K981651 · Biostar, Inc. · PSZ · Nov 25, 1998 · Microbiology
Device Facts
Record ID
K981651
Device Name
AB FLU OIA
Applicant
Biostar, Inc.
Product Code
PSZ · Microbiology
Decision Date
Nov 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3328
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BioStar® FLU OIA assay is an Optical ImmunoAssay (OIA) test for the qualitative, rapid detection of influenza A and B viral antigen (nucleoprotein) from nasal aspirate, nasopharyngeal swabs, throat swabs, or sputum specimens. This test is intended for in vitro use to aid in influenza A and B viral infections. The FLU OlA is not intended to detect influenza C. Negative test results should be confirmed by culture [isolation].
Device Story
The BioStar AB FLU OIA is an in vitro diagnostic test for rapid detection of influenza A and B viral nucleoprotein. The device utilizes Optical ImmunoAssay (OIA) technology, which detects physical changes in the optical thickness of molecular thin films on a silicon wafer. Extracted specimen is applied to the surface where immobilized antibodies capture the target antigen; a substrate is added to increase film thickness (mass enhancement). This change alters the reflected light path, producing a visible color change (purple spot on gold background) interpreted by the clinician. The test is performed using a kit containing reagents, extraction tubes, and swabs. It provides qualitative results to aid in clinical diagnosis of influenza infections. Results are intended to be confirmed by cell culture. The device is designed for use in clinical settings to assist healthcare providers in managing patients with influenza-like symptoms.
Clinical Evidence
Clinical performance was evaluated in a three-site study of 184 symptomatic patients of all ages. The study compared the AB FLU OIA assay against 14-day cell culture. Sensitivity and specificity varied by specimen type: Nasal Aspirate (88.4% sensitivity, 69.4% specificity), Nasopharyngeal Swab (83.3% sensitivity, 76.2% specificity), Throat Swab (62.1% sensitivity, 79.5% specificity), and Sputum (81.1% sensitivity, 51.5% specificity). Analytical testing confirmed no cross-reactivity with a panel of bacterial and viral organisms and no interference from common substances like blood or mouthwash.
Technological Characteristics
The device utilizes Optical ImmunoAssay (OIA) technology based on the detection of physical changes in the optical thickness of molecular thin films on a silicon wafer. It is a manual, qualitative diagnostic test. The kit includes specimen extraction reagents, conjugate, wash, substrate, and controls. It does not involve electronic components, software, or external energy sources. The detection principle relies on antigen-antibody binding causing a visible color change on the optical surface.
Indications for Use
Indicated for qualitative, rapid detection of influenza A and B viral nucleoprotein antigen in patients of any age or sex exhibiting influenza-like symptoms. Specimens include nasal aspirate, nasopharyngeal swab, throat swab, or sputum. Not indicated for influenza C detection. Negative results require confirmation by cell culture.
Regulatory Classification
Identification
An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
{0}------------------------------------------------
# NOV 2 5 1998
1981651
Response to 9/11/98 Review Summary: BioStar®, Inc. AB FLU OIA Test Kit - K981651
8.0 510(k) SUMMARY (page 1 of 5)
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
A. Safety and effectiveness information required per [§807.92(a)(1)]:
- SUBMITTER'S NAME: BioStar, Inc. .
- ADDRESS: 6655 Lookout Rd. Boulder, CO 80301 ●
- TELEPHONE: (303) 530-3888 ext. 603 ●
- FAX: (303) 530-6601 .
- CONTACT PERSON: Roger C. Briden .
- DATE 510(k) SUMMARY PREPARED: September 18, 1998 .
- B. Safety and effectiveness information required per [§807.92(a)(2)]:
- TRADE OR PROPRIETARY NAME: AB FLU OIA® ●
- COMMON NAME: Influenza Assay
- . CLASSIFICATION NAME: Direct Antigen Detection Assay, Influenza Virus A, B
C. Identification of legally marketed device to which we are claiming equivalence [§807.92(a)(3)]:
- TRADE OR PROPRIETARY NAME: Directigen® Flu A .
- . REGULATORY CLASS: Class I
- PRODUCT CODE: 83GNT .
- MANUFACTURER: Becton Dickinson® and Co. .
- . 510(k) NUMBER: K902609
Note: Performance of the AB FLU OIA product was established versus cell culture.
#### D. Description of device [§807.92(a)(4)}:
#### Principle of the Test:
The AB FLU OIA test involves the extraction and detection of a protein antigen unique to influenza A or B (nucleoprotein). The Opical ImmunoAssay technology enables the direction of a physical change in the optical thickness of molecular thin films. This change is a result of antigen-antibody binding on an optical surface (silicon wafer). When extracted specimen is placed directly on the optical surface, the immobilized specific antibodies capture the antigen. After washing, the substrate is added, increasing the thickness (Mass Enhancement) of the molecular thin film. This change in thickness alters the reflected light path and is visually perceived as a color change. Slight changes in optical thickness produce a distinct visible color change. A positive result appears as a purple spot on the predominant gold background. When antigen is not present in the specimen, no binding takes place. Therefore, the optical thickness remains unchanged and the original gold indicating a negative result
{1}------------------------------------------------
## 8.0 510(k) SUMMARY (page 2 of 5)
#### Device Components:
The AB FLU OIA test kit contains the following:
- Sample Diluent Reagent 1: Specimen Extraction Reagent Reagent 2: Conjugate Wash Substrate Positive Control Negative Control Extraction Tubes Test Devices Transfer Pipettes Sterile Rayon Swabs
## E. Intended use of device [§807.92(a)(5)]:
The BioStar® AB FLU OIA® assay is an Optical ImmunoAssay (OIA) test for the qualitative, rapid detection of influenza A and B viral antigen (nucleoprotein) from nasal aspirate, nasopharyngeal swab, throat swab, or sputum specimens. This test is intended for in vitro diagnostic use to aid in the diagnosis of influenza A and B viral infections. The AB FLU OIA test is not intended to detect influenza C. Negative test results should be confirmed by cell culture.
#### F. Technological characteristics [§807.92(a)(6)]:
| Technological Characteristic | Predicate Device (Directigen Flu A) | Our Device (AB FLU OIA) |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Directigen Flu A Test is an in vitro<br>enzyme immunoassay (EIA) membrane test<br>for the direct rapid and qualitative detection<br>of influenza A viral antigen from suitable<br>specimens of symptomatic patients.<br>Nasopharyngeal wash and aspirate<br>specimens have been shown to be superior to<br>nasopharyngeal and throat swab specimens<br>and are the specimens of choice with the<br>Directigen Flu A Test. | The BioStar® AB FLU OIA® assay is<br>an Optical ImmunoAssay (OIA) test<br>for the qualitative, rapid detection of<br>influenza A and B viral antigen<br>(nucleoprotein) from nasal aspirate,<br>nasopharyngeal swab, throat swab, or<br>sputum specimens. This test is<br>intended for <i>in vitro</i> diagnostic use to<br>aid in the diagnosis of influenza A and<br>B viral infections. The AB FLU OIA<br>test is not intended to detect influenza<br>C. Negative test results should be<br>confirmed by cell culture. |
| Detection | Detects Influenza A nucleoproteins. | Detects Influenza A and B<br>nucleoproteins. |
| Technology | Enzyme Immunoassay (EIA) Membrane | Optical Immunoassay (OIA) |
| Specimens Evaluated | Nasopharyngeal wash, nasopharyngeal<br>aspirate, nasopharyngeal swab, throat swab. | Sputum, nasal aspirate,<br>nasopharyngeal swab, throat swab. |
{2}------------------------------------------------
#### 8.0 510(k) SUMMARY (page 3 of 5)
#### G. Summary of nonclinical testing [§807.92(b)(1)]:
#### Analytical Sensitivity:
The analytical sensitivity was evaluated using 11 influenza strains: 7 influenza B strains. Duplicate dilutions of each viral strain were prepared in Phosphate Buffered Saline (PBS) and tested according to the AB FLU OIA Test Procedure. Viral detection limit tested in the AB FLU OIA test (TCID //test) was determined by inoculating dilutions in cell culture to estimate TCID
| Viral Strain | Viral Type | Detection Limit |
|-----------------------|------------|-----------------|
| Hong Kong/68 H3N2 | A | $2.6 x 10^3$ |
| Shangdong/9/93 H3N2 | A | $1.2 x 10^4$ |
| Texas/36/91 H1N1 | A | $7.1 x 10^3$ |
| Wuhan/359/95 H3N2 | A | $1.3 x 10^3$ |
| Bayern/7/95 H1N1 | A | $2.3 x 10^3$ |
| Singapore/1/57 H2N2 | A | $5.0 x 10^2$ |
| Hong Kong/156/97 H5N1 | A | $5.3 x 10^3$ |
| Panama/45/90 | B | $1.3 x 10^4$ |
| Beijing/184/93 | B | $9.4 x 10^2$ |
| Guangdong/5/94 | B | $6.0 x 10^3$ |
| Victoria/2/87 | B | $1.2 x 10^3$ |
#### Reactivity:
To demonstrate reactivity across a broad spectrum of influenza strains, the AB FLU OIA test was performed on a number of additional human and non-human strains.
> A/NWS/33 (H1N1) A/Swine/1976/31 A/New Jersey/8/76 (HswN1) A/Japan/170/62 (H2N2) A/Puerto Rico/8/34 (H1N1) A2/Taiwan/1/64 (H2N2) A/Port Chalmers/1/73 (H3N2) A/Victoria/3/75 (H3N2) A/Duck/Ukraine/1/63 (H3N8) A/Duck/Czechoslovakia/56 (H4N6) A/Chicken/Hong Kong/1203/97 (H5N1) A/Duck/Pennsylvania/10218/84 (H5N2) A/Shearwater/Australia/1/72 (H6N5) A/Equine 1/Prague/1/56 (H7N7) A/Turkey/Oregon/1/74 (H7N3) A/Turkey/Ontario/6118/68 (H8N4) A/Chick/Germany/N/48 (H10N7)
{3}------------------------------------------------
#### 8.0 510(k) SUMMARY (page 4 of 5)
A/Duck/England/56 (H11N6) A/Duck/Alberta/60/76 (H12N6) A/Gull/Maryland/704/77 (H13N6)
#### Analytical Specificity (Cross Reactivity):
To determine the analytical specificity of the AB FLU OIA test, the organisms and viruses listed in the bacterial and viral panels presented in the product package insert were grown in culture and tested at appropriate concentrations. A sample of each organism and virus was tested in duplicate. Cell density or viral titer was confirmed by plating an aliquot of suspension, growing the organism or virus, and counting the number of colonies or plaques formed. None of the organisms or viruses listed in the panels gave a positive result in the AB FLU OIA test.
#### Interfering Substances:
A variety of substances were obtained which interfere with the effectiveness of the AB FLU OIA test if encountered during the collection of specimens. Each substance was prepared in a manner and concentration so as to exceed the likely conditions and amounts found in situ. Each substance was tested in three independent analyses: In the absence of influenza virus, in the presence of only Influenza A (A/Hong Kong), and in the presence of only Influenza B (B/Panama). In addition, a saline control was tested in the same manner. All assays were run per the testing procedure described in the package insert. None of the specimens tested gave a positive signal when virus was not present in the sample. No decrease in signal strength was noted for either influenza A or B in the presence of either blood, or several commercially available mouth washes, throat sprays, cough lozenges, or cough/cold elixirs.
#### H. Summary of clinical testing [§807.92(b)(2)]:
The performance of the AB FLU OIA assay was compared to conventional methods for influenza culture in an cvaluation of clinical specimens. In a three-site study to compare the AB FLU OIA assay with culture, any combination of dual throat and dual nasopharyngeal swabs, nasal aspirate and sputum specimens were collected from 184 individuals exhibiting influenza-like symptoms. Patients of any age or sex were included in this study.
The influenza prevalence rate by site, as determined by confirmed culture from any speciment type, was 56.1% at site I (Rocky Mountain region), 56.9% at site 2 (Midwest region), and 41.0% at site 3 (Southwestern region). The table below summarizes the sensitivity and specificity of the AB FLU OIA test relative to 14-day culture for the four specimen types collected during the clinical study. The four specimen types identified in the table are nasal aspirate (NA), nasopharyngeal swab (NPS), throat swab (TS), and sputum (SP). For additional information regarding data collection and analysis, see the product package insert.
| D. Specimen Type | E. Relative Sensitivity | F. Relative Specificity |
|------------------|-------------------------|-------------------------|
| NA | 88.4 % | 69.4% |
| NPS | 83.3% | 76.2% |
| TS | 62.1% | 79.5% |
| SP | 81.1% | 51.5% |
1. Conclusions from nonclinical / clinical testing [§807.92(b)(3)].
In evaluating the cross reactivity of the AB FLU OIA test with organisms and viruses listed in the bacterial and viral panels presented in the product package insert, none of the organisms or viruses gave a positive result in the AB FLU OIA test. At levels exceeding those likely be encountered in situ during sample collection, neither blood, nor several commercially available mouthwashes, throat sprays, cough lozenges, or cough/cold elixirs interfered with the test. These results combined with the analytical sensitivity, along with the relative sensitivity and specificity calculated from the clinical study data demonstrate that the AB FLU OIA test is safe and effective.
{4}------------------------------------------------
Response to 9/11/98 Review Summary: BioStar®, Inc. AB FLU OIA Test Kit - K981651
## 8.0 510(k) SUMMARY (page 4 of 5)
- J. Additional information [§807.92(d)]:
:
No additional information has been requested by FDA at this time.
,
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem consists of three stylized human profiles facing right, with flowing lines beneath them, resembling a ribbon or fabric.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 25 1998
Roger Briden, Ph.D. Director, RA/OA BioStar, Inc. 6655 Lookout Road Boulder, CO 80301
Re: K981651 Trade Name: BioStar® AB FLU OIA Regulatory Class: I Product Code: GNX Dated: September 18, 1998 Received: September 22, 1998
Dear Dr. Briden:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
## Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K981651
Device Name: BioStar®, Inc. AB FLU OIA
Indications For Use: The BioStar® FLU OIA assay is an Optical ImmunoAssay (OIA) test for the qualitative, rapid detection of influenza A and B viral antigen (nucleoprotein) from nasal aspirate, nasopharyngeal swabs, throat swabs, or sputum specimens. This test is intended for in vitro use to aid in influenza A and B viral infections. The FLU OlA is not intended to detect influenza C. Negative test results should be confirmed by culture [isolation].
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Duboes.
sion of Clinical Laboratory Devices 510(k) Number.
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.