K981277 · Verimed Holdings, Inc. · HIR · Nov 20, 1998 · Obstetrics/Gynecology
Device Facts
Record ID
K981277
Device Name
PERIFORM
Applicant
Verimed Holdings, Inc.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Nov 20, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Biofeedback/monitoring vaginal probe intended for: - Treatment of stress, urge, and mixed incontinence. . - Facilitation in identification of the pubococcygeus muscles. . - . Strengthening of the pelvic floor muscles through biofeedback-assisted pelvic floor exercises.
Device Story
Periform® is a vaginal perineometric probe for acquiring superficial EMG signals from the human vaginal wall. It functions as a passive sensor; it does not deliver energy. The probe is inserted into the vagina to detect electrical action potentials generated by the pubococcygeus muscle during contraction. These signals are transported to an external biofeedback device for processing. Used in hospitals, clinics, doctor's offices, or home settings under clinician supervision. The output provides biofeedback to the patient, facilitating pelvic floor muscle exercises to improve voluntary control and strength. The device includes a Pelvic Floor Contraction Indicator (PFCI) for qualitative evaluation of musculature. It is a reusable, single-patient device.
Clinical Evidence
Bench testing only. Impedance testing confirmed the probe effectively transports electrical signals with minimal attenuation (approx. 10 ohms) across frequencies relevant to pelvic floor biofeedback (<200 Hz). Biocompatibility testing for body materials (Polystyrene type 564) met USP Class VI requirements; electrodes (Medical Grade Stainless Steel 316S31) were confirmed nontoxic for transient contact.
Technological Characteristics
Passive vaginal probe; no energy delivery. Materials: High Impact Polystyrene (type 564) body, Medical Grade Stainless Steel (316S31) electrodes. Dimensions: 76 mm shaft length, 34 mm width, 28.2 mm flange diameter. Sensing: 2-electrode longitudinal orientation for sEMG. Connectivity: Wired connection to external biofeedback device. Sterilization: Non-sterile; cleaning instructions provided. No software or active electronics.
Indications for Use
Indicated for adult female patients with stress, urge, or mixed urinary incontinence requiring pelvic floor muscle strengthening or identification of pubococcygeus muscles via biofeedback. Contraindicated for pregnant patients, those with menstrual, bladder, or vaginal infections, history of urinary retention, diminished mental/physical capacity, or anatomical structures preventing proper insertion. Do not use with electrical muscle stimulation.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Verimed International Inc. The logo is black and white and features a stylized "V" followed by the word "ERIMED" in bold, sans-serif font. Below the word "ERIMED" is the text "International Inc." in a smaller font size.
K98/277
510(k) Summary
# Summary of Safety and Effectiveness
## Periform® Perineometric Probe and Pelvic Floor Contraction Indicator
Manufactured by:
NEEN HealthCare Old Pharmacy Yard, Church Street Dereham, Norfolk NR19 1DJ England, United Kingdom.
Sold by:
Verimed International, Inc. 11950 NW 39 St., Suite D Coral Springs, FL 33065 {USA}
Telephone: (954) 344-2454 Fax: (954) 340-8812
Contact Person:
Signature
Kishore Puppala, Product Manager, Verimed International, Inc.
April 2, 1998.
Date Submitted
{1}------------------------------------------------
## Name of Device
Trade Name: Periform® Common name: Perineometric Probe Classification: Perineometer (per 21 CFR section 884.1425)
## Identification of Predicate Device
Verimed contends that the Periform® probe is substantially equivalent to the Verimed Electromvographic Perineometer™ [510(k) number K913912] and the PerryVaginal™ Sensor (also known as the PerryMeter Vaginal Perineometer) {510(k) number K911190].
## Description of Device
The Periform® is a vaginal perineometric probe designed specifically for the acquisition of the naturally present superficial EMG signals of the human vaginal wall. Surface electromyography (sEMG) is the technique of measuring and filtering the electrical impulse (action potential) signals generated by muscle fibers during contraction. The probe conducts these signals to a biofeedback device for further processing. The anatomy of the pelvic floor presents a difficulty in obtaining these signals. The use of a perineometric probe overcomes this difficulty.
The purpose of the Periform® is to improve the voluntary control and strength of the pelvic floor muscles with the aid of biofeedback equipment.
#### Intended Uses
#### Indications
- Treatment of stress, urge, and mixed incontinence. �
- Sensor for SEMG behavioral training. �
- Facilitation in identification of the pubococcygeus muscles. �
- � Strengthening of the pelvic floor muscles through biofeedbackassisted pelvic floor exercises.
- � Qualitative evaluation of the pelvic floor musculature using the Pelvic Floor Contraction Indicator.
## Contraindications/ DO NOT USE
- � Do not use during the menstrual period or when pregnant.
- � Do not use if symptoms of a bladder infection are present.
- Do not use if symptoms of a vaginal infection are present. �
- Do not use if patient has a history of urinary retention or symptoms � thereof.
- Do not use with patients who have diminished mental capacity or � physical competence or who cannot handle the device properly.
- Do not use while pregnant or while attempting to become pregnant. �
{2}------------------------------------------------
- ♦ Do not use if patient has an anatomical vaginal morphology and/or structure that does not permit proper insertion of the probe.
- ◆ Do not use this device in conjunction with electrical muscle stimulation.
- ♦ Do not use if patient is unable or unwilling to use device as directed and indicated.
| | Periform® | Verimed<br>Electromyographic<br>Perineometer™ | PerryVaginal™<br>Sensor |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| 510(k) number | | K913912 | K911190 |
| Mode of Use | Reusable for a single<br>patient | Same. | Same. |
| Parameter monitored | aggregate surface<br>electromyogram | Same. | Same. |
| User feedback | None | None | None |
| Intended Use | To produce<br>biofeedback signals<br>for processing by an<br>external biofeedback<br>device. | Same. | Same. |
| Indications for Use | Treatment of stress<br>and urge incontinence<br>and facilitation of<br>pelvic floor exercises. | Same. | Treatment of stress,<br>urge, and mixed<br>incontinence. |
| Performance<br>Standards | Currently no performance standards exist for sEMG biofeedback<br>perineometric probes. | | |
| Target Population | Adult female urinary<br>incontinence patients. | Same. | Same. |
| Anatomical Sites<br>designated for use. | Female<br>pubococcygeus<br>muscle area. | Same. | Same. |
| Energy used and/ or<br>delivered. | No energy delivered or<br>used, only<br>transported. | Same. | Same. |
| Compatibility with<br>environment and other<br>devices | Probe is not known to<br>conflict with other<br>devices or cause<br>environmental<br>hazards. | Same. | Same. |
| Where used | Hospitals, Clinics,<br>Doctor's offices, or<br>home use under<br>clinician's<br>supervision. | Same. | Same. |
| Sterility | Probe does not need<br>to be sterile.<br>Appropriate cleaning<br>procedure included<br>instructions for use. | Same. | Same. |
| # of Electrode Contacts | 2 | 2 | 3 |
| # of Lead Cables | 2 | 2 | 1 |
| Transducers | None | None | None |
| Body Material | High Impact<br>Polystyrene, type 564 | Plastic: Cycolac ABS-<br>GPM-5600-4500-FGE | N / A |
| Biocompatibility of<br>Body Material | Biocompatible | Biocompatible | Biocompatible |
| | | | 91 |
| Pelvic Floor<br>Contraction Indicator<br>(PFCI) included? | Yes. | No. | No. |
| PFCI Tip and End<br>Material | High Impact<br>Polystyrene 564 | N/A | N/A |
| PFCI Shaft Material | 'Stamylan' low density<br>polyethylene grade<br>2100 TNOO | N/A | N/A |
| Electrode Material | Medical Grade<br>Stainless Steel type<br>316S31 | Series 300 Passivated<br>Stainless Steel | Series 300 Passivated<br>Stainless Steel |
| Biocompatibility of<br>Electrode Material | Biocompatible | Biocompatible | Biocompatible |
| Electrode Orientation | Longitudinal | Circumferential | Longitudinal |
| Construction | 2-piece mold | 3-piece mold | 3-piece mold |
| Sensing Method | sEMG biofeedback<br>recording | Same. | Same. |
| Feedback Modes | No direct feedback to<br>user | Same. | Same. |
| Electrical Safety | Device is passive: Not<br>electrically powered | Same. | Same. |
| Electrodes and<br>Conductive Media | Not applicable. | Not applicable. | Not applicable. |
| Intentional Electrical<br>Current | Warning included in<br>Instructions for Use.<br>Socket type prevents<br>easy intentional<br>connection to AC<br>source. | Warning included in<br>Instructions for Use. | Warning included in<br>Instructions for Use. |
| Unintentional<br>Electrical Current | Probe is to be used<br>with appropriate<br>biofeedback device as<br>stated in Basic Design<br>Description, or with<br>PFCI only. | Probe is to be used<br>with biofeedback<br>devices satisfying<br>criteria listed in Basic<br>Design Description. | Probe is to be used<br>with biofeedback<br>devices satisfying<br>criteria listed in Basic<br>Design Description. |
| Risk of Mechanical<br>Injury | Risk prevention<br>considered in design<br>process (Attached). | Risk prevention<br>considered in design<br>process (Attached). | Unknown. |
| Risk of thermal or<br>Radiation Injury | None: probe does not<br>direct any form of<br>electromagnetic<br>energy or heat. | Same. | Same. |
| Chemical Safety | Body and electrodes<br>constructed of<br>chemically inert<br>materials. | Same. | Same. |
| Shaft Length | 76 mm | 81 mm | 55.9 mm |
| Shaft Diameter | Not applicable (N/A) | 19.1 mm | 17.8 mm |
| Probe Width | 34 mm | N/A | N/A |
| Basal Bulb Diameter<br>(Comparable to<br>Periform's Flange<br>Diameter) | N/A | 63.5 mm | 43.2 mm |
| Flange Diameter | 28.2 mm | N/A | N/A |
| Terminal Bulb<br>Diameter | N/A | 31.8 mm | 33.0 mm |
| Terminal Bulb Length | N/A | 22.7 mm | 26.7 mm |
| Electrode Spacing<br>Distance | N/A | 9.5 mm | 11.4 mm |
| Spacing Type | Axial Incremental | Radial Incremental | Axial Incremental |
| | | | C<br>U- |
| Electrode Surface Area <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 4.9 cm² x 2 | .5 cm x 2 | 1.2 cm4 x 3 |
#### Device Comparison
{3}------------------------------------------------
{4}------------------------------------------------
#### Non-Clinical Performance Data
An impedance test was performed to demonstrate that the Periform® is capable of transporting electrical signals with minimal attenuation. Various frequencies were used in order to verify that the Periform® would be effective in the range of frequencies encountered in pelvic floor biofeedback (typically less than 200 Hz). The test shows that the Periform® exhibits the low impedance values (around 10 ohms at all tested frequencies) needed for accurate recording of pelvic floor muscle activity.
## Biocompatibility Testing
The materials used in the Periform® have undergone safety tests and were found to be safe according to the required standards for each test. Tests performed and results were as follows:
| MATERIAL | TEST PERFORMED | RESULTS |
|-------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Polystyrene type 564.<br>Used in Periform® body. | USP tests for Class VI<br>plastics (Systemic Injection,<br>Intracutaneous Toxicity,<br>Implantation Test) | Meets the requirements of<br>USP Class VI Plastic |
| Medical Grade Stainless<br>Steel 316S31.<br>Used for Periform®<br>electrodes. | Standard Item used in the<br>manufacture of medical<br>device to contact human<br>tissue for transient periods<br>(less than 2 hours) | Nontoxic. |
| Polythene.<br>Used in Pelvic Floor<br>Contraction Indicator.<br>Not in contact with<br>patient. | No test performed as this<br>portion is not intended to<br>contact the body. | N/A |
## Conclusion
The Periform® is safe and effective for its intended use and is substantially equivalent to its predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure with three flowing lines representing the body and arms. The text "HUMAN SERVICES USA" is arranged in a circular pattern around the figure, indicating the department's focus on human services within the United States.
DEPARTMENT OF HEALTH & HUMAN SERVICES
NOV 20 1908
Kishore Puppala Product Manger Verimed International, Inc. 11950 N.W. 39th Street, Suite D Coral Springs, Florida 33065
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Re: K981277 Periform® Perineometric Probe Dated: August 28, 1998 Received: August 31, 1998 Regulatory Class: II 21 CFR884.1425/Procode: 85 HIR
Dear Mr. Puppala:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the ensctment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours
Lillian Yin, Ph.D.
Director. Division of Reproductive. Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K98 1277
Device Name: Periform® Perineometric Probe
Indications For Use:
Biofeedback/monitoring vaginal probe intended for:
- Treatment of stress, urge, and mixed incontinence. .
- Facilitation in identification of the pubococcygeus muscles. .
- . Strengthening of the pelvic floor muscles through biofeedback-assisted pelvic floor exercises.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seggern
(Division Sign-Off)
ion Sign-Off) Division of Reproductive, Abdominal, El and Radiological De 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.