← Product Code [GGP](/productcode/GGP) · K981200

# IL TEST PLASMINOGEN (K981200)

_Instrumentation Laboratory CO · GGP · May 26, 1998 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981200

## Device Facts

- **Applicant:** Instrumentation Laboratory CO
- **Product Code:** [GGP](/productcode/GGP.md)
- **Decision Date:** May 26, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.7290
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

IL Test™ Plasminogen is an in viro diagnostic test for the quantitative determination of plasminogen in human citrated plasma based on a synthetic chromogenic substrate. Dysplasminogenemia is associated with recurrent venous thrombosis. Acquired deficiencies of Plasminogen are associated with thrombolytic therapy, sepsis and Disseminated Intravascular Coagulation.

## Device Story

IL Test™ Plasminogen is an in vitro diagnostic reagent used for quantitative measurement of plasminogen levels in human citrated plasma. The assay utilizes a synthetic chromogenic substrate to determine plasminogen concentration. It is intended for use in clinical laboratory settings by trained personnel. The device output provides quantitative plasminogen activity levels, which assist clinicians in diagnosing dysplasminogenemia and monitoring acquired deficiencies related to sepsis, thrombolytic therapy, or Disseminated Intravascular Coagulation. The test is performed on automated coagulation analyzers (e.g., ACL 300, ACL Futura).

## Clinical Evidence

Bench testing only. Method comparison studies evaluated 51 plasma samples, showing correlation (r) of 0.983 (ACL 300) and 0.989 (ACL Futura) against the predicate. Within-run precision (CV) ranged from 1.7% to 4.4% across two levels of plasma activity.

## Technological Characteristics

In vitro diagnostic reagent using a synthetic chromogenic substrate for plasminogen quantification. Designed for use on automated coagulation analyzers (ACL 300, ACL Futura).

## Predicate Devices

- IL Test™ Plasminogen ([K864212](/device/K864212.md)/A)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## MAY 2 6 1998

K981200

# Section 3 IL Test™ Plasminogen - 510(k) SUMMARY (Summary of Safety and Effectiveness)

### Submitted by:

Carol Marble Senior Regulatory Affairs Specialist Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02173 Phone: (781) 861-4467 (781) 861-4464 Fax:

#### Contact Person:

Carol Marble Phone: (781) 861-4467

#### Summary Prepared:

April 1, 1998

## Name of the device:

IL Test™ Plasminogen

#### Classification name(s):

864.7290 Factor Deficiency Test Class II Test, Qualitative and Quantitative Factor Deficient 81GGP

#### Identification of predicate device(s):

IL Test™ Plasminogen K864212/A

#### Description of the device/intended use(s):

IL Test™ Plasminogen is an in vitro diagnostic test for the quantitative determination of plasminogen in human citrated plasma based on a synthetic chromogenic substrate. Dysplasminogenemia is associated with recurrent venous thrombosis. Acquired deficiencies of Plasminogen are associated with thrombolytic therapy, sepsis and Disseminated Intravascular Coagulation.

## Statement of How the Technological Characteristics of the Device Compare to the Predicate device:

The new IL Test™ Plasminogen uses the same test principle as the predicate II. Test™ Plasminogen and is substantially equivalent in performance, intended use and safety and effectiveness.

#### Summary of Performance Data:

In method comparison studies evaluating 51 plasma samples, the correlation (r) of the new IL Test™ Plasminogen to the predicate IL Test™ Plasminogen on the ACL 300 was 0.983 and on the ACL Futura was 0.989.

On the ACL 300, within run precision assessed over multiple runs using 2 levels of plasma gave a CV of 4.4% (at a mean of 50.1% activity) and 2.8% (at a mean of 95.7% activity). On the ACL Futura, within run precision assessed over multiple runs using 2 levels of plasma gave a CV of 1.8% (at a mean of 52.5% activity) and 1.7% (at a mean of 97.6% activity).

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 2 6 1998

Carol Marble Senior Requlatory Affairs Specialist Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, Massachusetts 02173-3190

K981200 Re : Il Test Plasminogen Regulatory Class: II Product Code: GGP April 1, 1998 Dated: Received: April 2, 1998

Dear Ms. Marble:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ಗ್ಗೆ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use Statement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: IL Test™ Plasminogen

## Indications for Use:

IL Test™ Plasminogen is an in viro diagnostic test for the quantitative determination of plasminogen in human citrated plasma based on a synthetic chromogenic substrate. Dysplasminogenemia is associated with recurrent venous thrombosis. Acquired deficiencies of Plasminogen are associated with thrombolytic therapy, sepsis and Disseminated Intravascular Coagulation.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                     |         |
|-----------------------------------------|---------|
| Division of Clinical Laboratory Devices |         |
| 510(k) Number                           | k981200 |

| Prescription Use     | <div> <span style="text-decoration: overline;">✓</span> </div> | OR | Over-The-Counter Use |
|----------------------|----------------------------------------------------------------|----|----------------------|
| (Per 21 CFR 801.019) |                                                                |    |                      |

| Section 2 | IL Test™ Plasminogen 510(k) | Page 1 of 1 |
|-----------|-----------------------------|-------------|
|-----------|-----------------------------|-------------|

---

**Source:** [https://fda-staging.innolitics.com/device/K981200](https://fda-staging.innolitics.com/device/K981200)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).