The HDI 1000 system is intended for opthalmic, fetal (OB/GYN), abdominal, intraoperative (abdominal organs, vasuclar, neurological,) pediatric, small organ, musculoskeletal (conventional, superficial,) neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, and peripheral vessel indications for use. Typical examinations shall include: General abdominal and pelvic studies including organ surveys and retroperitoneal cavity studies. Study of small parts including breasts, penis, testes, thyroid/parathyroid and the abdominal wall. Pediatric scans of organs. Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery. Monitoring procedures for infertility applications. First, second and third trimester pregnancy studies. Prostate, prostate biopsy guidance, and rectal wall studies. Neonatal cephalic studies. Transcranial studies of middle cerebral arteries, internal carotid artery, and vertebral arteries. Cardiac studies in adults and children. Biopsy guidance for tissue or fluid sampling. Ophthalmic studies of the eye and surrounding structures and studies to obtain blood flow information in the eye and surrounding structures. Conventional and superficial musculoskeletal studies.
Device Story
Mobile, software-controlled diagnostic ultrasound system; acquires ultrasound data via piezoelectric scanheads (linear, curved, phased arrays; 2.0-11.0 MHz). System processes reflected sound waves into 2-D, M-mode, Color Flow Doppler, Color M-mode, and Color Power Angio images. Operator selects scanheads via control panel; system software automatically optimizes operating conditions (frame rate, line density, frequency) based on user inputs. Output displayed on monitor for clinical diagnostic purposes; includes measurement and analysis packages. Used in clinical settings by healthcare professionals. Provides thermal/mechanical acoustic output indices (AIUM/NEMA 1992) to support ALARA principle. Benefits include non-invasive visualization of anatomical structures and blood flow for diagnostic assessment and biopsy guidance.
Clinical Evidence
Bench testing only. No clinical data provided. Device safety and effectiveness supported by compliance with electromechanical safety standards (EN 60601-1, UL 2601-1, IEC 1157) and adherence to AIUM/NEMA standards for real-time display of thermal and mechanical acoustic output indices.
Technological Characteristics
Mobile ultrasound system; piezoelectric transducer technology; 2.0-11.0 MHz linear, curved, and phased array scanheads. Modes: 2-D, M-mode, Color Flow, Color M-mode, Color Power Angio, Pulsed Doppler. Software-controlled optimization of frame rate, line density, and frequency. Biocompatible patient-contact materials. Compliance: EN 60601-1, UL 2601-1, C22.2 No. 601.1, IEC 1157, EN 60601-1-2. Acoustic output display per AIUM/NEMA 1992.
Indications for Use
Indicated for diagnostic ultrasound, pulsed Doppler, and pulsed echo imaging of the human body, including ophthalmic, fetal, abdominal, intraoperative, pediatric, small organ, neonatal/adult cephalic, cardiac, transrectal, transvaginal, peripheral vessel, and musculoskeletal (conventional/superficial) applications. Includes biopsy guidance.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
ATL HDI 3000 diagnostic ultrasound system
ATL Ultramark® 9 diagnostic ultrasound system
Atlantis Atlas 1.0 system
Storz Renaissance A/B Scan
HDI 5000 diagnostic ultrasound system
Submission Summary (Full Text)
{0}------------------------------------------------
K980860
510(k) Premarket Notification Musculoskeletal Indication for Use HDI® 1000 Ultrasound System
APR - 2 1998
### ATTACHMENT 1 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
#### Submitter's name, address, telephone number, contact person: 1)
Terrence J. Sweeney Vice President, Worldwide Quality and Regulatory Affairs Atlantis Diagnostics International, Inc. (ADII) 19015 North Creek Parkway, Suite 105 Bothell WA 98011 (425) 487-7602
Date prepared: March 3, 1998
- Name of the device, including the trade or proprietary name if applicable, the 2) common or usual name, and the classification name, if known:
Common/ Usual Name
Diagnostic Ultrasound System with Accessories
Proprietary Name
HDI® 1000 Diagnostic Ultrasound System
Classification Names:
Ultrasonic Pulsed Doppler Imaging System, Product Code 90 IYN, 21 CFR 892.1550
Diagnostic Ultrasonic Transducer, Product Code 90 ITX, 21 CFR 892.1570
Ultrasonic Pulsed Echo Imaging System, Product Code 90IYO, 21 CFR 892.1560
3) Identification of the predicate or legally marketed device:
> ADII believes that HDI 1000 system is substantially equivalent to the currently-marketed ATL HDI 3000 diagnostic ultrasound system, ATL Ultramark® 9 diagnostic ultrasound system, Atlantis Atlas 1.0 system, Storz Renaissance A/B Scan, and the HDI 5000 diagnostic ultrasound system.
### 4) Device Description:
This 510(k) adds the musculoskeletal applications (superficial and conventional) to the HDI 1000 system.
The HDI 1000 system is a general purpose, mobile, software-controlled, diagnostic ultrasound systems with an on-screen display for thermal and mechanical values related to potential bioeffect mechanisms. The HDI 1000 system functions by acquiring ultrasound data and displaying it on a monitor in
{1}------------------------------------------------
# K 980860
510(k) Premarket Notification Musculoskeletal Indication for Use HDI® 1000 Ultrasound System
2-D, M-mode, 2-D Color Flow Doppler, Color M-mode, Color Power Angio, Pulsed (PW) Doppler or in a combination of modes. An audio presentation of pulsed Doppler information is also available on the systems. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnostic purposes.
The HDI 1000 system is designed to accept a large selection of scanheads with up to two being connected to the system at any one time. The operator may select between the two scanheads by means of a control located on the system control panel. All actions affecting the performance of the scanhead are activated from the main system control panel.
The HDI 1000 system is designed to accept scanheads of the following types and frequency:
> frequency range: scanhead types:
2.0 - 11.0 MHz flat linear array curved linear array phased array
Specific operating conditions (frame rate, line density, center frequency, number of active elements etc.) are automatically optimized by the system software in response to user inputs such as field of view, focal depth, image quality, power etc.
The HDI 1000 system been designed to meet the following electromechanical safety standards:
- · EN 60601-1 European Norm, Medical Electrical Equipment
- · UL 2601-1, Underwriters Laboratories Standards, Medical Electrical Equipment
- · C22.2 No. 601.1, Canadian Standards Association, Medical Electrical Equipment
- · CEI/IEC 1157:1992, International Electrotechnical Commission, Requirements for the declaration of the acoustic output of medical diagnostic ultrasonic equipment
- · EN 60601-1-2, Collateral Standard: Electromagnetic Compatibility
5) Intended Use:
> The HDI 1000 system is intended for opthalmic, fetal (OB/GYN), abdominal, intraoperative (abdominal organs, vasuclar, neurological,) pediatric, small organ, musculoskeletal (conventional, superficial,) neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, and peripheral vessel indications for use.
Typical examinations shall include:
- · General abdominal and pelvic studies including organ surveys and retroperitoneal cavity studies.
- · Study of small parts including breasts, penis, testes, thyroid/parathyroid and the abdominal wall.
- Pediatric scans of organs. .
{2}------------------------------------------------
K980860
510(k) Premarket Notification Musculoskeletal Indication for Use HDI® 1000 Ultrasound System
- Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery.
- Monitoring procedures for infertility applications. .
- First, second and third trimester pregnancy studies. .
- Prostate, prostate biopsy guidance, and rectal wall studies. .
- Neonatal cephalic studies. .
- Transcranial studies of middle cerebral arteries, internal carotid artery, . and vertebral arteries.
- Cardiac studies in adults and children. .
- Biopsy guidance for tissue or fluid sampling. .
- Ophthalmic studies of the eye and surrounding structures and studies to . obtain blood flow information in the eye and surrounding structures.
- Conventional and superficial musculoskeletal studies. .
#### Technological Characteristics: 6)
This device operates identical to the predicate devices in that piezo material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as 2-D and M-mode images. Doppler shift caused by blood flow is displayed as Color Flow, or as spectrum analysis. The modes of these devices (2-D, M-mode, Color Flow, Color M-mode, Color Power Angio, and Pulsed Doppler) are the same as predicate devices identified in item 3. Scanhead patient contact materials are biocompatible and are also the same as the identified predicate devices.
These devices conform to the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment (AIUM/NEMA, 1992) for an on-screen display feature that provides information on potential thermal and cavitation bioeffect mechanisms. A user education program provides additional information so users may moderate the system's acoustic output in accordance with the ALARA (as low as reasonably achievable) principle.
The device's acoustic output limits are:
All Applications Other Than Ophthalmic:
| ISPTAd | 720 mW/cm2 | (Maximum) |
|-----------------------|---------------|-----------|
| TIS/TIB/TIC | 0.1 - 4.0 | (Range) |
| Mechanical Index (MI) | 1.9 | (Maximum) |
| ISPPAd | 0 - 700 W/cm2 | (Range) |
Ophthalmic Applications:
| ISPTAd | 50 mW/cm2 | (Maximum) |
|-----------------------|--------------|-----------|
| Thermal Index (TIC) | 0.1 - 1.0 | (Range) |
| Mechanical Index (MI) | .23 | (Maximum) |
| ISPPAd | 0 - 50 W/cm2 | (Range) |
The limits are same as predicate Track 3 devices.
510(k) Summary of Safety and Effectiveness
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 2 1998
Terrence J. Sweeney Vice President Worldwide Quality and Regulatory Affairs Atlantis Diagnostics International, Inc. 19015 North Creek Parkway Bothell, WA 98011
Re: K980860 HDI 1000 Diagnostic Ultrasound System Dated: March 3, 1998 Received: March 5, 1998 Regulatory class: II 21 CFR 892.1560/Procode: 90 IYO 21 CFR 892.1550/Procode: 90 IYN 21 CFR 892.1570/Procode: 90 ITX
Dear Mr. Sweeney:
We have reviewed your section 510/k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the HDI 1000 Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
CL10-5/10.0.5.0 MHz/Linear Array L11-5/11.0.5.0 MHz/38mm/Linear Array L19-5/9.0-5.0 MHz/38mm/Linear Array
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Terrence J. Sweeney
Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the investigational device exemptions (IDE) regulations, 21 CFR, Part 812. Therefore, your product labeling must be consistent with FDA's position on this use.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (2) CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Paul Gammell, Ph.D. at (301) 594-1212.
Sincerely yours,
Yaind h. Sypon
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal. Ear. Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
Section 4.3, page 1 of 4
Indications Statement for System
K980860
510(k) Number: Device Name:
HDI®1000 Ultrasound System
Indications for Use: Diagnostic ultrasound / Pulsed Doppler / Pulsed Echo imaging
(specify) of the human body as follows:
| Clinical Applications | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|-----------------------------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | P | | P | | P | P | | | |
| Fetal | | P | P | P | | P | P | | P (Note 1) | P (Note 2) |
| Abdominal | | P | | P | | P | P | | P (Note 1) | P (Note 2) |
| Intra-operative :<br>Abdominal organs &<br>vascular | | P | | P | | P | P | | P (Note 1) | P (Note 2) |
| Intra-operative<br>Neurological | | P | | P | | P | P | | P (Note 1) | |
| Pediatric | | P | | P | | P | P | | P (Note 1) | P (Note 2) |
| Small Organ<br>See Note 3 | | P | | P | | P | P | | P (Note 1) | P (Note 2) |
| Neonatal Cephalic | | P | | P | | P | P | | P (Note 1) | |
| Adult Cephalic | | P | | P | | P | P | | P (Note 1) | |
| Cardiac | | P | P | P | | P | P | | P (Note 1) | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | P | | P | | P | P | | P (Note 1) | P (Note 2) |
| Transvaginal | | P | P | P | | P | P | | P (Note 1) | P (Note 2) |
| Transurethral | | | | | | | | | | |
| Intavascular | | | | | | | | | | |
| Peripheral vessel | | P | | P | | P | P | | P (Note 1) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | | N | | N | N | | N (Note 1) | |
| Musculo-skeletal<br>Superficial | | N | | N | | N | N | | N (Note 1) | |
Mode of Operation (*includes simultaneous B-Mode)
Other Indications or Modes:
Note 1: PWD/CD, PWD/PAD
Note 2: Includes imaging for guidance of biopsy
Note 3: Thyroid, parathyroid, breast, scrotum, penis in adult, pediatric and neonatal patients
P indicates previously cleared indication in K961073. N indicates new indication.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON-ANOTHER PAGE IF ---NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Simid h. Segerson
(Division Sign-Off) (Division Sign-Ott)
Division of Reproductive, Abdominal, EN and Radiological L 510(k) Number
Prescription Use (Per 21 CFR 801.109)
{6}------------------------------------------------
Ultrasound Device Indications Statement
## K 980860
510(k) Number: Device Name: Transducer:
HDI®1000 Ultrasound System CL10-5/10.0-5.0 MHz/Linear Array
Indications for Use:
Diagnostic ultrasound / Pulsed Doppler / Pulsed Echo imaging (specify) of the human body as follows:
| Mode of Operation (*includes simultaneous B-Mode) | | | | | | | | | | |
|-----------------------------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Clinical Applications | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
| Ophthalmic | | P | | P | | P | P | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative :<br>Abdominal organs &<br>vascular | | P | | P | | P | P | | P (Note 1) | |
| Intra-operative | | | | | | | | | | |
| Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>See Note 2 | | P | | P | | P | P | | P (Note 1) | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intavascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | | N | | N | N | | N (Note 1) | |
| Musculo-skeletal<br>Superficial | | N | | N | | N | N | | N (Note 1) | |
Mode of Operation (*includes simultaneous B-Mode)
Other Indications or Modes:
Note 1: PWD/CD, PWD/PAD
Note 2: Thyroid, parathyroid, breast, scrotum, penis in adult, pediatric and neonatal patients
P indicates previously cleared indication in K961073. N indicates new indication.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Schmid G. Syverson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Dev
510(k) Number K980860
Prescription Use (Per 21 CFR 801.109)
{7}------------------------------------------------
Ultrasound Device Indications Statement
| 510(k) Number: | K980860 |
|----------------|--------------------------------------|
| Device Name: | HDI®1000 Ultrasound System |
| Transducer: | L11-5/11.0-5.0 MHz/38mm/Linear Array |
Indications for Use:
Diagnostic ultrasound / Pulsed Doppler / Pulsed Echo imaging (specify) of the human body as follows:
| Clinical Applications | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) |
|----------------------------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative :<br>Abdominal organs,<br>vascular | | P | | P | | P | P | | P (Note 1) | P (Note 2) |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | | P | | P | P | | P (Note 1) | P (Note 2) |
| Small Organ<br>See Note 3 | | P | | P | | P | P | | P (Note 1) | P (Note 2) |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intavascular | | | | | | | | | | |
| Peripheral vessel | | P | | P | | P | P | | P (Note 1) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | | N | | N | N | | N (Note 1) | |
| Musculo-skeletal<br>Superficial | | N | | N | | N | N | | N (Note 1) | |
Mode of Operation (*includes simultaneous B-Mode)
Other Indications or Modes:
Note 1: PWD/CD, PWD/PAD
Note 2: Includes imaging for guidance of biopsy.
Note 3: Thyroid, parathyroid, breast, scrotum, penis in adult, pediatric and neonatal patients.
P indicates previously cleared indication in K961073. N indicates new indication.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Syverson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT and Radiologicai D
510(k) Number K980860
Prescription Use (Per 21 CFR 801.109)
{8}------------------------------------------------
Section 4.3, page 4 of 4
Ultrasound Device Indications Statement
### K980860
| 510(k) Number: | K980860 |
|----------------|-------------------------------------|
| Device Name: | HDI®1000 Ultrasound System |
| Transducer: | LI9-5/9.0-5.0 MHz/38mm/Linear Array |
Indications for Use:
Diagnostic ultrasound / Pulsed Doppler / Pulsed Echo imaging (specify) of the human body as follows:
| Clinical Applications | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |
|---------------------------------------------------|------------|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------|--|
| | Ophthalmic | | | | | | | | | | |
| | Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intra-operative:<br>Abdominal organs,<br>vascular | | P | | P | | P | P | | P (Note 1) | P (Note 2) | |
| Intra-operative | | | | | | | | | | | |
| Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | N | | N | | N | N | | N (Note 1) | N (Note 2) | |
| Musculo-skeletal<br>Superficial | | N | | N | | N | N | | N (Note 1) | N (Note 2) | |
Mode of Operation (*includes simultaneous B-Mode)
Other Indications or Modes:
Note 1: PWD/CD, PWD/PAD
Note 2: Includes imaging for guidance of biopsy.
P indicates previously cleared indication in K961073. N indicates new indication.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) .
David G. Segerson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number L980860
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.