K981528 · Siemens Medical Solutions USA, Inc. · IYO · Oct 28, 1998 · Radiology
Device Facts
Record ID
K981528
Device Name
SONOLINE ELEGRA DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYO · Radiology
Decision Date
Oct 28, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SONOLINE Elegra ultrasound imaging system is intended for the following applications: General Radiology, Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Pelvic, Neonatal/Adult Cephalic, Urology, Vascular, Musculoskeletal. Superficial Musculoskeletal, and Peripheral Vascular applications. The addition of contrast agent imaging will not add any new applications or intended uses. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Device Story
The SONOLINE Elegra is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers and displays images in B-Mode, M-Mode, Color Mode, Pulsed (PW) Doppler, and Continuous (CW) Doppler. The system includes a patented Contrast Agent Imaging (CAI) technique (US patent 5,632,277) and harmonic imaging capabilities. It processes signals to enhance imaging of tissues and structures reflecting ultrasound at harmonic frequencies. The system provides on-screen thermal and mechanical indices for bioeffect monitoring. It is operated by clinicians in various clinical settings to perform anatomical measurements and analysis for diagnostic purposes. The output is displayed on a CRT, assisting clinicians in clinical decision-making across multiple specialties.
Clinical Evidence
Bench testing only. The submission relies on technical performance data and compliance with safety standards for diagnostic ultrasound equipment, including acoustic output measurements and adherence to AIUM/NEMA standards for thermal and mechanical indices.
Technological Characteristics
Mobile diagnostic ultrasound system; supports B-mode, M-mode, PW/CW Doppler, Color/Amplitude Doppler, and harmonic imaging. Transducers include phased, curved, and linear arrays. Connectivity includes on-screen display of thermal/mechanical indices. Complies with UL 2601-1, CSA 22.2 No. 601-1, and IEC 60601-1/1-2 safety/EMC standards. Software-controlled operation.
Indications for Use
Indicated for general radiology, abdominal, intraoperative, small parts, transcranial, OB/GYN, pelvic, neonatal/adult cephalic, urology, vascular, musculoskeletal, superficial musculoskeletal, and peripheral vascular applications. Contraindicated for in vitro fertilization and percutaneous umbilical blood sampling.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Acuson Sequoia Ultrasound System and Harmonic Imaging with Contrast Option (K973767)
{0}------------------------------------------------
#### OCT 28 1998 510(K) SUMMARY
K9815528
# Addition of Contrast Agent Imaging to SONOLINE Elegra Diagnostic Ultrasound system
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR, Part 807.92, Content and Format of a 510(k) Summary. Contrast Agent Imaging (CAI) is a patented technique (US patent number 5,632,277).
#### Submitted Bv: 1.
Siemens Medical Systems. Inc., Ultrasound Group 22010 S.E. 51st Street Issaquah, WA 98027-7002
# Contact Person:
Steve Hesler Manager of Regulatory Affairs (425) 557-1629
# Date Prepared:
March 31, 1998
#### 2. Proprietary Name:
SONOLINE Elegra Advanced SONOLINE Elegra
# Common/ Usual Name:
Diagnostic Ultrasound System with Accessories
# Classification Name:
| Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 Product Code 90-IYN |
|------------------------------------------|-----------------------------------|
| Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 Product Code 90-IYO |
| Diagnostic Ultrasound Transducer | FR # 892.1570 Product Code 90-ITX |
#### 3. Predicate Device:
- . Acuson Sequoia™ Ultrasound System and Harmonic Imaging with Contrast Option (K973767, 12/23/97)
- . SONOLINE Elegra (K945072, 11/21/95).
#### 4. Device Description:
The SONOLINE Elegra is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire ultrasound data and display it in B-Mode, M-Mode, Color Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, or in a combination of modes, on a CRT display.
The SONOLINE ® Eleara, has been designed to meet the following product safety standards:
- UL 2601-1. Safety Requirements for Medical Equipment
- CSA 22.2 No. 601-1, Safety Requirements for Medical Equipment
- Standard for Real Time Display of Thermal and Mechanical Indices on Diagnostic Ultrasound Equipment, AIUM/NEMA, 1992.
{1}------------------------------------------------
.
- 93/42/EEC Medical Devices Directive EN60601 = (IEC 601-1+ IEC 601-1-2). Safety and EMC Requirements for Medical Equipment
#### 5. Intended Uses:
The SQNOLINE Elegra ultrasound imaging system is intended for the following applications: General Radiology, Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Pelvic, Neonatal/Adult Cephalic, Urology, Vascular, Musculoskeletal. Superficial Musculoskeletal, and Peripheral Vascular applications. The addition of contrast agent imaging will not add any new applications or intended uses.
The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
#### 6. Technological Comparison to Predicate Device:
Both the SONOLINE Elegra and the predicate device from Acuson are full-featured, highend diagnostic ultrasound systems capable of B-mode, M-mode, CW Doppler, Color Doppler. Amplitude Doppler, and combined imaging modes utilizing a number of transducers with varying center frequencies.
Both systems are modified to optimize images obtained with the use of diagnostic ultrasound and ultrasound contrast agents.
Both systems employ modified transmit/receive functions which allow for enhanced imaging of tissues and structures which reflect the transmitted ultrasound at a harmonic, or multiple, of the transmit frequency.
# End of 510(k) Summary
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. In the center of the seal is an abstract image of an eagle with three lines representing the wings and body. The eagle is facing to the right.
OCT 28 1998
Steve Hesler Manager of Regulatory Affairs Siemens Medical Systems, Inc. Ultrasound Group
22010 S.E. 51st St. Issaquah, WA 98029-7002
Re: K981528 Trade Name: Sonoline Elegra and Elegra Advanced Diagnostic Ultrasound System (with Harmonic Imaging) Regulatory Class: II Product Code: 90 IYO/21 CFR 892.1560 Dated: October 7, 1998 Received: October 9, 1998
Dear Mr. Hesler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following probes intended for use with the Sonoline Elegra and Elegra Advanced Diagnostic Ultrasound System, as described in your premarket notification:
Probes Model Number
2.5PL20 Phased Array 3.5PL28 Phased Array 3.5C40 Curved Array 5.0HDPL40 Linear Array 5.0C50 Curved Array 6.5EC10 Curved Array 6.5EV13 Curved Array 7.5C30 Curved Array 7.5L40 Linear Array 7.5PL13 Phased Array
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
{3}------------------------------------------------
Page 2 - Steve Hesler
Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the investigational device exemptions (IDE) regulations, 21 CFR, Part 812. Therefore, your product labeling must be consistent with FDA's position on this use.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez 594-1212,
Sincerely yours,
Gerald li. Segman
Lillian Yin, Ph.D. Director, Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{4}------------------------------------------------
# Attachment 1
Ultrasound Device Indications Statement
510 (k) Number (if known) : Device Name : Intended Use:
K981528 SONOLINE Elegra/Elegra Advanced See below
Mode of Operation
| Mode of Operation | Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Harmonic Imaging |
|-------------------|----------------------------------|---|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|------------------|
| | Ophthalmic | | | | | | | | | | |
| | Fetal | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Abdominal | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Intraoperative<br>(Specify) * | P | P | P | P | P | P | P | | BMDC (P) | |
| | Pediatric | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Small Organ<br>(Specify) ** | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Neonatal<br>Cephalic | P | P | P | P | P | P | P | | BMDC (P) | |
| | Adult Cephalic | P | P | P | P | P | P | P | | BMDC (P) | |
| | Cardiac | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Trans-esophageal | | | | | | | | | | |
| | Transrectal | P | P | P | P | P | P | P | | BMDC (P) | |
| | Transvaginal | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Transurethral | | | | | | | | | | |
| | Intravascular | | | | | | | | | | |
| | Peripheral<br>vessel | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Laparoscopic | | | | | | | | | | |
| | Musculo-skeletal<br>Conventional | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Musculo-skeletal<br>Superficial | P | P | P | P | P | P | P | | BMDC (P) | N |
| | Other (specify) | | | | | | | | | | |
E = added under Appendix E N = new indication; P = previously cleared by FDA;
Additional Comments:
* = intraoperative (abdominal, neurological)
** = small organs (breast,testes, thryoid, penis,prostate)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Sypem
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devi
510(k) Number K981528
{5}------------------------------------------------
510 (k) Number (if known) :
Device Name :
K981528 2.0 CW probe for use with SONOLINE Elegra/Elegra Advanced
Intended Use:
| Device Name: | | 2.5 CW probe for use with Sonoline Glegra & Glegra Advanced | | | | | | | | |
|----------------------------------|---|-------------------------------------------------------------|---|-----|-----|--------------------------------|----------------------|------------------------------|----------------------------|---------------------|
| Intended Use: | | See below | | | | | | | | |
| Mode of Operation | | | | | | | | | | |
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | P | | | | | | |
| Abdominal | | | | P | | | | | | |
| Intraoperative<br>(Specify) | | | | P | | | | | | |
| Pediatric | | | | P | | | | | | |
| Small Organ<br>(Specify) | | | | P | | | | | | |
| Neonatal | | | | P | | | | | | |
| Cephalic | | | | P | | | | | | |
| Adult Cephalic | | | | P | | | | | | |
| Cardiac | | | | P | | | | | | |
| Trans-<br>esophageal | | | | | | | | | | |
| Transrectal | | | | P | | | | | | |
| Transvaginal | | | | P | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>vessel | | | | P | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | P | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
| N = new indication; | | | | | | P = previously cleared by FDA; | | | E = added under Appendix E | |
| Additional Comments: | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David b. Segner
(Divisio Sign-Off Division of Reproductive, Abdominal, ENT and Radiological Devi
510(k) Number
{6}------------------------------------------------
## 510 (k) Number (if known) :
K981528
| Device Name: 25PL20 Phased Array Transducer for use with SONOLINE Elegra/Elegra Advanced | | | | | | | | | | |
|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Intended Use: See below. | | | | | | | | | | |
| | Mode of Operation | | | | | | | | | |
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging |
| Ophthalmic | | | | | | | | | | |
| Fetal | E | E | E | E | E | E | E | | BMDC (E) | N |
| Abdominal | E | E | E | E | E | E | E | | BMDC (E) | N |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | E | E | E | E | E | E | E | | BMDC (E) | N |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | E | E | E | E | E | E | E | | BMDC (E) | |
| Cephalic<br>Adult | E | E | E | E | E | E | E | | BMDC (E) | |
| Cephalic<br>Cardiac | E | E | E | E | E | E | E | | BMDC (E) | N |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | N |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>vessel | E | E | E | E | E | E | E | | BMDC (E) | N |
| Laparoscopic | | | | | | | | | | |
| Musculo-<br>skeletal | | | | | | | | | | |
| Conventional<br>Musculo-<br>skeletal | | | | | | | | | | |
| Superficial | | | | | | | | | | |
| Other<br>(Specify) | | | | | | | | | | |
| | N = new indication; Additional Comments:<br>P = previously cleared by FDA; E = added under Appendix E | | | | | | | | | |
2.5PL20 Phased Array Transducer for use with SONOLINE Elegra/Elegra Advanced Device Name : See helow
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Jamil le. Sezmm
(Divis on Sign-Off) Reproductive, Abdominal, El Divisio and Radiological De
510(k) Number K981528
{7}------------------------------------------------
| Device Name: 3.5PL28 Phased Array Transducer for use with SONOLINE Elegra/Elegra Advanced<br>Intended Use: See below. | | | | | | | | | | |
|-----------------------------------------------------------------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Mode of Operation | | | | | | | | | | |
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging |
| Ophthalmic | | | | | | | | | | |
| Fetal | P | P | P | P | P | P | P | | BMDC (P) | N |
| Abdominal | P | P | P | P | P | P | P | | BMDC (P) | N |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | P | P | P | P | P | P | P | | BMDC (P) | N |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal<br>Cephalic | P | P | P | P | P | P | P | | BMDC (P) | |
| Adult Cephalic | P | P | P | P | P | P | P | | BMDC (P) | |
| Cardiac | | | | | | | | | | |
| Trans-<br>esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>vessel | P | P | P | P | P | P | P | | BMDC (P) | N |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added under Appendix E | | | | | | | | | | |
510 (k) Number (if known) : K981528
Device Name :
3.5PL28 Phased Array Transducer for use with SONOLINE Elegra/Elegra Advanced
Trans-
esophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral
vessel
Laparoscopic
Musculo-skeletal
Conventional
Musculo-skeletal
Superficial
Other (Specify)
| P | P | P | P | P | P | BMDC (P) | N |
|---|---|---|---|---|---|----------|---|
| | | | | | | | |
| | | | | | | | |
| | | | | | | | |
| | | | | | | | |
| | | | | | | | |
| | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David li. Leymm
(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological Dev 510(k) Number
{8}------------------------------------------------
510 (k) Number (if known) : K981528
Device Name :
Intended Use:
3.5C40 Curved Array Transducer for use with SONOLINE Elegra/Elegra Advanced See below
| Mode of Operation | | | | | | | | | | |
|---------------------------------------------|---|---|---|-----|-----|--------------------------------|----------------------|------------------------------|----------------------------|---------------------|
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging |
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | | E | E | | BMDC (E) | N |
| Abdominal | | E | E | E | | E | E | | BMDC (E) | N |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | E | E | E | | E | E | | BMDC (E) | N |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal<br>Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-<br>esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>vessel | | E | E | E | | E | E | | BMDC (E) | N |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | E | E | E | | E | E | | BMDC (E) | N |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
| N = new indication;<br>Additional Comments: | | | | | | P = previously cleared by FDA; | | | E = added under Appendix E | |
ﻪ: ﺑ
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seymann
Sign-Off Reproductive, Abdominal, EN and Ra
Radiological Devices
510(k) Number K981528
{9}------------------------------------------------
| Ultrasound Device Indications Statement | | | | | | | | | | |
|-----------------------------------------|-----------------------------------------------------------------------|---|---|-----|-----|---------------|----------------------------|------------------------|--------------------|------------------|
| 510 (k) Number (if known) : | K981528 | | | | | | | | | |
| Device Name : | 5.0HDPL40 Linear Array Transducer for use with SONOLINE Elegra/Elegra | | | | | | | | | |
| | Advanced | | | | | | | | | |
| Intended Use: | See below | | | | | | | | | |
| Mode of Operation | | | | | | | | | | |
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Harmonic Imaging |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | N |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | N |
| Intraoperative (Specify) | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | N |
| Small Organ (Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC (P) | N |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | P | P | P | | P | P | | BMDC (P) | N |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
| N = new indication; | P = previously cleared by FDA; | | | | | | E = added under Appendix E | | | |
| Additional Comments: | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Elnil A. Syverson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devic 510(k) Number
{10}------------------------------------------------
K981528
Device Name :
5.0C50 Curved Array Transducer for use with SONOLINE Elegra Elegra Advanced See below
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging |
|-----------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | N |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | N |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | N |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-<br>esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>vessel | | P | P | P | | P | P | | BMDC (P) | N |
| Laparoscopic | | | | | | | | | | |
| Musculo-<br>skeletal | | P | P | P | | P | P | | BMDC (P) | N |
| Conventional | | | | | | | | | | |
| Musculo-<br>skeletal | | | | | | | | | | |
| Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Segerson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological De 510(k) Number
510 (k) Number (if known) : Device Name : Intended Use:
{11}------------------------------------------------
#### 510 (k) Number (if known) : K981528
Device Name : Intended Use: 6.5EC10 Curved Array Transducer for use with SONOLINE Elegra/Elegra Advanced See below
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging |
|--------------------------------------|---|---|---|-----|-----|--------------------------------|----------------------|------------------------------|-----------------------|---------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) ** | | E | E | E | | E | E | | BMDC (E) | N |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-<br>esophageal | | | | | | | | | | |
| Transrectal | | E | E | E | | E | E | | BMDC (E) | |
| Transvaginal | | E | E | E | | E | E | | BMDC (E) | N |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-<br>skeletal | | | | | | | | | | |
| Conventional<br>Musculo-<br>skeletal | | | | | | | | | | |
| Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
| N = new indication; | | | | | | P = previously cleared by FDA; | | E = added under Appendix E | | |
| Additional Comments: | | | | | | | | | | |
** =small organ (breast, testes, thyroid, penis, prostate)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David h. Syverson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological 510(k) Number
{12}------------------------------------------------
## K981528
510 (k) Number (if known) : Device Name :
6.5EV13 Curved Array Transducer for use with SONOLINE Elegra/Elegra Advanced Intended Use: See below
| Mode of Operation | | | | | | | | | | | |
|------------------------------------------------------------|---|---|---|-----|-----|--------------------------------|----------------------|------------------------------|----------------------------|---------------------|--|
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative<br>(Specify) | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify) ** | | P | P | P | | P | P | | BMDC (P) | N | |
| Neonatal | | P | P | P | | P | P | | BMDC (P) | | |
| Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Trans- esophageal | | | | | | | | | | | |
| Transrectal | | P | P | P | | P | P | | BMDC (P) | | |
| Transvaginal | | P | P | P | | P | P | | BMDC (P) | N | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral<br>vessel | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-<br>skeletal | | | | | | | | | | | |
| Conventional | | | | | | | | | | | |
| Musculo-<br>skeletal | | | | | | | | | | | |
| Superficial | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | |
| N = new indication; | | | | | | P = previously cleared by FDA; | | | E = added under Appendix E | | |
| Additional Comments: | | | | | | | | | | | |
| ** =small organ (breast, testes, thyroid, penis, prostate) | | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Elmer A. Leyson
(Division Sign-Off) Division of Repreductive, Abdominal, ENT and Radiological De
510(k) Number K981528
{13}------------------------------------------------
## 510 (k) Number (if known) :
Device Name : Intended Use: K981528 7.5C30 Curved Array Transducer for use with SONOLINE Elegra Elegra Advanced See below
of Operatio
| Mode of Operation | | | | | | | | | | |
|--------------------------------------|---|---|---|-----|-----|--------------------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging |
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | | E | E | | BMDC (E) | N |
| Abdominal | | E | E | E | | E | E | | BMDC (E) | N |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | E | E | E | | E | E | | BMDC (E) | N |
| Small Organ<br>(Specify) ** | | E | E | E | | E | E | | BMDC (E) | N |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-<br>esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>vessel | | E | E | E | | E | E | | BMDC (E) | N |
| Laparoscopic | | | | | | | | | | |
| Musculo-<br>skeletal | | E | E | E | | E | E | | BMDC (E) | N |
| Conventional<br>Musculo-<br>skeletal | | | | | | | | | | |
| Superficial | | | | | | | | | | |
| Other<br>(Specify) | | | | | | | | | | |
| N = new indication; | | | | | | P = previously cleared by FDA; | | E = added under Appendix E | | |
Additional Comments:
** =small organ (breast, testes, thyroid, penis)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Lynam
(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiologic
510(k) Number K981528
{14}------------------------------------------------
#### 510 (k) Number (if known) : K981528
Device Name :
7.5L40 Linear Array Transducer for use with SONOLINE Elegra/Elegra Advanced See below
Intended Use: Ande of Operation
| Mode of Operation | | | | | | | | | | |
|-------------------------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Harmonic<br>Imaging |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | N |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | N |
| Intraoperative<br>(Specify) * | | P | P | P | | P | P | | BMDC (P) | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | N |
| Small Organ<br>(Specify) ** | | P | P | P | | P | P | | BMDC (P) | N |
| Neonatal | | P | P | P | | P | P | | BMDC (P) | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-<br>esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral<br>vessel | | P | P | P | | P | P | | BMDC (P) | N |
| Laparoscopic | | | | | | | | | | |
| Musculo-<br>skeletal<br>Conventional | | P | P | P | | P | P | | BMDC (P) | N |
| Musculo-<br>skeletal<br>Superficial | | P | P | P | | P | P | | BMDC (P) | N |
| Other (Specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added under Appendix E | | | | | | | | | | |
Additional Comments:
* = intraoper…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.