AMEDA DISPOSABLE VACUUM ASSIST CUP WITH FLUID RETENTION TRAP

K980976 · Hollister, Inc. · HDB · Aug 21, 1998 · Obstetrics/Gynecology

Device Facts

Record IDK980976
Device NameAMEDA DISPOSABLE VACUUM ASSIST CUP WITH FLUID RETENTION TRAP
ApplicantHollister, Inc.
Product CodeHDB · Obstetrics/Gynecology
Decision DateAug 21, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.4340
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap is a sterile, single use, device intended to be connected to an externally powered fetal vacuum extractor to aid in vaginal or Cesarean delivery of term or near term infants. The Fluid Retention Trap is intended to collect fluids aspirated during the delivery procedure.

Device Story

Device is a sterile, single-use vacuum assist cup for fetal extraction; connects to externally powered vacuum pump. New feature is a fluid retention trap consisting of high-impact polystyrene housing and polyurethane foam filter; trap connects to tubing between cup and pump to collect aspirated fluids during delivery. Used in clinical settings (labor/delivery) by obstetricians or qualified clinicians. Output is physical suction applied to fetal head to assist delivery; fluid trap prevents aspiration of fluids into pump. Benefits include safe fluid management during vacuum-assisted delivery.

Clinical Evidence

Bench testing only. Evaluated fluid handling capacity and resistance to air flow. Biocompatibility assessed per ISO 10993 and USP standards.

Technological Characteristics

Materials: High impact polystyrene housing, polyurethane foam filter. Connectivity: Connects to external vacuum pump via tubing. Form factor: Disposable vacuum cup with integrated fluid trap. Sterilization: Sterile, single-use. No software or electronic components.

Indications for Use

Indicated for term or near-term infants during vaginal or Cesarean delivery in cases of: delay in second stage labor with fetal head at outlet, pelvic floor, or mid-pelvis; delay in second stage associated with borderline cephalopelvic disproportion; fetal distress; shortening of second stage labor for fetal or maternal benefit; or umbilical cord prolapse with fully dilated cervix.

Regulatory Classification

Identification

A fetal vacuum extractor is a device used to facilitate delivery. The device enables traction to be applied to the fetal head (in the birth canal) by means of a suction cup attached to the scalp and is powered by an external vacuum source. This generic type of device may include the cup, hosing, vacuum source, and vacuum control.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a date, AUG 21 1998. The month is August, the day is the 21st, and the year is 1998. The text is in a simple, sans-serif font and is left-aligned. The image is a close-up of the date. llister K980976 Hollister Incorporated 2000 Hollister Drive Libertyville, Illinois 60048-3781 # Hollister Incorporated Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap # Safety and Effectiveness Summary ## 1. Submitter's name, Address and Contact Person Submitter Hollister Incorporated 2000 Hollister Drive Libertyville, IL 60048 Contact Person Joseph S. Tokarz Manager, Regulatory Affairs Phone (847)680-2849 (847)918-3860 Fax Date Summary Prepared - March 11, 1998 #### 2. Name of Device: Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap ## 3. Name of Predicate Device(s) Ameda Disposable Vacuum Assist Cup, K895700 Mitvyac Obstetrical Vacuum Extraction System by Neward Enterprises Tender Touch Obstetrical Vacuum Delivery Kit by Columbia Medical Inc. ## 4. Description of Device The purpose of this premarket notification is to add a fluid retention trap option to the current Ameda Disposable Vacuum Assist Cup in order to collect fluids that may be aspirated during the delivery procedure. The fluid retention trap consists of a housing segment that is constructed of high impact polystyrene and a polyurethane foam filter. The fluid retention trap, when used, is connected to the tubing of the device between the vacuum cup and the vacuum pump. ## 5. Statement of Intended Use The Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap is a sterile, single use, device intended to be connected to an externally powered fetal vacuum extractor to aid in vaginal or Cesarean delivery of term or near term infants. The Fluid Retention Trap is intended to collect fluids aspirated during the delivery procedure. #### 6. Statement of Technological Characteristics of the Device a. Laboratory Testing: The objective of the laboratory testing was to determine substantial equivalence of the Hollister Disposable Vacuum Assist Cup's Fluid Retention Trap to the predicate devices. Laboratory testing was performed to evaluate fluid handling capacity and resistance to air flow. The results of the laboratory testing support Hollister's claim of substantial equivalence to the predicate devices. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "Hollister" in a serif font, with a symbol to the left of the word. The symbol is a geometric design that looks like a stylized flower or star. The text and symbol are in black and the background is white. K98076 82092 # Hollister Incorporated Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap b. Biocompatibility: The biocompatibility for the Ameda Disposable Vacuum Assist Cup has been documented and cleared in previously submitted 510(k) K895700 "Ameda /Egnell Dolphin Dispo-Soft Vacuum Extractor." The biocompatibility of the Fluid Retention Trap was assessed based on principles and guidelines established by various governmental and standard setting organizations, such as: ISO 10993, International Standard Organization (ISO) Standard; General Program Memorandum #G95-1, U.S. FDA Office of Device Evaluation; United States Pharmacopeia (USP). Material biocompatibility issues have been addressed based upon biomaterial history or in separate in vitro or in vivo laboratory evaluations using licensed commercial reference laboratories. These evaluations have been contracted either by Hollister or the suppliers of the materials. Based upon the results of this assessment, the materials used to fabricate the Fluid Retention Trap are considered biocompatible and appropriate for their intended use. c. Comparison to Predicate Devices: The Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap is substantially equivalent to other devices that are in commercial distribution. This is based on the following: · The Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap is substantially equivalent in its intended use as a fetal vacuum extractor as described in 21 CFR 884.4340. · The predicate device, the Ameda Dolphin Dispo-Soft Vacuum Extractor, K895700 and the proposed device are identical in intended use, materials, and manufacturing processes with the exception of the addition of the fluid retention trap option on the proposed device. · The predicate devices, the Mityvac Extractor Cup System by Neward Enterprises, Inc., and the Tender Touch by Columbia Medical Inc.; and the proposed device are identical in intended use. The predicate and proposed devices include fluid collection capabilities. ### 7. Conclusion Based upon the information presented above it is concluded that the proposed Ameda Vacuum Assist Cup with Fluid Retention Trap is safe and effective for its intended use and is substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo for the Department of Health & Human Services. The logo consists of a stylized representation of three human figures, with their heads connected and bodies flowing into each other. The figures are depicted in black against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular pattern around the logo. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Re: K980976 Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap Dated: July 29, 1998 Received: July 30, 1998 Regulatory Class: II 21 CFR 884.4340/Procode: 85 HDB Dear Mr. Tokarz: AUG 21 1998 Mr. Joseph S. Takarz Hollister Incorporated Libertyville, IL 60048 2000 Hollister Drive Manager, Regulatory Affairs We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmadsmamain.html". Sincerely yours. Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Hollister Incorporated Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap #### Statement of Indications for Use b. 510(k) Number (if Known): _ K980976 Ameda Disposable Vacuum Assist Cup w/ Fluid Retention Trap Device Name: The Ameda Disposable Vacuum Assist Cup with Fluid Retention Trap is a sterile, single use, device intended to be connected to an externally powered fetal vacuum extractor to aid in vaginal or Cesarean delivery of term or near term infants. The Fluid Retention Trap is intended to collect fluids aspirated during the delivery procedure. ### Indications For Use: - Delay in second stage with the fetal head stationed at the outlet; on the pelvic floor; or in the . mid-pelvis. - Delay in the second stage associated with borderline cephalopelvic disproportion . - Fetal distress . - Shortening of second stage labor for fetal or maternal benefit . - Prolapse of the umbilical cord when the cervix is fully dilated ● (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-the-Counter-Use (Optional Format 1-2-96) Robert D. Salling/ (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K980976
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...