# PROFLEX POWDER FREE NITRILE EXAMINATION GLOVES (K981159)

_Latexx Partners Bhd. · LZA · Jun 1, 1998 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981159

## Device Facts

- **Applicant:** Latexx Partners Bhd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jun 1, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Proflex Powderfree Nitrile Examination Glove is a single use device intended for medical purposes that is worn on the hand of health care and similar personnel to prevent contamination between the health care personnel and the patient.

## Device Story

Proflex Powder Free Nitrile Examination Gloves are single-use, blue or natural white nitrile rubber gloves. Manufactured by washing and chlorinating nitrile rubber to achieve a powder-free surface. Used by healthcare and similar personnel in clinical settings as a protective barrier to prevent cross-contamination between the wearer and the patient. The device is worn on the hands; it does not involve electronic inputs, algorithms, or software. Clinical benefit is derived from the physical barrier properties of the nitrile material, which has been tested for biocompatibility (primary skin irritation and delayed contact sensitization) and physical performance standards.

## Clinical Evidence

No clinical tests were conducted. Evidence is based on non-clinical laboratory and animal testing, including biocompatibility (primary skin irritation in rabbits, delayed contact sensitization in guinea pigs) and physical performance testing per ASTM D 3578-95 and draft ASTM standards. Final product confirmed negative for starch via USP iodine test.

## Technological Characteristics

Nitrile rubber compound; powder-free surface achieved via washing and chlorination. Dimensions and physical properties tested per ASTM D 3578-95 and draft ASTM Standard Specification for Nitrile Examination Gloves. Biocompatible; non-irritating and non-sensitizing. No electronic components, software, or energy sources.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Class 1 patient examination glove (Nitrile) 80LZA

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K981159

#### 1 1998 JUN

### 510(k) SUMMARY

# PROFLEX* POWDER FREE NITRILE EXAMINATION GLOVES

| Submitter's Name :                                                 | Latex Partners Bhd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address :                                              | PT 5054, Kamunting Industrial Estate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                    | 34600 Taiping Perak                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                    | Malaysia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Submitter's Phone Number                                           | 605 891 5555                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Submitter 's Fax Number :                                          | 605 891 2688                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Name of Contact Person :                                           | Goh, See Khek                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Date of Preparation :                                              | February 25, 1998                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Name of Device :                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Trade Name :                                                       | PROFLEX* POWDER FREE NITRILE<br>EXAMINATION GLOVES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Common Name :                                                      | Nitrile examination gloves                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Name :                                              | Patient Examination Gloves                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Legally Marketed Device to Which<br>Equivalency is Being Claimed : | PROFLEX* Powder Free Nitrile Examination<br>Gloves as described in the 510(k) notification are<br>substantially equivalent to the Class 1 patient<br>examination glove (Nitrile) 80LZA, that meets<br>the current draft ASTM Standard Specification<br>for Nitrile Examination Gloves for Medical<br>Application. It also meets all the current<br>specifications listed under the ASTM<br>Specification D 3578 - 95, Standard Specification<br>for Rubber Examination Gloves, EXCEPT for<br>tensile strength and ultimate elongation. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Attachment 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                    | Description of the Device :                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | PROFLEX* Powder Free Nitrile Examination<br>Gloves meet the current draft ASTM Standard<br>Specification for Nitrile Examination Gloves for<br>Medical Application. It also meets all the current<br>specifications listed under the ASTM<br>Specification D 3578 – 95, Standard Specification<br>for Rubber Examination Gloves, <b>EXCEPT</b> for<br>tensile strength and ultimate elongation<br>requirement. They are blue or natural white in<br>colour and are powder free.              |
|                                                                    | Intended Use of the Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | PROFLEX* Powder Free Nitrile Examination<br>Gloves are intended for single use for medical<br>purposes that is worn on the hand of health care<br>and similar personnel to prevent contamination<br>between the health care personnel and the<br>patients.                                                                                                                                                                                                                                   |
|                                                                    | Summary of Technological<br>Characteristics Compared to the Predicate<br>Device :                                                                                                                                                                                                                                                                                                                                                                                                                                                      | There is no different technological characteristics.<br>Gloves are made from Nitrile rubber compound<br>and the initial products are powdered Nitrile<br>rubber gloves. These gloves are then further<br>processed into powder free gloves using the<br>existing technology, i.e. washing and then<br>chlorinating the surfaces of the gloves.                                                                                                                                               |
|                                                                    | Brief Discussion of Nonclinical Tests :                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Testing performed per ASTM D 3578-95 (also<br>per current draft ASTM Standard Specification<br>for Nitrile Examination Gloves for Medical<br>Application) and 21 CFR 800.20. Gloves meet<br>the current draft ASTM Standard Specification<br>for Nitrile Examination Gloves for Medical<br>Application. Test results also show that the<br>product meets all the current specifications listed<br>under the ASTM Specification D 3578 – 95,<br>Standard Specification for Rubber Examination |
| K981159                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Attachment 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Gloves, EXCEPT for tensile strength and<br>ultimate elongation requirements.<br><br>Primary skin irritation testing in the rabbit and<br>delayed contact sensitization testing in the guinea<br>pig indicate no irritation of sensitization.<br><br>Final product is negative for the presence of<br>starch using the USP iodine test.                                                                                                                                                       |
|                                                                    | Brief Discussion of Clinical Tests :                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | No new clinical tests were conducted under this<br>510(k).                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                    | Conclusions Drawn for the Nonclinical<br>and Clinical Tests :                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Nonclinical laboratory and animal data indicate<br>that the powder free product meets all<br>performance and biocompatability requirements.                                                                                                                                                                                                                                                                                                                                                  |
| Other Information Deemed Necessary by<br>FDA :                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Not applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{1}------------------------------------------------

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

{2}------------------------------------------------

. . . . . .

... ... ... ...

:

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

1 1998 JUN

Mr. Goh See Knek · Latexx Partners Berhad PT 5054, Kamunting Industrial Estate 34600 Taiping Perak MALAYSIA

Re : K981159 Proflex* Powder Free Nitrile Examination Trade Name: Gloves Regulatory Class : I Product Code: LZA Dated: March 21, 1998 Received: March 31, 1998

Dear Mr. Goh See Knek:

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory ..... action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{4}------------------------------------------------

Page 2 - Mr. Goh See Khek

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

thy A. Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## INDICATIONS FOR USE

#### Applicant

LATEXX PARTNERS BHD. PT 5054, Kamunting Industrial Estate P.O. Box 9 34600 Taiping Perak MALAYSIA

K981159

510(k) Number (if known)

:

:

Device Name

### PROFLEX POWDERFREE NITRILE EXAMINATION GLOVE

ж

Indications For Use :

Proflex Powderfree Nitrile Examination Glove is a single use device intended for medical purposes that is worn on the hand of health care and similar personnel to prevent contamination between the health care personnel and the patient.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Olim S. Lim

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number

Prescription Use Per 21 CFR 801.109 OR Over-The-Counter

X

---

**Source:** [https://fda-staging.innolitics.com/device/K981159](https://fda-staging.innolitics.com/device/K981159)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).