# TOXOGEN (K981263)

_Instrumentation Laboratory CO · GMM · Aug 6, 1998 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981263

## Device Facts

- **Applicant:** Instrumentation Laboratory CO
- **Product Code:** [GMM](/productcode/GMM.md)
- **Decision Date:** Aug 6, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3780
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

Toxogen is an in vitro diagnostic, rapid latex particle agglutination test for the qualitative and semiquantitative determination of total anti-toxoplasma antibodies in human serum. The test may be used with a single specimen to estimate the immunologic response of the individual or with paired acute and convalescent sera as an aid in the diagnosis of primary, acute or reactivated toxoplasmosis. This assay is not FDA cleared for use in blood or plasma donor screening.

## Device Story

Toxogen is an in vitro diagnostic latex particle agglutination test; detects total anti-toxoplasma antibodies in human serum. Device uses larger latex particles than predicate to increase sensitivity from 10 IU/mL to 5 IU/mL. Performed by laboratory personnel; provides qualitative or semiquantitative results. Clinicians use results from single specimens or paired acute/convalescent sera to estimate individual immunologic response and aid diagnosis of primary, acute, or reactivated toxoplasmosis. Output is visual agglutination observed by the operator.

## Clinical Evidence

Method comparison study of 100 serum samples showed 100% relative sensitivity and 91% relative specificity compared to the predicate. Reproducibility study using a 9-sample panel (negative, weak positive, high positive) tested by three operators over three days demonstrated 100% reproducibility (agreement within one 2-fold dilution).

## Technological Characteristics

In vitro diagnostic latex particle agglutination assay. Employs larger latex particles compared to predicate to achieve 5 IU/mL sensitivity. Manual test format.

## Regulatory Identification

Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.

## Predicate Devices

- Toxogen ([K913467](/device/K913467.md)/D)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SECTION 3 Toxogen - 510(k) SUMMARY (Summary of Safety and Effectiveness)

## Submitted by:

Carol Marble Senior Regulatory Affairs Specialist Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02173 Phone: (781) 861-4467 (781) 861-4464 Fax:

#### Contact Person:

Carol Marble Phone: (781) 861-4467

### Summary Prepared:

April 6, 1997

#### Name of the device:

Toxogen

#### Classification name(s):

866.3780 Toxoplasma gondii serological reagents Class II

## Identification of predicate device(s):

K913467/D Toxogen

### Description of the device/intended use(s):

Toxogen is an in vitro diagnostic, rapid latex particle agglutination test for the qualitative and semiquantitative determination of total anti-toxoplasma antibodies in human serum. The test may be used with a single specimen to estimate the immunologic response of the individual or with paired acute and convalescent sera as an aid in the diagnosis of primary, acute or reactivated toxoplasmosis. This assay is not FDA cleared for use in blood or plasma donor screening.

## Statement of how the Technological Characteristics of the Device compare to the Predicate device:

The new Toxogen is a modified version of the predicate Toxogen with a larger latex particle size to increase sensitivity from 10 IU/mL to 5 IU/mL. It is substantially equivalent in performance, intended use, safety and effectiveness to the predicate device as supported by the performance data and labeling.

#### Summary of Performance Data:

In a method comparison study evaluating 100 serum samples using the qualitative method, the relative sensitivity of the modified Toxogen as compared to the predicate Toxogen was 100% and the relative specificity was 91%.

In a reproducibility study, a panel of 9 serum samples (3 negative, 3 weak positive and 3 high positive) was tested by three different operators on 3 consecutive days following the semiquantitative procedure for Toxogen. The results, as defined by the ability to give agreement to within one 2-fold dilution on the replicates indicated 100% reproducibility.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG 6 1998

Carol Marble Senior Regulatory Affairs Specialist Instrument Laboratory 101 Hartwell Avenue Lexington, MA 02173-3190

K981263 Re: Trade Name: Toxogen Regulatory Class: II Product Code: GMM Dated: June 10, 1998 Received: June 11, 1998

Dear Ms. Marble:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely vours.

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Toxogen

## Indications for Use:

Toxogen is an in vitro diagnostic, rapid latex particle agglutination test for the qualitative and semiquantitative determination of total anti-toxoplasma antibodies in human serum. The test may be used with a single specimen to estimate the immunologic response of the individual or with paired acute and convalescent sera as an aid in the diagnosis of primary, acute or reactivated toxoplasmosis.

This assay is not FDA cleared for use in blood or plasma donor screening.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubois

510(k) Number

| Prescription Use     |  |
|----------------------|--|
| (Per 21 CFR 801.019) |  |

Over-The-Counter Use

Toxogen 510(k)

OR

---

**Source:** [https://fda-staging.innolitics.com/device/K981263](https://fda-staging.innolitics.com/device/K981263)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com