← Product Code [KGO](/productcode/KGO) · K981096

# SEMPERMED SURGEONS' GLOVES - SUPREME (K981096)

_Semperit Technische Produkte Gesellshaft · KGO · Oct 29, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K981096

## Device Facts

- **Applicant:** Semperit Technische Produkte Gesellshaft
- **Product Code:** [KGO](/productcode/KGO.md)
- **Decision Date:** Oct 29, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

This medical glove is worn on the hand of surgeons, nurses and similar personnel to prevent contamination between health care personnel and the patient.

## Device Story

Latex surgeons' gloves; powder-free; designed for use by surgeons, nurses, and healthcare personnel. Features single-side polymer coating, halogenation, and siliconization to facilitate donning. Provides barrier protection to prevent contamination between healthcare personnel and patients. Manufactured via multi-step washing process to ensure powder-free status. Used in clinical/surgical settings.

## Clinical Evidence

Bench testing only. Device meets ASTM D 3577-91 and EN 455 standards. Testing includes physical dimension tests, tensile strength, ultimate elongation, and stress at 500% elongation (before and after aging). Freedom from holes verified via sampling per ISO 2859 (AQL 1.0). Immunological RAST-inhibition, skin irritation (rabbits), and sensitization (guinea pig) tests performed.

## Technological Characteristics

Latex material; powder-free; single-side polymer coated; halogenated; siliconized. Dimensions: 270-280 mm length; 0.17-0.26 mm thickness. Meets ASTM D 3577-91 and EN 455/1-3 standards. Sterilization process validated periodically.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K981096

SEMPERIT TECHNISCHE PRODUKTE GES.M.B.H. Bundesstraße 26, A-2632 Wimpassing Tel.: ++43 2630 310 308, FAX: ++43 2630 310 479

## 510 (k) Summary

14.03.1998

1.0 APPLICANT Semperit Technische Produkte Ges.m.b.H. Modecenterstraße 22 A-1031 Vienna Phone: ++43 1 79777 490 ++43 1 79777 630 Fax:

- 2.0 CONTACT PERSON
Dipl.-Ing. R. Ehrenfeldner Semperit Technische Produkte Ges.m.b.H. Bundesstraße 26 A-2632 Wimpassing Phone: ++43 2630 310 216 ++43 2630 310 479 Fax:

Semperit Industrial Prod. Inc. Mr. Manfred Reichel 220 Kinderkamack Road Suite B. Westwood NJ Phone: (201) 664 5222 (800) 631 2566 (201) 664 6688 Fax:

- 3.0 Device Class: I Type 1 Product Code: 80 FPI/878.4460 (79KGO)
- Latex patient surgeons gloves 4.0 Specification: powederfree - Class I 80 FPI Type 1 meets all of the requirements of ASTM Standard D 3577-91 and EN 455/1-3
- 5.0 Device Description: Latex patient surgeons 'gloves powderfree
- 6.0 Intended use: This medical glove is worn on the hand of surgeons, nurses and similar personnel to prevent contamination between health care personnel and the patient. Supreme is a powderfree glove with an additional inner layer to make donning easily possible (MC = multilayer coating) and TUV-Productservice claim. (see attachement 1 and 2)
- 7.1 Surface treatment: Single side polymer coated, halogenation, siliconization and extensive washing with water in 8 steps. Inside and outside free from glove powder and free of talc. (see att. 9 - 12/6.1 and 6.2)

300/HAQ

{1}------------------------------------------------

# 7.2 Quality characteristics acc. ASTM D3577-91

| Description acc. ASTM items         |                                                                                                                        |       |                          |                                  |                            |                          |        |        |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------|--------------------------|----------------------------------|----------------------------|--------------------------|--------|--------|
| 8.2 Sterility Test                  | Periodical Validation of the process with<br>Bioservice scientific Laboratories - Germany<br>(attachement 8.2.1/8.2.2) |       |                          |                                  |                            |                          |        |        |
| 8.3 Freedom from holes              | Sampling acc. ISO 2859 (att. 9 - 14/13)<br>AQL 1,0                                                                     |       |                          |                                  |                            |                          |        |        |
| 8.4 Physical<br>Dimension<br>Test   | 2. Length 270 - 280 mm related to glove size                                                                           |       |                          |                                  |                            |                          |        |        |
|                                     | 3. Width                                                                                                               | 6     | 6,5                      | 7                                | 7,5                        | 8                        | 8,5    | 9      |
|                                     |                                                                                                                        | 79 ±2 | 84 ±3                    | 89 ±3                            | 97 ±3                      | 104 ±3                   | 110 ±3 | 114 ±3 |
|                                     | 4. Thickness: 017 - 0,26 mm                                                                                            |       |                          |                                  |                            |                          |        |        |
| 8.5 Physical<br>Requirement<br>Test | Before Aging                                                                                                           |       |                          | After Aging                      |                            |                          |        |        |
|                                     | Tensile<br>Strength<br>MPa, min                                                                                        |       | Ultimate<br>Elongation % | Stress at<br>500 %<br>Elongation | Tensile<br>Strength<br>MPa | Ultimate<br>Elongation % |        |        |
| Spec                                | 24                                                                                                                     |       | 750                      | 5.5                              | 18                         | 560                      |        |        |

Internal microscopic method 10 pow 

# 8 + 9.1 Immunological RAST-inhibition test

- 8 + 9.2 Skin irritation test on rabbits
- 8 + 9.3 Maximization test/Sensibilisation of skin on genuine-pig See attached test resultas (attachement 9 - 14/7, 9 - 14/1 - 6)
- 10. Conclusion

Sempermed powderfree surgeons gloves Supreme meets ASTM D 3577-91 and EN 455 standard but exceed FDA pinhole requirements (AQL 1,0). It meets the labelling claims as shown by attachements 1(2), 9 - 12/6.1 und 6.2, 9 - 14/13, 9 - 12/6.1/2

} 9-14 chapt.
1-3 p pages

{2}------------------------------------------------

Image /page/2/Picture/10 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which consists of three stylized human profiles facing to the right, stacked on top of each other. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 9 1998

Mr. Richard Ehrenfeldner Semperit Technische Produkte Gesellschaft m.b.H. Triester Bundesstrabe 26 A-2632 Wimpassing

Re : K981096 Sempermed Latex Surgeons' Gloves, Powder-Trade Name: Free Regulatory Class: I Product Code: KGO Dated: August 18, 1998 Received: August 21, 1998

Dear Mr. Ehrenfeldner:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

### Page 2 - Mr. Ehrenfeldner

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda/gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Semperit Technische Produkte Gesellschaft m.b.H.

A-2632 Wimpassing, Triester Bundesstraße 26 Telefon (02630) 310/216 Telex 16 óól semp a Telekopierer 02630/310/479

Image /page/4/Picture/2 description: The image shows the word "SEMPERIT" in large, bold, black letters on the top line. Below that, in a smaller font size, is the phrase "Technische Produkte" also in black letters. The text appears to be a logo or heading, possibly for a company or product line.

# INDICATIONS FOR USE

Image /page/4/Picture/4 description: The image shows a document with some text and a stamp. The text on the document reads "CRH/OCE/CHC" and "21 AUG 83 206 13". There is also a stamp on the document that says "RELEASED".

| Applicant:                | Semperit Technische Produkte Ges.m.b.H.                                                                                                                        |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known): | K 98 10 96                                                                                                                                                     |
| Device Name:              | Latex<br>Sempermed Surgeons' Gloves - Supreme, Powder Free                                                                                                     |
| Indications For Use:      | This medical glove is worn on the hand of surgeons, nurses and<br>similar personnel to prevent contamination between health care<br>personnel and the patient. |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrenze of CDRH Office of Device Evalution (ODE)

Olin S. L

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number

Prescription Use Per 21 CFR 081.109

OR

Over-The-Counter

* For a new submission, do NOT fill in the 510(k) number blank.

(Optional Format 1-2-96)

مزمد سنده

---

**Source:** [https://fda-staging.innolitics.com/device/K981096](https://fda-staging.innolitics.com/device/K981096)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
