← Product Code [KHE](/productcode/KHE) · K981661

# HEMAPROMPT (K981661)

_Aerscher Diagnostics · KHE · Sep 15, 1998 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981661

## Device Facts

- **Applicant:** Aerscher Diagnostics
- **Product Code:** [KHE](/productcode/KHE.md)
- **Decision Date:** Sep 15, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.6550
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

HemaPrompt is a guaiac-based in-vitro slide method for the qualitative detection of occult blood in feces by lay persons. It is a useful aid in the diagnosis or early detection of a number of gastrointestinal disorders, and is recommended for use in monitoring for recurrences of previously treated bowel conditions, and in screening for colorectal cancer or for other sources of gastro-intestinal bleeding.

## Device Story

HemaPrompt is a guaiac-based fecal occult blood test (FOBT) designed for home use. Input: stool specimen applied to test slide. Operation: guaiac-based chemical reaction causes color change on slide indicating presence of blood; developing solution is pre-contained in each slide to eliminate measurement/handling hazards. Output: visual color change interpreted by patient. Context: home setting; patient self-use. Healthcare provider receives results from patient; positive results trigger further clinical investigation. Benefit: increased patient compliance due to immediacy of results; early detection of gastrointestinal disorders/cancer; reduced cost. Safety: built-in control monitor ensures proper function.

## Clinical Evidence

Performance study involving 250 lay persons (age 50+, diverse demographics) and 50 nurses. Participants performed the test and interpreted results using provided instructions. Results showed no statistically significant variation in ability to perform the test or interpret results between lay persons and professionals. Bench testing confirmed chemical equivalence to professional-use systems.

## Technological Characteristics

Guaiac-based chemical test slide. Features individually contained developing solution within each slide to eliminate handling of hazardous reagents. Includes built-in control monitor for functional verification. Designed for over-the-counter (OTC) home use with simplified instructions (SMOG 5-6 grade level) and visual illustrations.

## Regulatory Identification

An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)

## Special Controls

*Classification.* Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(K) SUMMARY

# K 981661

## Preamble

HemaPrompt is presently approved and being used by health care professionals for the screening of stool for occult blood. It is a guaiac based system using a color change in the test slide to show the presence of a pathological amount of blood in the stool, which indicates the need for further investigation by medical personnel.

Screening of stool for occult blood has been recommended by the American Cancer Society and accepted by a number of medical societies as a worthwhile tool to detect a number of gastro-intestinal diseases, with occult blood often present in such conditions as ulceration, diverticulitis and cancer. It should be carried out on a yearly basis as a screening test for those over fifty, and for monitoring those with a family history of bowel disease, those on chemotherapy, or those with post bowel surgery for cancer.

At the present time the majority of such tests require the patient to apply a stool specimen to a test card which then is returned to the physician, usually by mail in a special envelope, for development and interpretation. It has been estimated however that compliance is poor ( less than 20% of test cards being returned to the physician by some estimates) possibly because of the lack of immediacy of results and the patients' unwillingness to send such material by mail.

# Risk/ Benefit Considerations.

## Benefits

The clinical benefit of early detection provided by HemaPrompt is a decrease in the death rate from complications caused by delayed diagnosis of bowel problems causing bleeding. By having the test performed at home one would expect increased compliance and decreased cost to the patient. The physician would still have to be notified of the results.

### Risks

Since the test is typically used in asymptomatic subjects, and assuming normal co-operation of the patient in notifying the physician, delays in seeking treatment are no greater than presently encountered. There is a recognized false positive and false negative rate which is no different from the professionally used test.

## Performance Considerations.

HemaPrompt for use at home is the same in configuration and chemical constitution as that currently used by professionals. HemaPrompt is much simpler and safer to use than similar home-use kits because the developing solution is individually contained in each slide thereby obviating the need to measure or handle a potentially hazardous solution. This design also obviates variations in user technique, and because of a "built - in" control monitor, proper function is obvious.

Utilizing the home - use kit and written instructions, the rate of a false reading of test results by the patient was similar to that obtained by professionals. This was demonstrated with a study involving a group of 250 lay persons (at different locations, over age 50, both genders, varying race and educational background) who were asked, after reading instructions, to perform the test and interpret the results. A group of 50 nurses who perform fecal occult blood tests routinely in a hospital was given the same exercise with minimal instruction. The results from the two groups showed no statistically significant variation in ability to perform the test and interpret results.

## Labeling Considerations.

Each test kit (of three slides) has adequate directions for home-use purposes employing concise wording (SMOG at the 5 - 6 grade level), liberal use of step drawings, and a variety of colored illustrations demonstrating examples of positive results. A questionnaire given to the 50 lay persons who had read these instructions, demonstrated a greater than 95% understandability rate on the most important question: "Should you see a physician if your test is positive ?" Several other questions were more difficult to understand and the test instructions were revised accordinaly.

## Conclusion

Lay persons are able to adequately perform a Fecal Occult Blood test in the home setting when used in the format provided by the HemaPrompt system.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the symbol.

SEP 1 5 1998

Robert Schreiber, M.D. AERSCHER, Inc. 353 High Street Chestertown, Maryland 21620

ку81661/51 Re: Trade Name: HemaPrompt Regulatory Class: II Product Code: KHE Dated: June 23, 1998 Received: June 26, 1998

Dear Dr. Schreiber:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

{2}------------------------------------------------

## Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

# Enclosure

{3}------------------------------------------------

510(k) Number.....

Pending

HemaPrompt Device Name.......

Indications For Use

HemaPrompt is a guaiac-based in-vitro slide method for the qualitative detection of occult blood in feces by lay persons.

It is a useful aid in the diagnosis or early detection of a number of gastrointestinal disorders, and is recommended for use in monitoring for recurrences of previously treated bowel conditions, and in screening for colorectal cancer or for other sources of gastro-intestinal bleeding.

For Over-The-Counter Use

Peter E. Mayron

Division Si

---

**Source:** [https://fda-staging.innolitics.com/device/K981661](https://fda-staging.innolitics.com/device/K981661)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
