← Product Code [LYZ](/productcode/LYZ) · K981488

# GLOVE, OAKTEX EXAMINATION, GLOVE, PATIENT EXAMINATION (K981488)

_Oak Carolina, Inc. · LYZ · Jul 7, 1998 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981488

## Device Facts

- **Applicant:** Oak Carolina, Inc.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Jul 7, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

This patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner.

## Device Story

OAKTEX Examination Glove is a powder-free vinyl patient examination glove. Device is worn on the examiner's hand or finger during medical examinations to serve as a barrier, preventing contamination between the patient and the examiner. It is a disposable, single-use device intended for general medical settings. The glove provides a physical barrier to reduce the risk of infection transmission during clinical procedures.

## Clinical Evidence

Bench testing only. Testing included water leak tests (ASTM D5151-92), tensile strength, elongation, and dimensional measurements (palm/finger thickness, length, width) compared against the predicate device and ASTM 5250-92 standards.

## Technological Characteristics

Material: Vinyl. Powder-free. Dimensions: Length 241mm; various widths (S, M, L, XL). Standards: Meets ASTM D5151-92 (water leak) and ASTM 5250-92 (with exception for width requirements).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Maxxim powder free vinyl examination glove

## Reference Devices

- Sensicare powder free medical glove

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

7 1998 JUL

## DAY CAROLINA INC.

f

Page: 10.1

# 510(k) Summary K981488

### Summary of 510(k) Safety and Effectiveness Information

| Submitter:                  | Oak Carolina (Darrell Alford)<br>Division of Oak Technical                                                                                                                             |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitters Address:         | 100 Roe Road<br>Travelers Rest, SC 296910                                                                                                                                              |
| Date Prepared:              | April 24, 1998                                                                                                                                                                         |
| Submitters Phone Number:    | (864) 834-1239                                                                                                                                                                         |
| Submitters Fax Number:      | (864) 834-1219                                                                                                                                                                         |
| Trade Name:                 | OAKTEX Examination Glove                                                                                                                                                               |
| Common Name:                | Vinyl Examination Glove                                                                                                                                                                |
| Classification Name:        | Vinyl Patient Examination Glove                                                                                                                                                        |
| Substantial Equivalence:    | This device is substantially equivalent to the Maxxim<br>powder free vinyl examination glove.                                                                                          |
| Device Description:         | This device is a powder free vinyl examination glove.                                                                                                                                  |
| Intended Use of the Device: | This patient examination glove is a disposable device<br>intended for medical purposes that is worn on the<br>examiners hand to prevent contamination between patient<br>and examiner. |

{1}------------------------------------------------

INA INC.

### 510(k) Summary (98) 488

Technological characteristics and non-clinical performance data comparison of device to predicate device:

| TEST                             | OAK Vinyl Exam Glove<br>96-X61 Series | Maxxim<br>Vinyl Exam Glove |
|----------------------------------|---------------------------------------|----------------------------|
| Water Leak Test<br>ASTM D5151-92 | Pass                                  | Pass                       |
| Tensile MPa Ultimate             | 10.4                                  | 13.7                       |
| Elongation, Ultimate (%)         | 355                                   | 437                        |
| Thickness - Palm (mm)            | .151                                  | .130                       |
| Thickness - Finger (mm)          | .125                                  | .113                       |
| Length (mm)                      | 241                                   | 241                        |
| Width/mm                         | S, M, L, XL<br>95, 105, 118, 124      | L<br>105                   |
| Powdered                         | No                                    | No                         |
| ASTM 5250-92                     | Meets*<br>ASTM 5250-92                | Meets<br>ASTM 5250-92      |

* Except for (width) dimensional requirements.

Conclusion drawn from technological characteristics and non-clinical testing:

We conclude that the Oak vinyl exam glove is substantially equivalent to the legally marketed Sensicare powder free medical glove based on the performance data comparison and technological characteristics.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle faces left and is enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

7 1998 11 11

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Darrell W. Alford Quality Assurance Manager OAK Carolina, Incorporated Division of Oak Technical 100 Roe road Travelers Rest, South Carolina 29690

K981488 Re : OAKTEX PowderFree Vinyl Patient Examination Trade Name: Glove Regulatory Class: I Product Code: LYZ Dated: April 24, 1998 Received: April 27, 1998

Dear Mr. Alford:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major -----regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਜ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Mr. Alford

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.qov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure ________

{4}------------------------------------------------

Image /page/4/Picture/3 description: The image shows the logo for OAK CAROLINA INC. The word "OAK" is in a bold, sans-serif font and is enclosed in an oval shape. To the right of the logo is the text "CAROLINA INC." also in a bold, sans-serif font.

Revised Page 8.2

�

510(k) Number if known): K981488

Device Name: Oaktex Powder Free Vinyl Patient Examination Glove

Indications For Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Chen S. Lim

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital, Innection of

510(k) Number K981488

Prescription Use

OR

Over-The-Counter Use X
(Optional Format 1-2-96)

TRAVELERS REST, SQUTH CAROLINA 29690-9426 · 866-834-1239 · FAX 864-834-12 19

---

**Source:** [https://fda-staging.innolitics.com/device/K981488](https://fda-staging.innolitics.com/device/K981488)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).