ROCHE DIAGNOSTICS, BOEHRINGER MANNHEIM LIQUID TOTAL BILIRUBIN REAGENT
Applicant
Roche Diagnostics/Boehringer Mannheim GmbH
Product Code
CIG · Clinical Chemistry
Decision Date
Jun 15, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1110
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Roche Diagnostics, Boehringer Mannheim Liquid Total Bilirubin reagent is intended for use for the quantitative determination of total bilirubin in serum and plasma of adults and neonates. It is for use on automated clinical chemistry analyzers. According to the Code of Federal Regulations, Title 21 (Food and Drugs), Part 862.1110, a bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.
Device Story
In vitro diagnostic reagent for quantitative measurement of total bilirubin in serum/plasma; utilizes diazo reaction where bilirubin couples with diazonium ion in acidic medium (pH 1-2) to form azobilirubin; color intensity proportional to total bilirubin concentration; measured photometrically on automated clinical chemistry analyzers; used in clinical laboratory settings by trained personnel; results assist clinicians in diagnosing/treating liver, hemolytic, hematological, and metabolic disorders; provides improved operational convenience via liquid, ready-to-use reagent format compared to lyophilized predicate.
Clinical Evidence
No clinical data provided. Substantial equivalence based on bench testing and performance comparison to predicate device.
Technological Characteristics
Liquid, ready-to-use diazo reagent system; utilizes photometric measurement of azobilirubin; designed for use on automated clinical chemistry analyzers; two-point calibration (saline blank and ~2.6 mg/dL standard); 5-week on-board reagent stability.
Indications for Use
Indicated for quantitative determination of total bilirubin in serum and plasma of adults and neonates for diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.
Regulatory Classification
Identification
A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Predicate Devices
Roche Diagnostics, Boehringer Mannheim Total Bilirubin/DPD reagent system (K781921)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and consists of a stylized eagle. The U.S. Food & Drug Administration logo is on the right and consists of the FDA acronym in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
March 14, 2019
Roche Diagnostics/Boehringer Mannheim Gmbh Luann Ochs 9115 Hague Rd. Indianapolis, IN 46256
Re: k981632
Trade/Device Name: Liquid Total Bilirubin Reagent Regulation Number: 21 CFR 862.1110 Regulation Name: Bilirubin (total or direct) test system Regulatory Class: Class II Product Code: CIG, MQM Dated: May 7, 1998 Received: May 8, 1998
Dear Luann Ochs:
This letter corrects our substantially equivalent letter of June 15, 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): Device Name: Roche Diagnostics, Boehringer Mannheim Liquid Total Bilirubin Reagent
Indications for Use:
The Roche Diagnostics, Boehringer Mannheim Liquid Total Bilirubin reagent is intended for use for the quantitative determination of total bilirubin in serum and plasma of adults and neonates. It is for use on automated clinical chemistry analyzers.
According to the Code of Federal Regulations, Title 21 (Food and Drugs), Part 862.1110, a bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Clinical Laboratory Zevices
510(k) Number K981633
{3}------------------------------------------------
K981632
JUN 1 5 1998
## 510(k) Summary
| Introduction | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence. |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1) Submitter<br>name, address,<br>contact | Roche Diagnostics, Boehringer Mannheim Corporation<br>9115 Hague Rd.<br>Indianapolis, IN 46250<br>(317) 845-2000 |
| | Contact Person: Luann Ochs |
| | Date Prepared: May 7, 1998 |
| 2) Device name | Proprietary name: Roche Diagnostics, Boehringer Mannheim Liquid Total<br>Bilirubin Reagent |
| | Common name: bilirubin (total or direct) test system |
| | Classification name: Diazo colorimeter, bilirubin, 75CIG<br>Device Class II |
| 3) Predicate<br>device | We claim substantial equivalence to the currently marketed Roche Diagnostics.<br>Boehringer Mannheim Total Bilirubin/DPD reagent system, catalog number<br>1039034, a modification of the Single Vial DPD Total Bilirubin reagent<br>system, K781921. |
| 4) Device<br>Description | Total bilirubin, in the presence of a suitable solubilizing agent, is coupled with<br>a diazonium ion in a strongly acid medium (ph 1 - 2). |
| | Bilirubin + diazonium ion<br><span style="text-decoration:overline">acid</span><br>-> Azobilirubin |
| | The intensity of the color of the azobilirubin formed is proportional to the total<br>bilirubin concentration and can be measured photometrically. |
Continued on next page
{4}------------------------------------------------
### 510(k) Summary, Continued
The Roche Diagnostics, Boehringer Mannheim Liquid Total Bilirubin reagent 5) Intended use is intended for use for the quantitative determination of total bilirubin in serum and plasma of adults and neonates. It is for use on automated clinical chemistry analyzers.
#### The Roche Diagnostics, Boehringer Mannheim Liquid Total Bilirubin reagent 6) Comparison is substantially equivalent to other products in commercial distribution to predicate intended for similar use. Most notably it is substantially equivalent to the device currently marketed Roche Diagnostics, Boehringer Mannheim Total Bilirubin/DPD reagent system, catalog number 1039034, a modification of the Single Vial DPD Total Bilirubin reagent system, K781921.
The following table illustrates the similarities between the Roche Diagnostics, Boehringer Mannheim Liquid Total Bilirubin Reagent and the predicate device. Specific data on the performance of the system have been incorporated into the draft labeling in Section V of this submission. Labeling for the predicate device is provided in Section VI.
| Feature | New Liquid Total<br>Bilirubin Reagent | Predicate Total<br>Bilirubin Reagent |
|------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Intended Use | Measurement of total<br>bilirubin | Measurement of total<br>bilirubin |
| Sample Type | Serum or plasma, no<br>preparation required | Serum or plasma, no<br>preparation required |
| Use on Automated<br>Chemistry Analyzers? | Yes | Yes |
| Test Principle | Diazo reaction with<br>formation of an<br>azobilirubin product,<br>measured<br>spectrophotometrically | Diazo reaction with<br>formation of an<br>azobilirubin product,<br>measured<br>spectrophotometrically |
| Calibration | Two points, blank<br>(saline) and about 2.6<br>mg/dL total bilirubin | Two points, blank<br>(saline) and about 2.6<br>mg/dL total bilirubin |
#### Similarities:
Continued on next page
{5}------------------------------------------------
## 510(k) Summary, Continued
| Feature | New Liquid Total<br>Bilirubin Reagent | Predicate Total<br>Bilirubin Reagent |
|-------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Calibration Stability | Perform a new<br>calibration once a week,<br>or with a bottle or<br>reagent lot change | Perform a new<br>calibration once a week,<br>or with a bottle or<br>reagent lot change |
| Reagent On-board<br>Stability | 5 weeks | 5 days |
| Kit Configuration,<br>Reagent Preparation | R1, liquid, ready-to-use<br>R2, liquid, ready-to-use | R1a, lyophilized,<br>requires reconstitution<br>with buffer<br>R1, buffer |
6) Comparison to predicate device (continued)
#### 6) Comparison Differences: to predicate device, continued
There are no significant differences between the Roche Diagnostics, Boehringer Mannheim Liquid Total Bilirubin reagent and the predicate device for purposes of considering substantial equivalence.
#### Performance characteristics:
The performance of the Roche Diagnostics, Boehringer Mannheim Liquid Total Bilirubin reagent is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Roche Diagnostics, Boehringer Mannheim Total Bilirubin/DPD reagent system, catalog number 1039034, a modification of the Single Vial DPD Total Bilirubin reagent system, K781921.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.