← Product Code [DKZ](/productcode/DKZ) · K981197

# ONE STEP URINE DRUG OF ABUSE AMPHETAMINE TEST (K981197)

_Tcpi, Inc. · DKZ · Jun 16, 1998 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981197

## Device Facts

- **Applicant:** Tcpi, Inc.
- **Product Code:** [DKZ](/productcode/DKZ.md)
- **Decision Date:** Jun 16, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3100
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The TCPI One Step™ Urine Drug of Abuse Amphetamine Test is a rapid, qualitative, competitive binding immunoassay for the determination of Amphetamine in urine. The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The TCPI One Step™ Urine Drug of Abuse Amphetamine assay is not intended to monitor drug levels, but only to screen urines for the presence of amphetamine and its metabolites.

## Device Story

One Step™ Urine Drug of Abuse Amphetamine Test is a chromatographic absorbent device for qualitative urine screening. Principle: competitive binding immunoassay; drug/metabolites in sample compete with immobilized drug conjugate on porous membrane for limited antibody sites. If drug concentration exceeds 500 ng/ml, labeled antibody-dye conjugate binds to free drug, preventing magenta band formation in positive reaction zone. Control zone binds unbound dye to confirm device function. Used in professional medical/forensic settings. Results are preliminary; require GC/MS confirmation. Assists clinicians in identifying potential amphetamine presence to guide further diagnostic or forensic decision-making.

## Clinical Evidence

Bench testing and clinical evaluation performed. Laboratory studies against GC/MS on 260 samples showed 99.62% accuracy. Clinical trials on 300 patient samples showed 96.00% accuracy. Combined total accuracy was 97.68% (560 samples). Excluding 12 ephedrine-positive samples, accuracy was 99.82%.

## Technological Characteristics

Chromatographic absorbent device; porous membrane support; competitive binding immunoassay format. Detects amphetamine at 500 ng/ml cutoff. Visual readout via magenta color bands.

## Regulatory Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 981197

510k Submission for

JUN 16 1998 One Step™ Urine Drug of Abuse Amphetamine Test Technical Chemicals & Products, Inc.

Page 59 of 59

Revision A -3/15/98 Printed (3/31/98)

## Statement of Safety and Effectiveness

The sponsor, Technical Chemicals and Products, Inc. (3340 S.W. 15th Street, Pompano Beach, Florida, 33069), has developed, manufactured and tested under GMP/GLP guidelines a device for the qualitative testing of urine for the presence of Amphetamine and its metabolites in a screening format. This summary was originally written in May of 1995 and has been updated as of March, 1998.

The trade name of the device is One Step™ Urine Drug of Abuse Amphetamine Test having a designated common name of Amphetamine Test System and a classification as a Class II device per 21 CFR ¶ 862.3100. This device is intended for professional medical/forensic screening of urine for Amphetamine.

TCPI's One Step™ Urine Drug of Abuse Amphetamine Test consists of a chromatographic absorbent device in which the drug or drug metabolites in the sample compete with a drug conjugate immobilized on a porous membrane support for the limited antibody sites. As the test sample flows up through the absorbent device, the labeled antibody-dye conjugate binds to the free drug in the specimen forming an antibody:antigen complex. This complex competes with immobilized antigen conjugate in the positive reaction zone and will not produce a magenta color band when the drug is above the detection level of 500 ng/ml. Unbound dye conjugate binds to the reagent in the control zone, producing a magenta color band, demonstrating that the reagents and device are functioning correctly.

The sponsor tested the product in laboratory studies against GC/MS. The accuracy of this 260 sample population was 99.62%. Subsequent clinical trials on 300 patient samples vielded a calculated accuracy of 96.00%. When the results of all testing is combined the total number of comparisons was 560 and the calculated accuracy was found to be 97.68%. If the 12 ephedrine positive samples are dropped the accuracy of the TCPI test is 99.82%. The comparison of results clearly demonstrated no difference between the TCPI One Step™ Urine Drug of Abuse Amphetamine Test and other screening tests as noted in the presentation of results. Like all of the screening methods tested TCPI's will report false positives when ephedrine and phenyl propanolamine are present.

Additional information on this submission may be obtained by contacting Dr. Cleve W. Laird, President, Drial Consultants, Inc. at 805-522-6223(Ca) or by fax at 805-522-1526.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

发行办
: ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 1 6 1898

Dr. Jeffrey Bolts . Director of Quality Assurance Technical Chemicals & Products, Inc. P.O. Box 8726 Ft. Lauderdale, Florida 33310

Re: K981197 One Step™ Urine Drug of Abuse Amphetamine Test Requlatory Class: II Product Code: DKZ Dated: March 31, 1998 Received: April 2, 1998

Dear Dr.Bolts:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In additi n, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{2}------------------------------------------------

Page 2

11

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

な

{3}------------------------------------------------

#### 510(k) Number (if Known): As Yet to be Assigned

### Device Name: One Step™ Urine Drug of Abuse Amphetamine Test

Indications For Use:

#### INTENDED USE

The TCPI One Step™ Urine Drug of Abuse Amphetamine Test is a rapid, qualitative, competitive binding immunoassay for the determination of Amphetamine in urine. The test provides only preliminary data which should be confirmed by other methods such as gas chromatography/mass spectrophotometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated6. The TCPI One Step™ Urine Drug of Abuse Amphetamine assay is not intended to monitor drug levels, but only to screen urines for the presence of amphetamine and its metabolites.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANDOTHER PAGE IF NEEDED)

# (Division Sign-Off) Division of Clinical Laboratory Device 510(k) Number

Concurrence of CDRH, Office of Device Evaluation (ODE)

Perscription Use: 
(Per 21 CFR 801.109 or

Over The Counter Use: (Optional Format 1-2-96)

1244887
 세상남도출장걸

---

**Source:** [https://fda-staging.innolitics.com/device/K981197](https://fda-staging.innolitics.com/device/K981197)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com