K980970 · Sterling Diagnostic Imaging, Inc. · LLZ · Jun 8, 1998 · Radiology
Device Facts
Record ID
K980970
Device Name
IIRAD DIRECTRAY OPERATOR CONSOLE
Applicant
Sterling Diagnostic Imaging, Inc.
Product Code
LLZ · Radiology
Decision Date
Jun 8, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
To provide a network connection to various output (e.g. hardcopy, softcopy and archive) devices from any radiographic device which uses the DirectRay™ Image capture system. To synchronize the ready states of the DirectRay™ device and the radiographic equipment. When a single X-ray console is desired, to provide the X-ray console features (e.g. technique selection) of the radiographic equipment.
Device Story
DirectRay™ Operator Console (DROC) acts as an interface between X-ray generating equipment and the DirectRay™ digital image capture device (K973206). DROC synchronizes ready states between the capture device and X-ray generator; enables acquisition, preview, and transmission of images and text data; and provides exam administration reporting. In the 'Integrated' version, DROC replaces the traditional X-ray console by incorporating technique selection functions. Located in the control booth, usually within a cabinet, the device is operated by radiology staff. It optionally links digital images to hospital/radiology information systems (HIS/RIS). By centralizing networking and console controls, it streamlines workflow and image management for radiographic procedures.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
The device is a console system housed in a cabinet for control booth installation. It functions as a network interface and controller for X-ray equipment and digital image capture systems. It supports connectivity to HIS/RIS and various output devices (hardcopy/softcopy/archive).
Indications for Use
Indicated for use in clinical settings where transmission of radiographic images and associated patient text data is required from radiographic equipment utilizing the DirectRay™ image capture system to output devices including hardcopy, softcopy, or archive systems.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
JUN - 8 1998
Image /page/0/Picture/1 description: The image shows the text "K986970" at the top. Below that is the word "STERLING" in large, bold letters with a line underneath. Underneath that is the text "Diagnostic Imaging" in a smaller font.
Image /page/0/Picture/2 description: The image shows a vertical, rectangular shape with a white, curved line running through the center. The background of the rectangle is black. The white line starts at the bottom left corner, curves towards the right, and then curves back to the left as it reaches the top right corner.
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| SUBMITTED BY: | Jean E. Bartlett<br>Regulatory Affairs<br>STERLING DIAGNOSTIC IMAGING, INC.<br>PO Box 19048, Mail Drop 102<br>Greenville, SC 29602-9048 | |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--|
| PREPARATION DATE: | March 13, 1998 | |
| DEVICE NAME: | DirectRay™ Operator Console<br>Two versions: (a) Parallel | |
There are two versions of the device covered by this premarket notification, the Parallel version and the Integrated version, and the classification information is slightly different for the two versions:
(b) Integrated
## CLASSIFICATION and PROCODE:
| Parallel: | 21CFR 892.2050 (Proposed)- PACS Device (90 LMD)<br>and accessory to:<br>Unclassified - Solid-State Digital X-Ray Device(90MQB) |
|-------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Integrated: | 21CFR 892.2050 (Proposed) - PACS Device (90 LMD)<br>21CFR 892.1680 -Stationary X-Ray System (90KPR)<br>and accessory to:<br>Unclassified - Solid-State Digital X-Ray Device(90MQB) |
PREDICATE DEVICES: In both versions, the DROC is an accessory to the DirectRay™ digital image capture device (K#973206).
LINX™ Networking System - K964250 Parallel:
MicroX-80 HF - K944149 Integrated: (X-ray console functions only)
> Sterling Diagnostic Imaging, Inc. 10 South Academy Street P.O. Box 19545 Greenville, SC 29602-0248
36
{1}------------------------------------------------
DEVICE DESCRIPTION: The DirectRay™ Operator Console is that device, which when used with any X-ray generating equipment and the DirectRay™ device (Sterling Diagnostic Imaging Direct Radiography™ ' K#973206) allows the synchronization of the ready states of the DirectRay™ device and the X-ray generator, and allows the operator to acquire, preview and transmit image and associated text data along with exam administration report capabilities. In one version, this device will also integrate these functions with the typical functions of an X-ray console eliminating the need for a separate X-ray console. This device is located in the control booth and will usually be housed in a specially designed cabinet. An optional feature of the DROC is the ability to link the digital image with the hospital information/radiology information systems (HIS/RIS).
INTENDED USE: To provide a network connection to various output (e.g. hardcopy, softcopy and archive) devices from any radiographic device which uses the DirectRay™ Image capture system.
To synchronize the ready states of the DirectRay™ device and the radiographic equipment. When a single X-ray console is desired, to provide the X-ray console features (e.g. technique selection) of the radiographic equipment.
COMPARISON TO PREDICATE DEVICE(s): The DROC brings together the features of the LINX™ Networking System and the X-ray console functions of the Micro X-80 into a single device. With respect to the synchronization of the ready states of the DirectRay™ device and the radiographic equipment, the DROC is considered an accessory to the DirectRay™.
Signature JardBartlett
Date March 13, 1998
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized abstract design resembling an eagle or bird in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the central design. The text is in a simple, sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 8 1998
Jean E. Bartlett Regulatory Affairs Manager Sterling Diagnostic Imaging 10 South Academy Street Mailbox 120 P.O. Box 19048 Greenville, SC 29602-9048
Re:
K980970 iiRAD™ Operator Console Dated: March 13, 1998 Received: March 16, 1998 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Ms. Bartlett:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. -The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdr/dsmaldsmam.html".
Sincerely yours,
Zillian Yi
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS STATEMENT
510(k) Number (if known): __ K 9.30970
Device Name: DirectRay™ Operator Console
Indications for Use:
The DirectRay™ Operator Console has application whenever the transmission of radiographic images and associated patient text data is desired to take place from an input device, such as any radiographic equipment which uses the DirectRay™ device, to any output device, such as hardcopy, softcopy, or archive devices.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | ✓ |
|---------------------------------------|---|
| OR Over-The-Counter | |
(Optional Format 1-2-96)
David C. Severson
(Division Sign-Off) Division of Reproductive, Abdominal, B and Radiological Dev 510(k) Number
39
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.