← Product Code [FZP](/productcode/FZP) · K981239

# DEROYAL INDUSTRIES DISPOSABLE LAPAROSCOPIC CLIP APPLIER WITH IMPLANTABLE TITANIUM CLIPS (K981239)

_Deroyal Surgical · FZP · May 14, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981239

## Device Facts

- **Applicant:** Deroyal Surgical
- **Product Code:** [FZP](/productcode/FZP.md)
- **Decision Date:** May 14, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The DeRoyal Industries Disposable Laparoscopic Clip Applier with Implantable Titanium Clips is intended for use in a variety of laparoscopic procedures to occlude tubular structures and vessels. The tissue being ligated should be consistent with the size of the clip.

## Device Story

Disposable, single-patient-use laparoscopic clip applier; fabricated from biocompatible stainless steel and medical-grade polymers. Device contains up to 20 implantable titanium clips. Operated by surgeon during laparoscopic procedures; shaft sized for 10 mm cannula. Principle of operation: manual actuation of handle/trigger drives internal mechanism to close jaws around vessel/tubular structure, forming clip; release of handle opens jaws to release clip. Automatic feed feature advances next clip into jaws. Output is a securely formed titanium clip providing vessel occlusion. Benefits include efficient, disposable ligation during minimally invasive surgery.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Materials: biocompatible stainless steel and medical-grade polymers. Form factor: disposable handheld applier with shaft for 10 mm cannula; 33 mm shaft length. Mechanism: manual trigger-actuated mechanical clip closure and automatic feed. Clip: titanium, ~9 mm closed length. Sterility: sterile.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- United States Surgical Corporation
- Ethicon
- Pilling Weck (formerly Edward Weck)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K981239

MAY | 4 1998

# 510(k) SUMMARY: DeRoyal Industries, Inc. Disposable Laparoscopic Clip Applier with Implantable Titanium Clips

- DeRoyal Industries, Inc. (1) 200 DeBusk Lane Powell, TN 37849
Contact Person: Telephone: Date Summary Prepared:

- Trade or Proprietary Name: (2)
Common Name:

Classified Name:

(3) Predicates: Camille Matlock (423) 938-7828 April 1, 1998

> DeRoyal Industries Disposable Laparoscopic Clip Applier with Implantable Titanium Clips

> Disposable Laparoscopic Clip Applier with Implantable Titanium Clips

Implantable Clip (74FZP, 21 CFR & 878.4300), Class II

Endoscope and accessories, (78KOG, 21 CFR & 878.1500). Class II

United States Surgical Corporation (! Ethicon . Pilling Weck (formerlv Edward Weck)/ ﺗﻤﺖ ﺍ

{1}------------------------------------------------

### Description of Device: (4)

The DeRoyal Disposable Laparoscopic Clip Applier with Implantable Titanium Clips is a disposable single patient use device fabricated from biocompatible stainless steels and biocompatible medical grade polymers. The Disposable Clip Applier contains up to 20 implantable titanium clips. The closed clip length is approximately 9 mm. The intended use of the device is to occlude vessels and other tubular structures during laparoscopic surgical procedures. The clip applier jaws, with the open clip in the jaws, is placed around the vessel or other tubular structure. Actuation/compression of the clip applier handle/trigger drives a mechanism within the device to close the jaws, thereby forming the clip securely around the vessel. Release/decompression of the clip applier handle/trigger allows the jaws to open and release the clip. The automatic feed feature releases the next clip into the jaws for another application.

The shaft of the applier is sized to fit through a 10 mm cannula. The overall shaft length is approximately 33 mm and consistent with other endoscopic instruments.

- Intended Use: (5)
The DeRoval Disposable Clip Applier with Implantable Titanium Clips is intended for use during laparoscopic surgery to occlude a variety of vessels or other tubular structures.

| Characteristics          | Predicate Devices                                                                   | DeRoyal Industries |
|--------------------------|-------------------------------------------------------------------------------------|--------------------|
| Intended Use             | Endoscopic procedures to achieve occlusion of vessels and other tubular structures. | Same               |
| Clip Surface             | Textured                                                                            | Same               |
| Laparoscopic Intro. Size | $       \cong 10     $ MM                                                           | Same               |
| Clip Dimensions          | $       \cong 9     $ MM                                                            | Same               |
| Sterility                | Sterile                                                                             | Same               |
| Materials                | Polymers, stainless steel, and black chrome coating.                                | Same               |

#### (6) Technological Characteristics:

{2}------------------------------------------------

The DeRoyal Disposable Clip Applier with Implantable Titanium Clips has similar/same technological characteristics as the predicate devices in that they are comprised of similar design, similar materials, and are intended to be used to clip vessels or tubular structures during laparoscopic surgery.

## Conclusion: (7)

The proposed device has the same intended use and the same basic technology as the legally marketed predicate devices identified in the premarket notification submission. The proposed device contains, in some combination, similar/same features, materials, and design as the predicate devices and does not pose any new questions concerning safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three stripes representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# MAY | 4 1998

Ms. Camille Matlock ·Requlatory Affairs DeRoyal Industries, Incorporated 200 DeBusk Lane Powell, Tennessee 37849

K981239 Re: Trade Name: Disposable Laparoscopic Clip Applier with Implantable Titanium Clips Regulatory Class: II Product Code: FZP and KOG April 1, 1998 Dated: April 3, 1998 Received:

Dear Ms. Matlock:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Ms. Matlock

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Colia M. Witten, Ph.D., M.

Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

K981239

510(k) Number (if known):

Disposable Laparoscopic Clip Applier with Implantable

Titanium Clips

Indications for Use:

Device Name:

The DeRoyal Industries Disposable Laparoscopic Clip Applier with Implantable Titanium Clips is intended for use in a variety of laparoscopic procedures to occlude tubular structures and vessels. The tissue being ligated should be consistent with the size of the clip.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) ીન્દ્ર nei Ji Restorative Devic Olk) Number Prescription Use OR Over-The-Counter Use (Per 21 CFR § 801.109)

---

**Source:** [https://fda-staging.innolitics.com/device/K981239](https://fda-staging.innolitics.com/device/K981239)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
