K981289 · Crestat Diagnostics, Inc. · CIG · Apr 27, 1998 · Clinical Chemistry

Device Facts

Record IDK981289
Device NameN-ASSAY L T-BIL
ApplicantCrestat Diagnostics, Inc.
Product CodeCIG · Clinical Chemistry
Decision DateApr 27, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1110
Device ClassClass 2

Indications for Use

The intended use for the N-ASSAY L T-BIL Reagent is for the quantitative determination of serum Total Bilirubin in the diagnosis and treatment of various liver diseases. For in vitro diagnostic use only.

Device Story

N-ASSAY L T-BIL is a ready-to-use liquid reagent for quantitative measurement of total bilirubin in human serum. Used in clinical laboratory settings for liver function screening. Principle of operation: chemical oxidation method using sodium nitrite at pH 3.7. Serum sample mixed with reagent; total bilirubin oxidized to biliverdin; causes decrease in absorbance at 450 nm (main) and 546 nm (sub). Absorbance measured before and after oxidation to calculate concentration. Provides clinicians with quantitative data to assist in diagnosing and monitoring liver disease.

Clinical Evidence

Bench testing only. Comparison studies against predicate assay yielded correlation coefficient of 0.99922 and regression equation y = 1.0166x + 0.1192. Precision studies indicate acceptable day-to-day performance. Minimum detectable level is 0.05 mg/dl; linearity up to 50 mg/dl.

Technological Characteristics

Liquid reagent; chemical oxidation method using sodium nitrite; pH 3.7; spectrophotometric detection at 450 nm and 546 nm; linear range 0.05 to 50 mg/dl.

Indications for Use

Indicated for the quantitative determination of serum Total Bilirubin in patients requiring diagnosis and treatment of liver diseases. For in vitro diagnostic use.

Regulatory Classification

Identification

A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K981289 ## CRESTAT DIAGNOSTICS, INC. 25549 Adams Avenue Murrieta, CA 92562 1300/HAGG APR 2 7 1998 ### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92. The assigned 510(k) number is: Date: April 2, 1998 Colin Getty Submitted by: KAMIYA BİOMEDICAL COMPANY 910 Industry Drive, Seattle WA 98188 TEL: 206-575-8068; FAX: 206-575-8094 For: Murrieta, CA 92562 N-ASSAY L T-BIL (Total Bilirubin Assay Reagent) Product: Crestat Diagnostics, Inc. 25549 Adams Avenue Serum bilirubin measurement is widely used as a screening test for liver functions. The diazo coupling method and bilirubin oxidase enzymatic method are widely used; however, these methods have disadvantages such as interference by coexistent serum substances and poor reagent stability. The N-ASSAY L T-BIL reagent is based on a chemical oxidation method, utilizing sodium nitrite as an oxidizing agent, shows good correlation with similar Total Bilirubin reagents, practically no interference by coexistent substances, high sensitivity with good reproducibility, wide assay range, and is convenient ready-to-use liquid type reagent. In this method, a serum sample containing total bilirubin is mixed with the reagent containing sodium nitrite. Total bilirubin (direct bilirubin) is oxidized by nitrite at pH 3.7 to produce biliverdin which causes the absorbance at 450 nm (main) and 546 (sub), specific to bilirubin, to decrease. Therefore, the total billinubin concentration in the sample can be obtained by measuring the absorbance before and after the sodium nitrite oxidation. {1}------------------------------------------------ # CRESTAT DIAGNOSTICS, INC. ### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS (Continued) The safety and effectiveness of the liquid Crestat N-ASSAY L T-BIL Reagent is demonstrated by its substantial equivalence to the WAKO Chemicals Total Bilirubin liquid reagent (K970985) which is also based on chemical oxidation. Both test systems are intended to quantitatively measure Total Bilirubin in human serum. In comparison studies against the predicate assay, a correlation coefficient of 0.99922 and a regression equation y = 1.0166 x + 0.1192 was obtained with serum samples. Precision studies indicate acceptable values can be obtained on a day to day basis. The minimum detectable level of this method is 0.05 mg/dl. The N-ASSAY L T-BIL reagent is linear to 50 mg/dl. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized representation of a human figure with three heads in profile, facing to the right. The figure is composed of flowing lines, creating a sense of movement and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 APR 27 1998 Crestat Diagnostics, Inc. Colin Getty C/O KAMIYA Biomedical Company 910 Industry Drive Seattle, Washington 98188 K981289 Re: N-ASSAY L T-BIL (Total Bilirubin Assay Reagent) Requlatory Class: II Product Code: CIG Dated: April 2, 1998 Received: April 8, 1998 Dear Mr. Getty: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations . {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Sitman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation -----------Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### INDICATIONS FOR USE STATEMENT 510(k) Number (if known): _ N-ASSAY L T-BIL (Total Bilirubin Assay Reagent) Device Name: #### Indications For Use: The intended use for the N-ASSAY L T-BIL Reagent is for the quantitative determination of serum Total Bilirubin in the diagnosis and treatment of various liver diseases. For in vitro diagnostic use only. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number. k981289 Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use _ Optional Format 1-2-96) OR
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...