The Mobetron is a mobile electron linear accelerator that produces beams of electrons used in the radiation therapy treatment of both malignant (cancer) and benign conditions. The Mobetron has four electron beam treatment energies and a set of electron applicators which provide a range of field sizes from 3 to 10 cm in diameter. For delivery of electron beam radiation during a surgical procedure, known as Intraoperative Radiation Therapy (IORT), the Mobetron system includes a sterile cap and sterile drapes for each IORT procedure, IORT applicators, a QA system, a surgical table extender with added support leg, and bolus material.
Device Story
Mobile electron linear accelerator (Linac) for intraoperative radiation therapy (IORT) in operating rooms. System comprises modulator, electron gun, RF system, accelerator guide, and motor-driven gantry on a self-contained transport stand. Produces electron beams at 4, 6, 9, and 12 MeV. Electrons scattered by metallic foils, pre-collimated by primary tungsten collimator, and shaped by patient-inserted applicators (3-10 cm diameter). Operator uses handheld control to position accelerator head and monitor dosimetry, machine interlocks, and periscopic video. Enables precise delivery of radiation directly to surgical site. Benefits include mobility between ORs, allowing IORT without moving patient to fixed radiation facility. Requires post-move output/energy verification via included QA system.
Clinical Evidence
Bench testing only. Performance evaluation confirmed device consistently performed within design parameters and equivalently to the predicate device.
Technological Characteristics
Mobile electron linear accelerator. Components: modulator, electron gun, RF system, accelerator guide, motor-driven gantry. Energy: 4, 6, 9, 12 MeV. Collimation: primary tungsten collimator and electron applicators. Dosimetry: dual transmission ion chambers. Mobility: self-contained transportation system. Control: handheld unit with dosimetry readout and video interface. QA: includes dedicated output/energy verification system.
Indications for Use
Indicated for radiation therapy treatment of malignant (cancer) and benign conditions in patients requiring intraoperative radiation therapy (IORT).
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
K981112
Accelerator, 510(k) Device
Number K852907
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS as required by section 807.92(c)
| 1. | Submitter's Information: | Dated March 24, 1998<br>Intraop Medical, Inc.<br>3170 De La Cruz Blvd., Suite 108<br>Santa Clara, CA 95054 |
|----|---------------------------------------|------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Donald A. Goer, Ph.D.<br>President & CEO<br>(408) 986-6020 (phone)<br>(408) 986-0222 (fax) |
| 2. | Common or Usual Name: | Electron Linear Accelerator (Linac) |
| | Proprietary and Trade Name of Device: | MOBETRON |
| | Device Classification: | System, Radiation Therapy, Charged-Particle, Medical |
| 3. | Predicate Device: | Siemens Mevatron ME Linear |
The MOBETRON is a mobile electron beam accelerator designed 4. Description of Device: specifically for use in the operating room (O.R.). Once in the OR, the accelerator head can be positioned in the horizontal plane, rotated about the coronal plane, and moved along the beam axis direction to facilitate docking with an electron applicator that is inserted into the patient. Together with the C-arm rotation, the accelerator head motion is designed to allow the accelerator to be easily and safely positioned for intraoperative treatments. All motorized motions are variable speed and are controlled by a lightweight hand-held control. The operator control system contains the dosimetry readout parameters, accelerator controls, machine interlock status, and a video output of the periscopic viewing system.
When the electrons emerge from the accelerator, they are scattered by a set of thin metallic foils and are pre-collimated to a circular field size of 10 cm. (at the standard treatment distance of 50 cm. SSD) by a primary tungsten collimator. The scattered electron beam passes through two independent transmission ion chambers and is further collimated to the desired field size by electron applicators which are inserted into the patient. The scattering foil and collimation technique and applicator systems used in the Mobetron is similar to techniques used for IORT in other electron accelerator systems. The Mobetron provides four (4) electron beams of energy 4, 6, 9 and 12 MeV, and a set of electron applicators of various diameters from 3 cm to enable the delivery of Intraoperative Radiotherapy (IORT) treatment.
Statement of Intended Use: The Mobetron is a mobile electron linear accelerator that న. produces beams of electrons used in the radiation therapy treatment of both malignant (cancer) and benign conditions. The Mobetron has four electron beam treatment energies and a set of electron applicators which provide a range of field sizes from 3 to 10 cm in diameter.
For delivery of electron beam radiation during a surgical procedure, known as Intraoperative Radiation Therapy (IORT), the Mobetron system includes a sterile cap and sterile drapes for each IORT procedure, IORT applicators, a QA system, a surgical table extender with added support leg, and bolus material.
The intended use is the same as the predicate device.
{1}------------------------------------------------
i ર. Statement of Technological Characteristics: The basic operation of the Mobetron is similar to that of any electron linear accelerator system, in that it employs a modulator, electron gun, rf-system and accelerator guide system. In the Mobetron, a lightweight electron beam linear accelerator is mounted on a motor driven gantry. The gantry is attached to a stand which is supported on a selfcontained transportation system. A modulator and operator control console complete the system. This system produces electron beams of energies, field size, and dose rate equivalent to other electron accelerators used for IORT.
The intended use and performance characteristics are equivalent to the predicate device and therefore we believe it is substantially equivalent to it.
Differences: The differences between the systems are minor and do not effect the safety and 7. efficacy of the treatment. The major difference between the Mobetron and the predicate device is that the predicate device is fixed in location and limited to a single O.R. The Mobetron is mobile and can be used in more than one O.R. Once it is in a specific O.R., its use is essentially identical to the predicate device. To assure that the Mobetron is performing correctly after it is moved to a new location, output and energy checks are made using a quality assurance device provided with the unit. The predicate device requires similar testing of output and energy prior to use even though the device is not moved. The Mobetron is thus substantially equivalent to the predicate device.
Performance Evaluation: Performance tests were conducted and the results indicated that 8. the Mobetron consistently performed within the design parameters and equivalently to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized image. The image is of a caduceus, a symbol of medicine, with a single snake winding around a staff. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 4 1998
Donald A. Goer, Ph.D. President Intraop Medical 3170 De La Cruz Blvd., Suite 108 Santa Clara, CA 95054
Re:
K981112 Mobetron (Electron Linear Accelerators) Dated: March 24, 1998 Received: March 26, 1998 Regulatory class: II 21 CFR 892.5050/Procode: 90 LHN
Dear Dr. Goer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal. Food, Drug, and Cosmenc Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Kilian Yin
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## STATEMENT OF INDICATIONS FOR USE
The Mobetron is a mobile electron linear accelerator that produces beams of electrons used in the radiation therapy treatment of both malignant (cancer) and benign conditions. The Mobetron has four electron beam treatment energies and a set of electron applicators which provide a range of field sizes from 3 to 10 cm in diameter.
For delivery of electron beam radiation during a surgical procedure, known as Intraoperative Radiation Therapy (IORT), the Mobetron system includes a sterile cap and sterile drapes for each IORT procedure, IORT applicators, a QA system, a surgical table extender with added support leg, and bolus material.
Daniel G. Symm
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devi 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.