The containment rings are for use with screws which are indicated for sacral fixation, or for limited pedicular fixation. When used as a pedicle screw system, the system is intended only for use with autogenous bone graff in order to facilitate fusions of the L5-S1 joint in patients with grade 3 or 4 spondylolisthesis. Screws are not intended for placement in the pedicles above L3, and are intended for removal after development of a solid fusion mass. The specific indications for the Osteonics® Long Arm Containment Ring, as a component of the legally marketed Osteonics® Spinal System, fall within the same indications as all components of the Osteonics® Spinal System: Non-Pedicular Use; fixation of the T4-S2 spine: - Long and short curve scollosis - Vetebral fracture or dislocation - Spondylolisthesis - Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the t disc confirmed by history and radiographic studies). - Previously failed fusion - Spinal tumor. Pedicular Use: - When used as a pedicle screw system, the system is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. Pedicle screws are not intended for placement in pedicles above L3.
Device Story
Osteonics® Long Arm Containment Ring is a component of the Osteonics® Spinal System; used by surgeons during spinal fixation procedures. Device placed over long arms of spinal screws; aids in achieving clean break of scored screw arms; protects soft tissues; prevents screw arm spreading during screw blocker tightening. Locking mechanism identical to predicate Osteonics® Spinal System Low Profile Cap. Provides surgeons an alternative to the Low Profile Cap. Used in clinical settings; manual operation by physician.
Clinical Evidence
No clinical data provided. Substantial equivalence based on design, material, and functional similarities to the predicate device, supported by pre-clinical data incorporated by reference to a prior submission.
Technological Characteristics
Material: ASTM F-136-96 Titanium Alloy (Ti6Al-4V ELI). Form factor: Containment ring for spinal screw system. External diameter: 15 mm. Sterilization: Standard methods. Connectivity: None (mechanical component).
Indications for Use
Indicated for patients with T4-S2 spinal conditions including scoliosis, vertebral fracture/dislocation, spondylolisthesis, degenerative disc disease, failed fusion, or spinal tumor. Pedicular use restricted to patients with severe (Grade 3-4) spondylolisthesis at L5-S1 requiring fusion with autogenous bone graft; device must be removed after fusion. Contraindicated for pedicle placement above L3.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
Predicate Devices
Osteonics® Spinal System Low Profile Cap
Submission Summary (Full Text)
{0}------------------------------------------------
07:49 FROM MAY-87-1998
OSTEDNICS-REG AFFAIRS TO
510(k) Premarket Notification
Osteonics® Long Arm Containment Ring
MAY 20 1998
## 510(k) Premarket Notification Summary of Safety and Effectiveness for the Osteonics® Long Arm Containment Ring
#### Submission Information
Name and Address of the Sponsor of the 510(k) Submission:
Contact Person:
Osteonics Corporation 59 Route 17 Allendale, NJ 07401-1677
Kate Sutton Regulatory Affairs Specialist
Date of Summary Preparation:
May 5, 1998
Device Identification
Proprietary Name:
Osteonics® Long Arm Containment Ring (a component of the Osteonics® Spinal System)
Common Name:
Classification Name and Reference:
Spinal fixation appliance
Spinal Interlaminal Fixation Orthosis 21 CFR §888.3050
#### Predicate Device Identification
The Osteonics® Long Arm Containment Ring is substantially equivalent to the following competitive and/or Osteonics devices, which have previously been determined substantially equivalent by FDA:
Osteonics® Spinal System Low Profile Cap ●
#### Device Description
The Osteonics® Long Arm Containment Ring is placed over the long arms of the screw to aid the physician in achieving a clean break along the scored sections of the screw arms. Additionally, placement of the Osteonics® Long Arm Containment Ring over the body of the screw protects soft tissues and prevents the arms of the screw from spreading as the screw blocker is tightened. The locking mechanism of the containment ring is identical to that of the Osteonics® Spinal System Low Profile Cap. The containment ring does not replace the Low Profile Cap, rather, it offers surgeons an alternative to the Low Profile Cap, if so desired.
{1}------------------------------------------------
MHY-VY-1996
### Intended Use:
The containment rings are for use with screws which are indicated for sacral fixation, or for limited pedicular fixation. When used as a pedicle screw system, the system is intended only for use with autogenous bone graff in order to facilitate fusions of the L5-S1 joint in patients with grade 3 or 4 spondylolisthesis. Screws are not intended for placement in the pedicles above L3, and are intended for removal after development of a solid fusion mass.
The specific indications for the Osteonics® Long Arm Containment Ring, as a component of the legally marketed Osteonics® Spinal System, fall within the same indications as all components of the Osteonics® Spinal System:
#### Non-Pedicular Use; fixation of the T4-S2 spine:
- Long and short curve scollosis �
- Vetebral fracture or dislocation .
- Spondylolisthesis .
- Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the t disc confirmed by history and radiographic studies).
- . Previously failed fusion
- Spinal tumor .
### Pedicular Use:
- When used as a pedicle screw system, the system is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. Pedicle screws are not intended for placement in pedicles above L3.
## Statement of Technological Comparison:
The substantial equivalence of the Osteonics® Long Arm Containment Ring to the predicate Osteonics® Spinal System Low Profile Cap-in terms of intended use, material, and design features-is based on the following:
#### Intended Uses:
The intended uses of the subject Osteonics® Long Arm Containment Ring are identical to those of the predicate device.
## Material:
Both subject and predicate devices use the identical material, ASTM F-136-96 Titanium Alloy (Ti6Al-4V ELI).
## Design:
The design and function of the Osteonics® Long Arm Containment Ring is consistent with that of the predicate Osteonics® Spinal System Low Profile Cap and differs only in the following:
{2}------------------------------------------------
| | | | | | 17 07 |
|--|--|--|--|--|-------|
|--|--|--|--|--|-------|
510(k) Premarket Notification
- The diameter of the hole that provides access to the screw blocker has been increased.
TO
- The external diameter of the subject device is 1 mm larger than that of the predicate device (15 mm vs. 14 mm).
None of these design differences raise any new questions of safety or effectiveness.
#### Summary
Based on the similarities presented above, the supporting testing summary, the pre-clinical data incorporated by reference to a prior submission, and the fact that the Osteonics® Long Arm Containment Ring employs standard sterilization and packaging methods, the substantial equivalence of the Osteonics® Long Arm Containment Ring to the legally marketed Osteonics® Spinal System Low Profile Cap is demonstrated.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three faces in profile, with a stylized wing above them. The faces are facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 20 1998
Ms. Kate Sutton Requlatory Affairs Specialist Osteonics Corporation 59 Route 17 Allendale, New Jersey 07401-1677
Re : K981452 Long Arm Containment Ring - part of the Osteonics® Spinal System Regulatory Class: II Product Codes: KWP and MNH Dated: April 20, 1998 Received: April 22, 1998
Dear Ms. Sutton:
We have reviewed your Section 510(k) notification of intent to market the device system referenced above and we have determined the device system is substantially equivalent (for the indications for use stated in the enclosure) to device systems marketed in interstate commerce prior to May 28, 1976 or to devices that have been reclassified in accordance with the provisions of the Federal, Food, Drug, and Cosmetic Act This decision is based on your device system being (Act) . . found equivalent only to similar device systems labeled and intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. You may, therefore, market your device system subject to the general controls provisions of the Act and the limitations identified below.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that beling or otherwise promoting this device system for pedicular screw fixation/attachment for indications other than severe spondylolisthesis, as described above, would cause the device system to be adulterated under 501(f) (1) of the Act.
{4}------------------------------------------------
Page 2 - Ms. Kate Sutton
This device system, when intended for pedicular screw fixation/attachment to the spine for indications other than severe spondylolisthesis, as described above, is a class III device under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly:
- All labeling for this device, including the package 1. label, must state that there are labeling limitations. The package insert must prominently state that the device system using pedicle screws is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.
- You may not label or in anyway promote this device system 2. for pedicular, screw fixation/attachment to the cervical, thoracic or lumbar vertebral column for intended uses other than severe spondylolisthesis, as described above. The package insert must include the following statements:
#### WARNINGS :
- When used as a pedicle screw system, this device . system is intended only for grade 3 or 4 spondylolisthesis at the fifth lumbar - first sacral (L5-S1) vertebral joint.
- The screws of this device system are not intended . for insertion into the pedicles to facilitate spinal fusions above the L5-S1 vertebral joint.
- Benefit of spinal fusions utilizing any pedicle . screw fixation system has not been adequately established in patients with stable spines.
- Potential risks identified with the use of this . device system, which may require additional surgery, include :
device component fracture, loss of fixation, non-union, fracture of the vertebra, neurological injury, and vascular or visceral injury.
{5}------------------------------------------------
See Warnings, Precautions, and Potential Adverse Events sections of the package insert for a complete list of potential risks.
- Any pedicular screw fixation/attachment for intended uses 3. other than severe spondylolisthesis, as described by item 1, for this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR, Part 812. All users of the device for pedicular screw fixation/attachment for intended uses other than severe spondylolisthesis, as described above, must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conducting an investigation.
- Any previous warning statements identified as part of 4 . previous 510(k) clearances or required by OC/Labeling and Promotion which stated that a component/system was not approved for screw fixation into the pedicles of the spine must be replaced by the warnings of items 1 and 2 above.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
FDA advises that the use of your device system with any other device components but those identified in this 510(%)>>would require submission of a new 510(k) providing documentation of design, material, and labeling compatibility between the device components. Mechanical testing of a spinal system consisting of the subject device components and other device components, whether yours or those of other manufacturers, may also be required.
{6}------------------------------------------------
Page 4 - Ms. Kate Sutton
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general' information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
P.13
Page 1 of 1
## 510(k) Number (if known): K981452
# Device Name: Osteonics® Long Arm Containment Ring
Indications For Use:
The uses for the Osteonics® Long Arm Containment Ring, as part of the legally marketed Osteonics® Spinal System, are as follows:
# Non-Pedicular Use; fixation of the T4-S2 spine:
- Long and short curve scoliosis .
- Vetebral fracture or dislocation .
- Spondylolisthesis .
- Degenerative disc disease (defined as back pain of discogenic origin with degeneration of . the disc confirmed by history and radiographic studies).
- Previously failed fusion .
- Spinal tumor
#### Pedicular Use:
- When used as a pedicle screw system, the system is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. Pedicle screws are not intended for placement in pedicles above L3.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF T NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|-----------------------------------------|--------------------------------------------------------|
| Prescription Use | X |
| (Per 21 CFR 801.109) | |
| (Division Sign-Off) | |
| Division of General Restorative Devices | |
| 510(k) Number | K981452 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.