# STABLEMAPR STEERABLE INTRACARDIAC CATHETERS (K981642)

_Medtronic Cardiorhythm · DRF · Aug 5, 1998 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981642

## Device Facts

- **Applicant:** Medtronic Cardiorhythm
- **Product Code:** [DRF](/productcode/DRF.md)
- **Decision Date:** Aug 5, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Medtronic CardioRhythm StableMapr Catheter is intended for use in diagnostic electrophysiologic procedures. These catheters are designed for recording intracardiac electrograms and temporary pacing associated with electrophysiology studies.

## Device Story

StableMapr is a steerable, closed-lumen diagnostic electrophysiology catheter; 7 French diameter; 60-125 cm length. Features platinum alloy electrodes fixed around shaft; distal tip electrode; additional mapping electrodes. Manipulator handle allows physician to vary tip curvature radius and lateral deflection. Device connects to external stimulator/recorder equipment to capture analog intracardiac electrical signals and deliver temporary pacing. Used by physicians in clinical electrophysiology studies to assess arrhythmia prognosis, guide therapy selection, and evaluate therapeutic intervention effectiveness. Operates as a passive conduit for electrical signals; no internal processing or software.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing of electrical and mechanical performance and biocompatibility data leveraged from the predicate Marinr catheter.

## Technological Characteristics

7 French diameter; 60-125 cm length. Platinum alloy electrodes. Closed-lumen shaft. Mechanical manipulator handle for tip steering. Passive electrical signal transmission. Biocompatible materials identical to predicate.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Medtronic CardioRhythm Marinr EP catheter
- Invine Biomedical Steerable Atrial Mapping Inquire H
- Cordis Webster Halo XP Tricuspid Mapping Catheters

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Medironic &

# SECTION 6. 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

#### Device Name

StableMapr Steerable Intracardiac Electrode Catheters

### Device Description

The StableMapr Series of EP Diagnostic Catheters are used to record electrical activity from within the heart and its vasculature and to stimulate the heart in accordance with various programmed electrical stimulation protocols incorporated into an EP diagnostic study. All catheters connect physician stimulator/recorder equipment to the desired region of the heart and carry analog signals. Data from the EP study is used to assess arrhythmia patient prognosis, guide therapy selection and evaluate the effectiveness of previously selected therapeutic interventions.

The StableMapr Catheters are closed-lumen diagnostic electrophysiology catheters with a variable number of platinum allov recording/stimulating electrodes fixed around the catheter shaft. The most distal electrode is located at the catheter tip. The StableMapr Catheters include addition mapping electrodes and a preformed distal tip section, as well as other minor modifications for improved manufacturability. The electrode bands are welded to electrical wires incorporated inside the shaft of the catheter that run from each electrode to the electrical connector at the proximal end of the catheter. The outside diameter of the catheter is 7 French and the usable length ranges from 60 to 125 cm. A manipulator handle at the proximal end of the catheter permits the physician to vary the angle of curvature of the catheter tip. All models have controls in the handle to vary the radius of curvature, and the distal end of some models may be laterally deflected.

#### Intended Use

The Medtronic CardioRhythm StableMapr Catheter is intended for use in diagnostic electrophysiologic procedures. These catheters are designed for recording intracardiac electrograms and temporary pacing associated with electrophysiology studies.

#### Substantial Equivalence

The StableMapr series of EP catheters is substantially equivalent to the Medtronic CardioRhythm Marinr EP catheter, as well as the Invine Biomedical Steerable Atrial Mapping Inquire H and Cordis Webster Halo XP Tricuspid Mapping Catheters.

{1}------------------------------------------------

### Medtronic @

### Catheter Testing Results and Conclusion

The StableMapr catheter is constructed of similar materials to those found in the Medtronic CardioRhythm Marinr Catheter. The patient contacting materials utilized in the fabrication of the StableMapr are identical to those used in the Marinr catheter. Therefore, all biocompatibility testing was fulfilled via the predicate Marinr Catheter device.

The non-clinical testing were conducted in accordance with applicable FDA guidance. The tests quantified and confirmed the adequacy of electrical and mechanical performance and reliability of the Marinr catheters. These tests support the substantial equivalence of the StableMapr catheters with the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, overlaid with three human silhouettes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.

# AUG - 5 1998

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Kristen Honl Senior Product Requlation Manager Medtronic CardioRhythm 1312 Crossman Avenue Sunnyvale, CA 94089-1113

Re : K981642 StableMapr Steerable Intracardiac Electrode Catheter Regulatory Class: II (two) Product Code: DRF Dated: May 6, 1998 Received: May 8, 1998

Dear Ms. Honl:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Please note: this response to your premarket Federal Register. notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: StableMapr Steerable Intracardiac Electrode Catheters

- The Medtronic CardioRhythm StableMapr catheter is Indications For Use: intended for use in diagnostic electrophysiology procedures. These catheters are designed for recording intracardiac pacing associated with electrograms and temporary electrophysiology studies.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

y Oalk Jell
15- Offl

a: Devices

Prescription Use

OR (Per 21 CFR 801.109) Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K981642](https://fda-staging.innolitics.com/device/K981642)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com