← Product Code [LHN](/productcode/LHN) · K981358

# ARTOSCAN S (K981358)

_Biosound Esaote, Inc. · LHN · Jul 13, 1998 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981358

## Device Facts

- **Applicant:** Biosound Esaote, Inc.
- **Product Code:** [LHN](/productcode/LHN.md)
- **Decision Date:** Jul 13, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Artoscan S is intended for diagnostic nuclear magnetic resonance imaging of the hip, leg, knee, ankle, foot, shoulder, elbow, forearm, wrist and hand. The device produces transverse, sagittal, coronal and oblique cross-sectional images, displaying the internal structure of the limbs and joints being imaged. The images that are produced correspond to the spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon the MR parameters (spin-lattice relaxation time (T1), spinspin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.

## Device Story

Artoscan S is a dedicated extremity MRI system using a 0.18 T permanent Neodymium alloy magnet. It captures MR signals from limbs (hip, leg, knee, ankle, foot, shoulder, elbow, forearm, wrist, hand) via solenoidal or DPA receiver coils. The system transforms these signals into cross-sectional images (transverse, sagittal, coronal, oblique) based on proton density and relaxation times (T1, T2). Operated by clinicians in a fixed installation, the device features a large gantry opening to reduce patient claustrophobia. Output images are displayed on a 17-inch monitor and used by physicians to visualize internal limb structures for diagnostic purposes. The system supports DICOM3 networking for printing and image transmission.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, technical specifications, and comparison of imaging parameters (e.g., pulse sequences, gradient strength, FOV) against predicate devices.

## Technological Characteristics

0.18 T 'C'-shaped Neodymium alloy permanent magnet; 20 mT/m gradient strength; 0.5 msec rise time; solenoidal and DPA (quadrature) receiver coils; DICOM3 connectivity; Pentium 200 MHz CPU; 32 MB RAM; 2.6 GB hard disk; 640 MB optical disk; self-contained RF shielding.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Esaote Artoscan M ([K963262](/device/K963262.md))
- Hitachi AIRIS ([K945155](/device/K945155.md))
- Elscint PRIMA 1TG ([K970990](/device/K970990.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUL 1 3 1898

K981358

Safety and Effectiveness Summary Artoscan S Biosound Esaote

# Safety and Effectiveness Summary

The following safety and effectiveness summary has been prepared pursuant to The following surcey and commaries specified in 21CFR 807.92(a).

807.92(a)(1)

# Submitter Information

Submitter Information

| Colleen Hittle, Official Correspondent |
|----------------------------------------|
| 8000 Castleway Drive                   |
| Indianapolis, IN 46250                 |

| Phone:     | (317) 849-1916 |
|------------|----------------|
| Facsimile: | (317) 577-9070 |

| Contact Person: | Colleen Hittle |
|-----------------|----------------|
| Date:           | March 25, 1998 |

807.92(a)(2)

FDA/CDRH/ODE/DMC

14 APR 98 09 57

Artoscan S Trade Name: Magnetic resonance diagnostic device Common Name: System, Nuclear Magnetic Resonance Imaging Classification Name(s):

90LNH Classification Number:

807.92(a)(3)

| Esaote  | Artoscan M | K963262 |
|---------|------------|---------|
| Hitachi | AIRIS      | K945155 |
| Elscint | PRIMA 1TG  | K970990 |

Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table.

Predicate Device(s)

{1}------------------------------------------------

#### Imaging System 1.

ERR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1

,在

| Characteristic | ARTOSCAN S                                                                | AIRIS<br>K945155                                                  | Comments                                                                                         |
|----------------|---------------------------------------------------------------------------|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Clinical use   | Knee, leg, ankle, foot,<br>hand, wrist, forearm,<br>elbow, shoulder, hip. | Whole body, central<br>nervous system and<br>orthopaedic regions. | Two new joints<br>(hip and<br>shoulder) have<br>been introduced<br>for examination.              |
| Field of view  | From 100 up to 300 mm,<br>step 10 mm (140 mm are<br>displayed).           | From 50 up to 350 mm in<br>diameter.                              | A larger FOV<br>has been realized<br>to better cover<br>the anatomical<br>parts to be<br>imaged. |

#### 1. Magnetic System

| Characteristic            | ARTOSCAN S                                                     | AIRIS<br>K945155                                               | Comments |
|---------------------------|----------------------------------------------------------------|----------------------------------------------------------------|----------|
| Type of magnet            | "C" shaped Neodymium<br>alloy permanent magnet.                | "C" shaped Neomax™<br>(Neodymium alloy)<br>permanent magnet.   |          |
| Static field<br>strength  | 0.18 T (perpendicular to<br>patient's body).                   | 0.3 T (perpendicular to<br>patient's body).                    |          |
| Fringe field              | 5 G line at 1.4 m<br>maximum from the centre<br>of the magnet. | 5 G line at 2.5 m<br>maximum from the centre<br>of the magnet. |          |
| Gantry opening            | 620 mm x 240 mm (w x h)                                        | 1000 x 380 mm(w x h)                                           |          |
| Magnetic<br>system weight | 1700 kg                                                        | 10000 kg                                                       |          |

#### Radiofrequency System 1.

| Characteristic               | ARTOSCAN S                                                                                                | AIRIS<br>K945155                                                                                                   | Comments                                                                                                                                              |
|------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Solenoidal<br>receiver coils | Shoulder/Knee Coil<br>17.5x7.6x14.7 cm<br>Hip Coil<br>20 x 23 x 14 cm<br>(internal dimensions: w x d x h) | Large joint Coil (shoulder, knee)<br>18.4x23 cm wxh<br>Flexible body Coil (hip and other regions)<br>20x150 cm wxd | The technology of the solenoidal (or linear) coils is unchanged compared to the technology used for the ARTOSCAN M coils. Coils with different shapes |

{2}------------------------------------------------

| Characteristic | ARTOSCAN S             | AIRIS<br>K945155        | Comments                                                                                                                            |
|----------------|------------------------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
|                |                        |                         | have been<br>developed to<br>adapt to the<br>geometrical<br>characteristics of<br>the gantry and to<br>the new limbs to<br>examine. |
| RF power       | up to 600 W peak pulse | up to 5000 W peak pulse |                                                                                                                                     |

#### Patient Positioning 1.

、

| Characteristic         | ARTOSCAN S                                                                                                                                                                                                                                                     | AIRIS K945155                                                                                                                                                                         | Comments                                                                                                                                |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Patient comfort        | The large magnet opening<br>and the device design, that<br>allows the patient to be<br>almost completely out of<br>the magnet, except the<br>examined portion, virtually<br>eliminate claustrophobia<br>and other related reactions<br>during the examination. | The large magnet opening<br>virtually eliminates the<br>anxiety and claustrophobia<br>often experienced by<br>patients in the narrow bore<br>of conventional MR imag-<br>ing systems. |                                                                                                                                         |
| Patient<br>positioning | An adjustable couch is<br>used to position the<br>patient's limb inside the<br>gantry.<br>The couch is pushed into<br>the gantry manually.                                                                                                                     | An adjustable motorized<br>table is used to position the<br>patient inside the gantry.                                                                                                | The same<br>positioning<br>philosophy as<br>ARTOSCAN M,<br>suitable for the<br>new gantry<br>shape, has been<br>used for<br>ARTOSCAN S. |

:

{3}------------------------------------------------

#### Installation Area Conditions 1.

: 上一篇:

,

| Characteristic | ARTOSCAN S                                                                                        | AIRIS<br>K945155                                                                                                                | Comments |
|----------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------|
| Area necessary | 3.7 m x 3.7 m                                                                                     | 4 m x 5 m (gantry and<br>patient table)<br>4 m x 2.5 m (operator<br>console, MRI unit and<br>other components)                  |          |
| RF shielding   | Self contained RF<br>shielding (up to 30<br>dBμV/m).<br>External shielding (up to<br>100 dBμV/m). | The scan room must be RF<br>shielded to 60 dB. RF<br>noise inside the scan room<br>should meet 0.5-30 MHz,<br>0 dBμV/m or less. |          |

#### 2. Magnetic System

| Characteristic               | ARTOSCAN S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | PRIMA 1TG<br>K970990                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Comments                                                                                                                                                                                                         |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Gradient<br>strength         | 20 mT/m                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | whole-body set<br>21 mT/m                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | With higher<br>gradient strength<br>it is possible to<br>minimize at the<br>same time FOV<br>and slice<br>thickness, also<br>for sequences<br>with very short<br>TE, improving<br>the spatial<br>resolution.     |
|                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                  |
| Rise time                    | 0.5 msec from 0 to 20 mT<br>or from -20 to 0 mT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 0.5 msec to full amplitude<br>supplementary set<br>30 mT/m<br>0.41 msec to full amplitude                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                  |
|                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                  |
| Characteristic               | ARTOSCAN S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ARTOSCAN M<br>K963262                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comments                                                                                                                                                                                                         |
| Data<br>acquisition<br>modes | Spin warp 2D/3D<br>Half Scan<br>Half Echo                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Spin warp 2D/3D<br>Half Scan<br>Half Echo                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Unchanged                                                                                                                                                                                                        |
| Pulse sequences              | Multiplanar Ortho-Scout<br>T1-weighted Spin Echo<br>(SET1)<br>T2-weighted Spin Echo<br>(SET2)<br>Multiple Spin-Echo<br>(MEDE)<br>Gradient Echo (GE)<br>Inversion Recovery (IR)<br>Short TI Inversion<br>Recovery (STIR)<br>Half Echo T1-weighted<br>Spin Echo (SET1HE)<br>Half Fourier T1-weighted<br>Spin Echo (SET1HF)<br>Turbo T2-weighted Spin-<br>Echo (TSE)<br>Turbo Multiple Spin-Echo<br>(TME)<br>Turbo Gradient Echo<br>(TGE)<br>3D Gradient Echo (GE3D)<br>Turbo 3D T1-weighted<br>Gradient Echo (T3D_T1)<br>Turbo 3D T2-weighted<br>Gradient Echo (T3D_T2)<br>Short TI Inversion<br>Recovery Gradient Echo<br>(GEFS)<br>Magnetization Transfer<br>(GEMT)<br>3D Magnetization Transfer<br>(GE3DMT) | Multiplanar Ortho-Scout<br>T1-weighted Spin Echo<br>(SET1)<br>T2-weighted Spin Echo<br>(SET2)<br>Multiple Spin-Echo<br>(MEDE)<br>Gradient Echo (GE)<br>Inversion Recovery (IR)<br>Short TI Inversion<br>Recovery (STIR)<br>Half Echo T1-weighted<br>Spin Echo (SET1HE)<br>Half Fourier T1-weighted<br>Spin Echo (SET1HF)<br>Turbo T2-weighted Spin-<br>Echo (TSE)<br>Turbo Multiple Spin-Echo<br>(TME)<br>Turbo Gradient Echo<br>(TGE)<br>3D Gradient Echo (GE3D)<br>Turbo 3D T1-weighted<br>Gradient Echo (T3D_T1)<br>Turbo 3D T2-weighted<br>Gradient Echo (T3D_T2)<br>Short TI Inversion<br>Recovery Gradient Echo<br>(GEFS)<br>Magnetization Transfer<br>(GEMT)<br>3D Magnetization Transfer<br>(GE3DMT) | Unchanged                                                                                                                                                                                                        |
| Slice thickness              | 2D FT: from 2 mm to 10<br>mm, step 0.5 mm<br>3D FT: from 0.6 mm to 10<br>mm, step 0.1 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 2D FT: from 2 mm to 10<br>mm, step 0.5 mm<br>3D FT: from 0.6 mm to 10<br>mm, step 0.1 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Unchanged                                                                                                                                                                                                        |
| Interslice<br>spacing        | Not necessary: contiguous<br>slicing is available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Not necessary: contiguous<br>slicing is available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Unchanged                                                                                                                                                                                                        |
| Characteristic               | ARTOSCAN S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ARTOSCAN M K963262                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Comments                                                                                                                                                                                                         |
| Slice orientation            | Transverse<br>Sagittal<br>Coronal<br>Oblique<br>Multislice packages                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Transverse<br>Sagittal<br>Coronal<br>Oblique<br>Multislice packages                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Unchanged                                                                                                                                                                                                        |
| Acquisition matrix           | 2D FT: from 192x128 up to 256x256; phase encoding step 8<br>3D FT: from 192x128 up to 256x256; slice encoding from 8 to 128, step 8; phase encoding step 8                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 2D FT: from 192x128 up to 256x256; phase encoding step 8<br>3D FT: from 192x128 up to 256x256; slice encoding from 8 to 128, step 8; phase encoding step 8                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Unchanged                                                                                                                                                                                                        |
| Spatial resolution           | Up to 0.4 mm.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Up to 0.4 mm.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Unchanged                                                                                                                                                                                                        |
| Multi-slice                  | 128 slices max (for each echo)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 128 slices max (for each echo)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Unchanged                                                                                                                                                                                                        |
| Multi-echo                   | 2 echoes max                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 2 echoes max                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Unchanged                                                                                                                                                                                                        |
| Scan parameters              | TR<br>SET1: 50 ~ 5000 ms, step 10 ms<br>SET2: 200 ~ 5000 ms, step 10 ms<br>MEDE: 200 ~ 5000 ms, step 10 ms<br>GE: 60 ~ 5000 ms., step 10 ms<br>IR: 260 ~ 5000 ms, step 10 ms<br>STIR: 100 ~ 5000 ms, step 10 ms<br>SET1HE: 50 ~ 5000 ms, step 10 ms<br>SET1HF: 50 ~ 5000 ms, step 10 ms<br>TSE: 200 ~ 5000 ms, step 10 ms<br>TME: 200 ~ 5000 ms., step 10 ms<br>TGE: 30 ~ 5000 ms, step 5 ms<br>GE3D: 40 ~ 5000 ms, step 10 ms<br>T3D_T1: 25 ~ 5000 ms, step 5 ms                                                                                                                                                                                                                                            | TR<br>SET1: 50 ~ 5000 ms, step 10 ms<br>SET2: 200 ~ 5000 ms, step 10 ms<br>MEDE: 200 ~ 5000 ms, step 10 ms<br>GE: 60 ~ 5000 ms., step 10 ms<br>IR: 260 ~ 5000 ms, step 10 ms<br>STIR: 100 ~ 5000 ms, step 10 ms<br>SET1HE: 50 ~ 5000 ms, step 10 ms<br>SET1HF: 50 ~ 5000 ms, step 10 ms<br>TSE: 200 ~ 5000 ms, step 10 ms<br>TME: 200 ~ 5000 ms., step 10 ms<br>TGE: 30 ~ 5000 ms, step 5 ms<br>GE3D: 40 ~ 5000 ms, step 10 ms<br>T3D_T1: 25 ~ 5000 ms, step 5 ms                                                                                                                                                                                                                                            | Unchanged                                                                                                                                                                                                        |
| Characteristic               | ARTOSCAN S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ARTOSCAN M<br>K963262                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comments                                                                                                                                                                                                         |
|                              | T3D_T2 : 35 ~ 5000 ms,<br>step 1 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | T3D_T2 : 35 ~ 5000 ms,<br>step 1 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                  |
|                              | GEFS: 70 ~ 5000 ms., step<br>10 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | GEFS: 70 ~ 5000 ms., step<br>10 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                  |
|                              | GEMT: 40 ~ 5000 ms,<br>step 5 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | GEMT: 40 ~ 5000 ms,<br>step 5 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                  |
|                              | GE3DMT: 45 ~ 5000 ms,<br>step 5 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | GE3DMT: 45 ~ 5000 ms,<br>step 5 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                  |
| TE                           | SET1: 16~34 ms, step 2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | SET1: 18 ~ 34 ms, step 2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | The sequences<br>are substantially<br>unchanged. For<br>some sequences<br>we used shorter<br>TE, to obtain a<br>more T1-<br>weighted<br>contrast. That is<br>possible thanks<br>the higher<br>gradient strength. |
|                              | SET2: 80~ 120 ms, step<br>10 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | SET2: 80~ 120 ms, step<br>10 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                  |
|                              | MEDE: 1st echo 38 ms;<br>2nd echo 90 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | MEDE: 1st echo 38 ms;<br>2nd echo 90 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                  |
|                              | GE: 12 ~ 24 ms, step 2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | GE: 12 ~ 24 ms, step 2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                  |
|                              | IR: 16 ~ 34 ms, step 2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | IR: 18~34 ms, step 2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                  |
|                              | STIR: 16 ~ 34 ms, step 2<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | STIR: 18 ~ 34 ms, step 2<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                  |
|                              | SET1HE: 10 ~ 24 ms, step<br>2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | SET1HE: 12 ~ 24 ms, step<br>2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                  |
|                              | SET1HF: 16 ~ 34 ms, step<br>2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | SET1HF: 18 ~ 34 ms, step<br>2 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                  |
|                              | TSE: 80 ~ 120 ms, step 10<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | TSE: 80 ~ 120 ms, step 10<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                  |
|                              | TME: 1st echo 38 ms; 2nd<br>echo 90 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | TME: 1st echo 38 ms; 2nd<br>echo 90 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                  |
|                              | TGE: 8 ~ 24 ms, step 1 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | TGE: 8 ~ 24 ms, step 1 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                  |
|                              | GE3D: 8 ~ 24 ms, step 1<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | GE3D: 14~24 ms, step 2<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                  |
|                              | T3D_T1: 8~24 ms, step<br>1 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | T3D_T1: 8~24 ms, step<br>1 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                  |
|                              | T3D_T2 : fixed at 16 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | T3D_T2 : fixed at 16 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                  |
|                              | GEFS: 8 ~ 24 ms, step 1<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | GEFS: 12 ~ 24 ms, step 2<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                  |
|                              | GEMT: 8 ~ 24 ms, step 1<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | GEMT: 8 ~ 24 ms, step 1<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                  |
|                              | GE3D MT: 8 ~ 24 ms, step<br>1 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | GE3D MT: 8 ~ 24 ms, step<br>1 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Unchanged                                                                                                                                                                                                        |
| TI                           | IR: 200 ~ 800 ms, step 10<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | IR: 200 ~ 800 ms, step 10<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                  |
|                              | STIR: 50~200 ms, step 5<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | STIR: 50~200 ms, step 5<br>ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                  |
| Characteristic               | ARTOSCAN S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ARTOSCAN M<br>K963262                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comments                                                                                                                                                                                                         |
|                              | GEFS : 20 ~ 200 ms, step<br>5 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | GEFS : 20 ~ 200 ms, step<br>5 ms…

---

**Source:** [https://fda-staging.innolitics.com/device/K981358](https://fda-staging.innolitics.com/device/K981358)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com