CLEANING SOLUTION (1) MODEL NUMBER A943-795, SALT BRIDGE SOLUTION MODEL NUMBER A943-794, CALIBRATION SOLUTION, PH 7.4 MO
Applicant
Thermo Bioanalysis Corp.
Product Code
JIX · Clinical Chemistry
Decision Date
Jun 12, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The products encompassed by this request are intended for in vitro diagnostic use and are intended for use in calibrating the electrodes and cleaning / flushing the sample flow path of the equivalent ABL™ pH/Blood Gas Systems. Radiometer Medical/Copenhagen is the Original Equipment Manufacturer of the analyzers and of the predicate Reagents. The ABL™ pH/Blood Gas Analyzers perform a broad array of blood gas and co-oximetery tests. ALKO manufactures the calibration reagents for the analyzer's analyte pH, (concentration of hydrogen ions) which is measured by glass membrane electrodes. ALKO also manufactures the Cleaning, Rinse, Hypochlorite and Salt Bridge Solution. These Reagents are intended to be used with equivalent ABL™ pH/Blood Gas Analyzers. As such, ALKO products are intended to serve as direct replacements to like named products manufactured by Radiometer Medical/Copenhagen. The Calibration Solution, pH 7.4/ 3 and Calibration Solution, pH 6.8/ 4 are intended to provide calibration points for the pH electrode on the analyzer. The Salt Bridge Solution /2 provides a liquid junction through the reference electrode during calibration and measurement. The Cleaning Solution /1, Rinse Solution /5 and Hypochlorite Solution are intended for maintenance of the analyzers electrodes and sample flow path. The products encompassed are to be handled using normal laboratory precautions.
Device Story
ALKO Reagents serve as direct replacements for Radiometer Medical/Copenhagen reagents used in ABL™ pH/Blood Gas Analyzers. Input: chemical solutions (calibration buffers, salt bridge, cleaning, rinse, hypochlorite) applied to analyzer electrodes and flow paths. Operation: solutions facilitate electrode calibration (pH 7.4/3, pH 6.8/4) and maintenance of the reference electrode junction and sample path. Used in clinical laboratory settings by trained personnel. Output: analyzer-generated blood gas and co-oximetry test results. Benefit: enables continued operation of existing ABL™ analyzers using compatible, equivalent reagents; ensures accurate calibration and system cleanliness for reliable diagnostic testing.
Clinical Evidence
Bench testing only. Precision study: 3 levels of control materials measured 1 run/day for 20 days (n=60 total per analyte). Results showed CV% ranging from 0.02% to 2.31%. Method comparison: 51 blood samples analyzed on two ABL™ 500 analyzers (one with ALKO reagents, one with Radiometer reagents). Linear regression analysis yielded slopes between 0.9961 and 1.0106 and R-squared values > 0.999 for pH, pCO2, and pO2.
Technological Characteristics
In vitro diagnostic chemical solutions. Includes buffered calibration solutions (pH 7.4, 6.8), salt bridge solution, cleaning solution, rinse solution, and hypochlorite solution. Designed for use with glass membrane electrodes in ABL™ pH/Blood Gas Analyzers. No electronic or software components; purely chemical reagents.
Indications for Use
Indicated for in vitro diagnostic use in clinical laboratories to calibrate electrodes and maintain (clean/flush) the sample flow path of Radiometer ABL™ pH/Blood Gas Analyzers.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}------------------------------------------------
JUN 1 2 1998
K9811795
# 510(k) Summary for ALKO Reagents on ABL™ 50, 500, 510, 520, 600. 610 and 620 pH/ Blood Gas Analyzers
The products encompassed by this 510(k) submission are Class II (75JIX) In Vitro Diagnostic Solutions manufactured by ALKO Diagnostic Corporation, 333 Fiske Street, Holliston, MA 01746. The Reagents are intended for use on equivalent ABL™ pH/ Blood Gas Analyzers. Radiometer Medical/Copenhagen is the original equipment manufacturer (OEM) of the analyzers and of predicate reagents which are necessary for the continued operation and use of the analyzers.
Information herein will support ALKO's position for the intended use of these products to the equivalent ABL™ pH/ Blood Gas Analyzers. The ABL™ pH/Blood Gas Analyzers perform a broad array of blood gas and co-oximetery tests. ALKO manufactures the calibration reagents for the analyzer's analyte pH (concentration of hydrogen ions) which is measured by the glass membrane electrode. ALKO also manufactures the Cleaning, Rinse, Hypochlorite and Salt Bridge Solution. The ALKO Reagents are intended to serve as direct replacements to like named products manufactured by Radiometer Medical /Copenhagen. The Calibration Solutions, pH 7.4/3 and Calibration Solution, pH 6.8/4 are buffered solutions for calibration of the pH electrode. The Salt Bridge Solution /2 provides a liquid junction through the reference electrode and the sample during calibration and measurement. The Cleaning Solution /S and Hypochlorite olution are all used to clean and rinse the analyzers electrodes, and sample flow path.
- . ALKO product A943-791 (Calibration Solution, pH 7.4/ 3) is equivalent to Radiometer Medical/Copenhagen product S1565 / 943-791 (Calibration Solution, pH 7.4/ 3).
- ALKO Product A943-792 (Calibration Solution, pH 6.8/ 4), is equivalent to Radiometer . Medical/Copenhagen S1575 / 943-792 (Calibration Solution, pH 6.8/ 4).
- ALKO product A943-794 (Salt Bridge Solution /2) is equivalent to Radiometer Medical/Copenhagen product S4915 / 943 -794 (Salt Bridge Solution /2).
- . ALKO product A943-795 (Cleaning Solution /1) is equivalent to Radiometer Medical/Copenhagen product S5345 / 943 -795 (Cleaning Solution /1).
- . ALKO product A943-793 (Rinse Solution /5) is equivalent to Radiometer Medical/Copenhagen product S4901 / 943 -793 (Rinse Solution /5).
- . ALKO product A943-906 (Hypochlorite Solution) is equivalent to Radiometer Medical/Copenhagen product S5362 / 943 -906 (Hypochlorite Solution).
{1}------------------------------------------------
## Page 2 / ALKO 510(k) Summary for ABL™ Equivalent Reagents.
ALKO uses a similar composition, description and packaging design as that used by Radiometer fedical/Copenhagen in its products. Equivalence is explained in the packaging section of this submission. (C.C.) L. Shown performance equivalence of its products to Radiometer Medical/Copenhagen products in the following manner:
- Through a method comparison where results obtained on an equivalent ABL™ p/H Blood Gas O Analyzer, calibrated with ALKO products and compared with results obtained on the same analyzer calibrated with Radiometer Medical/Copenhagen products; and
- Through a precision study where ALKO products were installed on an equivalent ABL™ p/H O Blood Gas Analyzer and samples were measured one run per day for twenty days. A summary of the results of these studies follows:
## PERFORMANCE CHARACTERISTICS
#### Precision Data
Precision data were collected from the analyses of three levels of control materials measured one run per day for twenty days on an ABL™ 500 pH/pCO2/pO2 analyzer calibrated with all ALKO reagents.
| Analyte | | N | Mean | STD | CV% | Min | Max |
|---------|------------|----|-------|--------|--------|-------|-------|
| pH | Total | 60 | 7.155 | 0.0025 | 0.0344 | 7.149 | 7.159 |
| | Run to Run | 20 | 7.155 | 0.0023 | 0.0325 | 7.149 | 7.157 |
| pCO2 | Total | 60 | 73.5 | 0.8096 | 1.1022 | 72.2 | 75.8 |
| | Run to Run | 20 | 73.5 | 0.7140 | 0.9720 | 72.5 | 75.6 |
| pO2 | Total | 60 | 76.4 | 1.2870 | 1.6848 | 73.9 | 79.2 |
| | Run to Run | 20 | 76.4 | 0.9898 | 1.2951 | 74.8 | 78.7 |
## Level 1
#### Level 2
| Analyte | | N | Mean | STD | CV% | Min | Max |
|---------|------------|----|-------|--------|--------|-------|-------|
| pH | Total | 60 | 7.370 | 0.0022 | 0.0301 | 7.365 | 7.381 |
| | Run to Run | 20 | 7.370 | 0.0015 | 0.0209 | 7.367 | 7.374 |
| pCO2 | Total | 60 | 44.2 | 0.5873 | 1.3296 | 43.4 | 47.5 |
| | Run to Run | 20 | 44.2 | 0.3177 | 0.7193 | 43.8 | 45.1 |
| pO2 | Total | 60 | 102.2 | 1.8308 | 1.7906 | 98.7 | 109.9 |
| | Run to Run | 20 | 102.2 | 1.5290 | 1.4955 | 99.5 | 105.3 |
{2}------------------------------------------------
Level 3
| Analyte | | N | Mean | STD | CV% | Min | Max |
|---------|------------|----|-------|--------|--------|-------|-------|
| pH | Total | 60 | 7.576 | 0.0026 | 0.0345 | 7.571 | 7.586 |
| | Run to Run | 20 | 7.576 | 0.0020 | 0.0267 | 7.572 | 7.579 |
| pCO2 | Total | 60 | 20.3 | 0.4511 | 2.2171 | 17.4 | 20.8 |
| | Run to Run | 20 | 20.3 | 0.3096 | 1.5214 | 19.2 | 20.8 |
| pO2 | Total | 60 | 157.4 | 3.6420 | 2.3146 | 149.0 | 165.3 |
| | Run to Run | 20 | 157.4 | 3.1385 | 1.9946 | 151.0 | 163.3 |
#### Accuracy by Correlation with Radiometer ABL™ Reagents
Correlation data were obtained from 51 blood samples for pH/pCO2/pO2, measured on two ABL™ 500 analyzers, one calibrated with all ALKO reagents as compared with the other one calibrated with all Radiometer reagents. A Linear Regression Analysis was performed using the Radiometer data as the Independent X Variable and ALKO Data as the Dependent Y Variable in the equation Y = a + bX.
| Analyte | N | Slope | Intercept | R Squared * | Range |
|---------|----|--------|-----------|-------------|---------------|
| pH | 51 | 1.0068 | -0.04888 | 0.9998 | 6.816 - 7.987 |
| CO2 | 51 | 1.0106 | -0.14954 | 0.9995 | 5.1- 152.7 |
| pO2 | 51 | 0.9961 | 0.46989 | 0.9995 | 12.6 - 285.0 |
*R Squared= Correlation Coefficient Squared
I hope you find this information useful and informative.
Jack C. Smith
Janet A, McGrath Regulatory Affairs
## 5|19|98
(date prepared)
(date prepared)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 1 2 1998
Janet A. McGrath · Requlatory Affairs Specialist Thermo BioAnalysis 333 Fiske Street Holliston, Massachusetts 01746
Re : K981795 Calibration Solutions pH 7.4, pH 6.8 and Cleaning, Salt Bridge, Rinse and Hypochlorite Solutions Regulatory Class: II Product Code: JIX, CHL May 19, 1998 Dated: Received: May 21, 1998
Dear Ms. McGrath:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸੈ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
#### 510(k) Number (if known):
## Device Name: Reagents on Equivalent ABL™ pH/Blood Gas Analyzers
#### Indication For Use:
The products encompassed by this request are intended for in vitro diagnostic use and are intended for use in calibrating the electrodes and cleaning / flushing the sample flow path of the equivalent ABL™ pH/Blood Gas Systems. Radiometer Medical/Copenhagen is the Original Equipment Manufacturer of the analyzers and of the predicate Reagents. The ABL™ pH/Blood Gas Analyzers perform a broad array of blood gas and co-oximetery tests. ALKO manufactures the calibration reagents for the analyzer's analyte pH, (concentration of hydrogen ions) which is measured by glass membrane electrodes. ALKO also manufactures the Cleaning, Rinse, Hypochlorite and Salt Bridge Solution. These Reagents are intended to be used with equivalent ABL™ pH/Blood Gas Analyzers. As such, ALKO products are intended to serve as direct replacements to like named products manufactured by Radiometer Medical/Copenhagen.
The Calibration Solution, pH 7.4/ 3 and Calibration Solution, pH 6.8/ 4 are intended to provide calibration points for the pH electrode on the analyzer. The Salt Bridge Solution /2 provides a liquid junction through the reference electrode during calibration and measurement. The Cleaning Solution /1, Rinse Solution /5 and Hypochlorite Solution are intended for maintenance of the analyzers electrodes and sample flow path. The products encompassed are to be handled using normal laboratory precautions.
( PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory diag
510(k) Number. 282775
Prescription Use *✓*
(Per 21 CFR 801.109)
OR
Over-The-Counter Use *_*_
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.