INNOVASIVE 3.5MM METAL ROC XS SUTURE BONE FASTENER

K981193 · Innovasive Devices, Inc. · MBI · Jun 16, 1998 · Orthopedic

Device Facts

Record IDK981193
Device NameINNOVASIVE 3.5MM METAL ROC XS SUTURE BONE FASTENER
ApplicantInnovasive Devices, Inc.
Product CodeMBI · Orthopedic
Decision DateJun 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The 3.5mm Metal ROC XS Suture Bone Fasteners are intended for the reattachment of soft tissue to bone for the following indications: SHOULDER Bankart repair Repair of rotator cuff tears Capsular instability Slap lesion repair Acromio-clavicular separation Capsule shift/capsulolabral reconstruction Biceps tenodesis Deltoid repair KNEE Extra-capsular repairs Reattachment of medial collateral ligament Reattachment of lateral collateral ligament Reattachment of posterior oblique ligament Joint capsule closure Patellar ligament and tendon avulsion repair Extra-capsular reconstruction ITB tenodesis ANKLE Lateral and medial instability Achilles tendon reconstruction and repair Mid-foot reconstructions FOOT Hallux valgus reconstruction ELBOW Tennis elbow repair Biceps tendon reattachment Medial and lateral repairs BLADDER NECK SUSPENSION Bladder neck suspension for female urinary incontinence due to urethral hypermobility

Device Story

The 3.5mm Metal ROC XS Suture Bone Fastener is a sterile, single-use orthopedic implant designed for soft tissue-to-bone fixation. The device consists of a shear pin, crown, and anvil. During deployment, a deployment handle pulls the crown upward against the anvil, forcing the crown to expand and contact the surrounding bone. Once seated, the shear pin breaks, leaving the fastener fixated in the pre-drilled bone hole. The device is used by surgeons in clinical settings to secure sutures for various ligament and tendon repairs. It is provided with reusable stainless steel instrumentation, including a drill guide and hole finder, which are autoclavable. The device offers a metal alternative to existing plastic fasteners, utilizing the same mechanical principle of radial compression for fixation.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Material: Stainless steel. Principle: Mechanical radial compression via crown expansion. Form factor: 3.5mm fastener assembly with shear pin, crown, and anvil. Sterilization: Sterile, single-use. Instrumentation: Reusable stainless steel drill guide, deployment handle, and hole finder (autoclavable).

Indications for Use

Indicated for soft tissue-to-bone reattachment in shoulder, knee, ankle, foot, and elbow procedures, and bladder neck suspension for female urinary incontinence due to urethral hypermobility.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JUN 1 6 1998 # 510(K) SUMMARY ## 1. SUBMITTER: Innovasive Devices, Inc. 734 Forest St. Marlborough, MA 01752 Telephone: 508-460-8229 Contact: Stephen M. Page, Manager of Regulatory Affairs Date Prepared: March 31, 1998 # 2. DEVICE: Innovasive 3.5mm Metal ROC XS Suture Bone Fastener Classification Name: Single/multiple component bone fixation appliances and accessories. Trade Name: Innovasive Devices 3.5mm Metal ROC XS Suture Bone Fastener ## 3. PREDICATE DEVICES: The predicate devices used to determine substantial equivalence for the modified Innovasive Devices 3.5mm Metal ROC XS Suture Bone Fastener were: - The 3.5mm Innovasive Devices ROC XS Suture Bone Fastener, a. marketed by Innovasive Devices, Marlborough, MA, and - b. The 3.5mm Innovasive Devices ROC EZ Suture Bone Fastener marketed by Innovasive Devices, Marlborough, MA. # 4. DEVICE DESCRIPTION: The 3.5mm Metal ROC XS suture bone fastener implant tip portion consists of a shear pin, crown and anvil. The crown and anvil are fitted onto the shear pin component such that the crown component is located below the anvil on the shear pin. As the crown is pulled up during the device deployment, the anvil forces it open. The anvil is tapered such that it will fit into the crown component as the deployment action progresses. As the crown is forced open by the anvil, the crown expands to make contact with the surrounding bone. This expanding crown results in the final fixation properties of the device. Once the anvil is completely seated inside the crown, the shearing pin shears from the fastener assembly resulting in the device being fixated into the bone hole. In addition to the Fastener, a stainless steel drill guide is available to establish the proper hole in the bone for the Fastener along with a deployment handle and hole finder. All of the instrumentation except the Fastener will be offered as reusable devices and can be autoclaved in the sterilization tray provided for this purpose. {1}------------------------------------------------ The 3.5mm Metal ROC XS will be available as a sterile, single use device. Both sutured and sutureless versions will be marketed. ## 5. INTENDED USE: The 3.5mm Metal ROC XS Suture Bone Fasteners are intended for the reattachment of soft tissue to bone for the following indications: #### SHOULDER #### KNEE Bankart repair Repair of rotator cuff tears Capsular instability Slap lesion repair Acromio-clavicular separation Capsule shift/capsulolabral reconstruction Biceps tenodesis Deltoid repair Extra-capsular repairs Reattachment of medial collateral ligament Reattachment of lateral collateral ligament Reattachment of posterior oblique ligament Joint capsule closure Patellar ligament and tendon avulsion repair Extra-capsular reconstruction ITB tenodesis #### ANKLE Lateral and medial instability Achilles tendon reconstruction and repair Mid-foot reconstructions FOOT Hallux valgus reconstruction #### ELBOW Tennis elbow repair Biceps tendon reattachment Medial and lateral repairs #### BLADDER NECK SUSPENSION Bladder neck suspension for female urinary incontinence due to urethral hypermobility # 6. COMPARISON OF CHARACTERISTICS: The existing Innovasive Devices 3.5mm ROC XS and 3.5mm ROC EZ Suture Bone Fasteners are comprised of plastic. The devices are used to secure a suture in a predrilled hole in bone. They remain fixed in the bone through radial compression as the device is deployed. The proposed 3.5mm Metal ROC XS, as the name suggests, is made of stainless steel. The proposed Indications for Use are identical to those of the existing 3.5mm ROC EZ. As with the existing devices, the proposed 3.5mm Metal ROC XS will be offered with #2 Suture, both with and without stainless steel needles. 290 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three human profiles incorporated into its design. ్లాల్లో వైద్యశాల గ్రామం నుండి 10 కి.మ Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 1 6 1998 Mr. Stephen M. Page Manager of Requlatory Affairs Innovasive Devices, Inc. 734 Forest Street 01752 Marlborough, Massachusetts Re: K981193 Innovasive 3.5 mm Metal ROC XS Suture Bone Fastener Regulatory Class: II Product Codes: MBI and HWC Dated: March 31, 1998 April 2, 1998 Received: #### Dear Mr. Page: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ದ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA)* will verify such assumptions. Failure to comply with the GMF regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ #### Page 2 - Mr. Stephen M. Page This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact Also, please note the Office of Compliance at (301) 594-4639. the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other gener Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, fu Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Office of Device Evaluation Center for Devices and Radiological Health Enclosure ার পর {4}------------------------------------------------ # INDICATIONS FOR USE ## SHOULDER Bankart repair Repair of rotator cuff tears Capsular instability Slap lesion repair Acromio-clavicular separation Capsule shift/capsulolabral reconstruction Biceps tenodesis Deltoid repair # KNEE ﺔ ﺍﻟﻤﺘﺤﺪﺓ Extra-capsular repairs Reattachment of medial collateral ligament Reattachment of lateral collateral ligament Reattachment of posterior oblique ligament Joint capsule closure Patellar ligament and tendon avulsion repairs Extra-capsular reconstruction ITB tenodesis. # ANKLE Lateral and medial instability Achilles tendon reconstruction and repair Mid-foot reconstructions ## FOOT Hallux valgus reconstruction #### ELBOW Tennis elbow repair Biceps tendon reattachment Medial and lateral repairs # BLADDER NECK SUSENSION - - - - - Bladder neck suspension for female urinary incontinence due to urethral hypermobility. Prescription Use (Per 21 CFR 801.109) 241 A (Division Sign-Off) Division of # Division of General Restorative Devices 510(k) Number _ K9811 # ੀਤੇ 1. 21
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...