← Product Code [CDO](/productcode/CDO) · K981766

# URIC ACID (K981766)

_Abbott Laboratories · CDO · Jun 19, 1998 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981766

## Device Facts

- **Applicant:** Abbott Laboratories
- **Product Code:** [CDO](/productcode/CDO.md)
- **Decision Date:** Jun 19, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1775
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Uric Acid assay is used for the quantitation of uric acid in serum, plasma, or urine. Measurements of uric acid are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.

## Device Story

In vitro diagnostic assay for quantitative determination of uric acid in human serum, plasma, or urine. Principle: uricase oxidizes uric acid to allantoin, producing hydrogen peroxide; peroxide reacts with 4-aminoantipyrine and TBHB via peroxidase to form a quinoneimine dye. Change in absorbance at 548 nm is measured to determine uric acid concentration. Used in clinical laboratory settings; operated by laboratory technicians. Output provides quantitative uric acid levels to healthcare providers to assist in diagnosing and monitoring renal and metabolic conditions.

## Clinical Evidence

Bench testing only. Method comparison studies performed on the AEROSET System against the predicate (Hitachi 717 Analyzer). Serum correlation: r=0.9990, slope=1.074, intercept=0.134 mg/dL. Urine correlation: r=0.9973, slope=0.999, intercept=0.184 mg/dL. Precision: Serum total %CV 1.3-2.3%; Urine total %CV 1.8-3.1%. Linearity: up to 33.1 mg/dL (serum) and 433.8 mg/dL (urine). Limit of quantitation: 0.2 mg/dL.

## Technological Characteristics

In vitro diagnostic clinical chemistry assay. Principle: Uricase/peroxidase enzymatic reaction producing a quinoneimine dye measured via spectrophotometry at 548 nm. Designed for use on the AEROSET System. No specific materials of construction or software algorithm class provided.

## Regulatory Identification

A uric acid test system is a device intended to measure uric acid in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.

## Predicate Devices

- Boehringer Mannheim Uric Acid assay ([K873363](/device/K873363.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K981766

AUN 1 9 1998

# 510(k) Summary

#### Submitter's Name/Address

Abbott Laboratories 1920 Hurd Drive Irving, Texas 75038

## Contact Person Mark Littlefield Section Manager MS 1-8 Regulatory Affairs (972) 518-6062 Fax (972) 753-3367

| Date of Preparation of this Summary:             | May 18, 1998  |
|--------------------------------------------------|---------------|
| Device Trade or Proprietary Name:                | Uric Acid     |
| Device Common/Usual Name or Classification Name: | Uric Acid     |
| Classification Number/Class:                     | 75CDO/Class I |

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is:

#### Test Description:

Uric Acid is an in vitro diagnostic assay for the quantitative determination of uric acid in human serum, plasma, or urine. The Uric Acid assay is a clinical chemistry assay in which uric acid is oxidized to allantoin by uricase with the production of hydrogen peroxide. The peroxide reacts with 4-aminoantipyrine and TBHB in the presence of peroxidase to vield a quinoneimine dye. The resulting change in absorbance at 548 nm is proportional to the uric acid concentration in the sample.

Uric Acid 510(k) May 18, 1998 Unc Acid twp

Section II Page 1

UNDUM

{1}------------------------------------------------

#### Substantial Equivalence:

The Uric Acid assay is substantially equivalent to the Boehringer Mannheim® Uric Acid assay (K873363) on the Hitachi® 717 Analyzer for the serum and urine application.

These assays vield similar Performance Characteristics.

Similarities:

- . Both assays are in vitro clinical chemistry methods.
- Both assays can be used for the quantitative determination of uric acid. .
- Both assays vield similar clinical results. .

Differences:

- . There is a minor difference between the assay ranges.
#### Intended Use:

The Uric Acid assay is used for the quantitation of uric acid in human serum, plasma, or urine.

#### Performance Characteristics:

Comparative performance studies were conducted using the AEROSET™ System. The Uric Acid assay method comparison vielded acceptable correlation with the Boehringer Mannheim Uric Acid assay on the Hitachi 717 Analyzer. The correlation coefficient = 0.9990, slope = 1.074 and the Y-intercept = 0.134 mg/dL for serum. For the urine application, the correlation coefficient = 0.9973, slope = 0.999, and the Y-intercept = 0.184 mg/dL. Precision studies were conducted using the Uric Acid assay. Within-run, between-run, and between-day studies were performed using two levels of control material. For the serum application, the total %CV for Level 1/Panel 101 is 2.3% and Level 2/Panel 102 is 1.3%. For the urine application, the total %CV for Level 1/Panel 201 is 3.1% and Level 2/Panel 202 is 1.8%. The Uric Acid assay is linear up to 33.1 mg/dL for the serum application and 433.8 mg/dL for urine. The

Section II Page 2

# 0000012

{2}------------------------------------------------

limit of quantitation (sensitivity) of the Uric Acid assay is 0.2 mg/dL. These data demonstrate that the performance of the Uric Acid assay is substantially equivalent to the performance of the Boehringer Mannheim Uric Acid assay on the Hitachi 717 Analyzer.

### Conclusion:

The Uric Acid assay is substantially equivalent to the Boehringer Mannheim Uric Acid assay on the Hitachi 717Analyzer as demonstrated by results obtained in the studies.

Uric Acid 510(k) May 18, 1998

Uric Acid Iwp

Section II Page 3

-----

0000013

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird design, composed of three curved lines that suggest the shape of a bird in flight. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

This I g 1998

Mark Littlefield Section Manager, Regulatory Affairs Abbott Laboratories 1920 Hurd Drive 75038 Irvinq, Texas

K981766 Re: Uric Acid Requlatory Class: I Product Code: CDO Dated: May 18, 1998 May 19, 1998 Received:

Dear Mr. Littlefield:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set -forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Gutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):

Device Name: Uric Acid

Indications For Use:

The Uric Acid assay is used for the quantitation of uric acid in serum, plasma, or urine. Measurements of uric acid are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.

(Division Sign-Off)
Division of Clinical, Laboratory Devices
510(k) Number k981766

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use Over-The-Counter Use OR (Per 21 CFR 801.109)

(Optional Format 1-2-96)

00000000

Uric Acid 510(k) May 18, 1998 Uric Acid Iwp

---

**Source:** [https://fda-staging.innolitics.com/device/K981766](https://fda-staging.innolitics.com/device/K981766)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
