K980979 · Fukuda Denshi USA, Inc. · CCK · Jun 12, 1998 · Anesthesiology
Device Facts
Record ID
K980979
Device Name
FUKUDA DENSHI MODEL HC-500 END TIDAL CO2 MODULE
Applicant
Fukuda Denshi USA, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Jun 12, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Fukuda Denshi model HC-500 End Tidal CO2 Module is intended to be used to noninvasively monitor inspired and end tidal CO2 and respiration rate by trained medical professionals by or on the order of a physician. The Fukuda Denshi model HC-500 End Tidal CO2 Module will function with all Novametrix reusable and disposable main stream sensors. The Fukuda Denshi model HC-500 End Tidal CO2 Module is intended to be used in an adult, infant and/or neonate patient population who are under the care of a physician, within the confines of a health care facility. The intended environment is critical care monitoring situations, including ventilator support and anesthesia where the Fukuda Denshi model DS-5300 Patient Monitor is being utilized. The Fukuda Denshi model HC-500 End Tidal CO2 Module is not intended for home use. The Fukuda Denshi model HC-500 End Tidal CO2 Module is intended to be used as part of the Fukuda Denshi DynaScope™ DS-5000 series of bedside patient monitoring systems. The HC-500 will not function as a stand alone device.
Device Story
Module for Fukuda Denshi DS-5000 series bedside patient monitors; non-invasively measures inspired/expired CO2 and respiration rate. Uses Novametrix main stream sensors (Capnostat II/III) employing solid-state infrared technology. Operates only when installed in DS-5300 monitor; not a standalone device. Used by trained medical professionals in critical care, ventilator support, and anesthesia settings. Provides real-time data and alarms to clinicians for patient management. Benefits include continuous respiratory monitoring to assist in clinical decision-making and patient safety.
Clinical Evidence
Bench testing only. Validation and verification testing conducted per Novametrix Medical Systems Inc. protocol. Testing included environmental testing per FDA 1993 guidance and electrical safety compliance (ANSI/AAMI ES1-1993). No clinical data presented.
Technological Characteristics
Main stream infrared CO2 sensing; solid-state technology. Compatible with Novametrix Capnostat II/III sensors. Integrated into Fukuda Denshi CPU. Electrical safety compliant with UL 2601, CSA 22.2, and AAMI standards. Module form factor designed for DS-5000 series bedside monitors.
Indications for Use
Indicated for real-time end tidal or inspired carbon dioxide concentration and/or respiration rate monitoring of intubated adult, infant, and neonate patients under physician care in healthcare facilities.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Predicate Devices
Fukuda Denshi model HC-300 End Tidal CO2 Module (K950044)
Reference Devices
Fukuda Denshi DS-5300 Patient Monitoring System (K964187)
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 12 1998
K980979
# 510(k) Summary Fukuda Denshi model HC-500 End Tidal CO2 Module
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR part 807.92.
The assigned 510(k) number is :
#### Submitter:
FUKUDA DENSHI AMERICA CORP. 17725 NE 65th St., Bldg C Redmond WA 98052 425/881-7737 Tel: 425/869-2018 Fax:
Contact Person:
David J. Geraghty Regulatory Affairs Manager FUKUDA DENSHI AMERICA CORP. 17725 NE 65th St., Bldg C Redmond WA 98052 425/881-7737 Tel: 425/869-2018 Fax:
Date Prepared:
March 9, 1998
Device Name:
Proprietary Name:
model HC-500 End Tidal CO2 Module
Common Name:
ETCO, Gas Module
Classification Name:
Carbon Dioxide Gas Analyzer (§869.1400)
#### Legally Marketed Device:
FUKUDA DENSHI model HC-300 End Tidal CO2 Module 510(k) number K950044. (The HC-500 is an addition to the to the DS-5300, 510(k), K964187.)
{1}------------------------------------------------
#### Description:
The Fukuda Denshi model HC-500 End Tidal CO2 Module is a computer controlled medical device. It is designed to non-invasively provide respiration, inspired and expired CO2 values when used as part of the DS-5300 Patient Monitoring System (K964187). The HC-500 is a module, not a stand alone device. As a module, it will only operate when installed into a DS-5300 bedside patient monitor. The Fukuda Denshi model HC-500 ETCO2 module will function with all NOVAMETRIX main stream sensors.
The safety and efficacy of the HC-500 module has been established through various techniques. Review of the Fukuda Denshi model HC-300 predecessor product's history does not reveal any complaints related to safety and/or effectiveness. There are no reports of adverse effects or reportable incidents for the currently marketed HC-300 ETCO, product.
#### Statement of Intended Use:
The Fukuda Denshi model HC-500 End Tidal CO2 Module is intended to be used to non-invasively monitor inspired and expired CO, and respiration rate by trained medical professionals by or on the order of a physician. The Fukuda Denshi model HC-500 End Tidal CO2 Module will function with all NOVAMETRIX reusable and disposable main stream sensors.
The Fukuda Denshi model HC-500 End Tidal CO2 Module is intended to be used in all patient populations who are under the care of a physician, within the confines of a health care facility. The intended environment is critical monitoring situations, including ventilator support and anesthesia, where the Fukuda Denshi DS-5300 bedside patient monitor is being utilized with an IB-5006 Input Box.
The Fukuda Denshi model HC-500 End Tidal CO2 Module is NOT INTENDED FOR HOME USE.
The Fukuda Denshi model HC-500 End Tidal CO, Module is intended to be used as part of the Fukuda Denshi DS-5000 series of bedside patient monitoring systems. The HC-500 will not function as a stand alone device.
{2}------------------------------------------------
## Comparison to Predicate Device
In summary, the HC-500 module is an updated version of the HC-300 ETCO2 module designed for use in the DS-5000 series of patient monitors. Changes to the system include a reduction in size and weight to meet the requirements of the DS-5000 series modules. Control circuits have been incorporated into a Fukuda Denshi designed CPU. The front end design, as specified by Novametrix Medical Systems, Inc. has been relayed out to DS-5000 series module size specifications. The majority of the changes can be summarized as re-layout of circuits and a reduction in size.
## Technological Characteristics
The Fukuda Denshi model HC-500 End Tidal CO2 Module uses small, light weight main stream sensors, Capnostat IITM or Capnostat III™, manufactured by Novametrix Medical Systems Inc. The sensors employ Novametrix solid state infrared technology. All ETCO, sensing and calculation algorithms are similar to those used in the HC-300 (K950044) and are the same as those employed by Novametrix Medical Systems Inc.
Control circuits have been incorporated into a Fukuda Denshi designed CPU. The front end design, as specified by Novametrix Medical Systems, Inc. has been relayed out to DS-5000 series module size specifications. The majority of the changes can be summarized as re-layout of circuits and a reduction in size.
No changes to the DS-5300 Patient Monitoring System (K964187) hardware or software were required to accommodate the HC-500.
## Testing:
Laboratory testing was conducted to validate and verify that the Fukuda Denshi model HC-500 End Tidal CO2 Module met all design specifications and was substantially equivalent to the Fukuda Denshi model HC-300 (K950044). End Tidal CO2 validation and verification testing conducted according to Novametrix Medical Systems Inc. testing protocol
Product testing included of all environmental testing identified in the FDA's DCRND November 1993 "Reviewer Guidance Document for Premarket notification Submissions" Draft Guidance Document. Additional testing was performed to demonstrate compliance with the ANSI/AAMI standards ES1-1993, "Safe current limits for electromedical apparatus." Finally, testing was
HC-500 510(k) Submission Section 2.doc
{3}------------------------------------------------
performed to verify that the design addressed all hazards and to validate the systems overall operation.
Although the device is neither life supporting nor life sustaining, diagnostic information derived from the use of the device and alarms generated by the device may be critical to the proper management of the patient.
The primary areas of risk for this device are the same as the predicate device and other devices in this class, and are the following:
- Electrical shock .
Excessive electrical chassis leakage current can disturb the normal electrophysiology of the heart, and possibly leading to the onset of cardiac arrhythmias.
- Misdiagnosis
- Inadequate design of the signal processing and measurement circuitry or program can lead to generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
- Inadequate design of the device's software, used to make various measurements, can lead to generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
- Inadequate design of the systems ability to alert the users through audible and visual indicators, can lead to user mistrust and/or inadequate response to the patient's condition. If an inadequate response to the patient's condition should occur the patient may unnecessarily be placed at risk.
The design of the HC-500 has taken into account all the above. The device is designed to meet UL 2601, CSA 22.2 and AAMI standards for electrical safety for medical equipment to prevent the possibility of excessive electrical leakage current to the patient.
The conclusions drawn from clinical and laboratory testing of the Fukuda Denshi model HC-500 End Tidal CO, Module demonstrates that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the Fukuda Denshi model HC-300 End Tidal CO2 Module, K950044
## Conclusion:
{4}------------------------------------------------
## COMPLIANCE CERTIFICATE
I certify that in my capacity as Regulatory Affairs Manager and official FDA correspondent of FUKUDA DENSHI AMERICA CORP. that I have conducted a reasonable search of all information known or otherwise available about the types and causes of safety or effectiveness problems that have been reported for cardiac monitors and arrhythmia detectors and alarms. I further certify that I am aware of the types of problems to which cardiac monitors and arrhythmia detectors and alarms are susceptible to and that to the best of my knowledge, the attached summary of the types of causes of safety or effectiveness problems about these devices is complete and accurate.
I hereby certify that the attached software development process and quality assurance procedures were adhered to and that all testing performed demonstrates that the functional requirements were met and that the system specifications were fulfilled.
Dara J. Durag, Rq
Signature
David J. Geraghty
Name
Regulatory Affairs Manager Fukuda Denshi America Corp.
Title
HC-500 510(k) Submission Section 2.doc
\$\infty\$
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 2 1998
Mr. David J. Geraghty Fukuda Denski USA, Inc. 17725 N.E. 65th Street, Building C Redmond, WA 98052
Re: K980979 Fukuda Denshi Model HC-500 End Tidal CO2 Module Requlatory Class: II (two) Product Code: 73 CCK Dated: March 13, 1998 Received: March 17, 1998
Dear Mr. Geraghty:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the You provisions of the Federal Food, Druq, and Cosmetic Act (Act). may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{6}------------------------------------------------
#### Page 2 - Mr. David J. Geraghty
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Collebon
Thomas J. Collebon, Ph.D.
Thomas J. Callahan, Ph. D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Statement of Intended Use:
The Fukuda Denshi model HC-500 End Tidal CO2 Module is intended to be used to noninvasively monitor inspired and end tidal CO2 and respiration rate by trained medical professionals by or on the order of a physician. The Fukuda Denshi model HC-500 End Tidal CO2 Module will function with all Novametrix reusable and disposable main stream sensors.
The Fukuda Denshi model HC-500 End Tidal CO2 Module is intended to be used in an adult, infant and/or neonate patient population who are under the care of a physician, within the confines of a health care facility. The intended environment is critical care monitoring situations, including ventilator support and anesthesia where the Fukuda Denshi model DS-5300 Patient Monitor is being utilized.
The Fukuda Denshi model HC-500 End Tidal CO2 Module is not intended for home use.
The Fukuda Denshi model HC-500 End Tidal CO2 Module is intended to be used as part of the Fukuda Denshi DynaScope™ DS-5000 series of bedside patient monitoring systems. The HC-500 will not function as a stand alone device.
#### Statement of Indications for Use
The Fukuda Denshi model HC-500 End Tidal CO2 Module is indicated in those situation where it is desirable to perform real-time end tidal or inspired carbon dioxide concentration and/or respiration rate monitoring of intubated adult, infant and/or neonate patients who are under the care of a physician, within the confines of a health care facility.
M. Pugh
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 1880979 510(k) Number .
prescriptions use
✓
tc
HC-500 510(k) Submission Section 3.doc
March 16, 1998 Page 2 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.