K981410 · Philips Medical Systems (Cleveland), Inc. · LNH · Jun 30, 1998 · Radiology
Device Facts
Record ID
K981410
Device Name
HAMMERSMITH ENDOCAVITARY COILS
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
LNH · Radiology
Decision Date
Jun 30, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Hammersmith Endocavitary Coils do not change the intended use of the Picker MR system with which they are used. Picker MR systems are intended for use as NMR devices that produce images that: (1) correspond to the distribution of protons exhibiting NMR. (2) devend upon NMR parameters (proton density, flow velocity, spin-lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images vield information that can be useful in the determination of a diagnosis. The Hammersmith Endocavitary Coils are indicated for use in the following anatomic regions and with the designated nuclei: Anatomic Regions: The Anus or Prostate by insertion into the rectum, and the Cervix by insertion into the vagina Nuclei Excited: Hydrogen
Device Story
Hammersmith Endocavitary Coils are receive-only MRI accessory coils designed for high-resolution imaging of the anus, prostate, and cervix. Unlike predicate devices using inflatable latex balloons, these are rigid, reusable probes shaped like ultrasound transducers. The device is inserted into the rectum or vagina by a clinician. It functions as an antenna to capture hydrogen proton signals during MRI scanning, which are then processed by the host Picker MR system to produce soft tissue images. The rigid design allows for fixed tuning and matching, eliminating the need for patient-specific tuning and reducing susceptibility artifacts caused by air. The device is used in a clinical MRI environment; output is interpreted by a physician to assist in diagnosis. Benefits include reusability, improved image quality due to reduced artifacts, and simplified setup.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Receive-only MRI coil; rigid Acrylonitrile resin enclosure; linear antenna configuration; fixed factory-set tuning and impedance matching; active diode decoupling; single receive channel; reusable; compatible with Picker Edge, Vista, Asset, Eclipse, Polaris, and Apollo MRI systems.
Indications for Use
Indicated for imaging the anus, prostate (via rectal insertion), or cervix (via vaginal insertion) in patients undergoing MRI. Contraindicated for any patient excluded from a normal digital examination.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K98 14/10
JUN 3 0 1998
## SUMMARY OF SAFETY AND EFFECTIVENESS (As required by 21 CFR 807.92)
#### 1. General Information
Class II Classification: Magnetic Resonance Imaging (MRI) Accessory Common/Usual Name: Magnetic Resonance Imaging (MRI) Coil Hammersmith Endocavitary Coils Proprietary Name: Establishment Registration: Picker International, Inc. World Headquarters 595 Miner Road Highland Heights. Ohio 44143 FDA Owner Number: #1580240 FDA Registration Number: #1525965 Performance Standards: Not Applicable
Intended Uses
2.
The Hammersmith Endocavitary Coils do not change the intended use of the Picker MR system with which they are used.
Picker MR systems are intended for use as NMR devices that produce images that: (1) correspond to the distribution of protons exhibiting NMR. (2) devend upon NMR parameters (proton density, flow velocity, spin-lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images vield information that can be useful in the determination of a diagnosis.
The Hammersmith Endocavitary Coils are indicated for use in the following anatomic regions and with the designated nuclei:
Anatomic Regions: The Anus or Prostate by insertion into the rectum, and the Cervix by insertion into the vagina Nuclei Excited: Hydrogen
PICKER INTERNATIONAL, INC.
(ENDOCAV)
3/31/98
I - 1
{1}------------------------------------------------
#### Device Description 3.
These three separate receive-only coils are for imaging the anus, prostate and cervix. The coils are designed to be inserted in either the rectum or vagina before imaging. The Hammersmith Endocavitary Coils are similar to the Philips Endo-Cavitary Coil in that they are reusable. However, unlike the Philips Endo-cavitary MRI Coil, they are rigid structures that do not contain an inflatable latex balloon. This difference in the design eliminates the need for limiting the number of times the Hammersmith Coils can be used, allows for fixed tuning and matching, and reduces the image artifacts due to susceptibility effects from air. Instead, the shape and dimensions of the Hammersmith Endocavitary Coils are similar to ultrasound probes used in endocavitary applications.
### Safety and Effectiveness 4.
The Picker Hammersmith Endocavitary Coils are substantially equivalent in safety and effectiveness to the Philips Endo-Cavitary MRJ Coil. The following chart has been compiled to demonstrate the substantial equivalence of these devices.
| Parameter | Hammersmith Endocavitary<br>Coils | Predicate Device: Philips Endo-<br>Cavitary MRI Coil (K930193) |
|-------------------------------|------------------------------------------------------------|----------------------------------------------------------------------|
| Compatible MRI<br>Systems | Picker Edge, Vista, Asset, Eclipse,<br>Polaris and Apollo. | Philips Gyroscan ACS Series II<br>and Gyroscan T5 series II systems. |
| Mode of Operation | Receive-Only. | Receive-Only. |
| Antenna<br>Configuration | Linear. | Linear. |
| Tuning/Impedance<br>Matching | Fixed tuning and matching. Factory<br>set. | Tuned for each patient. |
| Method of<br>Decoupling | Active diode decoupling. | Decoupling diode switch |
| Coil Enclosure | Rigid Acrylonitrile resin. | Flexible coil loop inside a non-<br>permeable latex balloon. |
| Coil Stabilization | External immobilization positioning<br>device. | Inflate balloon on coil with luer-<br>lock syringe. |
| Number of Receive<br>Channels | One. | One. |
PICKER INTERNATIONAL, INC.
(ENDOCAV)
{2}------------------------------------------------
| Parameter | Hammersmith Endocavitary<br>Coils | Predicate Device: Philips Endo-<br>Cavitary MRI Coil (K930193) |
|-------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cleaning and<br>Disinfection Method | Cleaning: MetriZyme detergent.<br>Disinfectant: Cidex Dip.<br>Covered with sheath. | Cleaning: Detergent.<br>Disinfectant: Cidex Dip<br>Covered with sheath. |
| Use Limits | None. | Used up to 50 times within one<br>year of manufacturing. |
| Indications for use | Anus or Prostate by insertion into<br>the rectum and Cervix by insertion<br>into the vagina. | Prostate or Uterus by insertion<br>into the rectum. |
| Contraindications | Any exclusion for a normal digital<br>examination. | Severe hemorrhoids, recent<br>surgery in the area of the rectum,<br>inflammatory bowel disease,<br>Crohn's disease and any exclusion<br>for normal endorectal<br>examinations. |
.
I - 3
3/3 1/98
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines forming its head and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 3 0 1998
Elaine K. Keeler, Ph.D. Manager, MR Clinical Science Picker International, Inc. 595 Miner Road Highland Heights, Ohio 44143 Re:
K981410 Hammersmith Endocavitary Coils Dated: April 17, 1998 Received: April 20, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 MOS/90 LNH
Dear Dr. Keeler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrb/dsmaldsmamain.html".
Sincerely yours,
Kilian Yin
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
1198 1410 510(k) Number (if known):
Device Name: Hammersmith Endocavitary Coils
### Indications for Use:
The Hammersmith Endocavitary Coils do not change the intended use of the Picker MR system with which they are used.
Picker MR systems are intended for use as NMR devices that produce images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon NMR parameters (proton density, flow velocity, spin-lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The Hammersmith Endocavitary Coils are indicated for use in the following anatomic regions and with the designated nuclei:
Anatomic Regions:
The Anus or Prostate by insertion into the rectum, and the Cervix by insertion into the vagina
Nuclei Excited:
Hydrogen
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Javid A. Segrin
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT and Radiological Devi
Radiological Devices
510(k) Number K981410
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.