MENTOR PATHWAY INTRODUCER/DILATOR

K981527 · Mentor Corp. · QPB · Jul 9, 1998 · General, Plastic Surgery

Device Facts

Record IDK981527
Device NameMENTOR PATHWAY INTRODUCER/DILATOR
ApplicantMentor Corp.
Product CodeQPB · General, Plastic Surgery
Decision DateJul 9, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5040
Device ClassClass 2

Indications for Use

The Mentor Pathway Introducer is indicated for use as an operating port at the skin insertion site when instruments are utilized beneath the dermal layer for lipoplasty surgery applications. The Dilator is indicated for dilating the dermal layer and aiding in the insertion of the Pathway Introducer. The Plug is indicated for sealing the lumen of the Pathway Introducer when not in use.

Device Story

System comprises three components: introducer, dilator, and plug. Dilator facilitates insertion of introducer into skin. Introducer features threaded distal end for skin anchoring and conical external section; includes internal seal to wipe excess fluid from surgical instruments (cannulae/sheaths) during withdrawal. Plug seals introducer lumen when inactive. Used in clinical lipoplasty settings by surgeons to provide stable access port for sub-dermal instrumentation. Device protects skin insertion site and manages fluid during instrument exchange.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

System consists of introducer, dilator, and plug. Introducer features threaded distal end and internal seal. Mechanical device; no energy source or software.

Indications for Use

Indicated for patients undergoing lipoplasty surgery requiring instruments beneath the dermal layer. Used to dilate skin, provide access port, and seal lumen.

Regulatory Classification

Identification

A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.

Special Controls

*Classification.* Class II (special controls). Consensus standards and labeling restrictions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Mentor Corp. Donna Crawford Manager, Corporate Regulatory Affairs 5425 Hollister Ave. Santa Barbara, California 93111 June 8, 2021 Re: K981527 Trade/Device Name: Mentor Pathway Introducer/dilator Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: OPB Dear Donna Crawford: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 9, 1998. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov. Sincerely, # Cindy Chowdhury -S Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 0 1098 JUL Ms. Donna A. Crawford Manager, Corporate Regulatory Affairs Mentor Corporation 5425 Hollister Avenue Santa Barbara, California 93111 Re: K981527 > Trade Name: Mentor Pathway Introducer/Dilator Regulatory Class: II Product Code: MUU Dated: April 24, 1998 Received: April 28, 1998 Dear Ms. Crawford: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ### Page 2 - Ms. Donna A. Crawford This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known):_ 长 98 / 527 Device Name: MENTOR Pathway™ Introducer/Dilator Indications For Use: The Mentor Pathway Introducer is indicated for use as an operating port at the skin insertion site when instruments are utilized beneath the dermal layer for lipoplasty surgery applications. The Dilator is indicated for dilating the dermal layer and aiding in the insertion of the Pathway Introducer. The Plug is indicated for sealing the lumen of the Pathway Introducer when not in use. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--------------------------------------------------------| |--------------------------------------------------------| | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of General Restorative Devices | | | 510(k) Number | K981527 | Prescription Use V (Per 21 CFR 801.109) OR Over-The Counter Use (Optional Format 1-2-96) {4}------------------------------------------------ K 981527 # 510(k) SUMMARY This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92. The assigned 510(k) number is: Submitted by: Donna A. Crawford Manager, Corporate Regulatory Affairs Mentor Corporation 5425 Hollister Ave. Santa Barbara, CA 93111 > Telephone: (805) 681-6000 extension 304 FAX: (805) 681-6004 Date Prepared: April 24, 1998 #### Device Name Proprietary Name: Mentor Pathway™ Introducer/Dilator Common Name: Introducer/Dilator # Substantial Equivalence Claim The Mentor Pathway Introducer/Dilator is similar to devices used to dilate the skin and provide an access port for surgical procedures. Examples of equivalent devices include the Ethicon Non-shielded Surgical Trocar (K963760), the Ximed surgical trocars and cannulae (K933458) and the Applied Medical Technology Endoscopic Access Device (K960232). The Mentor Pathway Introducer/Dilator is also similar in design and intended use to the Lysonix Lipoplasty Access Port (K980763). #### Indications For Use The Mentor Pathway Introducer is indicated for use as an operating port at the skin insertion site when instruments are utilized beneath the dermal layer for lipoplasty surgery applications. The Dilator is indicated for dilating the dermal layer and aiding in the insertion of the Pathway Introducer. The Plug is indicated for sealing the lumen of the Pathway Introducer when not in use. {5}------------------------------------------------ ## 510(k) SUMMARY Page 2 of 2 ### Device Description The Mentor Pathway Introducer/Dilator system consists of three (3) primary components: Introducer, Dilator and Plug. The Dilator consists of a dilating stem and a handle. It is used to facilitate the insertion of the Introducer. The Pathway Introducer consists of a threaded distal end which is threaded into the skin at the insertion site, and a conical section which remains outside the skin. There is a longitudinal hole in the Introducer through which a cannula and sheath (or other similar surgical instrumentation) are inserted, and there is a seal inside the hole which wipes excess liquid from the cannula and sheath during withdrawal. The Introducer acts as an access port during the insertion of cannulae and other operating instruments to a desired surgical site beneath the dermal layer. The Plug is used to seal the lumen of the Pathway Introducer when not in use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...