Cleveland, OH 44143
Re:
K981959 Outlook Large Phased Array Neck Coil Dated: June 2, 1998 Received: June 4, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 MOS
Dear Dr. Keeler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vito diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
William Yn
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K981959
Large Phased Array Neck Coil Device Name:
## Indications for Use:
The Picker Outlook system is intended for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and 3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The Large Phased Array Neck Coil is indicated for use in providing coverage of the neck, from the cerebellum to the upper thoracic spine region.
Anatomic Regions: Neck Nuclei Excited: Hydrogen
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
William
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981959
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.