K981604 · Careside, Inc. · KNK · Jun 25, 1998 · Clinical Chemistry
Device Facts
Record ID
K981604
Device Name
CARESIDE URIC ACID
Applicant
Careside, Inc.
Product Code
KNK · Clinical Chemistry
Decision Date
Jun 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1775
Device Class
Class 1
Indications for Use
For in vitro diagnostic use with the CARESIDE™ Analyzer to quantitatively measure uric acid from whole blood, plasma or serum specimens to aid in the diagnosis and treatment of patients with numerous renal and metabolic disorders including renal failure, gout, leukemia, psoriasis, starvation or wasting conditions, and of patients receiving cytotoxic drugs. It is intended for professional laboratory use: not for point of care or physician office laboratory use.
Device Story
CARESIDE™ Uric Acid cartridge is a single-use disposable test for quantitative uric acid measurement in whole blood, plasma, or serum. Used with CARESIDE™ Analyzer; operator inserts cartridge containing specimen into analyzer. Device spins cartridge to separate blood cells from plasma/serum via centrifugal force; meters 10µL of sample onto multi-layer reagent film. Uricase hydrolyzes uric acid to generate hydrogen peroxide; peroxidase catalyzes reaction with leuco dye to form blue dye. Reflectance measured at 655nm via LEDs and photodiodes. Analyzer calculates concentration using lot-specific standard curve. Output provides quantitative uric acid levels to clinicians for diagnosing/managing renal and metabolic disorders. Intended for professional laboratory use.
Clinical Evidence
No clinical data provided; substantial equivalence is based on analytical performance and comparison to the predicate device's established methodology.
Technological Characteristics
Single-use disposable cartridge containing multi-layer reagent film. Sensing principle: reflectance photometry at 655nm. Energy source: CARESIDE™ Analyzer (centrifugal separation and optical measurement). Connectivity: barcode scanning of cartridge. Software: automated processing of reflectance data against lot-specific standard curves.
Indications for Use
Indicated for quantitative measurement of uric acid in whole blood, plasma, or serum to aid diagnosis/treatment of renal and metabolic disorders (e.g., renal failure, gout, leukemia, psoriasis, starvation, wasting) and monitoring patients on cytotoxic drugs. Professional laboratory use only.
Regulatory Classification
Identification
A uric acid test system is a device intended to measure uric acid in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.
Predicate Devices
Vitros URIC Slides for Johnson and Johnson's Vitros DT 60 (K912844/A)
Submission Summary (Full Text)
{0}------------------------------------------------
## JUN 2 5 1998
# 510(K) SUMMARY CARESIDE™ URIC ACID SAFETY AND EFFECTIVENESS
#### Applicant Information I.
- A. Applicant Name
- Applicant/Manufacturer Address B.
- Telephone Number C.
- D. Contact Person
- FAX Number E.
- e-Mail Address F.
- Date 510(k) Summary prepared G.
#### Device Information II.
- Device Name (Trade) A.
- B. Device Name (Classification)
- Device Classification C.
### CARESIDE, Inc.
6100 Bristol Parkway Culver City, CA 90230 310-338-6767 Kenneth B. Asarch, Pharm.D., Ph.D. 310-338-6789 asarchk @ worldnet.att.net April 27, 1998
### CARESIDE™ Uric Acid
Uric acid test system Clinical chemistry panel Uric acid test system Regulation Number: 21 CFR 862.1775 Regulatory Class I Classification Number: 75JHB None applicable
- Special controls and D. performance standards
#### III. Substantial Equivalence Claim
- General equivalency claim A.
The ability to monitor analyte-specific biochemical reactions in dry film and other formats is widely recognized and has gained widespread acceptance for use in chemistry assays.
Uric acid in vitro diagnostic products, in both dry film and other formats, are already on the U.S. market, including uric acid products which utilize uricase catalyzed generation of hydrogen peroxide which reacts with chromogens in a peroxidase catalyzed reaction to form a blue dye.
#### B. Specific equivalency claim
This CARESIDE™ Uric Acid test is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros slides for the quantitative measurement of uric acid on the Vitros DT 60 II.
| Name of Predicate Device: | Johnson and Johnson's (formerly Eastman Kodak, Inc.) Vitros URIC Slides for Johnson and Johnson's Vitros DT 60 (formerly Eastman Kodak's DT 60 II). |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device 510K number: | K912844/A |
| Product Code: | 75JHB |
{1}------------------------------------------------
#### IV. Device Description
CARESIDE™ Uric Acid cartridges are used with the CARESIDE™ Analyzer to quantitatively measure uric acid concentration in anti-coagulated whole blood, plasma or serum specimens. The CARESIDE™ Uric Acid cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma or serum to a dry film to initiate the measurement of uric acid concentration. The film cartridge (patent pending) contains all reagents necessary to measure uric acid concentration.
#### A. Explanation of Device Function
Each CARESIDE™ Uric Acid cartridge consists of a uric acid-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the whole blood, serum, or plasma specimen into the cartridge sample deposition well, closes the lid and inserts the cartridge into the CARESIDE™ Analyzer. Once loaded, the CARESIDE™ analyzer scans the cartridge barcode, spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers, and brings the cartridge and the contained specimen to 37℃. As the cartridge continues to spin, the blood cells are separated from the plasma/serum and the cells accumulate in the separation well. Approximately ten microliters of plasma (or serum, as applicable) remain in the metering passage. Any excess sample flows into an overflow well. The plasma (or serum, as applicable) is automatically dispensed onto the multi-layer reagent film. The spreading layer distributes the uric acid containing specimen uniformly and removes large molecular weight components, such as protein, as the specimen moves into the underlying reaction layer. Uric acid is hydrolyzed by uricase to generate hydrogen peroxide which in turn oxidizes the leuco dye in a peroxidase catalyzed reaction. The color intensity of the resulting blue dye, as measured by the amount of reflected light at 655 nanometers, is directly related to the specimen uric acid concentration.
Test Reaction Sequence:
Uric Acid + 2 H20 + O2 - Unicase > Allantoin + H2O2+ CO2 Peroxidase H2O2 + Leuco Dye Blue Dye + 2 H20
As the cartridges spin, photodiodes measure reflectance of light emitted by wavelengthspecific light emitting diodes (LEDs) at a fixed time. The analyzer uses the reflectance measurements and the lot-specific standard curve to calculate uric acid concentration.
#### B. Test Summary
Uric acid is the major product of the catabolism of purine nucleosides, adenosine and guanosine. Purines from catabolism of dietary nucleic acid are converted to uric acid directly. Approximately 75 percent of uric acid excreted is lost in the urine. Most of the remainder is secreted into the gastrointestinal tract, where it is degraded to allantoin and other compounds by bacterial enzymes'. Hyperuricemia is caused when serum or plasma uric acid concentration levels rise above 7 mg/dL in men and 6 mg/dL in women's Disorders of hyperuricemia can be divided into those due to increased intake, decreased intake excretion, and increased production". Hyperuricemia can result from uric acid overproduction or under excretion, renal retention due to renal failure, drug or chemical toxicity, hypothyroidism, hyperparthyroidism, increased turnover of nucleic acid due to myeloproliferative syndromes or cancer chemotherapy, or specific enzyme deficiencies such as hypoxanthine-guanine phosphoribosyl transferase or phosphoribosyl pyrophosphate synthetase.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 25 1998
Kenneth B. Asarch, Ph.D. . VP Quality Systems and Regulatory Affairs CareSide Inc. 6100 Bristol Parkway Culver City, California 90230
Re: K981604 CareSide™ Uric Acid Regulatory Class: I Product Code: JHB Dated: April 27, 1998 Received: April 28, 1998
Dear Dr. Asarch:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### INDICATIONS FOR USE VI.
510(k) Number: K981604
CARESIDE™ Uric Acid Device Name:
For in vitro diagnostic use with the CARESIDE™ Analyzer to Indications for use: quantitatively measure uric acid from whole blood, plasma or serum specimens to aid in the diagnosis and treatment of patients with numerous renal and metabolic disorders including renal failure, gout, leukemia, psoriasis, starvation or wasting conditions, and of patients receiving cytotoxic drugs. It is intended for professional laboratory use: not for point of care or physician office laboratory use.
| Division Sign-Off) Division of Clinical Laboratory Devices | |
|------------------------------------------------------------|---------|
| 510(k) Number | K981604 |
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.