J-FX BIPOLAR HEAD

K980801 · Johnson & Johnson Professionals, Inc. · KWY · May 29, 1998 · Orthopedic

Device Facts

Record IDK980801
Device NameJ-FX BIPOLAR HEAD
ApplicantJohnson & Johnson Professionals, Inc.
Product CodeKWY · Orthopedic
Decision DateMay 29, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3390
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The J-FX Bipolar Head is indicated for use in partial hip replacement procedures for patients suffering severe pain and disability due to: femoral fracture · avascular necrosis of the femoral head · osteoarthritis · other abnormalities: where the major pathology affects the femoral head, where the acetabular cavity is normal and not deformed or weakened, and where acetabular replacement is not required or desirable.

Device Story

J-FX Bipolar Head is a metal/polymer femoral head endoprosthesis used in partial hip replacement surgery. Device consists of a cobalt-chrome alloy shell (ASTM F75) and a UHMWPE liner assembly (bearing insert, inner retaining ring, outer locking ring). Designed for use in patients with femoral head pathology where the acetabular cavity remains healthy. Surgeon implants device to restore hip function and alleviate pain. Device provides a bearing surface for the femoral hip head. Available in various shell diameters (38-60 mm) and liner sizes to accommodate anatomical variations.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: Co-Cr-Mo alloy shell (ASTM F75) and UHMWPE liner. Design: Three-piece liner (bearing insert, inner retaining ring, outer locking ring). Shell diameters: 38-60 mm. Liner sizes: 22.225 mm and 28 mm. Mechanical prosthesis; no software or electronics.

Indications for Use

Indicated for partial hip replacement in patients with severe pain/disability due to femoral fracture, avascular necrosis, or osteoarthritis, provided the acetabular cavity is normal and acetabular replacement is not required.

Regulatory Classification

Identification

A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K980801 # Summary of Safety and Effectiveness Data for the J-FX Bipolar Head MAY 2 9 1998 # Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, MA 02767-0350 ### Contact Person___________ Mary E. Gray Associate Regulatory Affairs Specialist Phone: (508) 828-3545 (508) 828-3212 Fax: ## Name of Device________________________________________________________________________________________________________________________________________________________________ | Proprietary Name: | J-FXTM Bipolar Head | |----------------------|----------------------------------------------------------------------------------| | Common Name: | Metal/Polymer Femoral Head Endoprothesis (Bipolar Cup) | | Classification Name: | Hip joint femoral (hemi-hip) metal/polymer cemented or<br>uncemented prosthesis. | | Regulatory Class: | Class II by 21 CFR 888.3390 | | Product Code: | KWY | | Owner/Operator No.: | 9001269 | #### Device Classification This device has been placed in Class II for Hip joint femoral (hemi-hip) metal /polymer cemented < uncemented prosthesis per 21 CFR § 888.3390. ## Statement of Substantial Equivalence The J-FX Bipolar Head is substantially equivalent to the P.F.C.® Bipolar Hip System cleared for marketing under premarket notifications K945793 (April 10, 1995)/ K931655 (June 9, 1994). The devices are identical in material, with the outer shell composed of cobalt-chrome alloy and the inner liner composed of UHMWPE. Both devices are identical in function (indicated use). {1}------------------------------------------------ ## Indications for Use The J-FX Bipolar Head is indicated for use in partial hip replacement procedures for patients suffering severe pain and disability due to: - femoral fracture - · avascular necrosis of the femoral head - · osteoarthritis - · other abnormalities: - where the major pathology affects the femoral head, - where the acetabular cavity is normal and not deformed or weakened, and - where acetabular replacement is not required or desirable. ## Physical Description The J-FX Bipolar Head is composed of an Co-Cr-Mo alloy shell (ASTM F75) and a ultra-high-molecular weight polyethylene (UHMWPE) liner. The liner is a three piece design consisting of a bearing insert, inner retaining ring and an outer locking ring. The bearing insert and inner retaining ring provide the bearing surface for the femoral hip head while the locking ring holds the liner assembly in place in the shell. The outer diameter of the metal shell varies in size from 38 through 58 mm in 1 mm increments for use with the 22.225 mm UHMWPE liner. For the 28 mm UHMWPE liner, the size range for the metal shell is 42 through 58 mm in 1 mm increments. A 60 mm metal shell is also available for use with both the 22.225 mm and 28 mm UHMWPE liner. Five sizes of polyethylene liners are available for the 22.225 mm J-FX Bilpolar head and four sizes for the 28 mm J-FX Bipolar head. With this availability of liner sizes, the mating of the full range of outer shell sizes are met. The polyethylene thickness ranges from 4.1 to 12.1 mm for the 22.225 mm head and 3.7 to 9.2 mm for the 28 mm head. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES • USA Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 2 9 1998 Ms. Mary E. Gray Associate Requlatory Affairs Specialist Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, Massachusetts 02767-0350 Re : K980801 Trade Name: J-FX™ Bipolar Head Regulatory Class: II Product Code: KWY Dated: February 25, 1998 Received: March 2, 1998 Dear Ms. Gray: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with .... . the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ #### Page 2 - Ms. Mary E. Gray This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html". Sincerely yours, a M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # SECTION II ## Indications for Use ## Indications for Use The J-FX Bipolar Head is indicated for use in partial hip replacement procedures for patients suffering severe pain and disability due to: - femoral fracture - · avascular necrosis of the femoral head - · osteoarthritis - other abnormalities: - where the major pathology affects the femoral head, - where the acetabular cavity is normal and not deformed or weakened, and - where acetabular replacement is not required or desirable. Prescription Use (Per 21 CFR 801.109) Acilla (Division Sign-Off) ା : Division . . . 510(k) Number_1981) 80 J-FX Bipolar Head Premarket Notification February 25, 1998
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...