← Product Code [ETD](/productcode/ETD) · K981575

# TYMPANOSTOMY TUBE (K981575)

_Grace Medical, Inc. · ETD · Jul 15, 1998 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981575

## Device Facts

- **Applicant:** Grace Medical, Inc.
- **Product Code:** [ETD](/productcode/ETD.md)
- **Decision Date:** Jul 15, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.3880
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic

## Indications for Use

Indications for insertion of a tympanostomy tube into an incision in the tympanic membrane include any of the following: - . Chronic otitis media with effusion characterized as either serous, mucoid, or purulent - . Recurrent acute otitis media which fails to respond satisfactorily to alternative therapies - . A patient with a history of persistent high negative middle ear pressure which may be associated with conductive hearing loss that is symptomatic, persistant or recurrent otalgia, persistent or recurrent vertigo and/or tinnitus - . Atelectasis resultant from retraction pocket of the tympanic membrane or eustachian tube dysfunction

## Device Story

Tympanostomy tubes are small, hollow devices inserted into an incision in the tympanic membrane (eardrum). They facilitate ventilation of the middle ear space and drainage of fluid. Used by otolaryngologists in a clinical or surgical setting. The device acts as a conduit to equalize pressure and prevent fluid accumulation, addressing conditions like chronic otitis media, recurrent infections, and eustachian tube dysfunction. By maintaining middle ear aeration, the tubes help resolve conductive hearing loss and associated symptoms like otalgia, vertigo, and tinnitus, or prevent further tympanic membrane retraction.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on device design and intended use.

## Technological Characteristics

Tympanostomy tubes are passive, mechanical implants designed for insertion into the tympanic membrane. The product line includes various geometries (T-tubes, grommets, beveled, parasol/umbrella designs) with inner diameters ranging from 0.97 mm to 2.03 mm. Materials and sterilization methods are not specified in the provided text.

## Regulatory Identification

A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 5 1998

Anthony D. Prescott Grace Medical, Inc. 31 Highway 70 Suite 107 Bartlett, TN 38133

Re:

K981575 Tympanostomy Tube (see enclosure) Dated: April 30, 1998 Received: May 4, 1998 Regulatory class: II 21 CFR 874.3880/Procode: 77 ETD

Dear Mr. Prescott:

We have reviewed your Section 510%) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Fecteral Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmam.n.html".

Sincerely yours,

Kilian Yin

Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{1}------------------------------------------------

## ENCLOSURE

:

.

neammaalstop post the productions of

| Grace<br>Catalog # | Description                         | 510(k)<br>page | Inner<br>Diameter |
|--------------------|-------------------------------------|----------------|-------------------|
| 1                  | Goode T-Tube                        | 21             | 1.14              |
| 2                  | Goode T-Tube                        | 21             | 1.32              |
| 3                  | Butterfly Vent Tube                 | 23             | 1.27              |
| 4                  | Goode T-Grommet Vent Tube           | 25             | 1.25              |
| 5                  | Cohen T-Grommet Vent Tube           | 26             | 1.25              |
| 6                  | Baxter T-Grommet Vent Tube          | 27             | 1.25              |
| 7                  | Paparella Type Vent Tube            | 29             | 1.02              |
| 8                  | Paparella Type Vent Tube            | 29             | 1.27              |
| 9                  | Paparella Type Vent Tube            | 29             | 2.03              |
| 10                 | Per-Lee Vent Tube                   | 30             | 1.57              |
| 11                 | Donaldson Vent Tube                 | 32             | 1.14              |
| 12                 | Armstrong Beveled Grommet Vent Tube | 34             | 1.14              |
| 13                 | Armstrong Beveled Vent Tube         | 35             | 1.14              |
| 14                 | Parasol/Umbrella Vent Tube          | 37             | 1.02              |
| 15                 | Parasol/Umbrella Vent Tube          | 37             | 1.52              |
| 16                 | Parasol/Umbrella Vent Tube          | 37             | 2.03              |
| 17                 | Shepard Grommet Vent Tube           | 38             | 1.14              |
| 18                 | Baxter Beveled Vent Tube            | 40             | 1.27              |
| 19                 | Baxter Beveled Vent Tube            | 40             | 0.97              |

.

:

.

:

t

" →

{2}------------------------------------------------

510(k) Number (if known):

Device Name: Tympanostomy Tube

Indications For Use:

Indications for insertion of a tympanostomy tube into an incision in the tympanic membrane include any of the following:

Kax 1575

- . Chronic otitis media with effusion characterized as either serous, mucoid, or purulent
- . Recurrent acute otitis media which fails to respond satisfactorily to alternative therapies
- . A patient with a history of persistent high negative middle ear pressure which may be associated with conductive hearing loss that is symptomatic, persistant or recurrent otalgia, persistent or recurrent vertigo and/or tinnitus
- . Atelectasis resultant from retraction pocket of the tympanic membrane or eustachian tube dysfunction

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Perscription Use (Per 21 CFR 801.109) (Optional Format 1-2-96) OR

Over-The-Counter Use

David H. Seymour
(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number .

---

**Source:** [https://fda-staging.innolitics.com/device/K981575](https://fda-staging.innolitics.com/device/K981575)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
